Automatic classification of tissues on pelvic MRI based on relaxation times and support vector machine

  • Jorge Arturo Zavala Bojorquez,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliation Le2i, Université Bourgogne Franche-Comte, Dijon, France

  • Pierre-Marc Jodoin,

    Roles Conceptualization, Formal analysis, Methodology, Writing – review & editing

    Affiliation VITAL, Université de Sherbrooke, Sherbrooke, Canada

  • Stéphanie Bricq,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliation Le2i, Université Bourgogne Franche-Comte, Dijon, France

  • Paul Michael Walker,

    Roles Conceptualization, Writing – review & editing

    Affiliations Le2i, Université Bourgogne Franche-Comte, Dijon, France, Centre Hospitalier Universitaire, Dijon, France

  • François Brunotte,

    Roles Conceptualization, Writing – review & editing

    Affiliations Le2i, Université Bourgogne Franche-Comte, Dijon, France, Centre Hospitalier Universitaire, Dijon, France

  • Alain Lalande

    Roles Conceptualization, Supervision, Writing – review & editing

    alain.lalande@u-bourgogne.fr

    Affiliations Le2i, Université Bourgogne Franche-Comte, Dijon, France, Centre Hospitalier Universitaire, Dijon, France

Abstract

Tissue segmentation and classification in MRI is a challenging task due to a lack of signal intensity standardization. MRI signal is dependent on the acquisition protocol, the coil profile, the scanner type, etc. While we can compute quantitative physical tissue properties independent of the hardware and the sequence parameters, it is still difficult to leverage these physical properties to segment and classify pelvic tissues. The proposed method integrates quantitative MRI values (T1 and T2 relaxation times and pure synthetic weighted images) and machine learning (Support Vector Machine (SVM)) to segment and classify tissues in the pelvic region, i.e.: fat, muscle, prostate, bone marrow, bladder, and air. Twenty-two men with a mean age of 30±14 years were included in this prospective study. The images were acquired with a 3 Tesla MRI scanner. An inversion recovery-prepared turbo spin echo sequence was used to obtain T1-weighted images at different inversion times with a TR of 14000 ms. A 32-echo spin echo sequence was used to obtain the T2-weighted images at different echo times with a TR of 5000 ms. T1 and T2 relaxation times, synthetic T1- and T2-weighted images and anatomical probabilistic maps were calculated and used as input features of a SVM for segmenting and classifying tissues within the pelvic region. The mean SVM classification accuracy across subjects was calculated for the different tissues: prostate (94.2%), fat (96.9%), muscle (95.8%), bone marrow (91%) and bladder (82.1%) indicating an excellent classification performance. However, the segmentation and classification for air (within the rectum) may not always be successful (mean SVM accuracy 47.5%) due to the lack of air data in the training and testing sets. Our findings suggest that SVM can reliably segment and classify tissues in the pelvic region.

Introduction

Tissue segmentation and classification is an important topic in Magnetic Resonance Imaging (MRI). It helps to study anatomical structures, to develop surgical planning, plan radiation therapy and perform quantitative analyses [1,2].

Most of the available segmentation and classification methods operate on conventional contrast-weighted MR images (T1-, T2-, and proton density-weighted images). However, the segmentation and classification of tissues from contrast-weighted MR images is fundamentally complex. This is due to the specific sequences used which impact the grayscale values of the resulting MR images. So, as opposed to CT-scan images, one cannot know a priori the grayscale values of a given tissue without acute knowledge of the exact acquisition protocol. As such, MR images are highly sensitive to acquisition parameters [3] such as the inversion time (TI), the echo time (TE), the repetition time (TR), the flip angle and the voxel size, as well as the sequences used to acquire the images [4]. MR images are also affected by various signal artifacts like partial volume effects, transverse coherences or spoiling, B0- and B1-inhomogeneities [5], as well as hardware characteristics (e.g. intra- and inter-scanner variations, magnetic field intensity, coil sensitivity [6]). Furthermore, the tissue properties themselves have an impact on the acquired MR images (e.g. relaxation times, proton density and physiological parameters). The interaction of these factors determines the image voxel intensity and, therefore, makes automatic tissue segmentation and classification challenging, especially for unsupervised methods [7].

Furthermore, the complexity of image formation leads to an unfortunate loss of physical meaning of the absolute voxel intensity [8]. This is because MR images are acquired in arbitrary units that are not comparable between studies [9]. Hence, segmentation and classification are performed by comparing contrast differences among arbitrary voxels’ intensities. As the intensity scaling is also arbitrary, MRI segmentation and classification methods require the implementation of complicated filtering and normalization techniques [10,11], which add additional steps to the segmentation and classification process. Moreover, these methods often have a strong heuristic flavor and do not generalize well to images that violate the hypotheses they are built upon.

To avoid these difficulties, some automatic methods work directly on parametric maps (quantitative MRI) or on generated synthetic images based on intrinsic properties of tissues such as T1 and T2 relaxation times. This is possible because in the T1 and T2 parametric maps, the voxel intensities represent individual tissue physical properties. These properties are scanner and pulse sequence independent, and have numerical meaning rather than representing signal intensity on an arbitrary scale [3,12]. i.e. quantitative MRI has the benefit of being independent of MRI settings and hardware imperfections [8]. A further advantage of the integration of imaging physics into the classification process is that it allows the optimization of the MR pulse parameters in a way that reduces the probability of misclassification [3].

