Workloads of radiological technologists under different conditions of heights of radiographic table and/or X-ray tube assembly were calculated using a software for preventing musculoskeletal complaint to investigate optimal working environment for general X-ray examinations. In the patient positioning, compressive force of lumbar disc decreased at higher radiographic table within the range of 45-90 cm. On the other hand, workload of the shoulder joint increased with increase in the height of radiographic table. Load of the shoulder joint similarly increased as the height of the X-ray tube assembly increased. Compressive force of lumbar disc reduced by approximately 10-30% as the height ratio of the radiographic table to body height increased by approximately 40%, compared to the lowest table of 45 cm. Muscle load of a 50-years-old woman was approximately double compared to a 30-year-old man, even in the same workload. It is important to keep suitable height of radiographic table for reduction of the workloads of lumbar rather than shoulder joint, because floating-type radiographic table is generally used.