Abstract

Needle electrodes, widely used in clinical procedures, are responsible for creating an electric field in the treated biological tissue. This is achieved by setting a constant voltage along the length of their metallic section. In accordance with Laplace’s equation, the electric field is spatially non-uniform around the electrode surface. Mathematical modelling can provide useful information on the spatial distribution of electrical fields. Indeed, exact solutions for the electrical problem are indispensable for validating numerical codes. All the analytical models developed to date to solve the needle electrode electrical problem have been one-dimensional models, which assumed an electrode of infinite length. We here propose the first analytical solution based on a two-dimensional model that considers the real length of the electrode in which the Laplace equation is solved through the method of separation of variables, dealing with the nonhomogeneous source term and boundary conditions by Green’s functions. On assuming a needle electrode of given length, the problem combines boundary conditions on the electrode boundary (of the first and second kind). Since this rules out using the Sturm-Liouville Theorem, the problem is decomposed into two different problems and the principle of superposition is used. The solution obtained can reproduce a reasonable electric field around the electrode, especially the edge effect characterized by an extremely high gradient around the electrode tip.

1. Introduction

Needle electrodes consist of a thin cylindrical metal piece partially coated with plastic. The electrode is really the exposed part (i.e., the non-insulated portion, also known as the conductive tip) beveled to a sharp point. During clinical use, the exposed part is completely surrounded by tissue. Single shaft needle electrodes are employed in many clinical procedures that require the creation of an electric field and the resulting current density field. To create this field a voltage difference has to be set between two electrodes, one of which is the needle electrode and the other is a large area electrode, also known as a patch, reference electrode, or dispersive electrode. A large variety of electrodes for soft tissue have been developed, such as the plain, single internally cooled, cluster internally cooled, expandable and wet electrodes [1]. The target tissue is the region near the needle because the current density there is high, producing the desired effect over the tissue. The accurate modeling of the electrical problem of current circulation between needle electrode and dispersive electrode through the tissue is important to determine if the desired effect can be obtained. Many clinical procedures employ needle electrodes of several types, such as electrical stimulation of excitable tissues [2, 3], characterization of tissue impedance [4], and thermal destruction of tissue by applying radiofrequency (RF) current [5, 6].

Mathematical modeling is able to provide useful information on the electrical behavior of needle electrodes, especially on the spatial distribution of the electrical field and current density produced in the tissue next to the electrode. The general electric behavior of any electrode is basically as follows: a voltage is uniformly set along the entire surface of the electrode, while both current density and electric field distributions are spatially non-uniform at the electrode surface, as they result from solving the Laplace equation (assuming electrical conductivity to be constant). This behavior has previously been correctly reproduced by mathematical models using numerical methods [710]. Since the numerical simulation has to be able to accurately reproduce the physical situation expressed by the partial differential equations that model the process, comparing their results with any existing benchmark analytical solutions would be one way of validating the numerical simulations. In other words, exact solutions for the electrical problem are indispensable to validate numerical codes, since the mesh refinements used in the codes should be able to capture the large gradients of voltage and electric field around the needle electrode, which the analytical solution must accurately determine. Closed form solutions may also be useful in designing control strategies [11], as well as for a quick check of operating conditions in clinical procedures.

To our knowledge, all the previous analytical models that aimed to solve the electrical problem produced by a needle electrode were one-dimensional models and assumed an electrode of infinite length [1214]. Since the real length of the electrode (non-insulated tip) was not considered, their solutions were not able to reproduce the edge effect characterized by an extremely high gradient around the tip and the joint between the insulated and non-insulated portions [9, 10]. Our goal was thus to find a two-dimensional analytical solution for the electric fields created by a needle electrode of a given length.

2. Materials and Methods

2.1. Description of the Model Geometry

The monopolar needle-like electrode is a small radius cylindrical device, which is inserted into a tissue. A voltage is applied between the electrode’s surface and a grounded patch located on the skin of the patient so that the current circulates through the tissue. The spatial distribution of voltage and current throughout all the tissues that comprise the electrical circuit is really very complex, but since the voltage decays rapidly as the radial distance from the electrode is increased, it is standard to model the electrical problem of monopolar cylindrical electrodes using a cylindrical coordinate system, by looking for planes or symmetry axes. This allows to model a three-dimensional physical problem with a simplified two-dimensional geometry (axisymmetric case) [15].

The analytical model represents a physical situation in which a needle electrode with radius is completely surrounded by homogeneous tissue (see Figure 1(a)). The reference electrode is assumed to be placed far from the needle electrode and has a comparatively much larger area. As the electrode radius is much smaller than the tissue, we assume the domain comprises an annular section of tissue of height = 150 mm with inner radius = 0.75 mm and outer radius = 60 mm (see Figure 1(b)). Since the conductive portion of the needle electrode is assumed to be in the middle of the tissue block, its length is .

