Review Open Access
Copyright ©The Author(s) 2015. Published by Baishideng Publishing Group Inc. All rights reserved.
World J Hepatol. Jun 8, 2015; 7(10): 1302-1311
Published online Jun 8, 2015. doi: 10.4254/wjh.v7.i10.1302
Hemodynamic monitoring during liver transplantation: A state of the art review
Mona Rezai Rudnick, Lorenzo De Marchi, Jeffrey S Plotkin, Department of Anesthesiology, Georgetown University Hospital, NW Washington, DC 20007, United States
Author contributions: De Marchi L and Plotkin JS designed research; Rudnick MR performed research; Rudnick MR and Plotkin JS analyzed data; Rudnick MR wrote the paper.
Conflict-of-interest: The authors have no conflict of interest to report.
Open-Access: This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Correspondence to: Jeffrey S Plotkin, MD, Department of Anesthesiology, Georgetown University Hospital, CCC Building Lower Level Rm CL-60, 3800 Reservoir Road, NW Washington, DC 20007, United States. plotkinj@gunet.georgetown.edu
Telephone: +1-202-4448640 Fax: +1-202-4448854
Received: September 19, 2014
Peer-review started: September 20, 2014
First decision: December 17, 2014
Revised: March 25, 2015
Accepted: April 8, 2015
Article in press: April 9, 2015
Published online: June 8, 2015

Abstract

Orthotopic liver transplantation can be marked by significant hemodynamic instability requiring the use of a variety of hemodynamic monitors to aide in intraoperative management. Invasive blood pressure monitoring is essential, but the accuracy of peripheral readings in comparison to central measurements has been questioned. When discrepancies exist, central mean arterial pressure, usually measured at the femoral artery, is considered more indicative of adequate perfusion than those measured peripherally. The traditional pulmonary artery catheter is less frequently used due to its invasive nature and known limitations in measuring preload but still plays an important role in measuring cardiac output (CO) when required and in the management of portopulmonary hypertension. Pulse wave analysis is a newer technology that uses computer algorithms to calculate CO, stroke volume variation (SVV) and pulse pressure variation (PPV). Although SVV and PPV have been found to be accurate predicators of fluid responsiveness, CO measurements are not reliable during liver transplantation. Transesophageal echocardiography is finding an increasing role in the real-time monitoring of preload status, cardiac contractility and the diagnosis of a variety of pathologies. It is limited by the expertise required, limited transgastric views during key portions of the operation, the potential for esophageal varix rupture and difficulty in obtaining quantitative measures of CO in the absence of tricuspid regurgitation.

Key Words: Intraoperative monitoring, Physiologic monitoring, Liver transplantation

Core tip: Accurate hemodynamic monitoring is essential to safely navigate orthotopic liver transplantation. Although specific indications for pulmonary artery catheters exist, their use has slowly been replaced by newer technologies which offer less invasive and more accurate measurement. The latest evidence on the strengths and limitations of arterial pulse wave form analysis, intraoperative transesophageal echocardiography, peripheral vs central arterial blood pressure monitoring and pulmonary arterial catheters are discussed.



INTRODUCTION

Orthotopic liver transplantation (OLT) has been performed for the past three decades with significant improvement in patient and graft survival. Despite improvements in the anesthetic and surgical techniques, it still is a challenging procedure, requiring dedicated, specifically trained providers and a collection of monitors not common to other operations.

Most classic hemodynamic monitors like radial and femoral arterial lines and a pulmonary artery catheter (PAC) are still part of the protocol at many institutions[1], but new technology has been emerging. These new devices and techniques along with evidence of the limitations of some of the classic monitors, are reshaping the way in which hemodynamics are monitored during anesthesia for liver transplantation in the 21st century.

HEMODYNAMICS DURING LIVER TRANSPLANTATION

Liver transplantation can be thought of as having 3 distinct stages: the dissection or pre-anhepatic phase, the anhepatic phase, and the neohepatic phase. Each stage has its own hemodynamic concerns.

The pre-anhepatic phase is when all the dissection occurs, and is marked by significant fluid shifts from drainage of ascites to the potential for significant blood loss in the presence of varices from portal hypertension. Additionally, manipulation of the liver and downward retraction of the inferior vena cava may intermittently obstruct venous return causing hemodynamically significant changes in preload[2].

The anhepatic stage is defined as the cessation of blood flow to the native liver until the time of reperfusion of the transplanted liver. With cross clamping of the portal vein and IVC, cardiac output (CO) may decrease by up to 50%[3]. To avoid this sudden loss of preload, volume loading should occur prior to crossclamping. An alternative is use of the “piggyback” technique by the surgeons where the inferior vena cava is only partially occluded. Other alternatives include the use of a temporary portocaval shunt or venovenous bypass. Some centers make use of one of these techniques routinely while others employ them as clinically indicated[4].

The neohepatic stage is defined as the beginning of reperfusion until the end of the case. Reperfusion is often marked by significant hemodynamic instability due to the rapid return of blood from the previously obstructed portal system and newly transplanted liver. This blood tends to be acidotic, hyperkalemic, cool, and contains a variety of inflammatory and vasodilatory mediators[3]. The result is often a transient but significant decrease in myocardial contractility, chronotropy and systemic vascular resistance[5]. Postreperfusion syndrome, defined as a decrease in mean arterial pressure by 30% for at least 1 min within 5 min of reperfusion, has been reported to occur in 12.1%-42% of patients[6,7]. After overcoming the instability of reperfusion, the remainder of the neohepatic stage tends to have relatively stable hemodynamics.

BLOOD PRESSURE

Invasive blood pressure monitoring is the standard of practice during liver transplantation. The number and location of these lines varies by center[1]. In healthy individuals, radial artery pressures have a higher systolic pressure as compared to femoral or more central pressures. This difference has been attributed to pulse amplification as a result of the impedance and harmonic resonance of the vasculature. However, the central and distal mean arterial pressures remain relatively unchanged[7,8].

Specific circumstances can create significant discrepancies between central and radial mean arterial pressures. When measuring radial arterial pressure, the presence of a proximal obstructive vascular lesions, rewarming after hypothermic cardiopulmonary bypass[9,10], and high dose vasopressor therapy in critically ill patients[7] are all known to underestimate central pressures. The cause for this discrepancy in obstructive vascular lesions is self-evident, but controversy exists regarding the etiology in cardiopulmonary bypass or vasopressor therapy. In the latter, it has been theorized vasoconstriction of the extremity conductance vessels, which are more sensitive to vasopressors than the central vasculature, significantly reduces the flow to the radial artery[10]. Regarding the use of cardiopulmonary bypass, it is theorized to cause an extreme vasodilatory state leading to proximal shunting, possibly in the splanchnic beds, in combination with distal vasoconstriction which both contribute to lower peripheral pressures[11].

