Next Article in Journal
Modification of the Properties of Polymer Composites in a Constant Magnetic Field Environment
Next Article in Special Issue
The Synthesis and Evaluation of RGD−Conjugated Chitosan Gel as Daily Supplement for Body Weight Control
Previous Article in Journal
Zinc Complexes with 1,3-Diketones as Activators for Sulfur Vulcanization of Styrene-Butadiene Elastomer Filled with Carbon Black
Previous Article in Special Issue
Interferon α2–Thymosin α1 Fusion Protein (IFNα2–Tα1): A Genetically Engineered Fusion Protein with Enhanced Anticancer and Antiviral Effect
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

In Vitro Analysis of Wearing of Hip Joint Prostheses Composed of Different Contact Materials

1
School of Biological Science and Medical Engineering, Beihang University, Beijing 100083, China
2
Beijing Institute of Medical Device Testing, Beijing 101111, China
3
Beijing Advanced Innovation Center for Biomedical Engineering, Beihang University, Beijing 100083, China
4
School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai 200240, China
*
Author to whom correspondence should be addressed.
Materials 2021, 14(14), 3805; https://doi.org/10.3390/ma14143805
Submission received: 15 May 2021 / Revised: 24 June 2021 / Accepted: 3 July 2021 / Published: 7 July 2021
(This article belongs to the Special Issue Biopolymer-Based Materials for Biomedical Engineering)

Abstract

:
Cobalt-chromium-molybdenum alloy (CoCrMo) and ceramic are the two most common materials for the femoral head in hip joint prostheses, and the acetabular liner is typically made from ultra-high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE), or highly cross-linked polyethylene blended with Vitamin E (VEXLPE). The selection of suitable materials should consider both wear performance and cost-effectiveness. This study compared the wear rate between different friction pairs using a hip joint simulator and then recommended a suitable prosthesis based on the corresponding processing technology and cost. All wear simulations were performed in accordance with ISO 14242, using the same hip joint simulator and same test conditions. This study found that when using the same material for the femoral head, the XLPE and VEXLPE liners had a lower wear rate than the UHMWPE liners, and the wear rate of the XLPE liners increased after blending with Vitamin E (VEXLPE). There was no significant difference in the wear rate of XLPE when using a CoCrMo or ceramic head. Considering the wear rate and cost-effectiveness, a CoCrMo femoral head with an accompanying XLPE liner is recommended as the more suitable combination for hip prostheses.

1. Introduction

Osteoarthritis (OA) is one of the most common diseases of the hip joint. It is clinically characterized by joint pain, deformation, and restricted movement [1]. Late-stage OA often requires surgical intervention, such as total hip arthroplasty (THA). However, postoperative complications after THA are common and can result in patient disability, implant failure, osteolysis, and prosthesis loosening [2]. It has been reported that over 50% of THA revisions are the result of implant loosening [3]. Wear particles generated between the femoral head and acetabular liner as the joint articulates are engulfed by macrophages to produce a large number of cytokines which activate osteoclasts and can cause osteolysis around the prosthesis, subsequently leading to prosthesis loosening [4,5]. Therefore, the wear performance of the hip joint prostheses is critical for long-term implant survival [6,7].
At present, the most widely used materials for the femoral head are cobalt-chromium-molybdenum alloy (CoCrMo) and ceramic, and the acetabular liner is typically made from ultra-high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE), or highly cross-linked polyethylene blended with Vitamin E (VEXLPE) [8]. Studies have shown that XLPE liners have a lower wear rate than UHMWPE [9,10] and that the addition of vitamin E to XLPE can further improve the anti-aging performance and wear resistance over UHMWPE liners [11,12]. It has also been reported that ceramic femoral heads offer superior wear resistance to CoCrMo heads [13].
Due to the large number of cycles required, the time-consuming nature of abrasion tests, and their associated costs, most studies to date conducted a comparative analysis of only one or two friction pairs. For example, Lizeth Herrera et al. only evaluated the wear performance of XLPE and UHMWPE liners [9], while Vesa Saikko assessed the impact of adding Vitamin E to XLPE [11]. Due to differences in test equipment, the methods, and the conditions simulated, there can be considerable variation in the results retrieved. Lizeth Herrera et al. reported an average wear rate of 1.35 ± 0.68 mm3/mc after 5 million cycles using MTS equipment and calf serum diluted to 20 g/L [9], whereas Hermida et al. reported a wear rate of 16.92 mg/mc using AMTI equipment and 90% calf serum [14]. Using AMTI equipment and a lubricant of undiluted bovine serum, Estok et al. performed abrasion tests on irradiated XLPE and recorded wear rates of −0.42 mg/mc for a 9.5-Mard gamma irradiated material and 1.2 mg/mc for a 5-Mard material [15]. In addition to the inconsistent results reported, these studies only considered a small selection of materials commonly used in acetabular liners. Few studies have evaluated the wear performance of different combinations of head and liner materials.
The purpose of this study is to evaluate how different combinations of materials for the femoral head and acetabular liner affect the wear rate of artificial hip joint prostheses. This study hypothesized that a ceramic head paired with a VEXLPE liner would have the lowest wear rate.

