Chitosan coating as an antibacterial surface for biomedical applications

  • Mélanie D’Almeida ,

    Contributed equally to this work with: Mélanie D’Almeida, Nina Attik

    Roles Investigation, Methodology, Visualization, Writing – original draft

    Affiliation Université Lyon, Université Claude Bernard Lyon 1, CNRS, Laboratoire des Multimatériaux et Interfaces, Villeurbanne, France

  • Nina Attik ,

    Contributed equally to this work with: Mélanie D’Almeida, Nina Attik

    Roles Supervision, Validation, Writing – original draft, Writing – review & editing

    nina.attik@univ-lyon1.fr

    Affiliations Université Lyon, Université Claude Bernard Lyon 1, CNRS, Laboratoire des Multimatériaux et Interfaces, Villeurbanne, France, Université Lyon, Université Claude Bernard Lyon 1, UFR d’Odontologie, Lyon, France

  • Julien Amalric,

    Roles Formal analysis, Methodology

    Affiliation Science et Surface, Ecully, France

  • Céline Brunon,

    Roles Formal analysis, Methodology, Writing – review & editing

    Affiliation Science et Surface, Ecully, France

  • François Renaud,

    Roles Investigation, Methodology, Validation

    Affiliation Université Lyon, Université Claude Bernard Lyon 1, CNRS, MATEIS (UMR 5510), Villeurbanne, France

  • Hazem Abouelleil,

    Roles Methodology, Writing – review & editing

    Affiliations Université Lyon, Université Claude Bernard Lyon 1, CNRS, Laboratoire des Multimatériaux et Interfaces, Villeurbanne, France, Université Lyon, Université Claude Bernard Lyon 1, UFR d’Odontologie, Lyon, France

  • Bérangère Toury,

    Roles Conceptualization, Formal analysis, Funding acquisition, Investigation, Supervision, Writing – original draft

    Affiliation Université Lyon, Université Claude Bernard Lyon 1, CNRS, Laboratoire des Multimatériaux et Interfaces, Villeurbanne, France

  • Brigitte Grosgogeat

    Roles Data curation, Funding acquisition, Project administration, Supervision, Writing – original draft, Writing – review & editing

    Affiliations Université Lyon, Université Claude Bernard Lyon 1, CNRS, Laboratoire des Multimatériaux et Interfaces, Villeurbanne, France, Université Lyon, Université Claude Bernard Lyon 1, UFR d’Odontologie, Lyon, France, Service de Traitements et de Consultations Dentaires, Hospices Civils de Lyon, Lyon, France

Abstract

Background and objectives

A current public health issue is preventing post-surgical complications by designing antibacterial implants. To achieve this goal, in this study we evaluated the antibacterial activity of an animal-free chitosan grafted onto a titanium alloy.

Methods

Animal-free chitosan binding on the substrate was performed by covalent link via a two-step process using TriEthoxySilylPropyl Succinic Anhydride (TESPSA) as the coupling agent. All grafting steps were studied and validated by means of X-ray Photoelectron Spectroscopy (XPS), Time-of-Flight secondary ion mass spectrometry (ToF-SIMS) analyses and Dynamic-mode Secondary Ion Mass Spectrometry (DSIMS). The antibacterial activity against Escherichia coli and Staphylococcus aureus strains of the developed coating was assessed using the number of colony forming units (CFU).

Results

XPS showed a significant increase in the C and N atomic percentages assigned to the presence of chitosan. A thick layer of polymer deposit was detected by ToF-SIMS and the results obtained by DSIMS measurements are in agreement with ToF-SIMS and XPS analyses and confirms that the coating synthesis was a success. The developed coating was active against both gram negative and gram positive tested bacteria.

Conclusion

The success of the chitosan immobilization was proven using the surface characterization techniques applied in this study. The coating was found to be effective against Escherichia coli and Staphylococcus aureus strains.

Introduction

In recent years, the development and application of biomaterials has greatly increased in several medical fields such as plastic surgery, ophthalmology, orthopedic and cardiac surgery, urology and odontology [1]. In order to respond to growing clinical needs, it is essential to offer superior performing new biomaterials. Devices based on titanium alloys are mostly acknowledged for their mechanical properties, their excellent corrosion resistance and their high biocompatibility [2,3]. Nevertheless, titanium implant surfaces do not prevent bacterial adhesion and proliferation [4]. The main challenge is to provide antibacterial properties to titanium alloy based implants in order to inhibit bacteria growth and/or disrupt biofilm formation on the implant surface. An efficient method therefore would be the functionalization of implant surfaces with natural bioactive polymers to prevent infections [5].