Various segmentation and classification methods using the tissues’ physical properties have been proposed. For example, Chen et al. [13] performed segmentation of white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF) using a region-based active contour method applied on T1 relaxation maps. Traynor et al. [14] utilized both T1 and T2 relaxation times to segment the thalamus in 16 regions using a genetic algorithm. Iglesias et al. [15] implemented an atlas-based method using generated synthetic images from the T1 relaxation times to segment the hippocampus. Knowing that the intensities of T1-weighted, T2-weighted or proton density images are the product of different physical variables, Iglesias et al. [15] computed synthetic T1-weighted images using the tissues’ physical properties to achieve image homogenization and perform atlas matching to segment the hippocampus. As mentioned by the authors, the drawback of their method is that the T1 value of at least one tissue must be known a priori. This is a major weakness considering that the true relaxation times of tissues are still not measured with sufficient accuracy and precision [5].

To the best of our knowledge, most previous methods have been designed for brain and thorax imaging, and little effort has been devoted to pelvic imaging, although prostate cancer is the most commonly diagnosed cancer for men worldwide [16]. For prostate cancer, MRI is currently the imaging modality of choice for early detection and classification of tumorous areas. An accurate segmentation and classification of prostate cancer is required to guide radiotherapy or surgery, to perform volume estimation and to track disease progression [17]. However, the segmentation and classification of prostate is particularly challenging in MRI because signal intensity is not standardized, and image appearance is for a large part determined by acquisition protocol, coil profile, scanner type and field strength [18]. These are major obstacles in the development of prostate segmentation and classification methods.

In this paper, we use a machine learning method to learn the relation between the T1/ T2 intrinsic parameters for six different tissues namely prostate, fat, muscle, bladder, bone marrow, and air (within the rectum). As such, a Support Vector Machine (SVM) is used to segment and classify the relevant structures within the pelvis. The MRI feature descriptors are the T1 relaxation time, the synthetic generated pure T1-weighted images (with different TI), the T2 relaxation time and synthetic generated pure T2-weighted images (with different TE), and a probabilistic shape prior which encodes the location of anatomical regions.

Materials and methods

To carry out the segmentation and classification of the different tissues present in an image, our approach exploits the information obtained from different standard MR sequences. The sequences are used to obtain T1-weighted and T2-weighted images in order to compute the T1 and T2 relaxation maps. These maps serve as the input feature vectors of the SVM used for tissue segmentation-classification. Furthermore, the relaxation times are utilized to generate pure synthetic T1- and T2-weighted images that are also used as input features. A shape probabilistic map is also used as a feature descriptor to include anatomical information in the segmentation-classification method. The segmented image contains the labels and boundaries of the various tissues within. The different steps of our method are illustrated in Fig 1 on one slice of a volunteer.

thumbnail
Fig 1. Pipeline of the segmentation and classification process using the SVM classifier.

After obtaining the T1- and T2-weighted images (a-b) from their corresponding sequences, the body region is found using a threshold method on the T2-weighted images (c). Then, the fit procedure is applied to obtain the relaxation maps and other different variables from the received signal model (d-e). These variables are then used to generate the synthetic T1- and synthetic T2-weighted images (f-g). Later, an independent process using the training data is used to generate tissues probabilistic maps, i.e. the probability of a voxel belonging to different tissues (h). At this point, all feature descriptors are normalized (i) to be used by the multi-class SVM classifier (j), which predicts the different classes (k).

https://doi.org/10.1371/journal.pone.0211944.g001

Image acquisition

The images were acquired with a 3T Trio TIM clinical scanner (Siemens Medical Solutions, Germany) using a phased-array body coil. Twenty-two men gave their informed consent to participate in the study, which had been approved by the local institutional review board and the local ethic committee (the IRB of the University Hospital of Dijon (France) approved the research). The French law and regulation was followed, and the management of the patient is endorsed by the University Hospital of Dijon, that assume having participant consent by default when the participant is supervised by our institution. The study procedures were in accordance with the ethical standards of the committees with responsibility for human experimentation and with the Helsinki Declaration of 1975, as revised in 2008. Certain pathologies may alter tissue relaxation values and so subjects with systemic pathologies such as hematological diseases (anemias, myelodysplasias) and muscular disorders (atrophy, Duchenne and Becker diseases, etc.) should be excluded from the study. The group was comprised of 22 healthy volunteers. The age range was from 22 to 63 years old (mean 30 ± 14). Four to five slices were acquired depending on the anatomy of the subject. MRI slices were acquired with the same FOV, resolution and position at the level of the pelvis for both T1- and T2-weighted images. Axial pelvic orientations were chosen as it is commonly used in prostate MRI.

An inversion recovery-prepared turbo spin echo sequence was used to obtain the T1-weighted images. These images were acquired at eight different inversion times (i.e. TIs were from 50 ms to 10000 ms). The TIs were chosen to assure maximum contrast among tissues and to cover the whole span of T1 relaxation times of tissues present in the slice section. Other nominal sequence parameters are repetition time (TR) of 14000 ms, echo time (TE) of 7.4, echo train length (ETL) of 11, field of view (FOV) from 180 x 280 mm to 220 x 380 mm, slice thickness of 5 mm, number of excitations (NEX) = 1 and matrix size of 168 x 256. The average acquisition time per TI was 3 min 33 s.

A 32-echo spin echo sequence was used to obtain the T2-weighted images. The T2-weighted images were acquired at different TE from 8.8 to 281.6 ms in 8.8 ms steps with a TR of 5000 ms. The same nominal values for FOV, NEX and matrix size were defined as for the T1 sequence of every volunteer. The acquisition time was 14 minutes for the T2-weighted spin echo sequence.