2. Materials and Methods

This study consisted of 5 groups with different contact materials: ceramic head and XLPE liner, ceramic head and VEXLPE liner, CoCrMo head and XLPE liner, and CoCrMo head and UHMWPE liner. With reference to ISO 14242-1:2014 [16], ISO 14242-2:2016 [17], five samples were used for the wear test of each group of contact materials (combination of a femoral head and acetabular liner). Three of the five samples in each group were randomly chosen as tested samples and the remaining two (or one) were control samples for the calculation of wear loss. The head and liner combinations chosen were commercially available total hip prostheses (Table 1). The materials used to manufacture the liners were produced by the same company (Quadrant, Switzerland) using the same manufacturing processes, such as polyethylene resins, crosslinking dose or crosslink density, post-irradiation thermal treatments, and incorporation methods. The unformed materials were then processed into liners by the Manufacturers stated in Table 1. Except for sizing, the ceramic heads were identical. All products conformed to YY 0118-2016 “Joint replacement implants—Hip joint prostheses” [18] (equivalent to the ISO 7206-2 “Implants for surgery—Partial and total hip joint prostheses—Part 2: Articulating surfaces made of metallic, ceramic, and plastics materials” [19]). The liners in this study had a surface roughness of 0.3–0.5 μm, a thickness of 5–7 mm, and a roundness of 50–70 μm. The femoral heads in this study had a surface roughness of 0.010–0.015 μm, and a roundness of 2–5 μm.

2.1. Samples

Different sizes of ceramic and CoCrMo femoral heads were paired with XLPE, VEXLPE, and UHMWPE liners to create 5 groups (Table 1). The wear rates were compared under the same test conditions developed in accordance with ISO 14242-1:2014 [16], ISO 14242-2:2016 [17].