Similarly, it has been suspected in OLT that radial arterial and more centrally measured pressures may not correlate well[12]. Our observations and demonstrated in unpublished data, over many years, have shown a consistent decrease in systolic, diastolic and mean pressures in radial vs femoral arterial pressures, most pronounced immediately after reperfusion. The theory behind this, similar to post-cardiopulmonary bypass physiology, is extreme peripheral vasodilation, seen especially during reperfusion, decreases distal pressures disproportionately to central pressure[13,14]. Studies attempting to demonstrate this effect have had conflicting results. Acosta et al[15] found no difference in mean arterial pressure at any point during OLT, while Arnal et al[16] found systolic but not mean arterial pressures differed during reperfusion which was exaggerated in patients receiving vasopressors. A study looking at pediatric OLT, on the other hand, did show a discrepancy is both systolic and mean arterial pressures when comparing femoral to radial pressures throughout most of the operation[17]. Interestingly, a separate study showed noninvasive blood pressure measurements of the upper extremities more closely correlated with femoral artery pressures than radial pressures 3-10 min after reperfusion. Presumably this is due to the more proximal location of the blood pressure cuff[18].

Larger trials are required to definitively determine the reliability of radial pressure monitoring during OLT, but currently the literature suggests central monitoring from femoral or brachial locations and mean arterial pressures should be followed over peripheral radial systolic pressures.

PAC

According to the Frank-Starling law, as end-diastolic volumes increase, myocardial fiber length increases which in turn increases the number of myosin-actin connections resulting in increased CO. Classically, central venous and pulmonary artery occlusion pressures (PAOPs) have been used as surrogates for volume measurements. However, numerous studies have established that these static preload measurements are poor predictors of end diastolic volume and fluid responsiveness in a wide variety of medical and surgical patients[19-26]. The compliance of the heart and vasculature, intrathoracic pressures, cardiac contractility and valvular pathologies significantly affect the pressure measured for a given preload making static pressure measurements an unreliable indicator of end diastolic volume[27,28].

Specifically during liver transplant, Costa et al[29] found stroke volume index did not correlate with either central venous pressure (CVP) or PAOP. Similarly, Rocca et al[28] looked at pre-anhepatic dissection, the anhepatic phase, and after reperfusion and found no correlation between cardiac index and CVP or PAOP during liver transplantation. A separate study by Rocca et al[30] again found in 244 patients undergoing liver transplantation that CVP and PAOP correlate poorly with stroke volume index.

PACs also allow for the measurement of CO via intermittent thermodilution. Although often used as the gold standard for CO measurements, its accuracy depends on several user-dependent techniques such as the speed, volume and temperature of the injectate as well as its timing with respect to the respiratory cycle[27]. Significant tricuspid regurgitation or intracardiac shunts also limit the accuracy of thermodilution[31,32]. Additionally, during liver transplantation, the rapid return of cooled blood during reperfusion and administration of large volumes of intravenous fluids generate thermal noise which may result in an underestimation of the true CO[33,34].

A new generation of PACs allow for continuous CO monitoring. This technology uses heat instead of cold thermodilution via a thermal filament connected to a specialized PAC and a distal thermistor. CO values are then continuously calculated. Several studies have shown continuous CO measurements correlate well with intermittent thermodilution in a variety of patient populations[35-37]. Although also seen to be accurate during liver transplantation[33], it has similar limitations to intermittent thermodilution with less accuracy during reperfusion and cross-clamping[34].

Mixed venous oxygen saturation (SvO2) can be measured via a PAC and used as an indirect measure of CO. However, changes in SvO2 are not very specific to CO and may be the result of changes in oxygen content, oxygen consumption or, in the case of reperfusion, return of desaturated blood[38]. Specifically, in liver transplantation SvO2 has shown to poorly correlate with CO[39].

Perhaps the most important use of PAC during liver transplantation is in patients with portopulmonary hypertension. Patients with mean PA pressures above 50 have typically been denied liver transplantation due to an unacceptably high risk of mortality ranging from 71%-100%[40,41]. While mild (mean PA pressures 25-35 mmHg) and moderate (mean PA pressures 35-45) pulmonary hypertension are not strict contraindications for liver transplantation[42,43], these patients still do have an elevated perioperative mortality rate as high as 33%-35%[40,41]. When pulmonary artery pressures are responsive to treatment, the mortality risk significantly decreases[44-46]. Preoperative workup demonstrating increased PA pressures on echocardiography suggest portopulmonary hypertension. Additionally, hypoxemia or exaggerated respiratory alkalosis also may suggest increased PA pressures[47]. However, a PAC is the only modality that directly measures pulmonary arterial pressures. At the start of the procedure, it may unveil worse pulmonary hypertension than suspected during preoperative evaluation leading to cancellation of the procedure. Intraoperatively, the sudden volume shifts and release of vasoactive mediators seen, especially during reperfusion, may result in significant right heart strain and failure[48]. Treatment of elevated PA pressures during OLT includes the use of venovenous bypass during the anhepatic phase, phosophodiesterase-5 inhibitors, endothelin receptor antagonists, and prostacyclins[48-50].

The use of PACs does have significant diagnostic limitations however they are still one of the most accurate tools to assess CO and an essential monitor in patients with significant pulmonary hypertension.

CO WAVEFORM ANALYSIS

Beginning in 1904, Joseph Erlanger and Donald R Hooker theorized that CO could be derived from the arterial pulse pressure[51]. Logically, as more blood is ejected from the heart, there should be a greater pressure transmitted to the arterial tree. This correlation is limited, however, due the vascular resistance which determines the runoff of blood from the conductance vessels into the arterioles. Even more complex is the nonlinear compliance of the arterial tree where a given volume of blood will decrease compliance more at higher pressures as compared to lower pressures[52]. Only recently has the technology been available to not only precisely analyze pressure waveforms but also the sufficient knowledge to create algorithms which account for the complex physiology of pulse wave morphology.

There are currently several commercially available products which calculate continuous CO from arterial waveform analysis. The PiCCO system (Pulsion Medical System, Munich, Germany) requires the placement of a thermodilution catheter into the axillary or femoral artery. A solution is then injected into any central venous catheter and CO is then determined by the arterial temperature probe. After this initial calibration, the system then continuously calculates CO based on arterial waveform analysis. Additionally, static preload indices such as global end diastolic volume and intrathoracic blood volume can be calculated. The LiDCO system (LiDCO Plus, Cambridge United Kingdom) is similar to the PiCCO system but uses lithium indicator dilution rather than thermodilution to calibrate the pulse wave form to the CO[53]. The Flowtrac/Vigileo system (Edwards Lifesciences, Irvine, CA United States) uniquely does not require intermittent CO bolus calibration. It accounts for changes in arterial compliance and resistance using a conversion factor Khi which factors in the standard deviation of the mean arterial pressure and the analysis of the skewness and kurtosis of the arterial waveform. Large vessel compliance is then estimated using patient demographics of age, gender and body surface area[54].

In patients undergoing liver transplant, PiCCO derived CO measurements were found to agree with the gold standard of PAC thermodilution CO[55]. Similarly Nissen et al[56] looking at dissection, anhepatic, early and late reperfusion phases during OLT also found arterial pulse wave CO measurements to correlate with thermodilution CO.

Despite a few reassuring studies, significant concerns have been raised concerning the validity of arterial waveform analysis in particular patient populations whose physiology may not be well represented by the algorithms. CO by waveform analysis was found to correlated poorly with PAC thermodilution CO in patients on a significant dose of vasopressor[57,58]. There was also poor correlation during cardiac surgery[59] and abdominal aortic aneurysm operations[60] with the uncalibrated FlowTrac system performing worse than the PiCCO system.