Various strategies based on titanium surface coating are reported to bring antibacterial properties to the implant surface [69]. One of the methods comprises of using chitosan (CS) as a natural bioactive biopolymer commonly employed due to its cytocompatibility, its antibacterial properties and its biodegradability [10]. It can be prepared in different forms such as gels, nanoparticles, fibers, membranes or sponges, allowing a large variety of biological and medical applications such as tissue engineering [11,12], wound healing [1315], drug delivery [16,17] or skin regeneration [18]. This polysaccharide is a copolymer of N-acetyl-D-glucosamine and D-glucosamine units with β binding at position 1 and 4 obtained by heterogeneous deacetylation of chitin [13,19] and is obtained from microorganisms, some fungi and the exoskeletons of crustaceans [20].

In order to bond chitosan to a titanium support, several methods are considered in the literature; such as the dopamine glutaraldehyde method [21], the electrodeposition of chitosan on substrates [22], or the use of catechol functional groups [23]. Another strategy consists of using an organosilane as a coupling agent; such as 3-AminoPropylTriEthoxySilane (APTES) associated with glutaraldehyde or succinic anhydride [2427] or TriEthoxySilylPropylSuccinic Anhydride (TESPSA) [28]. Which is often selected as an intermediate between biomolecules and titanium due to its inherent antibacterial and biological properties [29].

Recent in vitro studies focused on the surface characterization, especially on the interface between an animal-based chitosan coating and metal support using both bulk and outermost surface characterization techniques [27]. Our group demonstrated the cytocompatiblity of grafted animal chitosan using TESPSA as coupling agent in the presence of fibroblasts [28,30]. However, biological and antibacterial behaviors differ according to chitosan characteristics such as origin, acetylating degree, molar mass and the production conditions [3133]. Moreover, the biopolymer form, whether film, fiber or sponge, also has an impact on the biological properties [34].

In this study, we present the steps of a method used to graft ultra-pure fungi chitosan onto ultra-pure titanium alloy (TA6V). To react with the titanium alloy support, TESPSA is employed as a silane source to ensure the formation of a peptide bond between chitosan and the titanium surface. The presence of the chitosan coating is demonstrated using X-ray Photoelectron Spectroscopy (XPS) and Time-of-Flight Secondary Ion Mass Spectrometry (ToF-SIMS) in static and dynamic modes. Furthermore, the effects of the bioactive coating against two common bacterial strains Escherichia coli and Staphylococcus aureus were evaluated.

Materials and methods

Chemical functionalization of titanium alloy

Titanium alloys (TA6V) were supplied by Global D (France), pellets of 1cm machined by Phen’X (batch: 14010091) were used and the raw materials identification was HEPTAL (cast: 82706880–78770). The polymer immobilization on titanium surface using TESPSA was performed as previously reported with modifications [28,29]. After the cleaning and oxidizing steps, leading to samples respectively labeled TA6V and PiTA6V, the silanation step of the surface was achieved by immersion of the sample in a solution of TriEthoxySilylPropylSuccinic Anhydride (TESPSA) (ABCR, Germany) dissolved in pentane (v/v, 0.1/100) under argon. After 1 hour, the solvent was removed and the samples were then heated at 150°C for 12 hours. The samples were then cleaned in pentane, tetrahydrofuran and dichloromethane by ultrasound for 20 min each (TPiTA6Vl). After drying under inert gas, a biopolymer solution, containing 2 wt % chitosan (DA 21.7, Mw 222 000 g/mol, Kitozym, Belgium) and 3% (v/v) acetic acid in deionized water, was grafted on the samples surface by dip coating (v = 1 mm/s). The chitosan coated samples were then dried at 80°C for 24 h and labeled CSTPiTA6V.

Surface chemistry characterizations

X-ray Photoelectron Spectroscopy (XPS).

