Skip to content
Licensed Unlicensed Requires Authentication Published by De Gruyter August 24, 2010

Accuracy of the Precision® point-of-care ketone test examined by liquid chromatography tandem-mass spectrometry (LC-MS/MS) in the same fingerstick sample

  • Marcel J.W. Janssen , Ben H.E. Hendrickx , Carin D. Habets-van der Poel , Joop P.W. van den Bergh , Anton A.M. Haagen and Jaap A. Bakker

Abstract

Background: The Precision® (Abbott Diabetes Care) point-of-care biosensor test strips are widely used by patients with diabetes and clinical laboratories for measurement of plasma β-hydroxybutyrate (β-HB) concentrations in capillary blood samples obtained by fingerstick. In the literature, this procedure has been validated only against the enzymatic determination of β-HB in venous plasma, i.e., the method to which the Precision® has been calibrated.

Methods: In this study, the Precision® Xceed was compared to a methodologically different and superior procedure: determination of β-HB by liquid chromatography tandem-mass spectrometry (LC-MS/MS) in capillary blood spots. Blood spots were obtained from the same fingerstick sample from out of which Precision® measurements were performed. Linearity was tested by adding varying amounts of standard to an EDTA venous whole blood matrix.

Results: The Precision® was in good agreement with LC-MS/MS within the measuring range of 0.0–6.0 mmol/L (Passing and Bablok regression: slope=1.20 and no significant intercept, R=0.97, n=59). Surprisingly, the Precision® showed non-linearity and full saturation at concentrations above 6.0 mmol/L, which were confirmed by a standard addition experiment. Results obtained at the saturation level varied between 3.0 and 6.5 mmol/L.

Conclusions: The Precision® β-HB test strips demonstrate good comparison with LC-MS/MS. Inter-individual variation around the saturation level, however, is large. Therefore, we advise reporting readings above 3.0 as >3.0 mmol/L. The test is valid for use in the clinically relevant range of 0.0–3.0 mmol/L.

Clin Chem Lab Med 2010;48:1781–4.


Corresponding author: Marcel J.W. Janssen, Laboratory of Clinical Chemistry and Haematology, VieCuri Medical Center, P.O. Box 1926, 5900 BX Venlo, The Netherlands Phone: +31-77-3205201, Fax: +31-77-3205216,

Received: 2010-5-28
Accepted: 2010-6-25
Published Online: 2010-08-24
Published in Print: 2010-12-01

©2010 by Walter de Gruyter Berlin New York

Downloaded on 26.4.2024 from https://www.degruyter.com/document/doi/10.1515/CCLM.2010.351/html
Scroll to top button