Segmentation and classification

In our method, MR images are segmented by classifying each pixel with a multi-class linear SVM classifier [19]. As mentioned in the introduction, the grayscale values of a given tissue can vary significantly from one acquisition to another. As such, raw input data suffer from a large variance making it hardly separable. While non-linear classifiers such as random forests [20], Adaboost [21], neural networks [22] or Kernel-SVM [23] could be used, the large variance of the data can lead to overfitting and poor generalization, especially when the training set accounts for a limited number of subjects. In this paper, we use features with a much lower variance while making classes linearly separable. While linear models are less prone to overfitting (and thus ensure a lower generalization error) they are also much faster, easier to train and have fewer hyper-parameters. The features used are the T1 relaxation time, the synthetic T1-weighted images, the T2 relaxation time, the synthetic T2-weighted images, and a probabilistic shape prior, which are explained in the following section.

The labels of the training data were defined by manually selecting ROIs corresponding to the different tissues present in the imaged anatomical areas, i.e. prostate, fat, muscle, bone marrow, bladder and air. ROIs were outlined by a radiologist with 20 years of experience based on a T1- and/or T2-weighted image that best depicted the tissues. Selected ROIs and their corresponding labels are illustrated in Fig 2.

thumbnail
Fig 2. Selected ROIs on a T1-weighted image of the pelvic area.

1) air, 2) fat, 3) muscle, 4) prostate, 5) bone marrow, and 6) bladder.

https://doi.org/10.1371/journal.pone.0211944.g002

T1 relaxation time

The longitudinal relaxation time, T1 feature, is calculated from a magnitude monoexponential fit to the signal recovery data for each voxel using the bisquare weights nonlinear least squares fitting method [24]. This method was chosen for its capacity to eliminate the influence of noise by reducing the weight of outliers during the fit. A three-parameter model was used to describe the received signal at different inversion times (S(TI)): (1) where T1 is the relaxation time, S0 the equilibrium magnetization and α the inversion efficiency. These parameters represent the fitting variables of the model. S(TI) is the measured signal value for a given value of TI. The model describes a magnitude signal (only positive values) because the phase component of the received signal is not used during the reconstruction (i.e. the acquired MR images were magnitude images with no phase component).

The fitting interval restrictions on the model were set to the values defined in [24] (S0 [1, 65535], α[0, 2] and T1[0, 3000]). The interval on S0 considers any potential signal intensity expected in a MR DICOM image, the α intervals span all the potential inaccuracies of the scanner inversion pulse from 0° to 180°, and the T1 intervals span all the expected T1 relaxation times of tissues present in the image.

To reduce the fitting time, the background is automatically extracted and the fit is performed only with voxels that belong to the body region. This is achieved by creating a body mask from the T2-weighted image with the shortest TE (8.8 ms) with a threshold of 4% of the highest voxel intensity. At this TE, the tissues have almost full transversal magnetization and their intensity is maximum (Stissue≈25*Sbackground or noise), which allows an easy separation between the body region and the background. The body region is defined by performing connected component analysis, i.e. two regions are identified and the region with the largest number of voxels is labeled as the body region (Sizebody region≈ 4 Sizebackground).

The accuracy of the fit was measured by the Rsquare metric. Only the Rsquare values of voxels describing ≥85% of the data were considered as good fits. The fitting stops when a maximum number of iterations is reached or a good fit is found. Fig 1D shows the process of the T1 relaxation time calculation.

Since the fit was performed in a voxel by voxel fashion, the obtained relaxation times formed a T1 relaxation map which has the appearance of an image (structures can be identified), conceptually different from a T1-weighted image, in that individual pixel values now have a numerical meaning (i.e. T1 values in ms at each location in the anatomical area), rather than representing signal intensity on an arbitrary scale.

Once the parameters from the fit are calculated, the inverse process is performed to obtain a synthetic T1-weighted image.

Synthetic T1-weighted images

Synthetic MR images can be generated with arbitrary contrast weighting, if the appropriate MR signal model is used [25]. In this case, the synthetic T1-weighted images are generated using the following model: (2)

This model describes a signal with a perfect inversion pulse of 180° (i.e. α = 2) and depends on only three variables: S0, T1 and TI. An example of the signal modeling is shown in Fig 3A.

thumbnail
Fig 3. Model of a T1 and T2 relaxation signal.

The T1 relaxation curve is simulated with an ideal inversion pulse (α = 2 = 180°) and magnetic field homogenization (S0 = 4095) (a). The T2 signal is also simulated with magnetic homogenization (S0 = 4095). Reported T1 and T2 values at 3 Tesla from the literature were used to generate the signal: fat (385 ms, 121ms), bone marrow (585 ms, 127 ms), muscle (1295 ms, 40 ms), prostate (1700 ms, 74 ms), and bladder (3000 ms, 50 ms) [24,33,34], respectively. TIs and TEs were chosen to cover the complete span of relaxation times of the previous tissues.

https://doi.org/10.1371/journal.pone.0211944.g003

To generate the synthetic T1-weighted images, S0 is assigned a constant maximum intensity value commonly used by MR scanners [26] (i.e. 4095). This constant value simulates a perfect homogeneous B0 for every voxel. The T1 relaxation values are set from the values found from the T1 fit calculation. The TIs were defined to span all the relaxation times of tissues present in the image, and to sample the signal intensities where they could produce the maximum contrast among tissues (TIs = 25 ms, 200 ms, 400 ms, 700 ms, 900 ms, 1000 ms, 1100 ms, 1200 ms, 1500 ms, 2000 ms, 3000 ms, 4000 ms, 5000 ms). Examples of generated synthetic T1-weighted images are provided in Fig 4.

thumbnail
Fig 4. Process by which the synthetic T1-weighted image is computed.