Measurements were carried out using a PHI Quantera SXM instrument (Physical Electronics, Chanhassen, USA) equipped with a 180 hemispherical electron energy analyzer and a monochromatized Al Kα (1486.6 eV) source operated at 15 kV and 4 mA. The analysis spot had a diameter of 200 μm and the detection angle relative to the substrate surface was 45°. Standard deviations were calculated from measurements performed on two different areas. Data were analyzed using the Multipak software (version v.9.6.0, Physical Electronics). The depth probed by XPS analysis is between 5–10 nm.

Time-of-Flight secondary ion mass spectrometry (ToF-SIMS).

Measurements were performed using a ToF-SIMS 5 instrument (ION-TOF GmbH, Germany) following analysis and calibration conditions reported in our previous study [27]. A pulsed primary ion source of Bi3+ was operated at 25 KeV. The scanning area of secondary ions was 100 μm × 100 μm. The depth probed was about the first monolayer in the static mode. Data were analyzed using the IONTOF Measurement Explorer software. Standard deviations were calculated from measurements performed on three different areas. The spectra were acquired in high current bunched mode. When required, charge effects were compensated by means of a pulsed electron flood gun (Ek = 20 eV), the primary ion dose density being then 1.25 × 1012 Bi3+/cm2. Spectra calibration was achieved using positions of C-, C3-, C4- peaks in negative mode, and CH3+, C4H7+, C5H9+ peaks in positive mode. Spectra comparison was carried out after a normalization of the intensity, proportionally to the total intensity of each spectrum. Pure liquid TESPSA and self-supported chitosan membrane were used as reference samples in order to select characteristic fragments that could be compared with the covalently linked chitosane coating (Table 1).

thumbnail
Table 1. Specific positive and negative fragments of TESPSA and chitosan.

https://doi.org/10.1371/journal.pone.0189537.t001

Dynamic-mode Secondary Ion Mass Spectrometry (DSIMS).

ToF-SIMS 5 instrument (ION-TOF GmbH, Germany) was used with a pulsed primary ion source of Bi1+ operated at 25 KeV for analysis, the same operating conditions were selected as used in a previous study [27]. Scanning area of secondary ions was 100 μm × 100 μm in the center of a larger sputtered area, in order to minimize edge effects and redeposition linked to the formation of the crater. A primary ion source of O2+ was operated in non-interlaced mode at 2 KeV for sputtering with sputtered area of 300 μm × 300 μm. Charge effects were compensated by means of a pulsed electron flood gun (Ek = 20 eV). The depth probed is at the nanometer scale in the dynamic mode, with each point of the profiles obtained from a mass spectrum. Ion intensities were plotted versus the sputtering time, which was converted into depth after profilometry measurements.

Antibacterial assessments

Antibacterial activity of chitosan coated samples was assessed in vitro against two bacterial strains: Escherichia coli CIP54127 and Staphylococcus aureus CIP483. These strains were selected in order to represent gram-negative and gram-positive bacteria respectively, according to the ISO 22196: 2007 standard method for the measurement of antibacterial activity on plastics and extended to non-porous surfaces in 2011 [35]. The number of colony forming units (CFU) on PCA was counted immediately after inoculation and after 24 h incubation at 37°C for both control and tested samples. Uncoated titanium alloy was used as control. The antibacterial activity A was calculated using the following formula: A = (Ut–U0)–(Ct–C0) where ‘U’ and ‘C’ were the average between the logarithmic number of bacteria enumerated after removal from untreated surfaces and coated titanium, respectively; and ‘t’ corresponds to the 0 and 24 h inoculation times tested. A sample was considered as active only if A is higher than 1 and bactericidal when the number of viable bacteria is null. All experiments were performed in duplicate with three samples at each (n = 6).

Results and discussion

Control of functionalization of titanium alloy surface

X-ray Photoelectron Spectroscopy (XPS) analysis.

XPS is considered to be a very highly sensitive method for the characterization of the surface chemistry of materials used in different applications, especially in biotechnology [36] and to verify the chemical composition of different coatings [37,38]. Hence, this technique was used to analyze all samples after grafting. Atomic elemental compositions measured at the surface of all samples are reported in Table 2 and XPS survey spectra of PiTA6V (a), TPiTA6V (b) and CSTPiTA6V (c) are presented in Fig 1A. In addition, deconvolution of the carbon signal was observed in order to allow the differentiation of carbon chemical environment. Changes of chemical bonding nature in the carbon environment allowed the success of the grafting procedure. Fig 1B, 1C and 1D present deconvolutions of the carbon signals detected in samples PiTA6V (B), TPiTA6V (C) and CSTPiTA6V (D).

thumbnail
Fig 1.

