オートクレーブ法によるハリの損傷について (第2報)
滅菌, 100回 (2500分) 実施の場合
縄田 隆生中村 辰三小田原 良誠篠原 昭二池内 隆治北出 利勝南川 正純兵頭 正義
著者情報
ジャーナル フリー

1981 年 30 巻 2 号 p. 105-112

詳細
抄録

We previously reported on needle damage to needles sterilized 50 times. This time we increased the number of sterilizations to 100 times and investigated needle damage. damage.
Sterilization was performed using a small high pressure steam sterilizer at a temperature of 126°C for 25 minutes each time.
1. Non-gold or silver needles had a tendency to become finer but there was no remarkable weight change noted.
2. Resistance to bending tests showed that Chinese needles become more difficult to break.
3. Elasticity tests showed that non-gold needles become easier to restore to their original state.
4. Insertion tests showed that some needles, especially Chinese needles and stainless steel No. 3 needles, become more difficult to insert smoothly.
5. Hardness and sturdiness tests showed a general decrease in both factors with the exception of hard silver No. 3 needles and Chinese needles which become sturdier.