日本レーザー医学会誌
Online ISSN : 1881-1639
Print ISSN : 0288-6200
ISSN-L : 0288-6200
CO2レーザ手術のBIOPHYSICS (第3部)
CO2レーザ切開の定量的分析と問題点
菱本 久美郎
著者情報
ジャーナル フリー

1980 年 1 巻 1 号 p. 245-250

詳細
抄録

This series of reports entitled “Biophysical Studies on CO2 Laser Surgery” is based upon the CO2 laser surgical research which we performed from 1969 thru 1970 at the Laser Laboratory of the University of Cincinnati Medical Center.
Experiments were undertaken to quantitate depth of incision by CO2 laser system as a function of laser output, power density and cutting velocity. An American Optical CO2 laser with surgical manipulator was used for focal incision of liver and skin. Experimental data are shown in Figure 1, 3, 4, 5 & 6.
(1) Graphical analysis of the relationship among laser power (P), incision depth (X) and cutting speed (v) indicates functional relationship of the form: X=k√P/v, where k is a constant. (Hishimoto-Rockwell's formula)
(2) Further graphical analyses of the relationship of power density (H), cutting velocity (v) and incision depth (X) suggest a relationship of the form: H=KvNXn, where K is a constant.
(3) Non-linear curve, suggestive of a functional relationship of H=Aebv, at higher cutting velocities for skin incisions in Figure 5 must be attributed to increases in focal spot size and quasi-CW nature of energy delivery of the CO2 laser scalpel.
(4) A problem that arises in applying the above formulas for deeper incision depth is abrupt reduction of the power density beyond the focal waist of the laser beam as the beam diverges. This variation in power density in deeper layer of the tissue may lead to deviation in mathematically expected values from the true ones.

著者関連情報
© 特定非営利活動法人 日本レーザー医学会
前の記事 次の記事
feedback
Top