XPS general survey (A) of PiTA6V (a), TPiTA6V (b) and CSTPiTA6V (c) and XPS high resolution and deconvolution of carbon spectra of PiTA6V (B), TPiTA6V (C) and CSTPiTA6V (D).

https://doi.org/10.1371/journal.pone.0189537.g001

thumbnail
Table 2. Atomic elemental composition percent for the individual reaction steps determined by XPS.

The different surfaces obtained were labeled as follow: unmodified surface (TA6V), piranha treated titanium alloy surface (PiTA6V), TESPSA modified PiTA6V (TPiTA6V) and chitosan modified TPiTA6V surface (CSTPiTA6V).

https://doi.org/10.1371/journal.pone.0189537.t002

Comparison between elemental compositions of samples before and after piranha treatment demonstrated the positive effect of the oxidizing step since the oxygen content increased, meaning more Ti-OH sites were available at the sample surface. The relative content of atomic ratios of O/Ti of TA6V (3.53 ± 0.03) and PiTA6V (3.92 ± 0.32) samples were in agreement with these results. In the same way, a decrease in the carbon content may imply less surface contamination. Nevertheless, as previously noticed, it appears difficult to eliminate all contaminations, partially due to the high surface reactivity and atmospheric elements [25].

The silane grafting is highlighted by a significant increase of the C and Si atomic percentages. Meanwhile, absence of Ti peak is in agreement with a total masking of the surface after the silanation step (Fig 1A). These observations are in accordance with the total covering of the sample surface with a silane layer over 10 nm (XPS detection limit). On TPiTA6V, the C peak was deconvoluted into three main components at 289 eV (15.4 ± 0.6%), at 286.2 eV (6.5 ± 0.8%) and at 284.8 eV (77.3 ± 2.3%) which were attributed to O = C-O, C-O/C-N and C-C/C-H/C-Si respectively and the last component at 287.6 eV (0.8 ± 0.9%) was attributed to O-C-O/C = O. Therefore, due to the non-detection of silicium on PiTA6V sample in the general survey, the signal at 284.8 eV was assigned to C-C or C-H chemical groups. On TPiTA6V sample, an increase in this signal (Fig 1C) could be characteristic of the new C-Si environment brought by the silane. Here the silane signature and the masking of substrate were present still after all the washing steps. This result proved the strong link between titanium alloy and silane. In the same way, the signal at 289.0 eV was attributed to O = C-O environment characterizes the anhydride function of the silane. The effective covalent grafting of silane onto titanium-based support was previously observed [29].

Finally, following the biopolymer deposition step (CSTPiTA6V), a significant increasing of C and N atomic percentages assigned to the chitosan could be observed. In parallel, no detection of Ti and Si peaks are in agreement with the total masking of the silane. The deconvoluted peaks of C 1s gave four components: components at 289 eV (2.0 ± 0.1%), at 287.6 eV (15.9 ± 1.5%), at 286.2 eV (51.5 ± 2.8%) and at 284.8 eV (30.7± 4.0%) which were assigned to N-C = O, O-C-O/C = O, C-O/C-N and C-C/C-H respectively [39,40]. These results confirm the success of chitosan grafting as previously noticed [27,28].

Time-of-Flight secondary ion mass spectrometry (ToF-SIMS).

Analyses of positive and negative fragments allowed the monitoring of each step of the grafting process to the support (Fig 2). Following the silanation step, characteristic positive and negative fragments of TESPSA were investigated (Fig 2A and 2B) and compared to the uncoated sample PiTA6V. Results showed the complete masking of the substrate after silane deposition, proven by the decreasing amounts of Ti+, TiO+, TiO2- and Ti2O5H-. In addition, absence of molecular fragments corresponding to TESPSA alone such as SiC13H25O6+ [M+H]+ and SiC13H25O7- [M+OH] were in agreement with the effective reaction between the support and the silane. A cross-polymerization reaction could occur and strengthen the film via the Si-O-Si bond formation.

thumbnail
Fig 2.