2.2. Wear Tests

An AMTI joint prosthesis simulator (ADL-H12-01, AMTI, Watertown, MA, USA) (Figure 1) was used for all wear tests, which were performed in accordance with ISO 14242-1:2014 [16], as detailed below.
All samples were weighed on a high precision analytical balance (XS205DU, Mettler Toledo LLC, Columbus, OH, USA) with a maximum weight capacity of 220 g and a resolution of 0.01 mg.
The liners were first pre-soaked in calf serum for 2 days. The liners were then repeatedly washed, dried, and weighed until the mass change of each sample exceeded 24 h after being less than 10% of the cumulative mass change of the sample, as specified in ISO 14242-2:2016 [17].
After a constant weight was achieved, the femoral head was mounted in a custom rig. The focus of this study is the wear performance of the contact materials, and so only the wear produced by the articulating surfaces was considered. Therefore, the acetabular liner and the acetabular cup were directly secured in a rig (acetabular cup fixing tool) using a curing agent, and the rig was connected to an actuator on the simulator, which could simulate abduction, adduction, flexion, and extension of the hip joint (Figure 1a). As specified in ISO 14242-1:2014 [16], the acetabular component was tilted 30° ± 3° in the axial direction (Figure 1b). All three test specimens were subject to flexion/extension, abduction/adduction, inward/outward rotation, and an axial force using the input values recommended in ISO 14242-1:2014 [17]. Variations in the angular movement of the femoral component relative to the acetabular component were −18° to +25° for flexion/extension, −10° to +2° for inward/outward rotation, and −4° to +7° for abduction/adduction. The axial force ranged between 300 and 3000 N in one loading cycle.
Calf serum was diluted with deionized water to obtain a solution with a protein content of 30 g/L to simulate human synovial fluid. After every 500,000 cycles, the serum was completely replaced, and each test specimen was moved to a different simulator to minimize the impact of system errors.
The temperature of the serum was controlled at 37 °C using an antiseptic agent consisting of 0.2% NaN3 and 5 mmol/L EDTA.
5 million cycles of wear test were performed for each group.
The gravimetric wear and wear rate were calculated according to ISO 14242-2:2016 [17]. The gravimetric wear ( W n ) referred to the net loss of mass from each test specimen after n loading cycles and was calculated using Equation (1). The gravimetric wear of all the test specimens was calculated at each testing stage during the wear simulation (500,000, 1 million, 2 million, 3 million, 4 million, and 5 million cycles).
W n = m 0 m n + S n S n = m n ¯ m 0 ¯
  • m 0 —the mass of the test specimen before the wear test
  • m n —the mass of the test specimen after n loading cycles
  • S n —the increase in mass of the control specimen over the same period
  • m 0 ¯ —the mass of the control specimen before the wear test
  • m n ¯ —the mass of the control specimen after n loading cycles
The wear rate ( a G ) was calculated as the least-squares linear fit relationship between W n and the number of loading cycles (n) using Equation (2).
W n = a G × n + b
  • Wn—the net loss in quality of the test specimen after n load cycles
  • n—cycles
  • b—constant
The average wear rate for the three test specimens was taken as the wear rate for that group. The results of Group 1 and Group 2 were compared to understand how adding Vitamin E to XLPE affected the wear rate of the liner. Group 3 and Group 4 were analyzed to determine the relationship between the wear rate of the liner and the accompanying femoral head material. And the wear rate from Group 4 and Group 5 was compared to determine which material (XLPE and UHMWPE) offered superior wear resistance when using a femoral head of the same size and material.

3. Results

The gravimetric wear of each test sample during the wear test is given in Figure 2. The wear rate of the five groups is shown in Table 2.

3.1. Ceramic Head with XLPE and VEXLPE Liners

When combined with a 36 mm ceramic head, the wear rate of the XLPE liner with Vitamin E (VEXLPE, Group 2) had a considerably higher wear rate (10.45 ± 1.01 mg/million cycles) than the plain XLPE liner (Group 1) (−2.48 ± 0.64 mg/million cycles). The addition of Vitamin E had a detrimental effect on the wearing of the liner. The negative wearing may be explained by the acetabular liner absorbing calf serum during the wear process. While each of the liners will absorb a certain amount of calf serum, when the amount of serum absorbed by the acetabular liner exceeds the loss in material mass from the liner, such as with XLPE (Group 1), negative wear will appear.

3.2. Ceramic and CoCrMo Femoral Heads with XLPE Liner

Ceramic and CoCrMo femoral heads of diameter 28 mm were assembled with an XLPE acetabular liner and assigned to Groups 3 and 4, respectively. The wear rate from these two groups was very similar, with rates of 0.60 ± 0.21 mg/million cycles and 0.68 ± 0.11 mg/million cycles respectively.

3.3. CoCrMo Femoral Head with XLPE and UHMWPE Liners

A 28 mm CoCrMo femoral head was paired with an XLPE liner (Group 4) and UHMWPE liner (Group 5). The results showed that the XLPE liner had a much lower mean wear rate (0.68 ± 0.11 mg/million cycles) than the UHMWPE liner (24.97 ± 1.59 mg/million cycles).

