心房細動症例における320列Area Detector Computed Tomography(ADCT)を用いた冠動脈描出能の検討
高木 雅悠兼子 武士加藤 亮平安野 泰史吉見 聡井田 義宏三田 祥寛片田 和広
著者情報
ジャーナル フリー

2011 年 67 巻 4 号 p. 321-327

詳細
抄録

The aim of this study was to evaluate 320-row area detector CT (ADCT) for patients with atrial fibrillation (Af) based on simulated exposure using electrocardiogram RR intervals and comparison with the findings of coronary CT angiography (CCTA) using 64-row multi slice CT (MSCT). The probability of including RR intervals of 900 ms or more was calculated when the acquisition time was varied from 1 to 4 beats. Overall, 51 patients with Af who underwent CCTA were examined. The exposure time for CCTA, the total dose length product (DLP) for the examination, and the image quality (scored 0 to 3: poor to excellent) were compared between ADCT and MSCT. The probability of including RR intervals of 900 ms or more was highly significantly increased at 3 beats of acquisition time. The exposure time using ADCT was reduced by 75% compared with MSCT (ADCT/MSCT: 2.8/11.3 s), and the total DLP was reduced by 40% (ADCT/MSCT: 1398/2277 mGy·cm). Moreover, ADCT provided diagnosable images in all cases, and the mean image quality score for ADCT was significantly higher than that for MSCT (ADCT/MSCT: 2.8/2.4). Thus, 320-row ADCT at 3 beats of acquisition time can provide CCTA images of acceptable quality for patients with Af.