The hyperdynamic circulation in liver cirrhosis is characterized by low systemic vascular resistance, elevated CO and possible underlying cardiomyopathy. The transplant operation then adds large changes to preload and afterload, vasoactive mediators during reperfusion, myocardial contractility changes and the possibility of significant hemorrhage[14]. Likely as a result of this altered physiology and dynamic intraoperative changes, several studies have found poor correlation between waveform analysis CO calculations when compared to the PAC thermodilution. Krejci et al[59] found, in Child-Pugh class B and C cirrhotics, both FlowTrac and LiDCO systems correlated poorly with PAC thermodilution. Two separate studies found that in Child-Pugh B and C cirrhosis, the degree of FlowTrac’s inaccuracy was proportional to the patient’s SVR with lower resistances showing less correlation to reference thermodilution values[61,62]. Della Rocca et al[63] examined the effect of the high output cardiac state and found Flowtrac underestimated CO in liver transplant patients whose CO exceeded 8 L/min.

A recent software update to FlowTrac has been released (third generation) whose aim was to improve CO accuracy specifically in low systemic vascular resistance states. Some improvements have been seen in septic patients[64] and those undergoing cardiac surgery[65]. Despite these improvements, in liver transplantion the accuracy of FlowTrac CO measurements remain unreliable. Tsai et al[66] found a 55% discrepancy between the third generation software of FlowTrac COs as compared to PAC thermodilution. Likewise, Su et al[67] found a percentage error of 75% which was inversely related to the patient’s systemic vascular resistance index. CO derived from waveform analysis depends on the intrarterial peripheral catheter reflecting systemic conditions. CO in cirrhosis and during liver transplantation, however, is not evenly distributed[29]. Peripheral arterial waveform analysis, therefore, cannot be recommended for liver transplant intraoperative monitoring or in Child-Pugh class B or C patients.

With the currently available technology, arterial waveform analysis cannot reliably measure CO during OLT.

STROKE VOLUME AND PULSE PRESSURE VARIATION

Although CO by waveform analysis has proven unreliable in liver transplant, the technology also allows for the continuous measurements of stroke volume variation (SVV) and pulse pressure variation (PPV).

Pulse pressure changes proportionally to left ventricular stroke volume[25]. During positive pressure ventilation, blood return to the right heart decreases. After this lower volume of blood passes through the pulmonary vasculature, left ventricular end diastolic volume decreases. The overall result is lower stroke volumes and a smaller pulse pressure after positive pressure ventilation. The magnitude of this difference is proportional to preload. Patients whose CO would be supplemented by increased intravascular volume will have a larger difference in pulse pressure during positive pressure ventilation and exhalation. A review of twenty-nine studies studying this phenomenon found SVV less than 11.6% ± 1.9% and a PPV less than 12.5% ± 1.6% predicted volume responsiveness in critically ill patients[26].

Arterial waveform analysis of the SVV has been shown to predict fluid responsiveness during general anesthesia[68] and the use of vasoconstrictors does not change the variation[69]. In liver transplantation, poor SVV was found to be a better predictor of right ventricular end diastolic volume index than CVP[70]. Kim et al[71] confirmed these findings and found a PPV of greater than 9% predictive of lower RVEDVI which would likely be fluid responsive. Biais et al[72] found, in liver transplant, a SVV of 10% discriminated fluid responders from non-responders with a 93% sensitivity and 94% specificity.

The accuracy of SVV or PPV in predicting fluid responsiveness depends on the patient meeting a specific set of criteria. Stroke volume is less predictably dependent on preload during spontaneous breathing. Breathing effort causes irregularity in both pulse rate and intrathoracic pressures[73]. Not only must the patient be mechanically ventilated with no breathing efforts, but the pressure must be adequate to decrease preload. The degree of SVV is linearly related to tidal volume with tidal volumes less than 10 mL/kg showing a lesser degree of SVV[74]. Likewise, when driving pressures (defined as the difference between the plateau and positive end expiratory pressure) are less than 20 cm H2O, a PPV less than 13% does not rule out fluid responsiveness[75]. De Backer et al[76] found that PPV only reliably predicted fluid responsiveness at tidal volumes above 8 mL/kg. Finally, during cardiac arrhythmias, SVV is more dependent on the irregular diastolic time than on intravascular volume[77].

Although waveform analysis has limited utility in quantitative measurements of CO in cirrhosis and liver transplantation, it is a minimally invasive option of monitoring end diastolic volume and fluid responsiveness.

TRANSESOPHAGEAL ECHOCARDIOGRAPHY

Echocardiography has a variety of intraoperative uses. In experienced hands, it has the capability of diagnosing right ventricle (RV) or left ventricle (LV) systolic or diastolic dysfunction, volume overload, global or regional wall abnormalities, and intracardiac air or thrombosis[78-80]. In liver cirrhosis, numerous case reports have reported its usefulness in diagnosing portopulmonary hypertension[81], ischemic heart disease, cirrhotic cardiomyopathy[82], and intraoperative thromboembolic events[83]. More rare conditions such as pericardial tamponade[84], cardiomyopathy secondary to undiagnosed hemochromatosis[85], and Takosubo cardiomyopathy[86,87] have also been reported during OLT intraoperative transesophageal echocardiography (TEE).

One of the greatest strengths of TEE is its ability to directly visualize in real-time the preload of both the right and left sides of the heart. As previously discussed, the PAOP is a known poor measure of LV preload. The PAOP may differ dramatically despite no change in left end-diastolic volume (LVEDV) and fails to detect hypovolemia when compared to direct measurement of LVEDV by TEE[88]. The LVEDV can be directly visualized from the transgastric mid-short axis view (TG mid-SAX). Additionally, TEE has been shown to accurately determine stroke volume and left ventricular changes[88]. In a multicenter study of 244 patients undergoing OLT, stroke volume index was found to be more strongly correlated with right ventricular end diastolic volume index than either CVP or PAOP[30].

Similarly, CVP can be an unreliable indicator of stroke volume and intravascular volume[20,23,89]. During increases in intravascular volume, the RV dilates which results in no significant change in the CVP despite large increases in volumes[90]. The resulting fluid overload can result in congestion and injury to the newly transplanted liver. SVV can indicate the need for additional fluid, but depends on a regular heart rate and adequate tidal volumes without respiratory effort. These are not limitations for TEE as it directly measures end diastolic volume. Determination of preload by TEE has significant clinical consequence. Hofer et al[82] found intraoperative use of TEE influenced fluid therapy in 50% of OLT patients.

Hypotension in the presence of adequate preload may be a result of myocardial dysfunction. Coronary artery disease (CAD) is not uncommonly encountered in patients undergoing OLT, with one study reporting up to 32% of patients over the age of 50 having moderate to severe disease[91]. Patients with known CAD undergoing OLT have a mortality of 50%[92]. Due to this high mortality, reversible ischemia seen on stress testing is a contraindication to proceeding with OLT[93]. Intraoperative TEE allows for the real time detection of myocardial ischemia manifested by regional wall motional abnormalities. A study by Smith et al[94] found intraoperative TEE was more sensitive in detecting myocardial ischemia than EKG changes. A separate study showed TEE may be more sensitive than conventional monitors as 73% of patients with regional wall abnormalities had no detectable change in heart rate, blood pressure or PA pressures[95].

