Open Access
26 March 2022 Multispectral snapshot imaging of skin microcirculatory hemoglobin oxygen saturation using artificial neural networks trained on in vivo data
Author Affiliations +
Abstract

Significance: Developing algorithms for estimating blood oxygenation from snapshot multispectral imaging (MSI) data is challenging due to the complexity of sensor characteristics and photon transport modeling in tissue. We circumvent this using a method where artificial neural networks (ANNs) are trained on in vivo MSI data with target values from a point-measuring reference method.

Aim: To develop and evaluate a methodology where a snapshot filter mosaic camera is utilized for imaging skin hemoglobin oxygen saturation (SO2), using ANNs.

Approach: MSI data were acquired during occlusion provocations. ANNs were trained to estimate SO2 with MSI data as input, targeting data from a validated probe-based reference system. Performance of ANNs with different properties and training data sets was compared.

Results: The method enables spatially resolved estimation of skin tissue SO2. Results are comparable to those acquired using a Monte-Carlo-based approach when relevant training data are used.

Conclusions: Training an ANN on in vivo MSI data covering a wide range of target values acquired during an occlusion protocol enable real-time estimation of SO2 maps. Data from the probe-based reference system can be used as target despite differences in sampling depth and measurement position.

1.

Introduction

Skin microcirculation is noninvasively accessible using optical instrumentation, and its physiologically related parameters, such as hemoglobin oxygen saturation (SO2), red blood cell (RBC) tissue fraction, and melanin content can be monitored using diffuse reflectance spectroscopy (DRS).1 Most DRS systems consist of a broadband light source and a spectrally resolving detector, although spectral resolution can also be achieved by sweeping or tuning the light source emission spectra. For fiber-based systems, the source–detector distance is small, typically in the mm range, causing detected photons to traverse sub-mm to single mm in tissue depth depending on the wavelength range. For imaging systems, a slightly larger sampling depth can be assumed in the visible wavelength range.2 Regardless of modality, the detected backscattered light intensity is affected by tissue scattering and absorption optical properties. This effect is wavelength dependent and, hence, DRS contains information about concentrations of biological chromophores3,4 dominated by oxy- and deoxyhemoglobin, melanin, and bilirubin. Analyzing the backscattered diffuse spectra, DRS can estimate microcirculatory physiological parameters such as SO2 without having to assess intensity variations associated with arterial blood pulsations, as is done in pulse oximetry.1

Until recently, many DRS systems were fiber-based and limited to pointwise measurements. Recent advancements in multispectral imaging (MSI) technology have created possibilities to register spatially resolved images, where each pixel is spectrally resolved, a so-called data cube. This can be acquired either by sweeping the light source and/or detector over an imaged area, collecting one row of spectral data at a time, or by sweeping through the spectral dimension using optical filters.5 These methods are limited by sweep time when it comes to studying rapidly changing events. There are snapshot systems capable of instantaneous capture of a whole spectral data cube using on-chip optical filters placed in a repeating mosaic pattern. These systems overcome the temporal limitation but introduce new challenges due to a limited spectral resolution with complex filter characteristics.6

Measured DRS spectra reflect changes in both absorption and scattering and are often analyzed by fitting to an analytical or numerical model of skin diffuse reflectance to assess the microcirculatory status of skin tissue.7 Solving the inverse problem involves iterative updates of model parameters until measured and modeled spectra agree. Diffusion theory (DT) is an approximation valid for simplified geometries and some other assumptions. This theory allows for analytical calculation of DRS spectra given the optical and geometrical properties of the measurement set-up.812 A more generally applicable model that is considered the golden standard is Monte Carlo (MC) simulations of light transport.1318 With MC, statistical sampling of millions or billions of photon trajectories is performed until a desired accuracy is obtained. Different methods have been proposed to accelerate MC simulations,19 some employing graphics processing units (GPUs). Recently, a real-time analysis using inverse MC for pointwise measurements has been achieved, calculating 2 to 10 sets of microcirculation parameters per second on a lap-top with a built-in GPU.20 For imaging applications, where the amount of data vastly surpasses that of a pointwise technique, processing is much more demanding.2

To avoid computationally demanding search algorithms, some research groups have applied machine learning approaches to directly extract tissue properties.14,21,22 The use of an artificial neural network (ANN) for determining optical properties was first reported by Farrell et al., where the algorithm was trained using DT.21 Later Farrell et al. evaluated ANNs on phantom and in vivo data.23 Since then, ANNs have demonstrated in several publications to accurately estimate the optical properties21,2426 and SO2.20,27 The computational efficiency of machine learning algorithms makes them well suited for real-time imaging applications, where the calculation load can be exactly determined by the number of inputs and number of nodes in the ANN.

Relevant and representative training data is of outmost importance when using ANNs. It must cover the full range of optical properties, types of tissues, measurement sites, etc., that may be targeted by the application. Training data may be generated using forward MC simulations when determining optical properties from reflectance measurements25,28,29 or chromophore concentrations directly.20,27 This approach requires an accurate model of the system properties such as spectral sensitivity, noise, and possible drift as well as a model of light propagation in tissue, and the measurement setup (skin type, location, etc.). Training data for snapshot imaging sensors must take the complex characteristics of the filter mosaic into account. However, we know from experience that it is difficult to determine the filter characteristics with such accuracy that reliable training data can be produced from MC simulations.2 Another approach, which to our knowledge has not previously been applied, is to train the algorithm using in vivo data that was coregistered with a validated reference system. This eliminates the need for a calibration procedure to handle the hardware characteristics and extensive light transport theory modeling.

The aim of this study was to develop and evaluate a methodology using a fast ANN method trained on in vivo data to estimate skin SO2 images from data captured by a snapshot MSI camera. The target data were concurrently measured by a validated point-measuring optical spectroscopy system. Measurements were continuously performed on healthy subjects during arterial and venous occlusion and release provocations, to capture SO2 values covering almost the complete 0% to 100% range. As a first proof-of-principle, this study is conducted in a population with a limited range of skin tones, focusing on demonstrating how variations in forearm skin SO2 can be imaged.