4. Discussion

The wear mechanisms of friction components depend on the conditions of articulation, material properties, and the surface topography of the co-acting parts [20]. The wear mechanisms of the polymeric socket resulted from several phenomena; including plastic deformation, abrasive wear, fatigue, and adhesion [20,21]. Therefore, it is important to examine these determinants to improve the durability of a friction pair.
The conditions of articulation of the samples selected in this study are kept the same, and the surface topography (surface roughness, roundness) of the friction joint articular surface is also kept within the manufacturing tolerance range, so the material is hypothesized to be the main factor affecting the wear performance in this study.
This study found that the XLPE and VEXLPE liners had a lower wear rate than the UHMWPE liners, and the wear rate of the XLPE liners increased after adding Vitamin E (Figure 3). However, changing the material of the femoral head had little effect on the wear rate.
It is known that the volume and rate of fluid uptake are affected by whether the bearing is in motion [15]. Therefore, the mass change of the absorbed serum by the control sample (only subjected to axial load) in each group does not accurately represent the amount of serum absorbed by the test sample (simulating human gait movement). Moreover, after the prosthesis absorbs serum, it is difficult to completely remove the serum through dehydration and deproteinization using the cleaning process described in ISO 14242-2:2016 [17].
Affatato et al. reported a wear rate of 16.1 ± 8.2 mg/million cycles for a VEXLPE liner combined with a CoCrMo femoral head [22]. In this current study, Group 4 combined an XLPE liner with a CoCrMo head but recorded a considerably lower wear rate of 0.68 ± 0.11 mg/million cycles. Although these studies used a different-sized femoral head, the results clearly show that adding Vitamin E to the liner can drastically increase the wear rate. The reason may be that the addition of VE inhibits the amount of cross-linking of XLPE [21,23], resulting in a loss of wear resistance. Meneghini et al. [24] recorded the friction torque on ceramic and CoCrMo heads when paired with XLPE liners with and without Vitamin E. With the addition of VE, the torque on the ceramic head increased by 58% and increased by 31% on the CoCrMo head. The higher torque produces greater friction, which can increase the degree of wear. This may be one of the reasons why the wear resistance of XLPE was reduced after adding VE.
As can be seen in Figure 4, changing the head material has little effect on the wear rate of the liner. Similarly, Merola et al. [21] reported a wear rate of 4.09 ± 0.64 mm3/106 cycles for a CoCrMo-XLPE combination and 3.35 ± 0.29 mm3/106 cycles for an Alumina-XLPE combination. Therefore, compared with a UHMWPE liner, XLPE offers better wear resistance [10].
Figure 5 clearly shows a much lower wear rate for the XLPE liner than the UHMWPE liner. When paired with a 32 mm CoCrMo head, Lizeth et al. [9] reported a wear rate of 1.35 ± 0.68 mm3/106 cycles for an XLPE liner and 46 ± 8.7 mm3/106 cycles for UHMWPE, similar to previous findings reported in clinical follow-up studies [25]. Tipper et al. [13] reported a wear rate of 31.0 ± 4.0 mm3/106 cycles (~29 mg/million cycles) for a UHMWPE liner paired with a 28 mm ceramic head, which is much greater than the wear rate observed in Group 3 of 0.60 ± 0.21 mg/million cycles. In conclusion, XLPE liners have a much lower wear rate than both VEXLPE and UHMWPE liners irrespective of the material used for the femoral head.
The cost of materials is also an important consideration when selecting a suitable prosthesis; In China, UHMWPE liners range from ¥2000 (approximately $310)~¥5000 (approximately $770), XLPE liners from ¥5000 (approximately $770)~¥10,000 (approximately $1550), CoCrMo heads from ¥3000 (approximately $460)~¥6000 (approximately $920), and ceramic heads from ¥10,000(approximately $1550)~¥20,000 (approximately $3100) each.
There are some limitations to this study that should be noted. Firstly, because of the long cycle of the wear test and high cost, representative sizes were chosen for this study. The aim of this study was to evaluate implant wear from different material combinations, therefore all comparisons were carried out using similarly sized implants. Secondly, the amount of serum absorbed by the control group was not necessarily the same as the test group. However, each group used the same method to set the control group and the test group, according to the methods detailed in ISO 14242, which also allowed comparison between groups. Thirdly, as with all in vitro research, the methods used in this study cannot fully replicate the in vivo conditions of the hip joint. However, the aim of this study was to compare the wear performance of different materials, and so the methods used are adequate for this purpose.