Pulmonary embolization can be a real and significant risk during OLT[96]. Paradoxical embolization to the systemic circulation may result in stroke and can occur as a result of a patent foramen ovale. Cirrhotic patients are at particular risk for paradoxical embolization as dilated pulmonary vasculature can allow the free passage of emboli to the systemic circulation. TEE can not only identify patients with right to left shunts but also distinguish between intracardiac and transpulmonary etiologies, thereby identifying those patients who are at higher risk for paradoxical embolism[97]. Furthermore, TEE Doppler is the most sensitive monitor for the detection of air embolism, able to detect 0.1 mL/kg of air[98]. The diagnosis of an acute pulmonary embolism, however, is more difficult with TEE, typically manifesting with signs of RV dysfunction such as RV dilation, hypokinesis and possible pulmonary hypertension[99]. However, a large burden of pulmonary embolism is required to see these effects, especially those over 30% of the pulmonary artery area, are more likely to show RV dysfunction[100]. The classic “McConnell sign” refers to hypokinesis of the RV with preservation of RV apical contractility[101]. This sign has been reported to be very specific for acute pulmonary embolism but with a sensitivity of only 19%[102,103].

Despite the many advantages of TEE, like every monitor, there are limitations and risks. The proficient use of TEE requires significant training and expertise with the American Society of Echocardiography recommending 300 transthoracic echocardiograms, 20 esophageal intubations and 50 transesophageal examinations within a 6 mo period[103]. However, the performance of “limited-scope examinations” by non-credentialed anesthesiologists is not uncommon with 88% of users lacking echocardiography certification[104].

Additionally, the presence of esophageal varices creates concern for rupture. However, while grade four esophageal varices may be a true contraindication[27], TEE has been safely performed in patients with grade I and II varices[105]. TEE is also limited in its ability to assess pulmonary artery pressures in the absence of tricuspid regurgitation with the far majority of centers using a PAC with or without the use of TEE[1]. As previously discussed, TEE is a very sensitive monitor for ischemia, however, the transgastric short axis view which best assesses the circumference of the left ventricle is largely unavailable during the operation due to posterior retraction of the stomach[27].

CONCLUSION

A variety of hemodynamic monitors are an essential part of the successful intraoperative management of patients undergoing OLT but each has their own indications and limitations (Table 1).

Table 1 Comparative summary table of hemodynamic monitors in orthotopic liver transplantation.
MonitorBenefits and usesLimitations
Invasive blood pressure monitoringBeat-to-beat monitoring of blood pressurePeripheral arteries possibly underestimate the central mean arterial pressure especially during reperfusion or use or high dose vasopressors
Pulmonary artery catheterAccurately determines cardiac output via intermittent thermodilutionInvasive
Directly measures PA pressuresStatic pressure measurements are imperfect indicators of fluid status or stroke volume
Arterial pulse wave analysis - COLess invasive option to calculate CODoes not reliably calculate CO in advanced cirrhosis or during OLT
Arterial pulse wave analysis - SVVPredicts fluid responsiveness in OLT populationRequires sinus rhythm
Requires patient does not make any spontaneous respiratory efforts
Most accurate during tidal volumes of 8-12 mL/kg
Transesophageal echocardiographyDirect assessment of cardiac fillingRequires advanced training
Monitors myocardial ischemia and strainLimited views intraoperatively
Potentially can diagnose pulmonary embolisms, shunts, effusions, and valvular pathologiesRisk of esophageal varix rupture or esophageal injury

Invasive blood pressure monitors is currently the standard of care during transplantation, however, the evidence suggests that peripheral measurements are possibly not representative of central perfusion pressures especially in instances of significant vasopressor use or in patients with unequal vasodilation as is the case in significant cirrhosis or during reperfusion. When possible, more central invasive monitors at the femoral or brachial artery have a theoretical advantage of representing central perfusion pressures.

PACs continue to have controversial indications in and outside the operating room. The available evidence is clear that static cardiac pressure measurements such as CVP and PAOP are imperfect predictors of fluid responsiveness and CO. However, CVP allows for the monitoring of the backpressure of the IVC into the newly transplanted liver and may still guide the transplant anesthesiologist in fluid management or need for vasodilators to prevent injury to the liver. When using a combination of data such as the heart rate, blood pressure, CVP, urine output, a more clearly picture of the patient’s hemodynamic status may emerge.

A role still exists for PACs for the accurate measurements of CO via intermittent or, more recently, continuous thermodilution which remains the current gold standard. In OLT, PACs also play an important role in the monitoring of patients with PA pressures in patients with pulmonary hypertension which, even in the setting of mild hypertension, carries a significant risk of morbidity and mortality.

Pulse pressure analysis is a newer monitoring technique. Although the promise of accurate and continuous CO analysis has not been delivered in the OLT patient population, continuous PPV monitoring does appear to predict fluid responsiveness and may serve as an invaluable guide.

TEE is the most direct measurement of cardiac filling that currently exists allowing for the real-time assessment of fluid status during these dynamic operations. Additionally, it also offers the unique benefit of diagnosing a variety of other intraoperative complications such as myocardial ischemia, pulmonary embolism, and pulmonary or pericardial effusions. It currently is limited by the expertise required to interpret the images, but as more anesthesiologists are trained in this technology it stands to supplant many of the indirect monitors currently in use.

The perfect hemodynamic monitor would be noninvasive, precise and accurate, and provide continuous data at all stages of transplantation. Until this device exists, adept intraoperative management requires knowledge of the applicability and known limitations of available technology. Perhaps the current best monitor is the experienced provider who can adeptly integrate the various pieces into a complete but adaptable perioperative treatment plan.

Footnotes

P- Reviewer: Shehata MMM, Shi ZJ S- Editor: Tian YL L- Editor: A E- Editor: Liu SQ