2.

Materials and Methods

The experimental setup and data acquisition were previously described in Ewerlöf et al. 201730 and 20212 and are summarized in the following sections. In this paper, the model used for analysis of MSI data is changed from inverse Monte Carlo to ANNs.

2.1.

Equipment

Diffuse reflectance data were acquired by a snapshot MSI camera (xiSpec MQ0022HG-IM-SM4X4-VIS, XIMEA®, Germany) containing a Fabry–Perot interference filter array (4×4  pixels) arranged in a recurring mosaic pattern. The 16 different filters have complex spectral sensitivities in the wavelength range 450 to 650 nm and have substantial spectral overlap. The mosaic filter array was positioned on top of the CMOS sensor (2048×1088  pixels), optically aligned to the CMOS sensor pixels. The camera was equipped with a 16-mm lens (C Series VIS-NIR Lens, Edmund Optics Inc., Barrington) and mounted side-by-side to an 8 LED ring light (R130, Smart Vision Lights) emitting light in the range 390 to 765 nm. A crossed polarizer/analyzer pair was placed in front of the light source (polarizer) and the detecting MSI camera (analyzer) to avoid specular reflections. Data were collected by the MSI system using MATLAB (MATLAB version 8.2.0, 2013b, computer software, The MathWorks Inc., Natick, Massachusetts). Sixteen consecutive snapshots, each with exposure time 20 ms, were acquired from the MSI system and summed to reduce noise and amount of data, resulting in an effective framerate of 1.4 frames per second. A PF-6000 EPOS instrument (Perimed AB, Järfälla, Sweden) was used as reference system.1 The probe-based method measures diffuse reflectance spectra at two source–detector fiber separations (0.4- and 1.2-mm center–center separation) in the wavelength interval 450 to 850 nm with a framerate of 1 Hz. Spectra were analyzed using an inverse Monte Carlo algorithm based on a three-layered tissue model that has been shown to assess absolute values of SO2 (%) to within 5% root-mean-square (RMS) deviation.31 The light source of the EPOS system was an AvaLight-HAL-S (Avantes, Apeldorn, The Netherlands) for the first 11 subjects and was changed to an AvaLight-HAL-S-Mini for the remaining 13 subjects. The arm cuff used for occlusion was controlled by the EPOS system.

2.2.

Measurement Protocols

Two separate occlusions (arterial and venous occlusion) were performed on 24 healthy subjects with a previously described protocol.2 Measurements were performed in seated position with one arm resting on a vacuum pillow (Germa Protec, AB Germa, Kristianstad, Sweden) for support and fixation. Left or right arm was chosen randomly and both protocols were performed on the same arm, starting with arterial occlusion. An arm cuff was placed around the upper arm and the subject was acclimatized in the room (24°C to 26°C room temperature) for 15 min before measurements started. For spatial reference in the region imaged by the MSI camera 3×3 ink dots were painted on the skin to mark, the corners of a quadrant of four 2×2  cm regions of interest (ROIs). The MSI camera was positioned 30 cm above the surface of the arm. The EPOS probe was applied to the skin using double-adhesive tape avoiding visible vessels, positioned distally and close to the area marked by the ink dots. The probe was repositioned between provocations.

For the arterial occlusion protocol, the subject’s forearm was completely occluded (arterial occlusion) by inflating the arm cuff to 250 mmHg for 5 min followed by a rapid release of pressure, causing reactive hyperemia reperfusion of the lower arm tissue. After a resting period of at least 45 min, the arm cuff was inflated once more, this time to 45 mmHg, causing a 5-min venous occlusion of the same arm. The venous occlusion was followed by a rapid release of pressure, again causing reperfusion of the lower arm tissue. Data were acquired for 5 min before occlusion (baseline), 5 min during occlusion, and for 10 min after release (reperfusion) for both protocols.

2.3.

Study Subjects

Measurements were performed on 24 healthy subjects and data from 20 of those were included in this study (10 women and 10 men, aged 21 to 39 years). They all had skin type I to III, self-reported using Fitzpatrick skin typing test.32 Exclusion criteria were known skin conditions or circulatory diseases, use of medication affecting the circulatory system and smoking. Extensive physical activity was not allowed for 24 h and coffee for 4 h prior to measurements. Written informed consent was obtained from all participants. The study was approved by the Regional Ethical Review Board in Linköping (D. No. 2015/392-31).

2.4.

Data Exclusion

Data from four subjects were excluded due to exclusion criteria (one subject had coffee prior to the measurements and one had a skin condition), nonadherence to the study protocol (one subject) and poor EPOS signal (one person). For the remaining subjects, data collected when the light source was turned off during dark measurements were excluded, as well as data from a few erroneous EPOS registrations.

2.5.

Data Preprocessing

Each 2048×1088-pixel image from the MSI camera contains intensity data shaped by the 16 unique bandpass filters overlaid on the sensor. Before further analysis, these raw images were dark-corrected by subtracting a dark raw image acquired interleaved during each measurement series. A multispectral data cube was then reconstructed using a weighted bilinear interpolation demosaicing algorithm33,34 followed by spatial binning of 4×4  pixels for each wavelength band. This resulted in a multispectral data cube with an effective resolution of 512×272  pixels for each of the 16 wavelength bands. Finally, the data cubes where white normalized using a multispectral data cube collected from a white reference Spectralon® tile (Labsphere, Inc., North Sutton, New Hampshire). The acquisition of dark raw images and white calibration raw images is further described in Ref. 2.

2.6.