Average normalized intensity of positive and negative fragments of PiTA6V and TPiTA6V (A) and (B) respectively, TPiTA6V and CSTPiTA6V (C) and (D).

https://doi.org/10.1371/journal.pone.0189537.g002

After chitosan deposition, characteristic positive and negative fragments of biopolymer defined previously were fully detected (Fig 2C and 2D) [27,41]. In parallel, the decrease of silane fragment intensities suggested a thick layer of deposited polymer. XPS and ToF-SIMS analyses proved the efficiency of chitosan deposition. However, it seems necessary to study the interface between each layer after the coating synthesis.

DSIMS analysis.

The results obtained by DSIMS measurements are in agreement with ToF-SIMS and XPS analyses and confirms the coating synthesis success. Each step of the process was characterized using DSIMS analysis [39]. ToF-SIMS positive ion depth profiles were recorded on chitosan-coated samples after completion of the process (Fig 3). As previously reported, the distribution of 46Ti+, C+, Si+, CH4N+ and Na+ ion intensities were profiled using a linear scale [27]. Three successive regions were observed on the CSAPiTA6V sample. The presence of C+ and CH4N+ was assigned to the presence of biopolymer on the support. The C+ fragment, a characteristic signal of the organic layer, decreased until a second area attributed to the polymer/silane/substrate interface became prominent. This second area was characterized by the presence of Si+ ions which corresponded to the TESPSA layer and by the enrichment and disappearance of CH4N+. Therefore, at this point, the bond formation between chitosan and silane also was suggested. Alkaline contaminations like Na+ ions were found at the organic and the interface layer. In parallel to the decrease of Si+ fragments, 46Ti+ fragments increased in the last area assigned to the support.

thumbnail
Fig 3. ToF-SIMS depth profiles of positive fragments recorded from chitosan coated sample (CSTPiTA6V).

https://doi.org/10.1371/journal.pone.0189537.g003

Antibacterial activity

The antibacterial properties of chitosan-coated titanium samples were evaluated by measurements of number of CFU (colony-forming unit). The results obtained for polymer-coated titanium samples versus uncoated samples against E. coli and S. aureus bacteria strains are detailed in Table 3 (Table 3). As expected, uncoated titanium samples had no antibacterial activity after 24 hours of exposure. On the other hand, no viable bacteria were found in the presence of TESPSA-chitosan titanium samples after 24 hours of contact. In the current experimental conditions, chitosan-coated samples showed effective bactericidal potential against the two strains tested.

thumbnail
Table 3. Antibacterial activity of uncoated and chitosan coated samples after 24 h of direct contact.

https://doi.org/10.1371/journal.pone.0189537.t003

Biofilm formation occurs after the implantation of medical devices on either hard or soft tissues, and is the major cause of implant failure and bacterial infections. It is recently reported that infections and inflammation are the most common forms of postoperative complications in the case of implantable titanium-based biomaterials. Because of the biocompatibility of titanium surface, titanium implants are suitable substrates for microbial colonization and biofilm formation, which is still a serious clinical problem [42,43]. The local and systemic application of antibiotics is the most common method used to treat and prevent bacterial infections. However, increased resistance to antibiotics in different bacterial strains is the main concern related to this type of therapy and was recently considered as a “global threat” to human health [44].

The development of biomaterials with antibacterial properties is the ultimate goal to decrease disease occurrence and improve health. In this context, chitosan immobilization on a titanium surface using TESPSA process was performed to obtain antibacterial activity of the implant tested. The developed chitosan-coated implant was able to inhibit E. coli and S. aureus bacteria growth; these two bacteria strains are commonly tested when the antibacterial ability of an implant coating has to be assessed [45]. In the present study, the chemical process TESPSA was used and did not affect the natural antibacterial properties of the animal-free chitosan used. The developed chitosan-coated implant proved to be an excellent bacterial agent against representative gram-positive and gram-negative bacteria. Furthermore, the use of high-purity non-animal chitosan should be highlighted in the current study as it is reported to be non-allergenic and better tolerated by the human organism which could make it easier to bring the developed medical devices to the market.