References
1.  Schumann R, Mandell MS, Mercaldo N, Michaels D, Robertson A, Banerjee A, Pai R, Klinck J, Pandharipande P, Walia A. Anesthesia for liver transplantation in United States academic centers: intraoperative practice. J Clin Anesth. 2013;25:542-550.  [PubMed]  [DOI]  [Cited in This Article: ]
2.  Robertson AC, Eagle SS. Transesophageal echocardiography during orthotopic liver transplantation: maximizing information without the distraction. J Cardiothorac Vasc Anesth. 2014;28:141-154.  [PubMed]  [DOI]  [Cited in This Article: ]
3.  Ozier Y, Klinck JR. Anesthetic management of hepatic transplantation. Curr Opin Anaesthesiol. 2008;21:391-400.  [PubMed]  [DOI]  [Cited in This Article: ]
4.  Schumann R. Intraoperative resource utilization in anesthesia for liver transplantation in the United States: a survey. Anesth Analg. 2003;97:21-28, table of contents.  [PubMed]  [DOI]  [Cited in This Article: ]
5.  Bukowicka B, Akar RA, Olszewska A, Smoter P, Krawczyk M. The occurrence of postreperfusion syndrome in orthotopic liver transplantation and its significance in terms of complications and short-term survival. Ann Transplant. 2011;16:26-30.  [PubMed]  [DOI]  [Cited in This Article: ]
6.  Fung SK, Hui TW, Wong AK, Lei GM. Anaesthesia for liver transplantation: experience at a teaching hospital. Hong Kong Med J. 1999;5:27-33.  [PubMed]  [DOI]  [Cited in This Article: ]
7.  Dorman T, Breslow MJ, Lipsett PA, Rosenberg JM, Balser JR, Almog Y, Rosenfeld BA. Radial artery pressure monitoring underestimates central arterial pressure during vasopressor therapy in critically ill surgical patients. Crit Care Med. 1998;26:1646-1649.  [PubMed]  [DOI]  [Cited in This Article: ]
8.  Remington JW, Wood EH. Formation of peripheral pulse contour in man. J Appl Physiol. 1956;9:433-442.  [PubMed]  [DOI]  [Cited in This Article: ]
9.  Stern DH, Gerson JI, Allen FB, Parker FB. Can we trust the direct radial artery pressure immediately following cardiopulmonary bypass? Anesthesiology. 1985;62:557-561.  [PubMed]  [DOI]  [Cited in This Article: ]
10.  Pauca AL, Hudspeth AS, Wallenhaupt SL, Tucker WY, Kon ND, Mills SA, Cordell AR. Radial artery-to-aorta pressure difference after discontinuation of cardiopulmonary bypass. Anesthesiology. 1989;70:935-941.  [PubMed]  [DOI]  [Cited in This Article: ]
11.  Mohr R, Lavee J, Goor DA. Inaccuracy of radial artery pressure measurement after cardiac operations. J Thorac Cardiovasc Surg. 1987;94:286-290.  [PubMed]  [DOI]  [Cited in This Article: ]
12.  Kang YG, Freeman JA, Aggarwal S, DeWolf AM. Hemodynamic instability during liver transplantation. Transplant Proc. 1989;21:3489-3492.  [PubMed]  [DOI]  [Cited in This Article: ]
13.  Krenn CG, De Wolf AM. Current approach to intraoperative monitoring in liver transplantation. Curr Opin Organ Transplant. 2008;13:285-290.  [PubMed]  [DOI]  [Cited in This Article: ]
14.  De Wolf AM. 6/2/06 Perioperative assessment of the cardiovascular system in ESLD and transplantation. Int Anesthesiol Clin. 2006;44:59-78.  [PubMed]  [DOI]  [Cited in This Article: ]
15.  Acosta F, Sansano T, Beltran R, Palenciano CG, Reche M, Roques V, Robles R, Bueno FS, Ramirez P, Parrilla P. Is femoral and radial artery pressure different during reperfusion in liver transplantation? Transplant Proc. 2000;32:2647.  [PubMed]  [DOI]  [Cited in This Article: ]
16.  Arnal D, Garutti I, Perez-Peña J, Olmedilla L, Tzenkov IG. Radial to femoral arterial blood pressure differences during liver transplantation. Anaesthesia. 2005;60:766-771.  [PubMed]  [DOI]  [Cited in This Article: ]
17.  Shin YH, Kim HY, Kim YR, Yoon JS, Ko JS, Gwak MS, Kim GS, Lee SK. The comparison of femoral and radial arterial blood pressures during pediatric liver transplantation. Transplant Proc. 2013;45:1924-1927.  [PubMed]  [DOI]  [Cited in This Article: ]
18.  Shin BS, Kim GS, Ko JS, Gwak MS, Yang M, Kim CS, Hahm TS, Lee SK. Comparison of femoral arterial blood pressure with radial arterial blood pressure and noninvasive upper arm blood pressure in the reperfusion period during liver transplantation. Transplant Proc. 2007;39:1326-1328.  [PubMed]  [DOI]  [Cited in This Article: ]
19.  Bendjelid K, Romand JA. Fluid responsiveness in mechanically ventilated patients: a review of indices used in intensive care. Intensive Care Med. 2003;29:352-360.  [PubMed]  [DOI]  [Cited in This Article: ]
20.  Kumar A, Anel R, Bunnell E, Habet K, Zanotti S, Marshall S, Neumann A, Ali A, Cheang M, Kavinsky C. Pulmonary artery occlusion pressure and central venous pressure fail to predict ventricular filling volume, cardiac performance, or the response to volume infusion in normal subjects. Crit Care Med. 2004;32:691-699.  [PubMed]  [DOI]  [Cited in This Article: ]
21.  Diebel L, Wilson RF, Heins J, Larky H, Warsow K, Wilson S. End-diastolic volume versus pulmonary artery wedge pressure in evaluating cardiac preload in trauma patients. J Trauma. 1994;37:950-955.  [PubMed]  [DOI]  [Cited in This Article: ]
22.  Hansen RM, Viquerat CE, Matthay MA, Wiener-Kronish JP, DeMarco T, Bahtia S, Marks JD, Botvinick EH, Chatterjee K. Poor correlation between pulmonary arterial wedge pressure and left ventricular end-diastolic volume after coronary artery bypass graft surgery. Anesthesiology. 1986;64:764-770.  [PubMed]  [DOI]  [Cited in This Article: ]
23.  Reuse C, Vincent JL, Pinsky MR. Measurements of right ventricular volumes during fluid challenge. Chest. 1990;98:1450-1454.  [PubMed]  [DOI]  [Cited in This Article: ]
24.  Sakka SG, Reinhart K, Meier-Hellmann A. Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients. Intensive Care Med. 1999;25:843-846.  [PubMed]  [DOI]  [Cited in This Article: ]
25.  Michard F, Boussat S, Chemla D, Anguel N, Mercat A, Lecarpentier Y, Richard C, Pinsky MR, Teboul JL. Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med. 2000;162:134-138.  [PubMed]  [DOI]  [Cited in This Article: ]
26.  Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37:2642-2647.  [PubMed]  [DOI]  [Cited in This Article: ]
27.  Feltracco P, Biancofiore G, Ori C, Saner FH, Della Rocca G. Limits and pitfalls of haemodynamic monitoring systems in liver transplantation surgery. Minerva Anestesiol. 2012;78:1372-1384.  [PubMed]  [DOI]  [Cited in This Article: ]