Selection of Data

A 9-min time interval (from 1 min before occlusion until 3 min after release) covering the full dynamic range of SO2 values was selected for all subjects, involving about 760 acquired MSI data cubes each for arterial and venous occlusion. The location of the four ROIs was continuously tracked for spatial reference during each measurement, as described in Sec. 2.5 in Ref. 2. Each pixel of the acquired data cube provided MSI intensities from 16 channels forming the spectrum used as training data. Each spectrum was normalized to its mean intensity. To increase the amount of training data and to minimize spatial dependencies, spectral data from 10 pixels in each data cube were randomly chosen from one ROI near the EPOS probe. This resulted in about 7600×2=15,200 spectra acquired from the two occlusions for each subject. When comparing results to the previously reported inverse Monte Carlo algorithm,2 the mean spectrum over the 2×2  cm ROI was used. For imaging purposes, the whole area of interest (the lower arm) was masked out from the background and each pixel analyzed separately.

The target SO2 values measured by the EPOS reference system (SO2,EPOS) during arterial and venous occlusion were unevenly distributed, as shown in Fig. 1. By including both provocations, a more even distribution between 0% and 100% was achieved.

Fig. 1

The combined distribution of SO2,EPOS values used as targets for training ANNs (all 20 subjects). Light blue bars show values acquired during arterial occlusion and dark blue represent values acquired during venous occlusion.

JBO_27_3_036004_f001.png

2.7.

Training of ANNs

Training of shallow ANNs (one hidden layer) was performed using the deep learning toolbox in MATLAB (MATLAB version 9.10.0, R2021a). ANNs were trained with MSI data (intensities from 16 wavelength bands) as input, using SO2,EPOS as target. Each ANN was initialized using the function fitnet where the number of hidden layers was stated, and the network training function set to Levenberg–Marquardt backpropagation using mean-square-error loss function. The hyperbolic tangent function was used as activation function of the nodes in the hidden layer. Using default settings, the input data for training were divided into training data (70%), validation data (15%), and test data (15%). The ANNs were trained with the MATLAB function train using parallel computing. Training parameters were set to default values. Before repeating the training, the ANN was re-initialized with new weight and bias values to avoid similar starting set-ups.

For the ANN training and evaluation, the principle of leave-one-subject-out cross-validation35 was applied, i.e., data from all subjects except one were used for training, validation, and testing. Untouched and independent data from the left-out subject were then used for the final evaluation of the ANN. A set contained about 19×15,200=288,800 spectra used for training, validation, and testing, while 15,200 unused spectra were used for the final evaluation. This procedure was repeated 20 times until all subjects had been used in the final evaluation. For each left-out subject, training was repeated 10 times with different starting parameters for each set of training data and for nine different sizes of the hidden layer (1, 2, 3, 4, 5, 7, 10, 15, and 20 nodes) resulting in 90 trained ANNs. The concurrently measured SO2,EPOS was used as target for each spectrum. A summary of the training procedure is shown in Fig. 2.

Fig. 2

An overview of the experiment set-up and main steps of training and evaluation procedures. Each training data set consisted of data from 19 out of 20 subjects. Data from the left-out subject, iteratively chosen, were used for evaluation.

JBO_27_3_036004_f002.png

The target SO2,EPOS values of the training data were unevenly distributed in the interval 0% to 100%, as shown in Fig. 1. To account for this, error weights were calculated and used for training and evaluation. For the entire data set (data from all 20 subjects), the number of targets were binned in their respective 10% interval. Error weights were calculated as 1/number of targets in each bin and normalized to a number between 0 and 1. The resulting array, W, containing a weight for each of the 10% bins, was applied to the data sets to level out the impact of data from the respective bin.

To investigate the importance of including data with a full span of SO2 target values, training was performed without using data from arterial occlusion. This training was performed using the same W as described above. The trained ANNs were evaluated for data from arterial occlusion protocols including both lower and higher SO2 values than those included in the training data set.

2.8.

Evaluation of ANN Results

The performance of a trained ANN was represented by a weighted RMS error, Erms,W, between target values from the EPOS system, SO2,EPOS, and values estimated by ANN, SO2,ANN, calculated according to Eq. (1). WEPOS was a vector containing the corresponding weight from W for each target value:

Eq. (1)

Erms,W=WEPOS·(SO2,EPOSSO2,ANN)2WEPOS.
ANNs chosen for each set of training data were applied to the 20 evaluation data sets containing data from arterial occlusion and to averaged data over a 2×2  cm ROI analyzed by the inverse MC algorithm presented in Ref. 2. To illustrate application of the ANN method for imaging, spatially resolved SO2 images, where a trained ANN was applied to spectral data from individual pixels, were also calculated for three points of time (during baseline, at end of occlusion, and during reperfusion).

2.9.

Statistics

Bland–Altman analysis36 was performed on the results to compare ANN results to target EPOS values. The analysis was performed for three time intervals: baseline (from 60 s before until 5 s before occlusion starts), end of occlusion (from 60 s before until 10 s before occlusion ends), and early after release (from 5 s before until 55 s after release). For the first two intervals, mean values were calculated for ANN and EPOS methods, respectively. During release the increase in SO2 is distinct. Since we wanted to capture the peak during release, we chose to use the median of nine samples around the peak EPOS value in the chosen interval. Mean bias was calculated, and upper and lower limits set to mean ±1.96 standard deviations (SD).

3.

Results

The performance of the trained ANNs during both training (a) and evaluation (b) for the 20 different data sets is shown in Fig. 3. For training, all data sets show a conformance in Erms,W, while it varies more when the trained ANNs are applied to unknown evaluation data. The number of nodes only had a minor influence on Erms,W. Mean Erms,W over all 20 evaluation data sets showed a lowest value for three nodes (9.5%). The variation in performance between 10 repeated trainings (shown as error bars) lies within 0.1% when calculated for data used during training and 0.7% for evaluation data. Therefore, the following results use only the first repetition of ANNs trained using three nodes in the hidden layer.

Fig. 3

Calculated Erms,W for ANNs trained with varying number of hidden nodes and applied to (a) the test data and (b) the leave-one-subject-out evaluation data, respectively. Each line represents the average ANN performance for 10 repetitions of training. Error bars show SD. Black dash-dotted line indicates the mean Erms,W over all data sets for each node size, respectively.

JBO_27_3_036004_f003.png

The time-resolved SO2,ANN for the evaluation data set with median Erms,W (9.1%) is shown in Fig. 4(a) together with corresponding SO2,EPOS. Bland–Altman analysis was performed for the 20 evaluation data sets analyzed with their respective trained ANN at three time intervals indicated by gray in Fig. 4(a). The result is shown in Fig. 4(b). ANNs give results similar to EPOS values in baseline with mean bias 0.08% with upper and lower limits of 17.4% and 17.3% deviation and at reperfusion with mean bias 0.7% with upper and lower limits of 7.8% and 6.4% deviation. Results at end of occlusion show a larger difference with mean bias 5.3% with upper and lower limits of 0.2% and 10.3% deviation, respectively.

Fig. 4

(a) Typical individual SO2,ANN (red) and corresponding SO2,EPOS (black) during arterial occlusion. Vertical lines indicate start and end off occlusion. Time intervals used for Bland–Altman analysis are marked in gray for baseline (0 to 55 s), end of occlusion (300 to 350 s), and reperfusion (355 to 410 s). (b) Agreement between SO2,ANN and SO2,EPOS as a Bland–Altman analysis of 20 subjects at baseline (squares/dashed lines), end of occlusion (circles/solid lines), and reperfusion (triangles/dash-dotted lines). The mean difference for each phase is represented by thin lines and the difference ±1.96*SD (95% prediction interval) by bold lines. Symbols filled with red represent data from the dataset in (a).

JBO_27_3_036004_f004.png

The ANNs were applied to the same MSI data analyzed by inverse MC in Ref. 2. The time-resolved result for a representative data set is shown in Fig. 5(a) and compared to the inverse MC result, SO2,invMC, and SO2,EPOS target data. In this case, the ANN method, compared to the inverse MC algorithm, gives a result closer to EPOS reference values in baseline and an even better match at reperfusion. At end of occlusion, the inverse MC algorithm is closer to EPOS. The Bland–Altman analysis comparing the inverse MC algorithm to ANN results for the 20 sets of ROI data in Fig. 5(b) shows higher values for the latter in all time intervals on average. The mean bias is 13.5% with upper and lower limits of 4.2% and 23.0% deviation in baseline, 5.1% with upper and lower limits of 1.0% and 9.2% deviation at end of occlusion, and 5.3% with upper and lower limits of 4.2% and 14.9% deviation for reperfusion.

Fig. 5

(a) The SO2 during arterial occlusion estimated by the inverse MC algorithm in Ref. 2 (SO2,invMC, blue), by ANN (SO2,ANN, red) as well as corresponding reference value (SO2,EPOS, black). Vertical lines indicate start and end off occlusion. Time intervals used for Bland–Altman analysis are marked in gray for baseline (0 to 55 s), end of occlusion (300 to 350 s), and reperfusion (355 to 410 s). (b) Agreement between SO2,ANN and SO2,invMC for the 20 subjects at baseline (squares/dashed lines), end of occlusion (circles/solid lines), and reperfusion (triangles/dash-dotted lines). The mean difference for each phase is represented by thin lines and the difference ±1.96×SD (95% prediction interval) by bold lines. Symbols filled with red represent data from the data set in (a).

JBO_27_3_036004_f005.png

Results from training including only data from venous occlusion, SO2,ANN,venOccl, is shown in Fig. 6(a) for a representative evaluation data set and clearly show the restricted ANN’s difficulties to estimate both low and high target SO2,EPOS values. Bland–Altman analysis in Fig. 6(b) confirms the observation. The mean bias in baseline is 4.0% with upper and lower limits of 15.1% and 23.1% deviation. At end of occlusion, the mean bias is 24.3% with upper and lower limits of 7.4% and 41.1% deviation and results at reperfusion show a mean bias of 13.1% with upper and lower limits of 23.7% and 2.5% deviation, respectively.

Fig. 6

(a) Typical individual SO2,ANN (red), when ANNs were trained using only venous occlusion data, and corresponding SO2,EPOS (black). Vertical lines indicate start and end off occlusion. Time intervals used for Bland–Altman analysis are marked in gray for baseline (0 to 55 s), end of occlusion (300 to 350 s), and reperfusion (355 to 410 s). (b) Agreement between SO2,ANN and SO2,EPOS for the 20 subjects at baseline (squares/dashed lines), end of occlusion (circles/solid lines), and reperfusion (triangles/dash-dotted lines). The mean difference for each phase is represented by thin lines and the difference ±1.96×SD (95% prediction interval) by bold lines. Symbols filled with red represent data from the data set in (a).

JBO_27_3_036004_f006.png

Table 1 presents the average SO2 for the three time intervals indicated by gray in Fig. 4(a) seen over all 20 evaluation data sets when estimated by EPOS, ANNs applied on evaluation data, ANNs and inverse MC algorithm applied on ROI data, and the ANNs trained on venous occlusion data applied on data from arterial occlusion.

Table 1

SO2 estimated by ANN and EPOS for arterial occlusion. Mean (SD) and [max min] are given at baseline, end of occlusion, and start of reperfusion. The SO2,ANN is calculated for evaluation MSI data and for the mean over ROI-data. The result for the case where ANNs were trained only on data from venous occlusion is presented in the last row.

