Skip to content
BY 4.0 license Open Access Published by De Gruyter June 5, 2023

Developments in reference measurement systems for C-reactive protein and the importance of maintaining currently used clinical decision-making criteria

  • Mauro Panteghini EMAIL logo

Of current inflammatory markers identified, C-reactive protein (CRP) has the assay and biological characteristics most conducive to use in clinical practice [1]. Its utility, strongly supported by several consensus reports [2], has been recently confirmed during the SARS-Cov-2 pandemic to both predict in-hospital death and exclude the need for admission in intensive care unit in coronavirus disease 2019 patients [3, 4]. As described in more detail in another contribution published on this journal issue [5], the road towards harmonization of CRP measurements has represented a challenging enterprise asking for about 20 years of efforts before to see a positive outcome on marketed assays [6]. Figure 1 displays the calibration hierarchy currently in place, using as common calibrator a commutable liquid frozen reference material with CRP entirely in the native status (ERM-DA474/IFCC), traceable to CRM 470 (later renamed to ERM-DA470) in turn calibrated with the WHO International Standard 85/506 (human CRP in pooled normal human serum) to which values had been assigned by using a pure protein as calibrant. Using this metrological traceability approach, the inter-assay variability, as witnessed by published data, is currently very good, with CV values in External Quality Assessment programs permanently below 10 %, and values close to 3 % if only the widely used commercial measuring systems are considered [7], [8], [9]. Importantly, this translates to the practical possibility to use common decision thresholds for the CRP clinical use minimizing result misinterpretation (Figure 2).

Figure 1: 
Metrological traceability chain currently adopted by the great majority of in vitro diagnostic manufacturers providing commercial assays for C-reactive protein (CRP).
Figure 1:

Metrological traceability chain currently adopted by the great majority of in vitro diagnostic manufacturers providing commercial assays for C-reactive protein (CRP).

Figure 2: 
Decision thresholds (in mg/L) for C-reactive protein as a marker of inflammation.
Figure 2:

Decision thresholds (in mg/L) for C-reactive protein as a marker of inflammation.

In the recent years, new CRP candidate reference materials (RM) were nominated to be listed in the Joint Committee for Traceability in Laboratory Medicine (JCTLM) database. JCTLM was however concerned about the impact of the materials for the laboratory medicine community that the JCTLM database serves. It was indeed unclear in which position of the current traceability chain for CRP shown in Figure 1 should these RM be inserted or if they belong to newly proposed traceability chains in which these materials are traced to the International System of Units (SI) through, e.g., the amino acids RM and the hydrolysis-isotope dilution mass spectrometry (IDMS) procedure. In other words, it was unclear if the new CRP candidate RM providers proposed an alternative calibration hierarchy to that currently used in the clinical community, already listed on the JCTLM database [10]. If this is the case, JCTLM recommended that further investigations should be carried out to understand the implications of introducing a new traceability chain for CRP for patient clinical results before proceeding with adoption of newly characterized RM and listing them on the JCTLM database.

The definition of the measurand, as quantity intended to be measured [11], is one of the major prerequisites needed for the application of the metrological traceability theory [12]. Once defined, the analytical selectivity towards the measurand should be maintained throughout procedures used in the entire traceability chain, and employed primary and secondary RM should contain the same ‘analytical target’. Concerning CRP, an agreement about the measurand definition can be easily reached: it is a non-covalent pentamer with identical subunits, each consisting of a single 206 amino acid chain and a relative molecular mass of ∼23,050, with minimal known post-translational modifications. However, problems observed for ‘CRP measurand’ in different type of candidate RM have been reported. Rzychon et al. showed that the lyophilization process resulted in a loss of ∼20 % of measurable CRP compared to the non-lyophilized material, even if both formats appeared commutable [13]. The presence of a monomeric form of CRP only in lyophilized materials, due to the partial dissociation of CRP, affected the signal response of different commercial immunoassays showing the centrality of the CRP aggregation status in the accurate measurand recovery. The lesson to be learned was clear: for RM to be used in standardization of CRP, changes in the processing steps may have a pronounced impact on the measurement results and would consequently lead to a biased calibration, even if the RM is commutable [14, 15]. Hence, a careful evaluation of the impact of different RM formats and processing steps on the properties of the RM and its suitability for calibration is a crucial requirement. Furthermore, the presence of a multimeric heterogeneity may affect different measurement procedures to a different extent: for instance, in IDMS-based procedures all polypeptide chains are measured, including ‘non-native’ ones [16]. This would not describe the properties of native CRP as ‘target analyte’ completely, leading to a significant bias if IDMS methods would have been used anywhere in the traceability chain, e.g., as secondary reference measurement procedure (RMP) [14, 17].

