日本化学会誌(化学と工業化学)
Online ISSN : 2185-0925
Print ISSN : 0369-4577
微分シンクロ蛍光法によるフェナントレン中の微量アントラセンの定量法
古澤 源久橘 正樹杉目 聡櫻井 新之介
著者情報
ジャーナル フリー

1998 年 1998 巻 3 号 p. 211-214

詳細
抄録

Since ordinarily available phenanthrene contains impurities such as phenanthrenequinone and fluoranthene, it is difficult to determine anthracene by spectrofluorimetry directly. Trace amounts of anthracene in phenanthrene can, however, be selectively determined by using the 4th-derivative synchronous fluorescence intensity without interference of these impurities. In 5 mL of cyclohexane, 25 mg of the sample is dissolved. The synchronous fluorescence spectrum of the solution is measured with wavelength interval (Δλ) of 24 nm in the range of 360-395 nm. Trace amounts of anthracene in phenanthrene can be determined from the 4th-derivative intensity at 376 nm, which is obtained by computation of the synchronous fluorescence spectrum. By this method, anthracene can be determined simply and rapidly. The determination limit is 0.5 ppm.

著者関連情報

この記事は最新の被引用情報を取得できません。

© The Chemical Society of Japan
前の記事 次の記事
feedback
Top