BaselineEnd of occlusionReperfusion
SO2,ANN (%)a55.8 (7.1)7.1 (2.9)87.3 (3.6)
[40.3 74.9][1.9 12.7][75.2 91.0]
SO2,ANN (%)b56.6 (7.1)6.5 (2.8)87.6 (3.7)
[41.0 75.8][0.8 11.0][75.2 91.5]
SO2,ANN,venOccl (%)a59.9 (6.0)26.1 (8.3)74.9 (4.6)
[46.3 71.5][17.7 48.6][64.3 83.6]
SO2,EPOS (%)55.9 (10.5)1.8 (1.6)88.0 (6.2)
[39.0 72.0][-0.5 5.3][66.8 96.2]
SO2,invMC (%)b43.1 (9.5)1.4 (2.3)82.2 (6.6)
[21.4 61.6][0.0 8.9][61.6 92.7]

aMean value over 10 pixels calculated from MSI data.

bCalculated from MSI data (mean intensity over ROI).

Our proposed and trained ANN was applied pixelwise to three intensity data cubes, resulting in SO2,ANN maps from three temporal points (during baseline, end of occlusion, and reperfusion) for arterial occlusion. The results are shown in Figs. 7(a)7(c) together with the corresponding SO2,invMC maps (d)–(f). Both algorithms show distinct differences for the three time points with SO2 values around 50% during baseline, close to 0% at end of occlusion and values between 90% and 100% for reperfusion. The computational time for an image (512×270  pixels) using a three node ANN was 0.056 s. For comparison, a 10 node ANN spent 0.066 s and an ANN with 20 nodes 0.082 s for the same image. The inverse MC algorithm took on average 0.051 s per pixel which adds up to 1 h 58 min for a 512×270-pixel image. Computations for both methods were performed by a stationary computer with an Intel® Xeon® CPU E5-1620 processor and 16 GB RAM.

Fig. 7

Spatially resolved SO2 estimated from data analyzed by (a)–(c) a trained ANN and (d)–(f) by the inverse MC algorithm described in Ref. 2. Data for three timepoints during arterial occlusion are presented: (a), (d) during baseline; (b), (e) end of occlusion; and (c), (f) reperfusion.

JBO_27_3_036004_f007.png

4.

Discussion

In this study, we investigated if data acquired in vivo from healthy subjects can be used for training ANNs that reliably estimate SO2 images from MSI data. Target data were SO2 values from a validated point measuring EPOS system. Evaluation with independent MSI data not included in the training data set, i.e., data from the left-out subject, showed excellent conformance between ANN estimates and EPOS values. Data used for training were measured by the same system and under similar conditions as the evaluation data, which eliminates the need to account for the system’s specific characteristics such as impinging light source spectrum, sensor filters, and polarization. The utilized xiSpec MSI camera has a sensor with complex transmission characteristics for each of the 16 different bandpass filters. Every xiSpec sensor is unique in terms of sensitivity to different wavelengths and training is consequently camera specific. Therefore, our presented method can be applied to other types of MSI systems if training data are acquired by the same system used for measurements.

Other groups have used training data generated with MC simulations of tissue models.7,27 This approach does not need a series of in vivo measurements, but instead needs modeling of both measurement system and light transport in the tissue under study. The noise induced during measurements by camera, light source and/or aspects of tissue and light transport model affects measured spectra and must be considered and accurately modelled.20 In our case noise is inherent from the in vivo measurements. The randomly chosen pixels in every measurement handle possible differences over the sensor area. Zherebtsov et al. use spectra simulated from a three-layered tissue model as training data with oxygen saturation ranging between 30% and 100%.7 Their ANN was then applied on spectra collected from the palm side of the hand during a 3-min finger occlusion protocol and a spectral range 675 to 825 nm was used for oxygen saturation estimations. Reported estimated SO2 values were 83% and 62% for baseline and at end of occlusion, respectively. Our corresponding values are 56% and 7% from ANN estimates and 56% and 2% measured with EPOS on the volar forearm. Differences might partly be explained by a difference in vessel beds with higher oxygenation in finger tissues. We investigated this further in a pilot experiment with a 5-min arterial finger occlusion using EPOS on the palm side of a finger. The results for SO2 ranged from 75% in baseline to 0% at end of occlusion. The longer wavelengths used by Zherebtsov et al. may give a larger penetration depth, but they report a value of 2  mm in tissue simulating phantoms. Light transport simulations using Monte Carlo technique show a sampling depth of 1.7 mm at 630 nm in skin tissue for our MSI system.2 Therefore, the differences probably origin from the lack of training data for low SO2 values, and small distinct spectral variations with oxygen saturation in their chosen wavelength range. From presented best-fit spectra their overall goodness-of-fit appear great, but finer distinct spectral variations (e.g., the 540 to 580 nm range and the 758 nm absorption peak) that are uniquely related to the oxygen saturation level is insufficiently accounted for in their inverse modeling. The high SO2 at end occlusion is hence likely an effect of an inverse solution that focuses more on the overall shape rather than the finer details. This demonstrates the importance of having accurate models that fully capture what is being measured by hyperspectral imaging or MSI when performing simulations.

Regardless of how training data are acquired, it must be relevant for the anticipated application and cover the full dynamics of possible outcomes. The arterial occlusion provocation provides a wide range of SO2 values from 0% at end of occlusion to values close to 100% during the reperfusion phase. Venous occlusion affects the SO2, but not to the same extent and mainly provide data in the range 10% to 70%. Applying ANNs trained on data from only venous occlusion on evaluation data from arterial occlusion show difficulties for ANNs to find accurate SO2 values for data with low and high target values, since those were not included in the training data set. Therefore, it is strongly recommended to include data from both arterial and venous occlusion in the training data and limit the impact of the still slightly uneven distribution of SO2 values by adding the error weight vector, W. One additional reason for including results from venous occlusion provocations in the training data set was to add data collected with varying levels of RBC tissue fraction.