The relaxation times of tissues and the selected inversion times (TI) are used to generate the synthetic T1-weighted images using the model describing a received signal from a perfect inversion pulse.

https://doi.org/10.1371/journal.pone.0211944.g004

The idea behind the chosen TIs is to sample the relaxation signal of tissues at precise moments where signal differences are maximal among tissues (e.g. at TI = 1200 ms, Fig 4), and to avoid sampling at those TIs where the signal of tissues are not easy to differentiate. However, it is difficult to define the perfect TIs a priori because the precise T1 relaxation times of tissues are still not measured with sufficient accuracy and precision [5]. A practical workaround is to use many TIs to cover a large span of T1 relaxation times (and by focusing on TI values that maximally differentiates tissues). This is because short TIs sample points favor the short T1 relaxation times, and long TIs sample times favor the long T1 relaxation times [8].

The synthetic T1-weighted images have important advantages with respect to the T1-weighted images: 1) The proton density influence and the effects of B0 and B1 inhomogeneities are removed; 2) The evolution of the signal has now a phase and not only a magnitude (i.e. it goes from negative to positive values). This is important because in magnitude images, tissues with different T1 relaxation times could be mapped with the same signal intensity (e.g. a signal intensity of 50 is equal to |-50| in magnitude images); 3) There is image homogenization through all the synthetic T1-weighted images (i.e. all intensities are in the range of [-4095, 4095]); 4) The dynamic range doubles with respect to the acquired T1-weighted images (i.e. from [0, 4095] to [-4095, 4095]); and 5) The synthetic images are independent of the scanner hardware and acquisition protocol.

T2 relaxation time and synthetic T2-weighted images

For the T2 relaxation time, a mono-exponential fit using a bisquare weights nonlinear least squares fitting method was applied to the data [24]. The following three-parameter model was used to describe the received signal at different echo times (S(TE)): (3) where T2, S0 and the y-offset (noise floor C) are the fitting parameters [27].

The fit was calculated only for voxels belonging to the region of the body defined during the T1 processing. The fits of voxels with Rsquare ≥ 85% were kept. The model interval restrictions were defined to reduce the fitting time (S0 [1, 65535], T2 [0, 3500] and C [0, 5000]). Fig 1E shows the process of the T2 relaxation time calculation.

Once the parameters from the fit are calculated, the inverse process is performed to obtain a synthetic T2-weighted image. This inverse process is similar to the one used to generate the synthetic T1-weighted image. The synthetic T2-weighted images are generated using the model described in Eq 3. S0 is kept constant (4095, a common maximum intensity value found in MR images), without y-offset (C = 0), and the T2 relaxation values are set from the values found from the T2 fit calculation. An example of the signal modeling is shown in Fig 3B.

As for the synthetic T1-weighted image, it is difficult to define the perfect TEs a priori, as many TEs were used to cover a large span of T2 relaxation times. This is because short TEs sample points favor the short T2 relaxation times, and long TEs sample points favor the long T2 relaxation times [8].

The synthetic T2-weighted images have important advantages with respect to the T2-weighted images: 1) The effects of B0 inhomogeneities, B1 inhomogeneities, proton density influence and noise (C) are removed; and 2) There is signal intensity homogenization (i.e. all the synthetic images are generated using the same signal scale: 0 to 4095).

Shape prior

In order to enforce anatomically plausible results, we created a probabilistic shape prior P(C|x) which encodes the posterior probability of having a tissue of a certain class C given a voxel position x. The process for computing this shape prior consists in the following steps:

  1. We localize the body center of mass for every volunteer in the training set.
  2. Every voxel in the ROIs drawn by the expert is translated into an accumulator grid using the body center of mass as the reference point. Each bin in the accumulator grid records the number of tissue instances for the different tissues. The accumulator grid has a size of 80 x 80 cm2 and covers all possible variations. The bin dimensions of the grid are set to 5 x 5 mm2, a trade-off between the partial volume effects (bigger bins increase the possibility of this effect) and the computation time (smaller bins increase the computation time and the amount of memory to handle the probabilistic maps).
  3. Finally, the probability that a certain voxel belongs to a particular tissue is computed for every bin by averaging out the ground truth labels of each training image.

The process is illustrated in Fig 5. The posterior probability values from the shape prior will later on be used as feature descriptors.

thumbnail
Fig 5. Probabilistic shape prior creation.

Starting from training images, the center of mass is localized (a). Then, every voxel in the ROIs is translated into an accumulator grid (b) and the instances of every different tissue in the different bins are counted to create the probabilistic map (c). From the probabilistic map, six-features descriptors are formed: air, fat, muscle, bone marrow, prostate and bladder.

https://doi.org/10.1371/journal.pone.0211944.g005

Feature normalization

At this point, we have 36 features: one T1 relaxation time, one T2 relaxation time, 12 synthetic T1-weighted images at different TIs, 16 synthetic T2-weighted images at different TEs, and six shape prior probabilities of the different tissue classes (fat, muscle, prostate, bone marrow, bladder, and air).

