Background

Computed tomography (CT) is the most frequently used cross-sectional clinical imaging modality in almost all European countries. In 2017, for example, 15 CT scans were performed per 100 inhabitants in Germany [1]. Since its introduction in the 1970s, CT technology has been continuously improved, as shown by increased scan speed, spatial resolution, and dose efficiency. One of the most promising developments in recent years is a new generation of X‑ray detectors. Conventional CT detectors (Fig. 1a) consist of distinct detector elements separated by reflecting lamellae [2]. The energy of an incoming X‑ray photon is converted into visible light by a scintillator (typically Gd2O2S). The optical photons are registered by photodiodes in the detector elements (detector pixels), and their number is proportional to the energy of the incoming X‑ray photons, i.e., the higher the energy of the X‑ray photons, the more optical photons are generated. Hence, the measured signal favors higher energy photons, which diminishes image contrast. Since this detection process is comparatively slow ([3]; Fig. 1b), the signals from several incoming photons are summarized, and these detectors are hence referred to as energy-integrating detectors (EIDs). Novel photon-counting detectors (PCDs) use a semiconductor (typically cadmium telluride [CdTe]) instead of a scintillator ([4]; Fig. 1c). Incoming X‑ray photons are converted into a measurable charge cloud that is transported to the detector pixels. This detection process is orders of magnitude faster compared to conventional EIDs. In combination with smaller detector pixels, this makes it possible to count single photons at clinical photon flux rates ([2, 5, 6]; Fig. 1d). Furthermore, photons can be sorted into several bins according to their energy, allowing for implicit dual- and multi-energy imaging with a single detector [5, 7,8,9].

Fig. 1
figure 1

Computed tomography detector types: a Conventional energy-integrating detector; an incoming X‑ray photon is converted to multiple visible photons, which are eventually detected by a light sensor resulting in a wide signal (c). b Direct conversion of incoming X‑ray photons into measurable charge clouds. d Photons reaching the detector can be almost simultaneously distinguished by a photon-counting detector. A detailed presentation of the detector element readout modes is provided in Fig. 5. CdTe cadmium telluride

Physical advantages

Spatial resolution and image noise.

As depicted in Fig. 1a, the scintillator in EIDs converts X‑ray photons into multiple optical photons. To achieve higher spatial resolution, reflecting lamellas are necessary to guide these photons to the used photodiodes, reducing the active detector area and dose efficiency.

The small detector pixels of PCDs allow for a higher spatial resolution, compared to conventional EIDs. Hence, current PCD CT scanners achieve a higher in-plane and cross-plane resolution [10] if image reconstruction is performed with appropriate voxel sizes and if dedicated ultra-high resolution (UHR) reconstruction kernels are used. Until recently, comparable UHR scans with conventional CT systems could only be performed by using dedicated high resolution hardware combs or grids that significantly reduce dose efficiency [11, 12]. Recently, EIDs with smaller pixels have been introduced into clinical practice, allowing for the acquisition of UHR data without additional grids or combs [13]. Having said that, the reflecting lamellae used in EIDs raise the question of the dose efficiency of these systems. However, publications evaluating this question are not available as yet.

Additionally, the combination of smaller detector elements and a reconstruction to the lower resolution of conventional EIDs allows for a reduction in noise and, conversely, radiation dose ([10, 14, 15]; Fig. 2). This effect has been theoretically predicted [16, 17], and recent measurements illustrated radiation dose reductions of up to 65% compared to EIDs at the same noise and spatial resolution [10, 18].

Fig. 2
figure 2

Dose- (ab) and noise-matched (cd) post-mortem thorax examinations using energy-integrating detector (EID) computed tomography (CT) (ac) and photon-counting detector (PCD) CT (bd) reconstructed to the same spatial resolution (CT window settings: center [C] = 60 HU, width [W] = 360 HU). The CT values (yellow) correspond to the standard deviation in the depicted regions of interest. ab A direct comparison reveals a 20% noise reduction for PCD over EID (computed tomography dose index volume = 23.0 mGy). cd The dose for the PCD CT was 14.6 mGy, while the dose for the EID CT acquisition was 23.0 mGy, i.e., the PCD acquisitions achieve the same noise level with 35% less dose compared to the EID acquisitions

Spectral properties.

Since PCDs count single photons, the energy weighting observed in EIDs is overcome. Hence, the effective energy of the detected X‑ray spectrum is shifted towards lower energies, increasing soft tissue and iodine contrast. Recent studies illustrate an increase in iodine contrast by up to 37%, which means that the contrast agent dose may be lowered by up to the same amount [19].

State-of-the-art PCD CTs simultaneously acquire at least two energy bins. Reconstructions are typically provided for the entire energy spectrum, i.e., corresponding to a conventional CT image, but individual energy bins can be reconstructed on demand. They might be used to compute images with optimal contrast and contrast-to-noise ratio (CNR), to perform material decomposition similar to dual energy imaging or to reduce beam-hardening and metal artifacts [20, 21].

Furthermore, bin images might be used in combination with potential novel contrast agents to exploit discontinuities in the absorption spectrum (k‑edges, see Fig. 3), i.e., photons of energies below the k‑edge (low bin) and above the k‑edge (high bin) of novel contrast agents containing chemical elements with high atomic numbers (high‑Z elements) can be separated. This approach promises greatly increased contrast compared to conventional iodinated contrast agents at lower injection volumes and potentially fewer undesired reactions to contrast media. Nonetheless, its effectiveness is also highly dependent on patient size. If more than two energy bins are measured, more than one contrast agent might be used simultaneously [22, 23]. However, high‑Z contrast agents are not yet available.

Fig. 3
figure 3

a Gadolinium (Gd), tungsten (wolfram, W), iodine (I), bismuth (Bi), and ytterbium (Yb) as potential contrast agents for photon-counting detector computed tomography (low threshold image, bone window). b X-ray mass attenuation coefficients of the elements in a [50]. Thresholds can be placed near the k‑edge of an element to maximize image contrast between low and high bins and hence in the final material decomposition. The k‑edge of I is at 33 keV, restricting the applicability of k‑edge imaging in clinical practice due to the poor photon statistics in the low energy bin. cd Low and high images at 80 kV using a threshold of 50 keV show a contrast improvement of Gd in the high bin. ef Low and high bin images using 140 kV and a threshold of 61 keV show an increase in contrast of W and Yb in the higher bin

Requirements

The SOMATOM CounT (Siemens Healthineers, Erlangen, Germany) [24] is the first available whole-body PCD CT based on a clinical dual-source CT scanner. It houses a PCD as well as a conventional EID, allowing for a comparison of both technologies in the same system (Fig. 4). The PCD used in this system provides a native detector pixel size of 225 µm. To limit data transfer demands, the detector usually summarizes the signal of 2 × 2 or 4 × 4 detector pixels (binning). The highest spatial resolution is achieved in the UHR mode using a 2 × 2 binning (Fig. 5a). Acquisitions in this mode result in a detector pixel size of 0.25 mm at the isocenter. Another mode referred to as Macro operates using a 4 × 4 binning, resulting in an overall detector pixel size of 0.50 mm and a spatial resolution that is similar to the EID in the SOMATOM CounT system (Fig. 5b). Macro and UHR modes simultaneously acquire two energy bins, while another mode referred to as Chess acquires four energy bins (Fig. 5c).

Fig. 4
figure 4

The SOMATOM CounT computed tomography (CT) system at the German Cancer Research Center (DKFZ). The position of the energy integrating detector (EID) and photon-counting detector (PCD) and the corresponding two X‑ray sources A and B are marked in blue and green. The housing is identical to a conventional clinical CT system

Fig. 5
figure 5

Readout modes of the SOMATOM CounT photon-counting detector as a combination of nearby detector elements. Each element is able to discriminate two different energy bins (Ei / Ej). a Ultra-high resolution mode with 2 × 2 binning and b macro mode with 4 × 4 binning discriminate two energy bins, while c chess mode provides four different energy bins

Potential clinical applications

The physical characteristics of PCDs have been demonstrated predominantly in well-defined experimental setups. This raises the question of how these findings may impact future patient examinations. The following sections present an overview of the first results obtained by PCD CT prototypes. Where appropriate, the authors provide PCD CT images of an ongoing explorative study at their department (Study number DRKS00017759). Acquisition parameters of this study were adapted from high-resolution EID CT protocols used for temporal bone and inner ear imaging [25, 26]. As such, they are not directly transferable to clinical whole-body or low-dose examinations. A recently published study focusing on sinus and temporal bone examination utilized comparable doses for the EID CT acquisitions and demonstrated dose reductions of more than 60% with PCD [27].