Data originating from the same subject will not be completely independent as the optical properties (e.g., scattering and absorption) within one subject only vary to some extent. Similarly, samples from one subject, taken at the same time point, but at different spatial locations, will not be completely independent as the local microcirculatory blood volume and blood oxygenation will only vary slightly, governed by the overall tissue status. We try to overcome this partly by randomly sampling different spatial points over time and partly using two different provocations where blood amount and oxygenation will vary over time. Still, the validation data, used for determining when to stop the ANN training, and the test data will not be completely independent since it originates from the same subjects as data used for training. The test performance in Fig. 3(a) also indicates that there is a dependency between training data and test data as all ANNs display a very similar behavior. However, with the low complexity of our ANNs and the large amount of training data, there is no apparent risk for a substantial overtraining. This is also supported by Fig. 3(b) where the final evaluation, using completely untouched and independent data, display an average performance over all left-out subjects that is only marginally larger than the test performance [Fig. 3(a)].

Optical properties vary for different skin types and probably at different skin locations. In this study, we included only subjects with Fitzpatrick skin types I to III which exclude strongly pigmented skin types with larger fractions of melanin. The reference EPOS system can measure darker skin up to at least Fitzpatrick skin type V with good signal-to-noise ratio and model these spectra with good fidelity.20 However, the need for training ANNs to the full range of Fitzpatrick skin types I to VI is also indicated by recent observations in pulse oximetry, where darker skin results in a small but larger number of undetected hypoxemia events.37 Furthermore, all measurements were performed on the volar forearm during occlusion provocations. Our results suggest that a training data set of 19 subjects is enough for training an SO2 algorithm that can estimate SO2 in low-pigmented forearm skin. The system can probably be adapted to a variety of situations if an ANN, possibly using more than three hidden nodes, is trained on additional data covering also darker skin tones and other skin locations.

To evaluate how an ANN should be structured for our MSI system, we trained ANNs with different number of nodes in the hidden layer and executed the training 10 times for each size. Using the leave-one-subject-out procedure, ANNs were trained on 20 training data sets and performance of our method was estimated for 20 sets of evaluation data not included in the respective training data set. Using leave-one-subject-out results in a reliable and unbiased estimate of the model performance.35 The Erms,W in Fig. 3(b) varies only slightly for the 20 evaluation sets, and the minimum mean value indicates that a node size of 3 is sufficient to estimate SO2. Repeated trainings showed marginal differences, which implies a stable algorithm. Based on this, we concluded that repeated trainings were not necessary for our purposes. If the training is extended to other tissue types, sites, and/or provocation protocols, this may change. A larger number of hidden nodes could possibly distinguish more multifaceted and complex spectra but may also be overtrained which could cause overfitting to certain properties or parts of the spectrum. Furthermore, the algorithm’s computational time increases with an increased size of the hidden layer.

Bland–Altman analysis of SO2,ANN compared to SO2,EPOS shows agreement at baseline and reperfusion, but slightly higher values of SO2,ANN at end of occlusion. The result at reperfusion shows that relatively few training spectra with high target SO2 values does not affect performance when evaluating data with target values in that range.

As discussed in Ref. 2, the measurement depth is slightly different between MSI system and EPOS. Training ANNs with data from our MSI system to target EPOS SO2 values diminish those systematic differences. Different measurement sites probably explore and compare tissues with different physiology, including vascularization. Despite site differences this study shows that data from the probe-based reference system can be used as target and enable spatially resolved SO2 maps.

When trained ANNs are applied to MSI data used by Ewerlöf et al., results show agreement between SO2,ANN and SO2,invMC in the reperfusion phase but less conformance in baseline and at end of occlusion. ANNs were trained to fit EPOS measurements while the inverse MC algorithm was unconnected to data from the reference system. Therefore, a better agreement between EPOS and ANN is expected. Comparing images analyzed with ANN and inverse MC, respectively, show similar values for the three time points. The ANN algorithm results in a more homogeneous distribution of SO2 values. The computational time for the two methods is essentially different in favor of ANNs.

5.

Conclusions

Our proposed ANN can estimate SO2 from data acquired by an MSI camera system with unknown spectral characteristics when it is trained to target SO2 values concurrently measured by a validated probe-based reference system. The high computational speed enables real-time imaging. Training data need to be chosen carefully to cover expected differences in SO2 and possibly also variation in skin type and skin location.

Disclosures

The authors declare no conflicts of interest.

Acknowledgments

The authors want to acknowledge Ingemar Fredriksson and Martin Hultman for help and advice regarding EPOS measurements (IF) and discussions on ANNs (both). This study was financially supported by VINNOVA Grants via the Swelife and MedTech4Health programs (Grant Nos. 2016-02211, 2017-01435, and 2019-01522).

References

1. 

I. Fredriksson et al., “Inverse Monte Carlo in a multilayered tissue model: merging diffuse reflectance spectroscopy and laser Doppler flowmetry,” J. Biomed. Opt., 18 (12), 127004 (2013). https://doi.org/10.1117/1.JBO.18.12.127004 JBOPFO 1083-3668 Google Scholar

2. 

M. Ewerlof et al., “Estimation of skin microcirculatory hemoglobin oxygen saturation and red blood cell tissue fraction using a multispectral snapshot imaging system: a validation study,” J. Biomed. Opt., 26 (2), 026002 (2021). https://doi.org/10.1117/1.JBO.26.2.026002 JBOPFO 1083-3668 Google Scholar

3. 

T. M. Bydlon et al., “Chromophore based analyses of steady-state diffuse reflectance spectroscopy: current status and perspectives for clinical adoption,” J. Biophotonics, 8 (1-2), 9 –24 (2015). https://doi.org/10.1002/jbio.201300198 Google Scholar

4. 