Each of the 36 features is normalized [28] so they have zero mean and unit variance: (4) where fold is the old feature value, fnorm is the normalized value, and μ and σ are the mean and standard deviation of the original feature range. The main advantage of normalization is to avoid features spanning across a large numeric range to dominate those with a smaller numeric range.

Classification

The tissue classification is performed using the multi-class linear SVM implementation from the libLinear software package [19]. A One-vs-All method was used to solve the multi-class classification problem. It consists in developing for each class a binary classifier that separates that class from the rest of the data. A linear SVM kernel was chosen due to the lack of prior data suggesting the use of a non-linear SVM kernel (with our features), and due to the lower vulnerability for overfitting compared with non-linear kernels [29]. The method then combines the classifiers for multi-class inference [30].

Evaluation

To evaluate the performance of the multi-class SVM classifier, a leave-one-out cross-validation was implemented. The overall database contains images from 22 subjects. We repeatedly trained our method on data from 21 subjects and tested on the remaining one. The reported cross validation accuracy is the percentage of data correctly classified pixel-by-pixel.

By its very nature, the dataset is highly imbalanced with far more pixels from the fat, muscle and bone marrow classes than from the prostate, bladder and air classes. To avoid biasing the classification toward the larger classes, the number of voxels from the fat, muscle and bone marrow classes were randomly down-sampled to match the number of voxels in the prostate class. Since the bladder and air ROIs are smaller than the prostate, we kept every pixel from those classes.

The mean accuracy was calculated for different combination of features to find the one with the higher prediction accuracy. Note that the classification process predicts a class label for each pixel, but accuracy is reported based on the ROIs manually outlined by the radiologist as shown in Fig 2.

Results

The segmentation-classification accuracy for the different tissues with different feature combinations is shown in Table 1. For most of the SVM models, the prostate classification accuracy was excellent (>80%), except for some models using a single feature descriptor (i.e. T1 relaxation time (≈ 72%), T2 relaxation time (0%) or synthetic T1-weighted images (≈ 75%)). The classification accuracy of fat and muscle was also excellent (>90%), except for the model using only the shape prior as feature descriptor (< 80%). The classification accuracy of bone-marrow only achieved good results in SVM models including the shape prior feature (> 85%), otherwise the SVM classification failed (< 25%). The bladder classification accuracy was excellent (> 80%) for those SVM models including most of the features (i.e. synthetic T1- and T2-weighted and shape prior; and all the features). Please note that the reported accuracy of the bladder was computed on the slices in which it is visible. The classification of air within the body was not always satisfactory (<48%) in all the SVM models, mainly due to the small size of this region and the difficulty to model air signal (classified as noise). The use of synthetic T1- and T2-weighted features resulted in an overall better segmentation-classification accuracy than just using T1 and T2 relaxation times as features descriptors for the SVM classifier.

Fig 6 presents an example of the computed T1 and T2 relaxation maps and the corresponding segmentation-classification results using all features. Visual inspection of the T1 map indicates a clear difference among the relaxation times of prostate (≈1700 ms), muscle (≈ 1500 ms), fat (≈ 380 ms) and bladder (≈ 2950 ms) so tissues from the various classes could easily be linearly separated. However, in the T2 map, the differences among the relaxation times of prostate (≈ 70 ms)—fat (≈ 90 ms)—bladder (≈ 60 ms) and fat—bone marrow (≈ 95 ms) are not visually perceptible. Nevertheless, the segmentation-classification accuracy for prostate, muscle, fat and bone marrow was excellent (≥ 90%). The bladder and air segmentation-classification accuracy was less accurate with around 10 and 45%, respectively.

thumbnail
Fig 6. Relaxation maps and segmentation map of volunteer 1.

Both T1 and T2 relaxations maps show homogeneous time values within the area of each tissue: prostate, muscle, fat, bladder and bone marrow. In the T1 and T2 maps, fat and bone marrow are not visually separable. The same happens in the T2 map for prostate and fat, muscle and bladder. However, the SVM classifier accuracy is rather excellent for prostate = 99.6%, fat = 92.5%, muscle = 99.3%, and bone marrow = 96.7%. Accuracy is lower for bladder (10.4%) and air (45%) mostly because of the small size of these regions.

https://doi.org/10.1371/journal.pone.0211944.g006

Another example of the T1 and T2 relaxation maps and segmentation-classification results is shown in Fig 7 for a volunteer affected with benign prostatic hyperplasia. The T1 relaxation map has considerably spurious relaxation times (≥2000 ms) in the muscle areas, and as expected fat and bone marrow are not easily differentiated. In the T2 relaxation map, only two areas are observable: 1) values below 70 ms (muscle) and 2) values above 71 ms (prostate, fat, bone marrow and feces). However, despite these difficulties the segmentation-classification accuracy for prostate (89.1%), muscle (98.8%), fat (100%) and bone marrow (89.7%) was high.

thumbnail
Fig 7. Relaxation maps and classification results of volunteer 3.