28.  Della Rocca G, Pompei L, Costa MG, Coccia C, Rossi M, Berloco PM, Pietropaoli P, Cortesini R. Hemodynamic-volumetric versus pulmonary artery catheter monitoring during anesthesia for liver transplantation. Transplant Proc. 2001;33:1394-1396.  [PubMed]  [DOI]  [Cited in This Article: ]
29.  Costa MG, Chiarandini P, Della Rocca G. Hemodynamics during liver transplantation. Transplant Proc. 2007;39:1871-1873.  [PubMed]  [DOI]  [Cited in This Article: ]
30.  Rocca GD, Costa MG, Feltracco P, Biancofiore G, Begliomini B, Taddei S, Coccia C, Pompei L, Di Marco P, Pietropaoli P. Continuous right ventricular end diastolic volume and right ventricular ejection fraction during liver transplantation: a multicenter study. Liver Transpl. 2008;14:327-332.  [PubMed]  [DOI]  [Cited in This Article: ]
31.  Balik M, Pachl J, Hendl J. Effect of the degree of tricuspid regurgitation on cardiac output measurements by thermodilution. Intensive Care Med. 2002;28:1117-1121.  [PubMed]  [DOI]  [Cited in This Article: ]
32.  Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth. 1993;40:142-153.  [PubMed]  [DOI]  [Cited in This Article: ]
33.  Greim CA, Roewer N, Thiel H, Laux G, Schulte am Esch J. Continuous cardiac output monitoring during adult liver transplantation: thermal filament technique versus bolus thermodilution. Anesth Analg. 1997;85:483-488.  [PubMed]  [DOI]  [Cited in This Article: ]
34.  Böttiger BW, Sinner B, Motsch J, Bach A, Bauer H, Martin E. Continuous versus intermittent thermodilution cardiac output measurement during orthotopic liver transplantation. Anaesthesia. 1997;52:207-214.  [PubMed]  [DOI]  [Cited in This Article: ]
35.  Boldt J, Menges T, Wollbrück M, Hammermann H, Hempelmann G. Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med. 1994;22:1913-1918.  [PubMed]  [DOI]  [Cited in This Article: ]
36.  Zöllner C, Polasek J, Kilger E, Pichler B, Jaenicke U, Briegel J, Vetter HO, Haller M. Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: a prospective criterion standard study. Crit Care Med. 1999;27:293-298.  [PubMed]  [DOI]  [Cited in This Article: ]
37.  Haller M, Zöllner C, Briegel J, Forst H. Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: a prospective criterion standard study. Crit Care Med. 1995;23:860-866.  [PubMed]  [DOI]  [Cited in This Article: ]
38.  De Wolf AM, Aggarwal S. Monitoring preload during liver transplantation. Liver Transpl. 2008;14:268-269.  [PubMed]  [DOI]  [Cited in This Article: ]
39.  Spiess BD, McCarthy RJ, Tuman KJ, Ivankovich AD. Bioimpedance hemodynamics compared to pulmonary artery catheter monitoring during orthotopic liver transplantation. J Surg Res. 1993;54:52-56.  [PubMed]  [DOI]  [Cited in This Article: ]
40.  Krowka MJ, Mandell MS, Ramsay MA, Kawut SM, Fallon MB, Manzarbeitia C, Pardo M, Marotta P, Uemoto S, Stoffel MP. Hepatopulmonary syndrome and portopulmonary hypertension: a report of the multicenter liver transplant database. Liver Transpl. 2004;10:174-182.  [PubMed]  [DOI]  [Cited in This Article: ]
41.  Ramsay MA, Simpson BR, Nguyen AT, Ramsay KJ, East C, Klintmalm GB. Severe pulmonary hypertension in liver transplant candidates. Liver Transpl Surg. 1997;3:494-500.  [PubMed]  [DOI]  [Cited in This Article: ]
42.  Steadman RH. Anesthesia for liver transplant surgery. Anesthesiol Clin North America. 2004;22:687-711.  [PubMed]  [DOI]  [Cited in This Article: ]
43.  Kuo PC, Plotkin JS, Gaine S, Schroeder RA, Rustgi VK, Rubin LJ, Johnson LB. Portopulmonary hypertension and the liver transplant candidate. Transplantation. 1999;67:1087-1093.  [PubMed]  [DOI]  [Cited in This Article: ]
44.  Krowka MJ, Plevak DJ, Findlay JY, Rosen CB, Wiesner RH, Krom RA. Pulmonary hemodynamics and perioperative cardiopulmonary-related mortality in patients with portopulmonary hypertension undergoing liver transplantation. Liver Transpl. 2000;6:443-450.  [PubMed]  [DOI]  [Cited in This Article: ]
45.  Sussman N, Kaza V, Barshes N, Stribling R, Goss J, O’Mahony C, Zhang E, Vierling J, Frost A. Successful liver transplantation following medical management of portopulmonary hypertension: a single-center series. Am J Transplant. 2006;6:2177-2182.  [PubMed]  [DOI]  [Cited in This Article: ]
46.  Ramsay M. Portopulmonary hypertension and right heart failure in patients with cirrhosis. Curr Opin Anaesthesiol. 2010;23:145-150.  [PubMed]  [DOI]  [Cited in This Article: ]
47.  Kuo PC, Plotkin JS, Johnson LB, Howell CD, Laurin JM, Bartlett ST, Rubin LJ. Distinctive clinical features of portopulmonary hypertension. Chest. 1997;112:980-986.  [PubMed]  [DOI]  [Cited in This Article: ]
48.  Safdar Z, Bartolome S, Sussman N. Portopulmonary hypertension: an update. Liver Transpl. 2012;18:881-891.  [PubMed]  [DOI]  [Cited in This Article: ]
49.  Plotkin JS, Kuo PC, Rubin LJ, Gaine S, Howell CD, Laurin J, Njoku MJ, Lim JW, Johnson LB. Successful use of chronic epoprostenol as a bridge to liver transplantation in severe portopulmonary hypertension. Transplantation. 1998;65:457-459.  [PubMed]  [DOI]  [Cited in This Article: ]
50.  Schroeder RA, Rafii AA, Plotkin JS, Johnson LB, Rustgi VK, Kuo PC. Use of aerosolized inhaled epoprostenol in the treatment of portopulmonary hypertension. Transplantation. 2000;70:548-550.  [PubMed]  [DOI]  [Cited in This Article: ]
51.  Hamilton WF, Remington JW. The measurement of the stroke volume from the pressure pulse. Am J Physiol. 1947;148:14-24.  [PubMed]  [DOI]  [Cited in This Article: ]
52.  Funk DJ, Moretti EW, Gan TJ. Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg. 2009;108:887-897.  [PubMed]  [DOI]  [Cited in This Article: ]
53.  Hofer CK, Cannesson M. Monitoring fluid responsiveness. Acta Anaesthesiol Taiwan. 2011;49:59-65.  [PubMed]  [DOI]  [Cited in This Article: ]
54.  Advanced Hemodynamic Monitoring: The FlowTrac Sensor. [Accessed 2014 Aug 28].  Available from: http://www.edwards.com/products/mininvasive/pages/flotracfaqs.aspx.  [PubMed]  [DOI]  [Cited in This Article: ]
55.  Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth. 2002;88:350-356.  [PubMed]  [DOI]  [Cited in This Article: ]
56.  Nissen P, Van Lieshout JJ, Novovic S, Bundgaard-Nielsen M, Secher NH. Techniques of cardiac output measurement during liver transplantation: arterial pulse wave versus thermodilution. Liver Transpl. 2009;15:287-291.  [PubMed]  [DOI]  [Cited in This Article: ]
