年次大会講演論文集
Online ISSN : 2433-1325
会議情報
638 Cable Ready System の初期固定に関する実験的研究
菅原 誠伊藤 雅之徳永 邦彦原 利昭
著者情報
会議録・要旨集 認証あり

p. 139-140

詳細
抄録

The transtrochanteric surgical approach to th e hip is commonly used, especially for revision hip surgery. Failure of greater trochanter osteotomy leads to hardware failure, persistent pain, and limp. The major cause of failure to heal is inadequate fixation of the greater trochanter. The objective of this study was to consider primary fixation characteristic by the tension of wire, pressure of cut surface of 2 fixation location. In this experiment, the model bone was used for the sample. As a result, when tension is the near 90 N, approximately 10 N decreased 60 minutes later. In addition, cut surface pressure decreases at greater trochanter top.

著者関連情報
© 2003 一般社団法人日本機械学会
前の記事 次の記事