Table 1 shows the requirements to be considered and correctly fulfilled in the preparation of a CRP RM to be used as common calibrator of commercial assays. In addition to control all relevant parameters discussed above that may modify the measurand definition, the commutability of candidate RM should be assessed by following the available recommended approaches [18], [19], [20]. Commutability is the ability of a RM to show inter-assay properties comparable to those of human samples. Only commutable materials can be used for direct value assignment to manufacturers’ calibrators, having large importance to ensure an unbroken traceability chain [21]. Non-commutability of a material means that using it for calibration will introduce a bias in the calibrated procedure, with wrong results for clinical samples and the risk of incorrect medical decisions. In developing higher-order RM, focusing on allowable measurement uncertainty (MU) is also important. As discussed elsewhere, the MU specifications of RM are defined by the performance specifications of the MU on clinical samples [10, 22, 23]. Finally, an investigation on possible effects for patient result interpretation with the introduction of new higher-order references for CRP is mandatory, and a comparison of measurement results obtained with the new calibration hierarchy to results of measurements obtained with the previous calibration should be established [24]. Already 15 years ago, I highlighted the importance of maintaining currently used clinical decision-making criteria and related clinical experience even when a new traceability chain incorporating theoretically better metrological aspects is proposed [12].

Table 1:

Characteristics to be considered in the preparation of a candidate reference material to be used for common calibration of commercial C-reactive protein assays.

  1. Avoid any influence that may modify the measurand definition and control all relevant parameters (e.g., matrix properties and interactions, structure, aggregation, oligomerization)

  2. Assess commutability

  3. Consider the allowable measurement uncertainty at the clinical sample level and use only a portion of it

  4. Provide evidence of the extent-of-equivalence making a comparison with reference materials already in use

In summary, new candidate RM or RMP for CRP cannot stand alone, but it must be integral parts of a reference measurement system, the current in use or a new one, and this must be clearly disclosed. CRP is a pentameric protein, and the pentamer is the clinical measurand. In its monomeric form CRP tends to aggregate. As most of the RM characterizations are carried out by sodium dodecyl sulphate-polyacrylamide gel electrophoresis (SDS-PAGE) and matrix-assisted laser desorption/ionization-time of flight (MALDI-ToF) on the monomeric form, it is recommended using other methods to check for the pentamer. The RM intended use requires validation of its usability for the purpose. If the intended use of new candidate RM includes the calibration of field methods used in medical laboratories, this requires commutability studies involving parallel measurements of the candidate RM and clinical samples for all commercial assays with which it may be potentially used as calibrator for implementing metrological traceability. However, commutability is an indication for the degree of harmonization achievable but not necessarily for the absence of bias between different calibration hierarchies, an aspect that should be tested by appropriate extent-of-equivalence studies.


Corresponding author: Prof. Mauro Panteghini, Research Centre for Metrological Traceability in Laboratory Medicine (CIRME), University of Milan, Via GB Grassi 74, 20157 Milan, Italy; and Joint Committee for Traceability in Laboratory Medicine (JCTLM) Task Force on Reference Measurement System Implementation, Milan, Italy, E-mail:

References

1. Plebani, M. Why C-reactive protein is one of the most requested tests in clinical laboratories? Clin Chem Lab Med 2023;61:1540–5. https://doi.org/10.1515/cclm-2023-0086.Search in Google Scholar PubMed

2. Lippi, G, Panteghini, M, Bernardini, S, Bonfanti, L, Carraro, P, Casagranda, I, et al.. Laboratory testing in the emergency department: an Italian society of clinical biochemistry and clinical molecular biology (SIBioC) and academy of emergency medicine and care (AcEMC) consensus report. Clin Chem Lab Med 2018;56:1655–9. https://doi.org/10.1515/cclm-2017-0077.Search in Google Scholar PubMed

3. Aloisio, E, Chibireva, M, Serafini, L, Pasqualetti, S, Falvella, FS, Dolci, A, et al.. A comprehensive appraisal of laboratory biochemistry tests as major predictors of COVID-19 severity. Arch Pathol Lab Med 2020;144:1457–64. https://doi.org/10.5858/arpa.2020-0389-sa.Search in Google Scholar PubMed

4. Aloisio, E, Colombo, G, Dolci, A, Panteghini, M. C-reactive protein and clinical outcome in COVID-19 patients: the importance of harmonized measurements. Clin Chem Lab Med 2023;61:1546–51. https://doi.org/10.1515/cclm-2023-0276.Search in Google Scholar PubMed

5. Borrillo, F, Panteghini, M. Current performance of C-reactive protein determination and derivation of quality specifications for its measurement uncertainty. Clin Chem Lab Med 2023;61:1552–7. https://doi.org/10.1515/cclm-2023-0069.Search in Google Scholar PubMed

6. Merlini, G, Blirup-Jensen, S, Johnson, AM, Sheldon, J, Zegers, I. IFCC Committee on Plasma Proteins (C-PP). Standardizing plasma protein measurements worldwide: a challenging enterprise. Clin Chem Lab Med 2010;48:1567–75. https://doi.org/10.1515/cclm.2010.314.Search in Google Scholar PubMed

7. Wojtalewicz, N, Schellenberg, I, Hunfeld, KP. Evaluation of INSTAND e.V.’s external quality assessment for C-reactive protein and procalcitonin. PLoS One 2019;14:e0221426. https://doi.org/10.1371/journal.pone.0221426.Search in Google Scholar PubMed PubMed Central