57.  Rödig G, Prasser C, Keyl C, Liebold A, Hobbhahn J. Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth. 1999;82:525-530.  [PubMed]  [DOI]  [Cited in This Article: ]
58.  Sakka SG, Kozieras J, Thuemer O, van Hout N. Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth. 2007;99:337-342.  [PubMed]  [DOI]  [Cited in This Article: ]
59.  Krejci V, Vannucci A, Abbas A, Chapman W, Kangrga IM. Comparison of calibrated and uncalibrated arterial pressure-based cardiac output monitors during orthotopic liver transplantation. Liver Transpl. 2010;16:773-782.  [PubMed]  [DOI]  [Cited in This Article: ]
60.  Kusaka Y, Yoshitani K, Irie T, Inatomi Y, Shinzawa M, Ohnishi Y. Clinical comparison of an echocardiograph-derived versus pulse counter-derived cardiac output measurement in abdominal aortic aneurysm surgery. J Cardiothorac Vasc Anesth. 2012;26:223-226.  [PubMed]  [DOI]  [Cited in This Article: ]
61.  Biais M, Nouette-Gaulain K, Cottenceau V, Vallet A, Cochard JF, Revel P, Sztark F. Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis. Anesth Analg. 2008;106:1480-1486, table of contents.  [PubMed]  [DOI]  [Cited in This Article: ]
62.  Biancofiore G, Critchley LA, Lee A, Bindi L, Bisà M, Esposito M, Meacci L, Mozzo R, DeSimone P, Urbani L. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery. Br J Anaesth. 2009;102:47-54.  [PubMed]  [DOI]  [Cited in This Article: ]
63.  Della Rocca G, Costa MG, Chiarandini P, Bertossi G, Lugano M, Pompei L, Coccia C, Sainz-Barriga M, Pietropaoli P. Arterial pulse cardiac output agreement with thermodilution in patients in hyperdynamic conditions. J Cardiothorac Vasc Anesth. 2008;22:681-687.  [PubMed]  [DOI]  [Cited in This Article: ]
64.  De Backer D, Marx G, Tan A, Junker C, Van Nuffelen M, Hüter L, Ching W, Michard F, Vincent JL. Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients. Intensive Care Med. 2011;37:233-240.  [PubMed]  [DOI]  [Cited in This Article: ]
65.  Zimmermann A, Steinwendner J, Hofbauer S, Kirnbauer M, Schneider J, Moser L, Pauser G. The accuracy of the Vigileo/FloTrac system has been improved--follow-up after a software update: a blinded comparative study of 30 cardiosurgical patients. J Cardiothorac Vasc Anesth. 2009;23:929-931.  [PubMed]  [DOI]  [Cited in This Article: ]
66.  Tsai YF, Su BC, Lin CC, Liu FC, Lee WC, Yu HP. Cardiac output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation. Transplant Proc. 2012;44:433-437.  [PubMed]  [DOI]  [Cited in This Article: ]
67.  Su BC, Tsai YF, Chen CY, Yu HP, Yang MW, Lee WC, Lin CC. Cardiac output derived from arterial pressure waveform analysis in patients undergoing liver transplantation: validity of a third-generation device. Transplant Proc. 2012;44:424-428.  [PubMed]  [DOI]  [Cited in This Article: ]
68.  Cannesson M, Musard H, Desebbe O, Boucau C, Simon R, Hénaine R, Lehot JJ. The ability of stroke volume variations obtained with Vigileo/FloTrac system to monitor fluid responsiveness in mechanically ventilated patients. Anesth Analg. 2009;108:513-517.  [PubMed]  [DOI]  [Cited in This Article: ]
69.  Hadian M, Severyn DA, Pinsky MR. The effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients. J Crit Care. 2011;26:328.e1-328.e8.  [PubMed]  [DOI]  [Cited in This Article: ]
70.  Su BC, Tsai YF, Cheng CW, Yu HP, Yang MW, Lee WC, Lin CC. Stroke volume variation derived by arterial pulse contour analysis is a good indicator for preload estimation during liver transplantation. Transplant Proc. 2012;44:429-432.  [PubMed]  [DOI]  [Cited in This Article: ]
71.  Kim SH, Hwang GS, Kim SO, Kim YK. Is stroke volume variation a useful preload index in liver transplant recipients? A retrospective analysis. Int J Med Sci. 2013;10:751-757.  [PubMed]  [DOI]  [Cited in This Article: ]
72.  Biais M, Nouette-Gaulain K, Roullet S, Quinart A, Revel P, Sztark F. A comparison of stroke volume variation measured by Vigileo/FloTrac system and aortic Doppler echocardiography. Anesth Analg. 2009;109:466-469.  [PubMed]  [DOI]  [Cited in This Article: ]
73.  Soubrier S, Saulnier F, Hubert H, Delour P, Lenci H, Onimus T, Nseir S, Durocher A. Can dynamic indicators help the prediction of fluid responsiveness in spontaneously breathing critically ill patients? Intensive Care Med. 2007;33:1117-1124.  [PubMed]  [DOI]  [Cited in This Article: ]
74.  Reuter DA, Bayerlein J, Goepfert MS, Weis FC, Kilger E, Lamm P, Goetz AE. Influence of tidal volume on left ventricular stroke volume variation measured by pulse contour analysis in mechanically ventilated patients. Intensive Care Med. 2003;29:476-480.  [PubMed]  [DOI]  [Cited in This Article: ]
75.  Muller L, Louart G, Bousquet PJ, Candela D, Zoric L, de La Coussaye JE, Jaber S, Lefrant JY. The influence of the airway driving pressure on pulsed pressure variation as a predictor of fluid responsiveness. Intensive Care Med. 2010;36:496-503.  [PubMed]  [DOI]  [Cited in This Article: ]
76.  De Backer D, Heenen S, Piagnerelli M, Koch M, Vincent JL. Pulse pressure variations to predict fluid responsiveness: influence of tidal volume. Intensive Care Med. 2005;31:517-523.  [PubMed]  [DOI]  [Cited in This Article: ]
77.  Guerin L, Monnet X, Teboul JL. Monitoring volume and fluid responsiveness: from static to dynamic indicators. Best Pract Res Clin Anaesthesiol. 2013;27:177-185.  [PubMed]  [DOI]  [Cited in This Article: ]
78.  Suriani RJ. Transesophageal echocardiography during organ transplantation. J Cardiothorac Vasc Anesth. 1998;12:686-694.  [PubMed]  [DOI]  [Cited in This Article: ]
79.  Plotkin JS, Johnson LB, Kuo PC. Intracardiac thrombus formation during orthotopic liver transplantation: a new entity or an old enemy? Transplantation. 1996;61:1131.  [PubMed]  [DOI]  [Cited in This Article: ]
80.  Fahy BG, Hasnain JU, Flowers JL, Plotkin JS, Odonkor P, Ferguson MK. Transesophageal echocardiographic detection of gas embolism and cardiac valvular dysfunction during laparoscopic nephrectomy. Anesth Analg. 1999;88:500-504.  [PubMed]  [DOI]  [Cited in This Article: ]
81.  Fukazawa K, Poliac LC, Pretto EA. Rapid assessment and safe management of severe pulmonary hypertension with milrinone during orthotopic liver transplantation. Clin Transplant. 2010;24:515-519.  [PubMed]  [DOI]  [Cited in This Article: ]
