S044012 水酸アパタイト溶射皮膜の4点曲げ密着強度試験におけるAE発生挙動
大塚 雄市小島 大輔野田 和宏宮下 幸雄武藤 睦冶
著者情報
会議録・要旨集 フリー

詳細
抄録

The purpose of this research is to propose a simpler method of evaluating adhesive strength for hydroxyapatite (HAp) coating using four-points bending test with a slit coated specimen. AE signals were recorded and delamination behavier from the edges of HAp coating was observed during fourpoints bending test in order to specify delamination points. AE signals from HAp coating specimen was observed earlier compared with that from Ti substrate. Delamination starting point of HAp coating is about 120MPa and energy release rate G_<max> is calcurated to be 8.59[N/m].