Even though this volunteer is affected by noncancerous prostatic hyperplasia, the T1 map presents a rather homogeneous range of values in the prostate area (1600–2000 ms). The T1 map also presents a considerable amount of misleading values (≥2000 ms) in the muscle areas. On the T2 map, the muscle areas are well defined (<70ms) and clearly differentiable from the rest of the tissues (prostate, fat, and bone marrow) which seem to merge in a single region. In any case, the SVM accuracy is excellent: prostate = 89.1%, muscle = 98.8%, fat = 100%, and bone marrow = 89.7%. Bladder and air are not present. All features were used to perform the classification.

https://doi.org/10.1371/journal.pone.0211944.g007

A particular example of the T1 and T2 relaxation maps and classification-segmentation results is shown in Fig 8. For this volunteer, the prostate does not present a homogeneous set of T1 relaxation times. Its values overlap those of muscle and bladder (1300 ms to 3000 ms). In the T2 relaxation map, the prostate values overlap those of fat and bladder (≥ 100 ms). However, the bladder is easily differentiated in the T1 relaxation map, but not in the T2 relaxation map (i.e. its values overlap with those of prostate and fat). Nevertheless, the good accuracy of our method for prostate (89.1%), bladder (71.9%) and fat (100%) shows that a linear classifier works well, even for challenging images.

thumbnail
Fig 8. Relaxation maps and classification results of volunteer 20.

The T1 relaxation map shows heterogeneous values for the bladder and prostate areas, and homogeneous non-distinctive areas for fat and bone marrow which complicate the classification process. The T2 map only shows two discernible areas: 1) fat-bone marrow-prostate-bladder (≥100ms) and 2) muscle (<100 ms) adding additional complexity to the classification process. Nevertheless, the SVM classifies accurately labeled the different areas: prostate = 89.1%, fat = 100%, muscle = 98.8%, bone marrow = 89.7%, bladder = 71.9% and air was not present. All features were used to perform the classification.

https://doi.org/10.1371/journal.pone.0211944.g008

Discussion

In this paper, we presented a new methodology to classify tissues within the pelvic. The approach uses a multi-class linear SVM and MRI relaxation times features to segment prostate, fat, muscle, bone marrow, bladder and air. The overall accuracy of the method indicates that our approach is a viable option to automatically classify and segment these tissues and that the proposed features allow to linearly separate tissues. The lower accuracy for bladder and air compared with the other tissues is due to the lack of enough training and testing data for those two tissues (besides, they were not always visible in all imaged slices). Since the performances of SVM depend on the availability of training data, the lack of enough samples of bladder and air affects its performance. Also, since air and bladder cover a small area, an error of just a few millimeters drastically reduces accuracy, especially when T1- and T2-weighted images are not rigorously aligned. A possible solution for the lack of samples is data augmentation, i.e. bladder and air samples could be cloned with random distortions to increase the number of training samples. As for the misalignment, it can be due to rectal distension and rectal contraction between acquisitions. The implementation of a non-rigid registration technique could be a solution for this problem. The air classification accuracy also depends on the capacity of the SVM classifier to model the air signal. However, if the air signal is present, it is classified as noise (air does not give any signal when exposed to the radio frequency pulse of MR imaging). Therefore, the SVM classifier is trying to model a noisy signal to classify air. Since noise is formed randomly during each acquisition, its modeling is challenging. Thus, this could be another reason for the low performance of the air classification accuracy.

The accuracy of the SVM classifier also depends on the chosen feature descriptors. The feature descriptors are T1 relaxation time, synthetic T1-weighted images, T2 relaxation time, synthetic T2-weighted images and the tissue prior probabilistic maps. The accuracy of the results indicates that the chosen feature descriptors characterize well the variability of the tissue characteristics.

The T1 and T2 relaxation times are optimal feature descriptors because they represent physical properties that can be compared between cohorts. They are independent of scanner’s hardware and acquisition protocol, and they can be used to generate synthetic weighted images.

The synthetic T1- and T2-weighted images have been used as a "perfect" tissue descriptor because they are independent of hardware and artifacts affecting MR image reconstruction. They can generate a specific TI and TE to achieve reliable differentiation of all desired tissues. This is unlike combining several contrast weighting techniques which is usually insufficient to distinguish tissues [6]. When generated with the correct TIs or TEs sample points, synthetic weighted-images bring to light the subtle differences between tissues with close relaxation times, which help to increase classification accuracy. In fact, the inclusion of the synthetic weighted images as features increased the bladder classification accuracy to satisfactory levels (>80%), which was not possible using only the T1 and T2 relaxation times (<68%). Moreover, synthetic images can be generated at the same relaxation sample point for every subject. This forms a common framework which increases the intra- and inter-scanner reproducibility, which could certainly help to carry out multi-center studies. Multi-center studies are difficult because merging data across scanners is problematic (i.e. the way tissues’ properties are changed to signal intensities is scanner dependent rather than tissue dependent [3], and these signal intensities are arbitrary). However, the standardized nature of synthetic weighted images facilitates comparison across sites and time points.

The diagnostic value of synthetic T1- and T2-weighted images was not thoroughly evaluated although their quality when compared to conventional MR images has been shown sufficient [31,32]. Future research could further explore the true diagnostic value of synthetic images. Also, while this work focuses on relaxation time features, it could be easily extended to work with other tissues’ physical properties such as the apparent diffusion coefficient, which could be used to generate synthetic diffusion weighted images.

Tissue segmentation and classification with machine learning and relaxation times features has received increasing interest recently. The work that relates the most to our study is a decision tree model method of tissue segmentation and classification based on the T1 and T2 relaxation times and anatomical knowledge proposed in [24]. In their study, prostate, fat, muscle and bone marrow are segmented and classified. However, their study does not take advantage of the relaxation time capacity to generate synthetic T1- and T2-weighted images neither does it go further to localize other body regions (i.e. bladder and air). Moreover, their method uses a special procedure to localize bone marrow, while in our method bone marrow is localized directly by the SVM classifier.