D. Kapsokalyvas et al., “Spectral morphological analysis of skin lesions with a polarization multispectral dermoscope,” Opt. Express, 21 (4), 4826 –4840 (2013). https://doi.org/10.1364/OE.21.004826 OPEXFF 1094-4087 Google Scholar

5. 

G. Lu and B. Fei, “Medical hyperspectral imaging: a review,” J. Biomed. Opt., 19 (1), 010901 (2014). https://doi.org/10.1117/1.JBO.19.1.010901 JBOPFO 1083-3668 Google Scholar

6. 

B. Geelen, N. Tack and A. Lambrechts, “A compact snapshot multispectral imager with a monolithically integrated per-pixel filter mosaic,” Proc. SPIE, 8974 89740L (2014). https://doi.org/10.1117/12.2037607 PSISDG 0277-786X Google Scholar

7. 

E. Zherebtsov et al., “Hyperspectral imaging of human skin aided by artificial neural networks,” Biomed. Opt. Express, 10 (7), 3545 –3559 (2019). https://doi.org/10.1364/BOE.10.003545 BOEICL 2156-7085 Google Scholar

8. 

T. J. Farrell, M. S. Patterson and B. Wilson, “A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,” Med. Phys., 19 (4), 879 –888 (1992). https://doi.org/10.1118/1.596777 MPHYA6 0094-2405 Google Scholar

9. 

S. L. Jacques, R. Samatham and N. Choudhury, “Rapid spectral analysis for spectral imaging,” Biomed. Opt. Express, 1 (1), 157 –164 (2010). https://doi.org/10.1364/BOE.1.000157 BOEICL 2156-7085 Google Scholar

10. 

A. Kienle and M. S. Patterson, “Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium,” J. Opt. Soc. Am. A, 14 (1), 246 –254 (1997). https://doi.org/10.1364/JOSAA.14.000246 JOAOD6 0740-3232 Google Scholar

11. 

A. Kienle et al., “Noninvasive determination of the optical properties of two-layered turbid media,” Appl. Opt., 37 (4), 779 –791 (1998). https://doi.org/10.1364/AO.37.000779 APOPAI 0003-6935 Google Scholar

12. 

A. Liemert et al., “P3 solution for the total steady-state and time-resolved reflectance and transmittance from a turbid slab,” Appl. Opt., 58 (15), 4143 –4148 (2019). https://doi.org/10.1364/AO.58.004143 APOPAI 0003-6935 Google Scholar

13. 

D. Boas et al., “Three dimensional Monte Carlo code for photon migration through complex heterogeneous media including the adult human head,” Opt. Express, 10 (3), 159 –170 (2002). https://doi.org/10.1364/OE.10.000159 OPEXFF 1094-4087 Google Scholar

14. 

A. Doronin and I. Meglinski, “Online object oriented Monte Carlo computational tool for the needs of biomedical optics,” Biomed. Opt. Express, 2 (9), 2461 –2469 (2011). https://doi.org/10.1364/BOE.2.002461 BOEICL 2156-7085 Google Scholar

15. 

Q. Fang, “Mesh-based Monte Carlo method using fast ray-tracing in Plucker coordinates,” Biomed. Opt. Express, 1 (1), 165 –175 (2010). https://doi.org/10.1364/BOE.1.000165 BOEICL 2156-7085 Google Scholar

16. 

B. H. Hokr et al., “Modeling focusing Gaussian beams in a turbid medium with Monte Carlo simulations,” Opt. Express, 23 (7), 8699 –8705 (2015). https://doi.org/10.1364/OE.23.008699 OPEXFF 1094-4087 Google Scholar

17. 

J. Ramella-Roman, S. Prahl and S. Jacques, “Three Monte Carlo programs of polarized light transport into scattering media: part I,” Opt. Express, 13 (12), 4420 –4438 (2005). https://doi.org/10.1364/OPEX.13.004420 OPEXFF 1094-4087 Google Scholar

18. 

L. Wang, S. L. Jacques and L. Zheng, “MCML--Monte Carlo modeling of light transport in multi-layered tissues,” Comput. Methods Programs Biomed., 47 (2), 131 –146 (1995). https://doi.org/10.1016/0169-2607(95)01640-F CMPBEK 0169-2607 Google Scholar

19. 

C. Zhu and Q. Liu, “Review of Monte Carlo modeling of light transport in tissues,” J. Biomed. Opt., 18 (5), 050902 (2013). https://doi.org/10.1117/1.JBO.18.5.050902 JBOPFO 1083-3668 Google Scholar

20. 

I. Fredriksson, M. Larsson and T. Stromberg, “Machine learning for direct oxygen saturation and hemoglobin concentration assessment using diffuse reflectance spectroscopy,” J. Biomed. Opt., 25 (11), 112905 (2020). https://doi.org/10.1117/1.JBO.25.11.112905 JBOPFO 1083-3668 Google Scholar

21. 

T. J. Farrell, B. C. Wilson and M. S. Patterson, “The use of a neural network to determine tissue optical properties from spatially resolved diffuse reflectance measurements,” Phys. Med. Biol., 37 (12), 2281 –2286 (1992). https://doi.org/10.1088/0031-9155/37/12/009 PHMBA7 0031-9155 Google Scholar

22. 

S. Panigrahi and S. Gioux, “Machine learning approach for rapid and accurate estimation of optical properties using spatial frequency domain imaging,” J. Biomed. Opt., 24 (7), 071606 (2019). https://doi.org/10.1117/1.JBO.24.7.071606 JBOPFO 1083-3668 Google Scholar

23. 

T. Farrell et al., “Charge-coupled device and neural-network-based instrument for the noninvasive determination of tissue optical properties in vivo,” Proc. SPIE, 2135 117 (1994). https://doi.org/10.1117/12.175987 PSISDG 0277-786X Google Scholar

24. 

