3 T MRIにおける3D Turbo SE法を使用した下垂体ダイナミック撮像法の検討
小川 匡史松村 善雄松本 剛土橋 俊男椿山 信吾
著者情報
ジャーナル フリー

2012 年 68 巻 3 号 p. 231-239

詳細
抄録

Dynamic magnetic resonance (MR) imaging for pituitary microadenomas is usually performed in 2-dimensional (2D) multi-slice method which used coronal T1-weighted imaging with turbo spin echo (SE) method. However, on MR images using 2D multi-slice method, the detectability of small lesions between slices may decrease. Therefore, the aim of our study is to investigate the influence that imaging parameters give to T1-weighted image with 3-dimensional (3D) turbo SE method, and to examine the use of 3D turbo SE method as the detection of pituitary microadenomas. We can plan the shortening of imaging time by shortening repetition time (TR), because the contrast to noise ratio (CNR) in the 3D turbo SE method was superior enough than that of the 2D turbo SE method. In addition, low refocusing flip angle induced the decrease of CNR, but it has the effect which decreases flow-induced artifacts. Dynamic MR imaging which used coronal T1-weighted imaging with 3D turbo SE method is feasible by utilizing the reduction of TR and low refocusing flip angle, as well as the combination of parallel imaging and radial sampling.