J Korean Orthop Assoc. 2004 Sep;39(5):566-571. Korean.
Published online Dec 29, 2017.
Copyright © 2004 The Korean Orthopaedic Association
Original Article

The Accuracy of Lower Extremity Alignment in a Total Knee Arthroplasty Using Computer-Assisted Navigation System

Jong Keun Seon and Eun Kyoo Song
    • Center for Joint Disease, Chonnam National University Hwasun Hospital, Jeonnam, Korea.

Abstract

Purpose

This study assessed the accuracy of the alignment of the lower extremity and the position of the components in a total knee arthroplasty (TKA), which was performed using a computer-assisted navigation system.

Materials and Methods

This study reviewed 40 TKA using a manual alignment system (group A) and 36 TKA using the Orthopilot. 4.0 navigation system (group B). The radiographs taken 2 months after surgery were used to measure the femoro-tibial angle, the mechanical axis deviation, and the inclination of the femoral and tibial component. A value within the optimum±3° was defined as an ‘acceptable’ and a value over the optimum±3° was befined as an ‘outlier’.

Results

The coronal inclination of the femoral component to the anatomical femoral axis showed outliers in 13 knees in group A and 1 knee in group B. The coronal inclination of the femoral component to the mechanical axis revealed 17 and 2 outliers in each group, respectively. The coronal inclination of the tibial component and sagittal inclination of femoral component showed no significant differences between two groups, but the sagittal inclination of tibial component showed 16 and no outlier in each group, respectively. Concerning the femoro-tibial angle, there were 9 outliers in group A and 2 outliers in group B. The mechanical axis deviation showed 9 and 2 outliers in groups A and B, respectively.

Conclusion

The navigation system helped to produce more ‘acceptable’ and less ‘outlier’ results than the manual alignment system. The alignment of the lower extremity after a TKA can be achieved more accurately using the navigation system.

Keywords
TKA; Navigation system; Alignment of lower extremity


Metrics
Share
PERMALINK