82.  Hofer CK, Zollinger A, Rak M, Matter-Ensner S, Klaghofer R, Pasch T, Zalunardo MP. Therapeutic impact of intra-operative transoesophageal echocardiography during noncardiac surgery. Anaesthesia. 2004;59:3-9.  [PubMed]  [DOI]  [Cited in This Article: ]
83.  Planinsic RM, Nicolau-Raducu R, Eghtesad B, Marcos A. Diagnosis and treatment of intracardiac thrombosis during orthotopic liver transplantation. Anesth Analg. 2004;99:353-356, table of contents.  [PubMed]  [DOI]  [Cited in This Article: ]
84.  Sharma A, Pagel PS, Bhatia A. Intraoperative iatrogenic acute pericardial tamponade: use of rescue transesophageal echocardiography in a patient undergoing orthotopic liver transplantation. J Cardiothorac Vasc Anesth. 2005;19:364-366.  [PubMed]  [DOI]  [Cited in This Article: ]
85.  Hughes CG, Waldman JM, Barrios J, Robertson A. Postshunt hemochromatosis leading to cardiogenic shock in a patient presenting for orthotopic liver transplant: a case report. Transplant Proc. 2009;41:2000-2002.  [PubMed]  [DOI]  [Cited in This Article: ]
86.  Eagle SS, Thompson A, Fong PP, Pretorius M, Deegan RJ, Hairr JW, Riedel BJ. Takotsubo cardiomyopathy and coronary vasospasm during orthotopic liver transplantation: separate entities or common mechanism? J Cardiothorac Vasc Anesth. 2010;24:629-632.  [PubMed]  [DOI]  [Cited in This Article: ]
87.  Tiwari AK, D’Attellis N. Intraoperative left ventricular apical ballooning: transient Takotsubo cardiomyopathy during orthotopic liver transplantation. J Cardiothorac Vasc Anesth. 2008;22:442-445.  [PubMed]  [DOI]  [Cited in This Article: ]
88.  Cheung AT, Savino JS, Weiss SJ, Aukburg SJ, Berlin JA. Echocardiographic and hemodynamic indexes of left ventricular preload in patients with normal and abnormal ventricular function. Anesthesiology. 1994;81:376-387.  [PubMed]  [DOI]  [Cited in This Article: ]
89.  De Wolf AM, Begliomini B, Gasior TA, Kang Y, Pinsky MR. Right ventricular function during orthotopic liver transplantation. Anesth Analg. 1993;76:562-568.  [PubMed]  [DOI]  [Cited in This Article: ]
90.  Burtenshaw AJ, Isaac JL. The role of trans-oesophageal echocardiography for perioperative cardiovascular monitoring during orthotopic liver transplantation. Liver Transpl. 2006;12:1577-1583.  [PubMed]  [DOI]  [Cited in This Article: ]
91.  Carey WD, Dumot JA, Pimentel RR, Barnes DS, Hobbs RE, Henderson JM, Vogt DP, Mayes JT, Westveer MK, Easley KA. The prevalence of coronary artery disease in liver transplant candidates over age 50. Transplantation. 1995;59:859-864.  [PubMed]  [DOI]  [Cited in This Article: ]
92.  Plotkin JS, Scott VL, Pinna A, Dobsch BP, De Wolf AM, Kang Y. Morbidity and mortality in patients with coronary artery disease undergoing orthotopic liver transplantation. Liver Transpl Surg. 1996;2:426-430.  [PubMed]  [DOI]  [Cited in This Article: ]
93.  Plotkin JS, Johnson LB, Rustgi V, Kuo PC. Coronary artery disease and liver transplantation: the state of the art. Liver Transpl. 2000;6:S53-S56.  [PubMed]  [DOI]  [Cited in This Article: ]
94.  Smith JS, Cahalan MK, Benefiel DJ, Byrd BF, Lurz FW, Shapiro WA, Roizen MF, Bouchard A, Schiller NB. Intraoperative detection of myocardial ischemia in high-risk patients: electrocardiography versus two-dimensional transesophageal echocardiography. Circulation. 1985;72:1015-1021.  [PubMed]  [DOI]  [Cited in This Article: ]
95.  Leung JM, O’Kelly B, Browner WS, Tubau J, Hollenberg M, Mangano DT. Prognostic importance of postbypass regional wall-motion abnormalities in patients undergoing coronary artery bypass graft surgery. SPI Research Group. Anesthesiology. 1989;71:16-25.  [PubMed]  [DOI]  [Cited in This Article: ]
96.  Ellis JE, Lichtor JL, Feinstein SB, Chung MR, Polk SL, Broelsch C, Emond J, Thistlethwaite JR, Roizen MF. Right heart dysfunction, pulmonary embolism, and paradoxical embolization during liver transplantation. A transesophageal two-dimensional echocardiographic study. Anesth Analg. 1989;68:777-782.  [PubMed]  [DOI]  [Cited in This Article: ]
97.  Krowka MJ, Tajik AJ, Dickson ER, Wiesner RH, Cortese DA. Intrapulmonary vascular dilatations (IPVD) in liver transplant candidates. Screening by two-dimensional contrast-enhanced echocardiography. Chest. 1990;97:1165-1170.  [PubMed]  [DOI]  [Cited in This Article: ]
98.  Furuya H, Suzuki T, Okumura F, Kishi Y, Uefuji T. Detection of air embolism by transesophageal echocardiography. Anesthesiology. 1983;58:124-129.  [PubMed]  [DOI]  [Cited in This Article: ]
99.  Chung T, Emmett L, Mansberg R, Peters M, Kritharides L. Natural history of right ventricular dysfunction after acute pulmonary embolism. J Am Soc Echocardiogr. 2007;20:885-894.  [PubMed]  [DOI]  [Cited in This Article: ]
100.  Chung T, Emmett L, Khoury V, Lau GT, Elsik M, Foo F, Allman KC, Kritharides L. Atrial and ventricular echocardiographic correlates of the extent of pulmonary embolism in the elderly. J Am Soc Echocardiogr. 2006;19:347-353.  [PubMed]  [DOI]  [Cited in This Article: ]
101.  McConnell MV, Solomon SD, Rayan ME, Come PC, Goldhaber SZ, Lee RT. Regional right ventricular dysfunction detected by echocardiography in acute pulmonary embolism. Am J Cardiol. 1996;78:469-473.  [PubMed]  [DOI]  [Cited in This Article: ]
102.  Kurzyna M, Torbicki A, Pruszczyk P, Burakowska B, Fijałkowska A, Kober J, Oniszh K, Kuca P, Tomkowski W, Burakowski J. Disturbed right ventricular ejection pattern as a new Doppler echocardiographic sign of acute pulmonary embolism. Am J Cardiol. 2002;90:507-511.  [PubMed]  [DOI]  [Cited in This Article: ]
103.  Wheeler AP, Bernard GR, Thompson BT, Schoenfeld D, Wiedemann HP, deBoisblanc B, Connors AF, Hite RD, Harabin AL. Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury. N Engl J Med. 2006;354:2213-2224.  [PubMed]  [DOI]  [Cited in This Article: ]
104.  Wax DB, Torres A, Scher C, Leibowitz AB. Transesophageal echocardiography utilization in high-volume liver transplantation centers in the United States. J Cardiothorac Vasc Anesth. 2008;22:811-813.  [PubMed]  [DOI]  [Cited in This Article: ]
105.  Spier BJ, Larue SJ, Teelin TC, Leff JA, Swize LR, Borkan SH, Satyapriya A, Rahko PS, Pfau PR. Review of complications in a series of patients with known gastro-esophageal varices undergoing transesophageal echocardiography. J Am Soc Echocardiogr. 2009;22:396-400.  [PubMed]  [DOI]  [Cited in This Article: ]