5. Conclusions

This study found that the wear rate of XLPE liners was lower than both VEXLPE and UHMWPE liners. Also, importantly, the wear rate of XLPE increased after adding Vitamin E (VEXLPE). The use of a CoCrMo or ceramic head had little effect on the wear resistance of the XLPE liners. Because ceramic femoral heads are more expensive and the production process is more complicated, and considering the superior wear resistance of XLPE, a CoCrMo femoral head with an accompanying XLPE liner is potentially superior to other contact materials in terms of wear resistance and cost-effectiveness.

Author Contributions

Conceptualization, J.S., M.Z., H.-Z.W. and C.-K.C.; Data curation, J.S., J.-J.W., S.-T.Y. and C.-K.C.; Formal analysis, J.S., J.-J.W., S.-T.Y. and C.-K.C.; Funding acquisition, C.-K.C.; Investigation, J.S., J.-J.W., S.-T.Y., H.-Z.W., N.-Z.Z. and Y.-C.L.; Methodology, J.S., J.-J.W., S.-T.Y., M.Z. and C.-K.C.; Project administration, C.-K.C.; Supervision, C.-K.C.; Validation, J.S., J.-J.W., S.-T.Y. and C.-K.C.; Visualization, J.S., N.-Z.Z., Y.-C.L. and C.-K.C.; Writing—original draft, J.S., J.-J.W., H.-Z.W. and C.-K.C.; Writing—review & editing, M.Z., H.-Z.W., N.-Z.Z., Y.-C.L. and C.-K.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Natural Science Foundation of China (Grant number: 2016YFC1101904).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data is available from the corresponding author.