The corner stone of the proposed method are the T1 and T2 relaxation times, for this reason an inversion recovery and spin echo sequences were used to compute them. These sequences are considered as the gold standard for MRI relaxometry. However, the trade-off is a long acquisition time required to acquire the T1- and T2-weighted images, which could be a difficulty in clinical practice. Hence, faster sequences could be adapted to compute the relaxation times in the pelvis (e.g. the Variable Flip Angle steady state spoiled gradient recalled echo (SPGR) imaging technique provides a series of high resolution T1-weighted images in a clinically feasible time [35]), and the method should work with these sequences thus it mostly depends on the relaxation times to perform the segmentation-classification, i.e. it is sequence independent.

Age related modifications in relaxation times are complex, but can be summarized as (i) a decrease in the degree of hydration of tissues with age and (ii) an increase in fat content in muscles [36,37]. A decrease in tissue hydration will lead to shorter T1 and especially T2 relaxation times in tissues, whereas an increase in tissue fat content will reduce T1, but increase T2. However, it could be infer that these variations are modest thus the SVM classification accuracy was excellent for muscle, fat, and bone marrow for the range of ages involved in the study (22–63 years old), since these tissues are the most affected by age related modifications in their relaxation times. Nevertheless, a bigger cohort with an ample range of ages is necessary to validate how well the SVM method can cape with these T1 and T2 age related variations.

Moreover, classification of prostate cancer tissue is not addressed in this study and an additional type of tissue (corresponding to prostate cancer) should be considered with a T1 slightly higher and a T2 considerably lower than the ones of normal prostate gland.

The main limitation of this study is the small number of volunteers. Twenty-two subjects are not enough to claim generalization, even though the fact that the SVM results came from a careful leave-one-out cross validation which reduces the bias of the reported accuracy values. Further validation on a much larger cohort that would include man of all ages, races and body masses is necessary.

Conclusion

We presented a reliable multi-linear SVM method to segment and classify the structures present in prostate MR examination: prostate, fat, muscle, bone marrow, bladder and air using the intrinsic T1 and T2 relaxation times of tissues. The SVM results provide solid information about these structures and could be potentially useful to measure the prostate volume for radiotherapy applications or for multi-modality registration (e.g. PET/MRI) to generate attenuation maps, and to perform quantitative analyses.