The boundary r = is divided into three sections in order to model the finite length of the electrode: 0 ≤ z is the insulated portion of the electrode, < z < is the conductive portion of the electrode, i.e., through which the electrical current flows, and finally z is the tissue below the electrode. The boundary conditions are insulation on the boundary r = , except for the conductive portion of the electrode, where we assume a uniform voltage . The top axially normal surface (z = 0) is considered electrically insulated, while both outer radius boundary (r = ) and the bottom surface (z = ) are held at zero voltage, mimicking the reference electrode.

The electrical problem is solved by the Laplace Equationwhere σ is the electrical conductivity of tissue (S/m), considered constant in the model, and is the electric voltage (V). The electric field and electric current density are obtained from Equation (1) is a quasi-steady simplification of the basic relationship between electric current density and charge, called continuity equation, which is obtained by manipulation of Maxwell’s equations. The continuity equation states that the net electrical current flowing in and out of a volume is equal to the rate of change of electrical charge inside the volume. Since the time scale of the electrical phenomenon is much shorter than that of other associated phenomena, such as thermal damage or electrostimulation, we can simplify in this case the continuity equation as Laplace equation (1). On the other hand the quasi-static approximation is suitable due to the low value of the frequencies, which implies that the wavelength is much larger than the physical dimensions, and hence no electromagnetic wave is present. In practical terms this means that the electrical problem is mathematically solved as if it was a DC problem, and not a problem with AC signals.

In practical terms, the electrical problem is analytically solved through the method of separation of variables, where the nonhomogeneous boundary conditions are handled using Green’s functions.

2.2. Calculation (Analytical Solution)

Considering that in this problem the voltage is assumed to be a function of the radial (r) and axial coordinates (z), Equation (1) can be stated assubject to the following boundary conditions: where is the voltage applied to the conductive portion of the needle electrode. The electrical conductivity of the electrode is very large compared to that of the tissue, and for that reason the active surface of the electrode can be considered at a constant voltage. Moreover, since the metallic electrode is not considered as part of the domain, the solution does not depend on the electrical properties of the electrode.

As the electrode radius is much smaller than the radial dimension of the mathematical domain, we considered the region and as a quasi-symmetric region. A symmetric condition is mathematically equivalent to an insulated boundary.

Since this is a problem with combined boundary conditions (of the first and second kind) on the boundary r = , it is not suitable for direct solution by the Sturm-Liouville Theorem. The option is to decompose it into two different problems and use the principle of superposition.The decomposition of into and was used to make it possible to solve a mathematical problem, which combines both Dirichlet and Neumann boundary conditions at r = . This decomposition is possible because of the linearity of the governing equation and boundary conditions, which allows the use of the principle of superposition.

is obtained solvingsubject to the following boundary conditions:where is a constant voltage with no physical significance (i.e., it does not correspond to the real applied voltage ), but was chosen to guarantee the final satisfaction of the original problem expressed by Equations (4)-(10). Details of how to choose the value of are given in Section 4. By using the method of separation of variables, the solution to Equations (12)-(18) can be written aswhere and are the modified Bessel functions of the first and second kind, respectively, and are the eigenvalues for this problemand is obtained solvingsubject to the following boundary conditions:

Boundary conditions (27) and (28) guarantee that, by superposing the solutions for and , the gradients at are zero for sections 0 ≤ z ≤ and ≤ z ≤ . The solution to Equations (22) to (28) iswhereandIn the special case = Once both and are calculated and superposed to obtain V, the magnitude of the electric field vector is calculated aswhile the magnitude of the current density vector J is calculated as After solving the equations analytically, Matlab (Mathworks, Natick, MA, USA) was used to compute each value at any particular point, so that all the plots were created by choosing a particular set of points.

3. Results

As mentioned, the value of must be suitably chosen to achieve a mathematical result compatible with the physics, i.e., a uniform voltage distribution along the electrode surface. Figure 2 shows the voltage distributions obtained for three different values of .

We found that if a small value is chosen for (=0.1×), the resulting voltage along the electrode surface is not constant and shows a convex shape, with higher voltage at both ends (see Figure 2(a)). In contrast, if a large value is chosen (=0.8×), the resulting voltage along the electrode surface is not constant and shows a concave shape, with lower voltage at both ends (see Figure 2(c)). In the case of a needle electrode with 1.5 mm diameter ( = 0.75 mm) and conductive portion of 30 mm, we found that a value = 0.378× produced a constant voltage along the electrode surface, which makes sense from a physical point of view (see Figure 2(b)). The effect of changing is clearer in Figure 3, which shows the voltage profile on the electrode surface (r = ).

However, the value of depends on the electrode dimensions. For instance, when the electrode diameter is 1 mm ( = 0.5 mm) has to be 0.335×, while that for 2 mm ( = 1 mm) has to be 0.435×. Figure 4 shows voltage plots for these electrode diameter values. Note that these values of produce a uniform voltage along the electrode surface.

Similarly, different values of have to be found for different lengths of the conductive portion of the electrode. Importantly, in the search for the suitable value, the dimensions of the domain must be sufficiently large in the direction; otherwise it is difficult to find a value that meets the requirement. Figure 5 shows the voltage distribution obtained for two different values of the length of the conductive portion. Note that the domain had to be enlarged for longer lengths.

Once had been obtained, the magnitude of the vectors (electric field) and (current density) were calculated from the distribution of and as expressed in Eqs. (33) and (34). For instance, Figure 6 shows the profile distribution computed on the symmetry axis. Note that the value of was suitably changed to achieve a uniform voltage on the electrode surface.

The combination of these two terms made it possible to plot the vector field. For instance, Figure 7 shows the vector distributions of the current density , in which the edge effect, characterized by an extremely high value of at the electrode edges, is easily appreciated. This behavior is even more marked when the Joule heating (q) is plotted. This term is the volumetric power density (W/m3) in the tissue caused by the electrical current flow and is given by . In some clinical applications, such as radiofrequency ablation, this term is responsible for creating a thermal lesion. Figure 8 shows an example of this distribution. It can be seen that higher power is applied at the electrode tip than at the proximal edge (where the insulated and exposed parts meet).

4. Discussion

In this study we found a novel analytical solution for the two-dimensional electrical problem associated with a limited-length needle electrode. To date, all the analytical solutions have been based on an oversimplified scenario in which the electrode was assumed to have an infinite length [1214] and did not reflect the characteristic performance of the electric field around the electrode, especially at the tips.

While the solution was obtained by using standard mathematical methods (separation of variables), an original approach based on the principle of superposition was used to handle the combined boundary conditions on the electrode-tissue boundary. The solution obtained has the minor problem of requiring the search for an appropriate value for the parameter in order to achieve a uniform voltage on the electrode surface, which is the only thing that makes sense from a physical point of view. In the future this search could be automated for specific electrode dimensions. For now, the proposed solution realistically represents the electric behavior of a needle electrode and provides a reliable and useful analytic solution that can be used by researchers to test their own codes before running extensive numerical experiments, in designing control strategies, as well as for a quick check of the operating conditions in clinical procedures. Moreover, even though the analytical solution is contextualized in the medical field of needle electrodes, it could also be useful for other scientific and industrial applications in which a voltage is applied between a thin metal cylinder and a remote reference electrode.

Although most needle electrodes have a real physical structure based on hollow rather than solid bars, the lumen being used for instance to circulate cooled liquid, place thermocouples, introduces guide-wires, etc. In this respect, although our mathematical model did not consider the lumen, the electrical behavior is exclusively associated with the distribution of the electrical variables in the tissue, and hence what is included or not included in the electrode is irrelevant in terms of these electrical variables, as the electrode surface acts as a Faraday cage.

It is important to point out that the analytical solution was obtained by assuming uniform tissue surrounding the electrode. In particular, a constant value for electrical conductivity σ had to be assumed in order to solve the electric problem using the Laplace Equation (Eq. (1)). Although the electrical conductivity changes significantly from tissue to tissue, it does not change much for a single type of well irrigated tissue, like the tissue around the electrode. The model herein presented is valid in situations where electrical conductivity is spatially and temporally constant, such as in electrostimulation. In the case of RF heating, the proposed solution is limited to represent the electric behavior (electric field, current density, and Joule heating) in an initial state, i.e., just before induced heating changes the σ distributions.

For validation purposes, a comparison between this analytical solution and a numerical simulation using COMSOL 4.2a (COMSOL, Burlington, MA, USA), for the same dimensions and parameters, was conducted. The numerical solution, based on Finite Element Method, does not include the electrode as part of the computational domain, but includes the region and , not included in the analytical solution because the method of separation of variables requires the use of a regular geometry. The results obtained by both methods are shown in Figures 9 and 10. Figures 9(a) and 9(b) compare vis-a-vis the voltage contour plots obtained by both methods. Figure 10 compares the voltage profiles along at a line where r=0.75 mm (interface electrode-tissue). Both figures show the very good agreement in results obtained by both methods.

5. Conclusions

A novel analytical solution is proposed, which allows accurate reproduction of the electric behavior of a needle electrode of given length surrounded by a homogenous space. Importantly, the solution also allows to reproduce the edge effect characterized by an extremely high gradient around the electrode tip. The analytic solution was finally validated against a numerical solution based on a commercial software based on Finite Element Method.

Data Availability

The datasets generated during the current study are available from the corresponding author on reasonable request.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Acknowledgments

This work was supported by the Universidad Autónoma de San Luis Potosí (México), which granted R. Romero-Méndez who is on a sabbatical leave to do research in the field of biomedical engineering. This work was supported by the Spanish Ministerio de Ciencia, Innovación y Universidades under “Programa Estatal de I+D+i Orientada a los Retos de la Sociedad" (grant number: RTI2018-094357-B-C21).