Conclusion

A covalent grafting of fungi chitosan to medical grade TA6V substrate was performed using a two-step process. The coupling agent TESPSA that presents an anhydride function was selected since it is commonly used in biomedical applications. The success of the chitosan immobilization was proved using the surface characterization techniques applied. An ideal coating for biomedical applications should have antibacterial activity against bacterial strains found in the implant vicinity, limiting bacterial adhesion and proliferation and therefore preventing implant-related infections. Furthermore, and due to its chemical resistance under acidic conditions, this chitosan coating design demonstrates potential as an antibacterial coating for different medical devices. This study represents one of the steps before further investigation for clinical use.

Acknowledgments

This work was partially supported by the Rhône-Alpes Region, DGCIS, FEDER, Global D and the Fonds Unique Interministériel program through the research project “Plastipolis”. The authors would like to thank Ms Aurelia Vettese for her assistance in the bacteriology experiments and Dr. Mark Cresswell from Lucideon Limited, UK for verification of English language.

References

  1. 1. Holzapfel BM, Reichert JC, Schantz J-T, Gbureck U, Rackwitz L, Nöth U, et al. How smart do biomaterials need to be? A translational science and clinical point of view. Adv Drug Deliv Rev. 2013; 65: 581–603. pmid:22820527
  2. 2. Mishnaevsky L Jr, Levashov E, Valiev RZ, Segurado J, Sabirov I, Nariman E, et al. Nanostructured titanium-based materials for medical implants: Modeling and development. Mater. Sci. Eng., R. 2014; 81: 1–19.
  3. 3. Geetha M, Singh AK, Asokamani R, Gogia AK. 2Ti based biomaterials, the ultimate choice for orthopaedic implants–A review. Prog Mater Sci. 2009; 54: 397–425.
  4. 4. Yeo I-S, Kim H-Y, Lim KS, Han J-S. Implant surface factors and bacterial adhesion: a review of the literature. Int J Artif Organs. 2012; 35: 762–772. pmid:23138699
  5. 5. Zhong X, Song Y, Yang P, Wang Y, Jiang S, Zhang X et al. Titanium Surface Priming with Phase-Transited Lysozyme to Establish a Silver Nanoparticle-Loaded Chitosan/Hyaluronic Acid Antibacterial Multilayer via Layer-by-Layer Self-Assembly. PLoS One. 2016; 11: e0146957. pmid:26783746
  6. 6. Zhao L, Chu PK, Zhang Y, Wu Z. Antibacterial coatings on titanium implants. J Biomed Mater Res B Appl Biomater. 2009; 91: 470–480. pmid:19637369
  7. 7. Michiardi A, Helary G, Nguyen PC, Gamble LJ, Anagnostou F, Castner DG et al. Bioactive polymer grafting onto titanium alloy surfaces. Acta Biomater. 2010; 6: 667–675. pmid:19733698
  8. 8. Tejero R, Anitua E, Orive G. Toward the biomimetic implant surface: Biopolymers on titanium-based implants for bone regeneration. Prog. Polym. Sci. 2014; 39: 1406–1447.
  9. 9. Gilabert-Chirivella E, Pérez-Feito R, Ribeiro C, Ribeiro S, Correia DM, González-Martín ML et al. Chitosan patterning on titanium implants. Prog Polym Sci 2017; 111: 23–28.
  10. 10. Dash M, Chiellini F, Ottenbrite RM, Chiellini E. Chitosan—A versatile semi-synthetic polymer in biomedical applications. Prog Polym Sci 2011; 36: 981–1014.
  11. 11. Grolik M, Szczubialka K, Wowra B, Dobrowolski D, Orzechowska-Wylegala B, Wylęgała E, et al. Hydrogel membranes based on genipin-cross-linked chitosan blends for corneal epithelium tissue engineering. J Mater Sci Mater Med. 2012; 23: 1991–2000. pmid:22569736
