Transactions of the Society of Instrument and Control Engineers
Online ISSN : 1883-8189
Print ISSN : 0453-4654
ISSN-L : 0453-4654
Estimation of the Differences between ITS-90 and IPTS-68 in the Range from 630°C to 960°C
Masaru ARAIHirohisa SAKURAI
Author information
JOURNAL FREE ACCESS

1992 Volume 28 Issue 11 Pages 1267-1273

Details
Abstract

The difference between the ITS-90 and the IPTS-68 in the temperature range from 630°C to 960°C is determined comparing an HTPRT and five Pt-Rh 10%/Pt thermocouples. The values of temperature based on the ITS-90 are obtained with an HTPRT calibrated at the freezing points of silver, aluminum, zinc and tin, and the triple point of water. Those of the IPTS-68 are obtained with thermocouples calibrated at the freezing points of gold, silver and aluminum. The HTPRT and the thermocouples are intercompared using a gas controlled sodium heat pipe, the temperature of which is stable within±10mK. The difference between the ITS-90 and the IPTS-68 is estimated from the intercomparison and is found to be 20mK±50mK near 800°C. The largest error in determination of the difference is attributed to the uncertainty of the thermocouples.

Content from these authors
© The Society of Instrument and Control Engineers (SICE)
Next article
feedback
Top