CT 画像の低コントラスト検出能に対する信号雑音比に基づく新たな指標の提案
橋本 純一阿部 慎司石森 佳幸門間 正彦圓谷 明男宮内 兼義
著者情報
ジャーナル フリー

2017 年 73 巻 7 号 p. 537-547

詳細
抄録

The low-contrast detectability of computed tomography (CT) images is commonly evaluated by the contrast-to-noise ratio (CNR) because of its convenience to measure. However, the correlation between CNR and visual detectability is poor because the CNR is a simple index determined by both the contrast of the object and the standard deviation of the image noise. On the other hand, the signal-to-noise ratio (SNR), especially SNR based on the statistical decision theory model (SNRS, D) and SNR based on the matched-filter model (SNRM) are considered superior to CNR. In this study, we investigated a new physical image quality index for evaluating low-contrast detectability (SNRA), which is approximately derived from SNRS, D and SNRM. The new index, which was calculated using the object size, contrast of the object and the noise power spectrum, provided good approximations when the diameter of the rod object was equal and >5 mm. The diameter dependency of the SNRA was also found to provide better sensitivity than the sensitivities of CNR and object-specific CNR, similar to SNRS, D and SNRM. The results suggested that the proposed convenient index should be useful for evaluating the low-contrast detectability of CT images.