8. Secchiero, S, Sciacovelli, L, Plebani, M. Performance evaluation of 14 specific proteins measurement checked by an external quality assessment scheme. Clin Chim Acta 2020;502:73–83. https://doi.org/10.1016/j.cca.2019.11.024.Search in Google Scholar PubMed

9. Braga, F, Panteghini, M. Derivation of performance specifications for uncertainty of serum C-reactive protein measurement according to the Milan model 3 (state of the art). Clin Chem Lab Med 2020;58:e263–5. https://doi.org/10.1515/cclm-2020-0532.Search in Google Scholar PubMed

10. Panteghini, M, Braga, F, Camara, JE, Delatour, V, Van Uytfanghe, K, Vesper, HW, et al.. JCTLM task force on reference measurement system implementation. Optimizing available tools for achieving result standardization: value added by Joint Committee on Traceability in Laboratory Medicine (JCTLM). Clin Chem 2021;67:1590–605. https://doi.org/10.1093/clinchem/hvab178.Search in Google Scholar PubMed

11. JCGM 200. International vocabulary of metrology – Basic and general concepts and associated terms (VIM), 3rd ed.; 2012.Search in Google Scholar

12. Panteghini, M. Traceability, reference systems and result comparability. Clin Biochem Rev 2007;28:97–104.Search in Google Scholar

13. Rzychon, M, Zegers, I, Schimmel, H. Analysis of the physicochemical state of C-reactive protein in different preparations including 2 certified reference materials. Clin Chem 2010;56:1475–82. https://doi.org/10.1373/clinchem.2010.147124.Search in Google Scholar PubMed

14. Schimmel, H, Zegers, I, Emons, H. Standardization of protein biomarker measurements: is it feasible? Scand J Clin Lab Invest Suppl 2010;242:27–33. https://doi.org/10.3109/00365513.2010.493362.Search in Google Scholar PubMed

15. Miller, WG, Greenberg, N, Budd, J, Delatour, V. IFCC working group on commutability in metrological traceability. The evolving role of commutability in metrological traceability. Clin Chim Acta 2021;514:84–9.10.1016/j.cca.2020.12.021Search in Google Scholar PubMed

16. Kilpatrick, EL, Bunk, DM. Reference measurement procedure development for C-reactive protein in human serum. Anal Chem 2009;81:8610–6. https://doi.org/10.1021/ac901597h.Search in Google Scholar PubMed

17. Boulo, S, Hanisch, K, Bidlingmaier, M, Arsene, CG, Panteghini, M, Auclair, G, et al.. Gaps in the traceability chain of human growth hormone measurements. Clin Chem 2013;59:1074–82. https://doi.org/10.1373/clinchem.2012.199489.Search in Google Scholar PubMed

18. Nilsson, G, Budd, JR, Greenberg, N, Delatour, V, Rej, R, Panteghini, M, et al.. IFCC working group recommendations for assessing commutability. Part 2: using the difference in bias between a reference material and clinical samples. Clin Chem 2018;64:455–64. https://doi.org/10.1373/clinchem.2017.277541.Search in Google Scholar PubMed PubMed Central

19. Budd, JR, Weykamp, C, Rej, R, MacKenzie, F, Ceriotti, F, Greenberg, N, et al.. IFCC working group recommendations for assessing commutability. Part 3: using the calibration effectiveness of a reference material. Clin Chem 2018;64:465–74. https://doi.org/10.1373/clinchem.2017.277558.Search in Google Scholar PubMed

20. Miller, WG, Budd, J, Greenberg, N, Weykamp, C, Althaus, H, Schimmel, H, et al.. IFCC working group recommendations for correction of bias caused by noncommutability of a certified reference material used in the calibration hierarchy of an end-user measurement procedure. Clin Chem 2020;66:769–78. https://doi.org/10.1093/clinchem/hvaa048.Search in Google Scholar PubMed PubMed Central

21. Braga, F, Panteghini, M. Commutability of reference and control materials: an essential factor for assuring the quality of measurements in Laboratory Medicine. Clin Chem Lab Med 2019;57:967–73. https://doi.org/10.1515/cclm-2019-0154.Search in Google Scholar PubMed

22. Braga, F, Infusino, I, Panteghini, M. Performance criteria for combined uncertainty budget in the implementation of metrological traceability. Clin Chem Lab Med 2015;53:905–12. https://doi.org/10.1515/cclm-2014-1240.Search in Google Scholar PubMed

23. Infusino, I, Panteghini, M. Measurement uncertainty: friend or foe? Clin Biochem 2018;57:3–6. https://doi.org/10.1016/j.clinbiochem.2018.01.025.Search in Google Scholar PubMed

24. Miller, WG, Panteghini, M, Wielgosz, R. Implementing metrological traceability of C-reactive protein measurements: consensus summary from the Joint committee for traceability in laboratory medicine workshop. Clin Chem Lab Med 2023;61:1558–60.10.1515/cclm-2023-0498Search in Google Scholar PubMed

Published Online: 2023-06-05
Published in Print: 2023-08-28

© 2023 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Downloaded on 28.4.2024 from https://www.degruyter.com/document/doi/10.1515/cclm-2023-0558/html
Scroll to top button