References

  1. 1. Kapur T, Grimson WEL, Wells WM, Kikinis R. Segmentation of brain tissue from magnetic resonance images. Med Image Anal. 1996;1(2):109–127. pmid:9873924
  2. 2. Yakar D, Debats OA, Bomers JG, Schouten MG, Lin EV, Fütterer JJ, et al. Predictive value of MRI in the localization, staging, volume estimation, assessment of aggressiveness, and guidance of radiotherapy and biopsies in prostate cancer. J MagnReson Imaging. 2012; 35(1): 20–31. pmid:22174000
  3. 3. Fischl B, Salat DH, van der Kouwe AJ, Makris N, Ségonne F, Quinn BT, et al. Sequence-independent segmentation of magnetic resonance images. Neuroimage. 2004;23:S69–S84. pmid:15501102
  4. 4. Bitar R, Leung G, Perng R, Tadros S, Moody AR, Sarrazin J, et al. MR pulse sequences: what every radiologist wants to know but is afraid to ask. Radiographics. 2006;26(2):513–537. pmid:16549614
  5. 5. Bojorquez JZ, Bricq S, Acquitter C, Brunotte F, Walker PM, Lalande A. What are normal relaxation times of tissues at 3 Tesla? MagnReson Imaging. 2016;35:68–80.
  6. 6. Schick F. Tissue segmentation: a crucial tool for quantitative MRI and visualization of anatomical structures. MagnReson Mater Phys Biol Med. 2016;29(2):89–93.
  7. 7. Wang L, Chitiboi T, Meine H, Günther M, Hahn HK. Principles and methods for automatic and semi-automatic tissue segmentation in MRI data. MagnReson Mater Phys Biol Med. 2016;29(2):95–110.
  8. 8. West J, Warntjes J, Lundberg P. Novel whole brain segmentation and volume estimation using quantitative MRI. Eur Radiol. 2012;22(5):998–1007. pmid:22113264
  9. 9. Shinohara RT, Sweeney EM, Goldsmith J, Shiee N, Mateen FJ, Calabresi PA, et al. Statistical normalization techniques for magnetic resonance imaging. NeuroImage Clin. 2014;6:9–19. pmid:25379412
  10. 10. Jäger F, Hornegger J. Nonrigid registration of joint histograms for intensity standardization in magnetic resonance imaging. IEEE Trans Med Imaging. 2009;28(1):137–50. pmid:19116196
  11. 11. Shinohara RT, Sweeney EM, Goldsmith J, Shiee N, Mateen FJ, Calabresi PA, et al. Statistical normalization techniques for magnetic resonance imaging. NeuroImage Clin. 2014;6:9–19. pmid:25379412
  12. 12. Tofts P. Quantitative MRI of the Brain: Measuring Changes Caused by Disease. 2005th ed. John Wiley & Sons; 2005. 675 p.
  13. 13. Chen P-F, Steen RG, Yezzi A, Krim H. Brain MRI T1-Map and T1-weighted image segmentation in a variational framework. In: IEEE International Conference on Acoustics, Speech and Signal Processing. Taipei, Taiwan; 2009. p. 417–420.
  14. 14. Traynor CR, Barker GJ, Crum WR, Williams SCR, Richardson MP. Segmentation of the thalamus in MRI based on T1 and T2. NeuroImage. 2011;56(3):939–950. pmid:21310246
  15. 15. Iglesias JE, Dinov I, Singh J, Tong G, Tu Z. Synthetic MRI signal standardization: application to multi-atlas analysis. In: Medical image computing and computer-assisted intervention: MICCAI. Beijing, China; 2010. p. 81–88.
  16. 16. Siegel RL, Miller KD, Jemal A. Cancer statistics, 2017. CA Cancer J Clin. 2017;67(1):7–30. pmid:28055103
  17. 17. Pasquier D, Lacornerie T, Vermandel M, Rousseau J, Lartigau E, Betrouni N. Automatic segmentation of pelvic structures from magnetic resonance images for prostate cancer radiotherapy. Int J Radiat Oncol Biol Phys. 2007;68(2):592–600. pmid:17498571
  18. 18. Litjens G, Toth R, van de Ven W, Hoeks C, Kerkstra S, van Ginneken B, et al. Evaluation of prostate segmentation algorithms for MRI: the PROMISE12 challenge. Med Image Anal. 2014;18(2):359–373. pmid:24418598
  19. 19. Fan R, Chang K, Hsieh C, Wang X, Lin C. LIBLINEAR: A Library for Large Linear Classification. J Mach Learn Res. 2008;9:1871–4.
  20. 20. Criminisi A, Shotton J, Konukoglu E. Decision Forests: A Unified Framework for Classification, Regression, Density Estimation, Manifold Learning and Semi-Supervised Learning. Found Trends® Comput Graph Vis. 2012;7(2–3):81–227.
  21. 21. Rätsch G, Onoda T, Müller K-R. Soft margins for AdaBoost. Mach Learn. 2001;42(3):287–320.
  22. 22. Demuth HB, Beale MH, De Jess O, Hagan MT. Neural network design. Martin Hagan; 2014.
  23. 23. Maji S, Berg AC, Malik J. Classification using intersection kernel support vector machines is efficient. In: Computer Vision and Pattern Recognition, CVPR. Anchorage, Alaska, USA: IEEE; 2008. p. 1–8.
  24. 24. Bojorquez JZ, Bricq S, Brunotte F, Walker PM, Lalande A. A novel alternative to classify tissues from T1 and T2 relaxation times for prostate MRI. MagnReson Mater Phys Biol Med. 2016;29(5):777–88. pmid:27160300
  25. 25. Callaghan MF, Mohammadi S, Weiskopf N. Synthetic quantitative MRI through relaxometry modelling. NMR Biomed. 2016;29(12):1729–1738. pmid:27753154
  26. 26. Kimpe T, Tuytschaever T. Increasing the number of gray shades in medical display systems—how much is enough? J Digit Imaging. 2007;20(4):422–432. pmid:17195900
  27. 27. Edden RAE, Smith SA, Barker PB. Longitudinal and Multi-Echo Transverse Relaxation Times of Normal Breast Tissue at 3 Tesla. J MagnReson Imaging. 2010;32(4):982–987.
  28. 28. Dunham MH. Data Mining: Introductory and Advanced Topics. 1st ed. Pearson; 2002.
  29. 29. Gatidis S, Scharpf M, Martirosian P, Bezrukov I, Küstner T, Hennenlotter J, et al. Combined unsupervised—supervised classification of multiparametric PET/MRI data: application to prostate cancer. NMR Biomed. 2015;28(7):914–922. pmid:26014883
  30. 30. Keerthi SS, Sundararajan S, Chang K-W, Hsieh C-J, Lin C-J. A Sequential Dual Method for Large Scale Multi-class Linear Svms. In: Proceedings of the 14th ACM SIGKDD International Conference on Knowledge Discovery and Data Mining. New York, NY, USA: ACM; 2008. p. 408–416.
  31. 31. Andersen AH, Zhang Z, Avison MJ, Gash DM. Automated segmentation of multispectral brain MR images. J Neurosci Methods. 2002;122(1):13–23. pmid:12535761
  32. 32. Nöth U, Hattingen E, Bähr O, Tichy J, Deichmann R. Improved visibility of brain tumors in synthetic MP-RAGE anatomies with pure T1 weighting. NMR Biomed. 2015;28(7):818–830. pmid:25960356
  33. 33. Pai A, Li X, Majumdar S. A comparative study at 3 Tesla of sequence dependence of T2 quantitation in the knee. MagnReson Imaging. 2008;26(9):1215–1220. pmid:18502073
  34. 34. De Bazelaire CMJ, Duhamel GD, Rofsky NM, Alsop DC. MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: preliminary results. Radiology. 2004;230(3):652–659. pmid:14990831
  35. 35. Ramos-Llordén G, Vegas-Sánchez-Ferrero G, Björk M, Vanhevel F, Parizel PM, Estépar RSJ, et al. NOVIFAST: A fast algorithm for accurate and precise VFA MRI T1 mapping. IEEE Transactions on Medical Imaging. 2018;1–1.
  36. 36. Cho S, Jones D, Reddick WE, Ogg RJ, Steen RG. Establishing norms for age-related changes in proton T1 of human brain tissue in vivo. Magn Reson Imaging. 1997; 15(10):1133–43. pmid:9408134
  37. 37. Kumar R, Delshad S, Woo MA, Macey PM, Harper RM. Age-related regional brain T2-relaxation changes in healthy adults. J Magn Reson Imaging JMRI. 2012; 35(2):300–8. pmid:21987489