Acknowledgments

We wish to thank Colin McClean for his assistance with editing the manuscript.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Mcalindon, T.E.; Lavalley, M.P.; Gulin, J.P.; Felson, D.T.J.J. Glucosamine and Chondroitin for Treatment of Osteoarthritis: A Systematic Quality Assessment and Meta-analysis. JAMA 2000, 283, 1469–1475. [Google Scholar] [CrossRef] [PubMed]
  2. Mayer, C.; Rommelmann, M.; Behringer, M.; Jäger, M.; Krauspe, R.; Zilkens, C. Wear Kinetics of Highly Cross-Linked and Conventional Polyethylene Are Similar at Medium-term Follow-Up After Primary Total Hip Arthroplasty. J. Arthroplast. 2018, 33, 2671–2676. [Google Scholar] [CrossRef] [PubMed]
  3. Jalali, O.; Scudday, T.; Fickenscher, M.-C.; Barnett, S.; Gorab, R. Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty. Arthroplast. Today 2020, 6, 316–321. [Google Scholar] [CrossRef]
  4. Purdue, P.E.; Koulouvaris, P.; Potter, H.G.; Nestor, B.J.; Sculco, T.P. The Cellular and Molecular Biology of Periprosthetic Osteolysis. Clin. Orthop. Relat. Res. 2007, 454, 251–261. [Google Scholar] [CrossRef]
  5. Gallo, J.; Raska, M.; Mrázek, F.; Petrek, M. Bone remodeling, particle disease and individual susceptibility to periprosthetic osteolysis. Physiol. Res. 2008, 57, 339–349. [Google Scholar] [CrossRef] [PubMed]
  6. Mattei, L.; Di Puccio, F.; Ciulli, E.; Pauschitz, A. Experimental investigation on wear map evolution of ceramic-on-UHMWPE hip prosthesis. Tribol. Int. 2020, 143, 106068. [Google Scholar] [CrossRef]
  7. Jäger, M.; Van Wasen, A.; Warwas, S.; Landgraeber, S.; Haversath, M.; VITAS Group. A multicenter approach evaluating the impact of vitamin E-blended polyethylene in cementless total hip replacement. Orthop. Rev. 2014, 6, 5285. [Google Scholar]
  8. Lachiewicz, P.F.; Kleeman, L.T.; Seyler, T. Bearing Surfaces for Total Hip Arthroplasty. J. Am. Acad. Orthop. Surg. 2018, 26, 45–57. [Google Scholar] [CrossRef] [PubMed]
  9. Herrera, L.; Lee, R.; Longaray, J.; Essner, A.; Wang, A. Hip simulator evaluation of the effect of femoral head size on sequentially cross-linked acetabular liners. Wear 2007, 263, 1034–1037. [Google Scholar] [CrossRef]
  10. Affatato, S.; Bersaglia, G.; Rocchi, M.; Taddei, P.; Fagnano, C.; Toni, A. Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions. Biomaterials 2005, 26, 3259–3267. [Google Scholar] [CrossRef]
  11. Saikko, V. Wear and friction of thin, large-diameter acetabular liners made from highly cross-linked, vitamin-E-stabilized UHMWPE against CoCr femoral heads. Wear 2019, 432–433, 202948. [Google Scholar] [CrossRef]
  12. Bracco, P.; Oral, E. Vitamin E-stabilized UHMWPE for Total Joint Implants: A Review. Clin. Orthop. Relat. Res. 2011, 469, 2286–2293. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  13. Tipper, J.; Firkins, P.; Besong, A.; Barbour, P.; Nevelos, J.; Stone, M.; Ingham, E.; Fisher, J. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator. Wear 2001, 250, 120–128. [Google Scholar] [CrossRef]
  14. Hermida, J.C.; Bergula, A.; Chen, P.; Colwell, C.W.; D’lima, D.D. Comparison Of The Wear Rates Of Twenty-Eight And Thirty-Two-Millimeter Femoral Heads On Cross-Linked Polyethylene Acetabular Cups In A Wear Simulator. J. Bone Jt. Surg.-Am. Vol. 2003, 85, 2325–2331. [Google Scholar] [CrossRef]
  15. Estok, D.M., 2nd; Burroughs, B.R.; Muratoglu, O.K.; Harris, W.H. Comparison of hip simulator wear of 2 different highly cross-linked ultra high molecular weight polyethylene acetabular components using both 32- and 38-mm femoral heads. J. Arthroplast. 2007, 22, 581–589. [Google Scholar] [CrossRef] [PubMed]
  16. ISO 14242-1. Implants for Surgery—Wear of Total Hipjoint Prostheses—Part 1: Loading and Displacement Parameters for Wear-Testing Machines and Corresponding Environmental Conditions for Test; ISO Copyright Office: Geneva, Switzerland, 2014. [Google Scholar]
  17. ISO 14242-2. Implants for Surgery—Wear of Total Hip-Joint Prostheses—Part 2: Methods of Measurement; ISO Copyright Office: Geneva, Switzerland, 2016. [Google Scholar]
  18. YY 0118-2016. Joint Replacement Implants-Hip Joint Prostheses; China Food and Drug Adminstration: Beijing, China, 2016.