  12. 12. Zhang Y-F, Cheng X-R, Chen Y, Shi B, Chen X-H, et al. Three-dimensional Nanohydroxyapatite/Chitosan Scaffolds as Potential Tissue Engineered Periodontal Tissue. J Biomater Appl. 2007; 21: 333–349. pmid:16543282
  13. 13. Deepthi S, Venkatesan J, Kim S-K, Bumgardner JD, Jayakumar R. An overview of chitin or chitosan/nano ceramic composite scaffolds for bone tissue engineering. Int J Biol Macromol. 2016; 93(Part B):1338–1353.
  14. 14. Anitha A, Sowmya S, Kumar PTS, Deepthi S, Chennazhi KP, Ehrlich H et al. Chitin and chitosan in selected biomedical applications. Prog Polym Sci. 2014; 39: 1644–1667.
  15. 15. Archana D, Singh BK, Dutta J, Dutta PK. In vivo evaluation of chitosan–PVP–titanium dioxide nanocomposite as wound dressing material. Carbohydr Polym. 2013; 95: 530–539. pmid:23618304
  16. 16. Khor E, Lim LY. Implantable applications of chitin and chitosan. Biomaterials. 2003; 24: 2339–2349. pmid:12699672
  17. 17. Jiang T, James R, Kumbar SG, Laurencin CT. Chapter 5—Chitosan as a Biomaterial: Structure, Properties, and Applications in Tissue Engineering and Drug Delivery. In: Deng SGKTL, editor. Natural and Synthetic Biomedical Polymers. Oxford: Elsevier. 2014; 91–113.
  18. 18. Behera SS, Das U, Kumar A, Bissoyi A, Singh AK. Chitosan/TiO2 composite membrane improves proliferation and survival of L929 fibroblast cells: Application in wound dressing and skin. regeneration. Int J Biol Macromol. 2017; 98:329–340. pmid:28174079
  19. 19. Singla AK, Chawla M. Chitosan: some pharmaceutical and biological aspects—an update. ‎J. Pharm. Pharmacol. 2001; 53: 1047–1067.
  20. 20. Vázquez J, Rodríguez-Amado I, Montemayor M, Fraguas J, González M, Murado MA. Chondroitin Sulfate, Hyaluronic Acid and Chitin/Chitosan Production Using Marine Waste Sources: Characteristics, Applications and Eco-Friendly Processes: A Review. Mar Drugs. 2013; 11: 747–774. pmid:23478485
  21. 21. Shi Z, Neoh KG, Kang ET, Poh C, Wang W. Bacterial adhesion and osteoblast function on titanium with surface-grafted chitosan and immobilized RGD peptide. J Biomed Mater Res A. 2008; 86: 865–872. pmid:18041731
  22. 22. Wang Z, Zhang X, Gu J, Yang H, Nie J, Ma G. Electrodeposition of alginate/chitosan layer-by-layer composite coatings on titanium substrates. Carbohydr Polym. 2014; 103: 38–45. pmid:24528698
  23. 23. Marie H, Barrere A, Schoenstein F, Chavanne MH, Grosgogeat B, Mora L. PEM anchorage on titanium using catechol grafting. PLoS One. 2012; 7(11): e50326. pmid:23226262
  24. 24. Martin HJ, Schulz KH, Bumgardner JD, Walters KB. XPS Study on the Use of 3-Aminopropyltriethoxysilane to Bond Chitosan to a Titanium Surface. Langmuir. 2007; 23: 6645–6651. pmid:17488131
  25. 25. Xu X, Wang L, Guo S, Lei L, Tang T. Surface chemical study on the covalent attachment of hydroxypropyltrimethyl ammonium chloride chitosan to titanium surfaces. Appl Surf Sci. 2011; 257: 10520–10528.
  26. 26. Gunda NSK, Singh M, Norman L, Kaur K, Mitra SK. Optimization and characterization of biomolecule immobilization on silicon substrates using (3-aminopropyl)triethoxysilane (APTES) and glutaraldehyde linker. Appl Surf Sci. 2014; 305: 522–530.
  27. 27. D’Almeida M, Amalric J, Brunon C, Grosgogeat B, Toury B. Relevant insight of surface characterization techniques to study covalent grafting of a biopolymer to titanium implant and its acidic resistance. Appl Surf Sci. 2015; 327: 296–306.
  28. 28. Campos D, Toury B, D’Almeida M, Attik GN, Ferrand A, Renoud P, Grosgogeat B. Acidic pH resistance of grafted chitosan on dental implant. Odontology. 2015;103(2):210–7. pmid:24972881
  29. 29. Godoy-Gallardo M, Guillem-Marti J, Sevilla P, Manero JM, Gil FJ, et al. Anhydride-functional silane immobilized onto titanium surfaces induces osteoblast cell differentiation and reduces bacterial adhesion and biofilm formation. Mater Sci Eng C Mater Biol App. 2016; 59: 524–532.
  30. 30. Toury B, Grosgogeat B, Renoud P, Desroches. Substrate on which chitosan or collagen is grafted via a covalent bond. Inventors; Patent WO2013110887 A1. 2013; UCBL CNRS. Lyon. France
  31. 31. Raafat D, Sahl HG. Chitosan and its antimicrobial potential-a critical literature survey. Microb Biotechnol. 2009; 2: 186–201. pmid:21261913
  32. 32. Kumirska J, Weinhold MX, Thöming J, Stepnowski P. Biomedical Activity of Chitin/Chitosan Based Materials—Influence of Physicochemical Properties Apart from Molecular Weight and Degree of N-Acetylation. Polymers. 2011; 3: 1875–1901.
  33. 33. Wenling C, Duohui J, Jiamou L, Yandao G, Nanming Z, Xiufang Z. Effects of the Degree of Deacetylation on the Physicochemical Properties and Schwann Cell Affinity of Chitosan Films. ‎J Biomater Appl. 2005; 20: 157–177. pmid:16183675
  34. 34. Tchemtchoua VT, Atanasova G, Aqil A, Filée P, Garbacki N, Vanhooteghem O et al. Development of a Chitosan Nanofibrillar Scaffold for Skin Repair and Regeneration. Biomacromolecules. 2011; 12: 3194–3204. pmid:21761871
  35. 35. ISO22196:2007-Plastics-Measurement of antibacterial activity on plastics surfaces.
  36. 36. McArthur SL, Mishra G, Easton CD. Applications of XPS in Biologyand Biointerface Analysis. SV Smentkowski. Surface Analysis and Techniques in Biology. Springer, Cham. 2014; chap 2: 9–36.
  37. 37. Xie Youneng, Zuo Jun, Zhou Bo, Mac Li, Yu Z.M., Qiuping , et al. Sandblast-free double-etched titanium for dental implants application. Materials Letters. 2016; 176: 74–77
  38. 38. Cordeiro Jairo M., Beline Thamara, Ribeiro Ana Lúcia, Rangele Elidiane C., da Cruz Nilson C. Cordeiro et al. Development of binary and ternary titanium alloys for dental implants. Dent Mater. 2017; 33: 1244–1257 pmid:28778495
  39. 39. Héquet A, Humblot V, Berjeaud J-M, Pradier C-M. Optimized grafting of antimicrobial peptides on stainless steel surface and biofilm resistance tests. Colloids and Surfaces B: Biointerfaces. 2011; 84: 301–309. pmid:21310597
  40. 40. Zemljič LF, Tkavc T, Vesel A, Šauperl O. Chitosan coatings onto polyethylene terephthalate for the development of potential active packaging material. Appl Surf Sci. 2013; 265: 697–703.
  41. 41. Renoud P, Toury B, Benayoun S, Attik GN, Grosgogeat B. Functionalization of titanium with chitosan via silanation: evaluation of biological and mechanical performances. PLoS One. 2012; 7(7): e39367. pmid:22859940
  42. 42. Pérez-Anes A, Gargouri M, Laure W, Van Den Berghe H, Courcot E, Sobocinski J, et al. Bioinspired Titanium Drug Eluting Platforms Based on a Poly-β-cyclodextrin-Chitosan Layer-by-Layer Self-Assembly Targeting Infections. Appl Mater Interfaces. 2015, 7 (23)12882–12893.
  43. 43. Ma K, Cai X, Zhou Y, Wang Y, Jiang T. In Vitro and In Vivo Evaluation of Tetracycline Loaded Chitosan-Gelatin Nanosphere Coatings for Titanium Surface Functionalization. Macromol Biosci. 2017; 17(2).
  44. 44. Ferraris S, Venturello A, Miola M, Cochis A, Rimondini L, Spriano S. Antibacterial and bioactive nanostructured titanium surfaces for bone integration. Appl Surf Sci. 2014; 311: 279–29139.
  45. 45. Tian Y, Cao H, Qiao Y, Meng F, Liu X Antibacterial activity and cytocompatibility of titanium oxide coating modified by iron ion implantation. Acta Biomater. 2014;10(10):4505–17. pmid:24914826