Neuroradiology

The comparison of cranial EID and PCD CT scans of 21 human volunteers revealed that image noise was found to be 13–21% lower in PCD images at the same dose [28], which is in accordance with the results demonstrated in Fig. 2. Contrast between gray and white matter was 16% lower in the EID images compared to the PCD images, since EIDs favor high energy photons resulting in a reduction of image contrast.

Nasal sinus and temporal bone

To investigate the imaging performance for very fine anatomical structures, 10 cadaveric temporal bones were scanned using a PCD CT system in UHR mode at clinical dose levels [29]. The reconstructed images demonstrated superior delineation of temporal microstructures. The three readers, each with at least 10 years of professional experience, preferred these images over the corresponding EID CT images. In combination with a tin filter, PCD UHR acquisitions result in average dose reductions of up to 67% in sinus images obtained in actual patients [27].

Pulmonary and cardiac imaging

Low-dose CT of the thorax is the most important imaging modality in lung cancer screening programs [30]. As a consequence, dose-reduced protocols and additional findings in high-resolution images are of particular interest.

In the case of 20 synthetic pulmonary nodules of up to 12 mm in diameter, the shape and volume quantification as well as the CT number accuracy could be improved by the use of PCD UHR scanning [31]. Superiority to EID CT systems in the assessment of lung nodules and small pulmonary vessels was also demonstrated with a preclinical PCD CT system based on a cadmium zinc telluride (CdZnTe) detector [4]. In this study, in-vivo scans of a rabbit as well as a three-dimensional printed lung phantom with noduli of different sizes and shapes were assessed. For same dose levels, edges and boundaries of sphere-shaped nodules with spikes were more prominent in PCD CT images than in EID CT and high-resolution EID CT images (HR-CT). Improvements in CT-value accuracy and reproducibility of nodules could also be demonstrated in low dose protocols [32]. Furthermore, noise levels in lung parenchyma could be reduced by 11% and CNR in ground-glass nodules and emphysema was improved in PCD CT images, implying prospects to decrease radiation dose for lung cancer screening programs.

These results are consistent with first thorax PCD CT examinations in patients. In the assessment of thoracic PCD CT images of 30 asymptomatic patients, three experienced radiologists voted in favor of PCD CT images regarding lung nodule image quality and beam hardening artifacts [33]. In a group of 22 patients screened for pulmonary nodules, malignancy, pneumonia, and other pathologies, the detection rate for bronchi of higher order as well as the visualization of small bronchiolar walls could be increased by PCD CT using a 1024 × 1024 matrix [34]. This matrix was also used to visualize microcystic honeycombing in Fig. 6e.

Fig. 6
figure 6

Clinical energy-integrating detector (ac) and photon-counting detector computed tomography image (bde) in lung window (center [C] = −500 HU, width [W] = 1500 HU) and axial reconstruction demonstrating findings of non-specific interstitial pneumonia in a 71-year-old patient; 1: honeycombing, 2: traction bronchiectasis. For an accurate diagnosis, the exclusion of lung parenchyma in these spaces, which requires increased signal-to-noise levels, is of great importance. Acquisition and reconstruction parameters as follows: ac SOMATOM Definition AS, 10/2019, pixel size (0.7 mm)3, computed tomography dose index volume (CTDIvol) = 4.34 mGy. bde SOMATOM CounT acquisition in the context of an explorative study 6 days after ac CTDIvol = 24.3 mGy, bd pixel size 0.5 mm × 0.5 mm × 0.75 mm (transversal [tv] × anterior posterior [ap] × cranio-caudal [cc]), e pixel size 0.27 mm × 0.27 mm × 1 mm(tv × ap × cc); all images: I70f reconstruction kernel

First in-vivo experiments towards cardiac imaging were conducted in a canine model of myocardial infarction by injecting a gadolinium-based contrast agent followed by an iodinated contrast agent 10 min later [35]. Material decomposition of acquired PCD acquisitions allowed for the visualization of first-pass and late myocardial enhancement. Further studies in humans will need to demonstrate the feasibility of a combined first-pass and late-enhancement scan in a single acquisition.

Abdomen and retroperitoneum

Standard abdominal protocols are restricted to one contrast phase per acquisition. This clinically relevant restriction may be overcome by PCD CT and new contrast agents. A comparison of the performance of PCDs and EIDs for abdominal imaging was performed in 15 patients undergoing a contrast-enhanced examination using a SOMATOM CounT [36]. At 2 min after injection of an iodinated contrast agent, an EID scan and an additional 6 s later a dose matched PCD scan was performed using the same patient position. PCD images were comparable to EID images and provided additional multi-energy data. Another study using a canine model demonstrated that these multi-energy data can be used to generate material maps of gadolinium, iodine, and bismuth contrast agents in a single acquisition [23].

Since the location, size, morphology, and composition of kidney stones determine the treatment strategy [37], PCD CT might be able to provide all relevant radiological information to the physician and patient to reach a definitive treatment decision. Ex-vivo examination of 40 kidney stones showed that PCD CT can distinguish stones consisting of uric acid and cysteine, while calcium oxalate monohydrate and apatite could not be reliably distinguished [38]. Even though a direct comparison was not performed with these samples, studies using Dual Energy CT (DECT) resulted in similar conclusions [39, 40]. Additionally, PCD CT made it possible to automatically detect and characterize stones of less than 3 mm [41].

Crystal arthritis

In a preclinical small animal PCD CT system, a gouty finger and meniscus of the knee from a tissue bank were examined [42], and DECT reconstructions from a conventional clinical CT served as a reference. Both detectors succeeded in detecting monosodium urate crystals. The PCD CT, however, allowed for the delineation of finer structures of aggregated crystals than did conventional CT. In the case of the meniscus, aggregates composed of calcium pyrophosphate could be distinguished from hydroxyapatite crystal deposits, which was validated using polarized light microscopy and X‑ray diffraction.

Oncologic imaging of bone marrow

Although research in PCD CT acquisitions of primary bone tumors or metastases is still at a very early stage, first images show qualitative improvements in the visualization of osteoblastic metastases ([43]; Fig. 7). For multiple myeloma, early stages are correlated with a heterogeneous tumor infiltration of bone marrow, which, if located in cancellous bone, can only be detected by magnetic resonance imaging (MRI). For intermediate and advanced stages, however, PCD CT may provide additional information on fine structure, density, and fat content of lesions (Figs. 8 and 9).

Fig. 7
figure 7

Photon-counting detector (PCD) computed tomography (CT) image (ace) and clinical energy-integrating detector image (bd) in bone window (center [C] = 500 HU, width [W] = 1500 HU) and axial reconstruction demonstrating osseous metastases in a 68-year-old patient suffering from breast cancer. PCD CT images show improved visualization of the lesions’ content with sharper margins. bc Osteoplastic lesion in the right iliac bone; d, e lesion with sclerotic margins ventral of the first sacral foramen. Acquisition and reconstruction parameters as follows: ace SOMATOM CounT acquisition in the context of an explorative study, computed tomography dose index volume (CTDIvol) = 24.3 mGy, pixel size 0.27 mm × 0.27 mm × 1 mm (transversal [tv] × anterior posterior [ap] × cranio-caudal [cc]), U70f reconstruction kernel. bd SOMATOM Definition Flash, CTDIvol 7.28 mGy, pixel size 0.7 mm × 0.7 mm × 2 mm (tv × ap × cc), B70f reconstruction kernel

Fig. 8
figure 8

Multimodal imaging of the 5th lumbar vertebra in a 64-year-old patient suffering from multiple myeloma. Photon-counting detector (PCD) computed tomogrpahy (CT) images (a, b) demonstrate diffuse and focal lesions, which are only vaguely visualized in magnetic resonance imaging (MRI) and positron emission tomography (PET) images (e-h) and cannot be reliably identified in energy-integrating detector CT acquisitions at clinical dose levels and reconstruction parameters (cd). In MRI, diffuse T2w-hyperintense (e), T1w-hypointense (f) infiltrates with increased signal intensity in the diffusion-weighted imaging sequence (g) are visualized in accordance with an increased tracer uptake in the F‑18 fluorodeoxyglucose (FDG)-PET CT (h). CT images in bone (ac) and soft tissue window (bd), acquisition and reconstruction parameters as follows: ab SOMATOM CounT acquisition in the context of an explorative study, pixel size 1 mm × 0.54 mm × 0.54 mm (transversal [tv] × anterior posterior [ap] × cranio-caudal [cc]), I40f, computed tomography dose index volume (CTDIvol) = 24.3 mGy. cd SOMATOM Emotion, pixel size 1.36 mm × 1.36 mm × 5 mm (cc × tv × ap), B41s, CTDIvol = 4.08 mGy. ah In coronal reconstruction. cd Courtesy of Dr. Andreas Winterberg from Gemeinschaftspraxis Radiologie und Nuklearmedizin am Beethovenplatz, Saarbrücken. h Courtesy of Prof. Dr. Antonia Dimitrakopoulou-Strauss from the Clinical Cooperation Unit Nuclear Medicine at DKFZ, Heidelberg

Fig. 9
figure 9

Photon-counting detector (PCD) computed tomography (CT) images reconstructed with conventional sharp (a) and ultra-high resolution (UHR) U70f kernel (b) available for PCD CT, demonstrating the 5th lumbar vertebral body in a 59-year-old patient suffering from multiple myeloma. In the UHR image, the isolated trabeculae and compacta bone do not suffer from over-estimation in terms of CT values and size, enabling an improved assessment of bone marrow and tumorous tissue. Acquisition and reconstruction parameters as follows: SOMATOM CounT, computed tomography dose index volume 24.3 mGy, pixel size 0.27 mm × 0.27 mm × 1 mm (transversal × anterior posterior × cranio-caudal), bone window and axial reconstruction

Radiation therapy

In proton therapy, stopping power ratio (SPR) estimations are crucial for predicting the effectiveness of any treatment and quantifying the potential harm to surrounding tissues. In an experimental setup of nine organic tissue samples, PCD CT showed equal or better results for SPR estimations compared to single energy CT or DECT [44]. Preliminary results in image-guided radiation therapy indicate that PCDs are capable of accurately tracking the motion of lung tumors with only 50 photons per acquisition, if an initial four-dimensional CT scan is provided beforehand [45].

Reduction of metal artifacts

Photons of high energy are more likely to penetrate high density materials compared to low energy photons. Consequently, high energy bin data in PCD CT shows a reduction in beam hardening artifacts and calcium blooming ([46,47,48,49]; Fig. 10). In an assessment of 18 different coronary stent types filled with diluted contrast agent, PCD CT images demonstrated a 37% lower increase in lumen attenuation, potentially improving assessment of stent occlusions [49].

Fig. 10
figure 10

Post-mortem energy-integrating detector (EID) computed tomography (CT) (ab) and photon-counting detector (PCD) CT (cd) acquisitions, reconstructed with a sharp kernel, bone window, computed tomography dose index volume = 34.3 mGy for EID CT and 23.2 mGy for PCD CT. a Widespread metal artifacts from a total endoprosthesis (TEP) of the left hip affecting the assessment of the adductor muscles (1). b Blooming surrounding the TEP makes it difficult to detect signs of loosening in EID images (2). cd In the high threshold bin of PCD CT data, these two sorts of artifacts are substantially reduced

Outlook

PCDs for CT have demonstrated superior capabilities when compared to state-of-the-art EIDs. Examples include, but are not limited to, reduced noise, increased spatial resolution, and the acquisition of multi-energy data. These advantages will affect future clinical CT protocols. Contrast-enhanced examinations will benefit from a reduction in administered contrast agent volume and higher image contrast.