人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
生物弁の力学的挙動について
佐藤 正明大島 宣雄前田 肇岡村 健二
著者情報
ジャーナル フリー

1979 年 8 巻 1 号 p. 108-111

詳細
抄録

Static tensile test, stress relaxation test and fluid dynamical test were carried out to investigate the relatioships between the mechanical deformation or stiffness of leaflets and the opening behaviour of bioprosthetic valves. The specimens used were fresh and glutaraldehyde (GA)-treated canine aortic valves. The deformability was found in the fresh specimen, medium in the 0.05% GA-treated tissue, while the 0.1-5.0% GA-treated tissues had the minimum deformability with almost the same material properties. In the fluid dynamical tests, the opening resistances of fresh and 0.05% GA-treated valves were less than those of 0.1-5.0% GA-treated valves. It was elucidated that the valve functions were closely related to the material properties of aortic valve leaflets.

著者関連情報
© 一般社団法人 日本人工臓器学会