  19. ISO 7206-2. Implants for Surgery—Partial and Total Hip Joint Prostheses —Part 2: Articulating Surfaces Made of Metallic, Ceramic and Plastics Materials; ISO Copyright Office: Geneva, Switzerland, 2011. [Google Scholar]
  20. Niemczewska-Wójcik, M.; Piekoszewski, W. The surface topography of a metallic femoral head and its influence on the wear mechanism of a polymeric acetabulum. Arch. Civ. Mech. Eng. 2017, 17, 307–317. [Google Scholar] [CrossRef]
  21. Merola, M.; Affatato, S. Materials for Hip Prostheses: A Review of Wear and Loading Considerations. Materials 2019, 12, 495. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  22. Affatato, S.; Freccero, N.; Taddei, P. The biomaterials challenge: A comparison of polyethylene wear using a hip joint simulator. J. Mech. Behav. Biomed. Mater. 2016, 53, 40–48. [Google Scholar] [CrossRef] [PubMed]
  23. Krushell, R.J.; Fingeroth, R.J.; Cushing, M.C. Early femoral head penetration of a highly cross-linked polyethylene liner vs a conventional polyethylene liner: A case-controlled study. J. Arthroplast. 2005, 20 (Suppl. 3), 73–76. [Google Scholar] [CrossRef] [PubMed]
  24. Meneghini, R.M.; Lovro, L.R.; Wallace, J.M.; Ziemba-Davis, M.J.J.A. Large Metal Heads and Vitamin E Polyethylene Increase Frictional Torque in Total Hip Arthroplasty. J. Arthroplast. 2016, 31, 710–714. [Google Scholar] [CrossRef] [PubMed]
  25. Krushell, R.J.; Fingeroth, R.J. Paper #1 Early Wear of a Crosslinked Polyethylene Liner Versus a Conventional Polyethylene Liner: A Case-Controlled Study. J. Arthroplast. 2005, 20, 265. [Google Scholar]
Figure 1. (a) Fixation in the hip joint simulator. (b) Fixation of the acetabular cup.
Figure 1. (a) Fixation in the hip joint simulator. (b) Fixation of the acetabular cup.
Materials 14 03805 g001
Figure 2. Gravimetric wear of the five test groups: (a) Group1; (b) Group2; (c) Group3; (d) Group4; (e) Group5 (see Table 2).
Figure 2. Gravimetric wear of the five test groups: (a) Group1; (b) Group2; (c) Group3; (d) Group4; (e) Group5 (see Table 2).
Materials 14 03805 g002aMaterials 14 03805 g002b
Figure 3. Wear rate for Group 1 and Group 2.
Figure 3. Wear rate for Group 1 and Group 2.
Materials 14 03805 g003
Figure 4. Wear rate for Group 3 and Group 4.
Figure 4. Wear rate for Group 3 and Group 4.
Materials 14 03805 g004
Figure 5. Wear rate for Group 4 and Group 5.
Figure 5. Wear rate for Group 4 and Group 5.
Materials 14 03805 g005
Table 1. Prosthesis combinations used in this study.
Table 1. Prosthesis combinations used in this study.
Group NumberMaterialHead Diameter (mm)Manufacturer
1Ceramic head & XLPE liner36Company A, Beijing Economic and Technological Development Zone, Beijing
2Ceramic head & VEXLPE liner36Company B, Shunyi, Beijing
3Ceramic head & XLPE liner28Company C, Haidian, Beijing
4CoCrMo head & XLPE liner28Company D, Haidian, Beijing
5CoCrMo head & UHMWPE liner28Company E, Tuttlingen, Germany
Table 2. Wear rate after 5 × 106 cycles.
Table 2. Wear rate after 5 × 106 cycles.
Group NumberHead Diameter (mm)MaterialsWear Rate of Liner (mg/million Cycles)Mean Wear Rate (mg/million Cycles)
136Ceramic−XLPE−3.000−2.48 ± 0.64
−2.660
−1.770
236Ceramic−VEXLPE11.60010.45 ± 1.01
10.070
9.690
328Ceramic−XLPE0.6400.60 ± 0.21
0.610
0.540
428CoCrMo−
XLPE
0.5990.68 ± 0.11
0.808
0.630
528CoCrMo−UHMWPE26.33324.97 ± 1.59
25.362
23.223
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Su, J.; Wang, J.-J.; Yan, S.-T.; Zhang, M.; Wang, H.-Z.; Zhang, N.-Z.; Luan, Y.-C.; Cheng, C.-K. In Vitro Analysis of Wearing of Hip Joint Prostheses Composed of Different Contact Materials. Materials 2021, 14, 3805. https://doi.org/10.3390/ma14143805

AMA Style

Su J, Wang J-J, Yan S-T, Zhang M, Wang H-Z, Zhang N-Z, Luan Y-C, Cheng C-K. In Vitro Analysis of Wearing of Hip Joint Prostheses Composed of Different Contact Materials. Materials. 2021; 14(14):3805. https://doi.org/10.3390/ma14143805

Chicago/Turabian Style

Su, Jian, Jian-Jun Wang, Shi-Tong Yan, Min Zhang, Hui-Zhi Wang, Ning-Ze Zhang, Yi-Chao Luan, and Cheng-Kung Cheng. 2021. "In Vitro Analysis of Wearing of Hip Joint Prostheses Composed of Different Contact Materials" Materials 14, no. 14: 3805. https://doi.org/10.3390/ma14143805

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop