日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
臨床技術
膝下三分枝を対象とした非造影 MRA における臨床に向けたエコートレイン数の最適化
藤本 綾子青羽 南臣町田 好男
著者情報
ジャーナル フリー

2020 年 76 巻 7 号 p. 689-698

詳細
抄録

Purpose: Non-contrast magnetic resonanse angiography (MRA) using the three-dimensional electrocardiogram-synchronized fast spin echo method uses systolic and diastolic arterial signal differences. The method relies on the flow void signal of the arterial flow because of dephasing during systole. However, depiction of slow flow such as that in a calf artery was degraded because of insufficient dephasing during systole. In this study, we optimized echo train length (ETL) using a flow phantom and normal volunteers for clinical examination of the calf arteries. Methods: Flow phantom and normal volunteer images were obtained with various ETLs (40, 50, 60, and 70). An averaged profile across the tube in the phantom was used for detailed investigation of flow dephasing. Visual evaluation was performed and signal intensity change along vessels was measured using normal volunteer images. Comparison with peak systolic velocity (PSV) measured using ultrasound equipment was also conducted. Results: Results of the flow phantom and normal volunteer study indicated that the overall depictability was improved with ETL 60 and 70, which was higher than the standard value. Additionally, the visualization of the peroneal artery with low PSV of ETL 70 had better depictability than ETL 60. Conclusion: This study suggested that ETL 70 might be better for clinical examination of the calf arteries.

著者関連情報
© 2020 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback