1932

Abstract

Hydrodynamic chromatography (HDC) has experienced a resurgence in recent years for particle and polymer characterization, principally because of its coupling to a multiplicity of physical detection methods. When coupled to light scattering (both multiangle static and quasi-elastic), viscometric, and refractometric detectors, HDC can determine the molar mass, size, shape, and structure of colloidal analytes continuously and as a function of one another, all in a single analysis. In so doing, it exposes the analytes to less shear force (and, hence, less potential for flow-induced degradation) than in, for instance, size-exclusion chromatography. In this review, we discuss the fundamental chromatographic underpinnings of this technique in terms of retention, band broadening, and resolution, and we describe the power of multidetector HDC with examples from the recent literature.

Loading

Article metrics loading...

/content/journals/10.1146/annurev-anchem-062011-143107
2012-07-19
2024-04-28
Loading full text...

Full text loading...

/content/journals/10.1146/annurev-anchem-062011-143107
Loading
/content/journals/10.1146/annurev-anchem-062011-143107
Loading

Data & Media loading...

  • Article Type: Review Article
This is a required field
Please enter a valid email address
Approval was a Success
Invalid data
An Error Occurred
Approval was partially successful, following selected items could not be processed due to error