2615 電子顕微鏡におけるコンタミネーション原因ガス成分の分析
佐藤 修橘内 浩之清水 実
著者情報
会議録・要旨集 認証あり

p. 473-474

詳細
抄録

Carbon deposition on specimens exposed to electron beams in an electron-microscope chamber, which is offer used to measure the size of ULSI patterns on silicon wafers, is very serious problem. We measured the increase of aluminum wiring-line width on wafers due to carbon deposition by directly spreading oils on the wafers. This method indicated that, linewidth increase varied from 1 to 200 nm. This method is a useful way to estimate carbon deposition since it is highly sensitive and does not get contaminated.

著者関連情報