M. Jager, F. Foschum and A. Kienle, “Application of multiple artificial neural networks for the determination of the optical properties of turbid media,” J. Biomed. Opt., 18 (5), 057005 (2013). https://doi.org/10.1117/1.JBO.18.5.057005 JBOPFO 1083-3668 Google Scholar

25. 

A. Kienle et al., “Spatially resolved absolute diffuse reflectance measurements for noninvasive determination of the optical scattering and absorption coefficients of biological tissue,” Appl. Opt., 35 (13), 2304 –2314 (1996). https://doi.org/10.1364/AO.35.002304 APOPAI 0003-6935 Google Scholar

26. 

Q. Wang et al., “Broadband ultraviolet-visible optical property measurement in layered turbid media,” Biomed. Opt. Express, 3 (6), 1226 –1240 (2012). https://doi.org/10.1364/BOE.3.001226 BOEICL 2156-7085 Google Scholar

27. 

V. Dremin et al., “Skin complications of diabetes mellitus revealed by polarized hyperspectral imaging and machine learning,” IEEE Trans. Med. Imaging, 40 (4), 1207 –1216 (2021). https://doi.org/10.1109/TMI.2021.3049591 ITMID4 0278-0062 Google Scholar

28. 

Q. Wang, K. Shastri and T. J. Pfefer, “Experimental and theoretical evaluation of a fiber-optic approach for optical property measurement in layered epithelial tissue,” Appl. Opt., 49 (28), 5309 –5320 (2010). https://doi.org/10.1364/AO.49.005309 APOPAI 0003-6935 Google Scholar

29. 

L. S. Zhang, Z. Z. Wang and M. Y. Zhou, “Determination of the optical coefficients of biological tissue by neural network,” J. Mod. Opt., 57 (13), 1163 –1170 (2010). https://doi.org/10.1080/09500340.2010.500106 JMOPEW 0950-0340 Google Scholar

30. 

M. Ewerlöf, M. Larsson and E. G. Salerud, “Spatial and temporal skin blood volume and saturation estimation using a multispectral snapshot imaging camera,” Proc. SPIE, 10068 1006814 (2017). https://doi.org/10.1117/12.2251928 PSISDG 0277-786X Google Scholar

31. 

I. Fredriksson et al., “Evaluation of a pointwise microcirculation assessment method using liquid and multilayered tissue simulating phantoms,” J. Biomed. Opt., 22 (11), 1 –9 (2017). https://doi.org/10.1117/1.JBO.22.11.115004 JBOPFO 1083-3668 Google Scholar

32. 

T. B. Fitzpatrick, “The validity and practicality of sun-reactive skin type-I through type-VI,” Arch Dermatol., 124 (6), 869 –871 (1988). https://doi.org/10.1001/archderm.1988.01670060015008 Google Scholar

33. 

J. Brauers and T. Aach, “A color filter array based multispectral camera,” (2006). Google Scholar

34. 

S. Mihoubi et al., “Multispectral demosaicing using pseudo-panchromatic image,” IEEE Trans. Comput. Imaging, 3 (4), 982 –995 (2017). https://doi.org/10.1109/TCI.2017.2691553 Google Scholar

35. 

D. Gholamiangonabadi, N. Kiselov and K. Grolinger, “Deep neural networks for human activity recognition with wearable sensors: leave-one-subject-out cross-validation for model selection,” IEEE Access, 8 133982 –133994 (2020). https://doi.org/10.1109/ACCESS.2020.3010715 Google Scholar

36. 

J. M. Bland and D. G. Altman, “Applying the right statistics: analyses of measurement studies,” Ultrasound Obstet. Gynecol., 22 (1), 85 –93 (2003). https://doi.org/10.1002/uog.122 Google Scholar

37. 

A.-K. I. Wong et al., “Analysis of discrepancies between pulse oximetry and arterial oxygen saturation measurements by race and ethnicity and association with organ dysfunction and mortality,” JAMA Network Open, 4 (11), e2131674 (2021). https://doi.org/10.1001/jamanetworkopen.2021.31674 Google Scholar

Biography

Maria Ewerlöf is a PhD student with an interest in biomedical optics at the Department of Biomedical Engineering, Linköping University, Sweden. Her research includes snapshot multi spectral imaging ( https://liu.se/en/employee/marew49) for spatial and temporal estimation of microcirculatory properties of human skin tissue.

Tomas Strömberg is head of the Department of Biomedical Engineering at Linköping University. His research is within biomedical optics ( https://liu.se/en/employee/tomst50), with a focus on light transport modeling in pointwise and imaging laser Doppler Flowmetry, laser speckle and diffuse reflectance spectroscopy. The work is done in close collaboration with clinical researchers and with industry.

Marcus Larsson is deputy head of the Department of Biomedical Engineering, Linköping University, Sweden. His research is focused on theoretical and applied biomedical optics ( https://liu.se/en/employee/marla29) for tissue and microcirculatory characterization using speckle-based techniques and steady-state spectroscopic techniques. This includes collaborative work with industry and clinical researchers.

E. Göran Salerud is active at the Department of Biomedical Engineering at Linköping University. His research is within biomedical optics ( https://liu.se/en/employee/gorsa75), focusing on hyper spectral imaging and microscopy, investigating spatial and temporal microcirculation physiology. The translational work of biomedical optics involves healthy subjects to rare diseases and point of care for new methods and technology.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Maria Ewerlöf, Tomas Strömberg, Marcus Larsson, and E. Göran Salerud "Multispectral snapshot imaging of skin microcirculatory hemoglobin oxygen saturation using artificial neural networks trained on in vivo data," Journal of Biomedical Optics 27(3), 036004 (26 March 2022). https://doi.org/10.1117/1.JBO.27.3.036004
Received: 7 October 2021; Accepted: 3 March 2022; Published: 26 March 2022
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Oxygen

Multispectral imaging

Skin

Data acquisition

Data modeling

In vivo imaging

Cameras

Back to Top