難消化性デキストリンおよび水溶性カルシウムを用いた生体臓器T1値,T2値近似ファントムの作成
山城 晶弘神谷 直紀大塚 薫駒津 和浩伊東 洋一久保田 展聡小林 正人
著者情報
ジャーナル フリー

2013 年 69 巻 2 号 p. 163-169

詳細
抄録

In magnetic resonance imaging (MRI), the ideal phantom should have similar T1 and T2 values to those of organs of interest for measuring the change in signal intensity, contrast ratio and contrast noise ratio. There have been several reports to develop such a phantom using materials with limited availability or complex methods. In this study, we have developed a simple phantom using indigestible dextrin and soluble calcium at 1.5-tesla MRI. The T1 and T2 values have been reduced by dissolving indigestible dextrin and soluble calcium in distilled water. The similar T1 and T2 values to those of organs (i.e., kidney cortex, kidney medulla, liver, spleen, pancreas, bone marrow, uterus myometrium, uterus endometrium, uterus cervix, prostate, brain white matter, and brain gray matter) have been obtained by varying the concentration of indigestible dextrin and soluble calcium. This phantom is easy to develop and has a potential to increase the accuracy of MRI phantom experiments.

著者関連情報
© 2013 公益社団法人 日本放射線技術学会