DOI: 10.4244/EIJV9I5A105

A plea for caution: transcatheter aortic valve implantation using the Medtronic CoreValve™ prosthesis without balloon predilatation

Ralf Zahn1*, MD; Timm Bauer1, MD; Falk-Udo Sack1, MD; Andreas Lehman2, MD; Bernd Cornelius3, MD; Rudolf Schiele1, MD

An 87-year-old woman with severe symptomatic aortic stenosis was selected for TAVI by the Heart Team because of a high surgical risk (log. EuroSCORE: 24.2%). The mean transvalvular gradient was 56 mmHg and the diameter of the aortic annulus 21 mm. Therefore a 26 mm CoreValve™ prosthesis (Medtronic, Minneapolis, MN, USA) was selected. We decided to perform the procedure without prior balloon valvuloplasty. After femoral access with an 18 Fr sheath, a stiff guidewire was placed in the left ventricle. The prosthesis was advanced over the stenotic valve without problems (Figure 1A). The sheath cover was then retracted under fluoroscopic guidance and the valve released. It became apparent that the valve was highly compressed in its distal part and the position seemed optimal, with the distal end of the prosthesis reaching one cell (8 mm) into the left ventricular outflow tract (Figure 1B). We decided to perform a balloon dilatation with a 22 mm Nucleus™ balloon (NuMED, Hopkinton, NY, USA). In the meantime, before the balloon was advanced over the valve, we saw that the valve had moved upwards, reaching now only half a cell below the annulus (Figure 1C). We rapidly performed the balloon valvuloplasty (Online Figure 1) which led to an appropriate expansion of the distal part of the valve, with the distal part of the prosthesis now ending only one to two millimetres below the annulus (Online Figure 2). Angiography showed that there was no residual aortic insufficiency (Online Figure 3).

Figure 1. A) Positioning of the Medtronic CoreValve™ prosthesis. B) Result after release of the prosthesis: good placement but compression of the distal part of the prosthesis. C) Result after positioning of a balloon: spontaneous cranial dislodgement of the prosthesis.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Online Figure 1. Balloon valvuloplasty with a 22 mm Nucleus™ balloon.

Online Figure 2. Result after balloon valvuloplasty.

Online Figure 3. Final angiography, demonstrating no residual insufficiency.

Volume 9 Number 5
Sep 27, 2013
Volume 9 Number 5
View full issue


Key metrics

On the same subject

Editorial

10.4244/EIJ-E-24-00022 Jun 3, 2024
Epicardial inflow versus myocardial distribution: average regional transmural coronary flow is not enough
Johnson NP and Gould K
free

Editorial

10.4244/EIJ-E-24-00018 Jun 3, 2024
Left atrial appendage occlusion: can CCT see the leaks?
Reddy M and Shahab A
free

Editorial

10.4244/EIJ-E-24-00011 Jun 3, 2024
The assessment of ANOCA: coming of age?
De Bruyne B and Mahendiran T
free

Viewpoint

10.4244/EIJ-D-24-00137 Jun 3, 2024
Left atrial appendage occlusion: are we targeting the right patients?
Holmes DR and Alkhouli M
free

Original Research

10.4244/EIJ-D-23-00802 Jun 3, 2024
Long-term cardiac computed tomography follow-up after left atrial appendage occlusion
Tiroke LH et al

Original Research

10.4244/EIJ-D-23-00778 Jun 3, 2024
Effects of age on microvascular function in patients with normal coronary arteries
Jansen TP et al

Original Research

10.4244/EIJ-D-23-00713 Jun 3, 2024
Impact of hyperaemic stenosis resistance on long-term outcomes of stable angina in the ILIAS Registry
Boerhout C et al
Trending articles
159.38

State-of-the-Art

10.4244/EIJ-D-22-00725 Jan 23, 2023
Mitral valve transcatheter edge-to-edge repair
Hausleiter J et al
free
42.75

Clinical Research

10.4244/EIJ-D-21-01091 Aug 5, 2022
Lifetime management of patients with symptomatic severe aortic stenosis: a computed tomography simulation study
Medranda G et al
free
30.35

EXPERT REVIEW

10.4244/EIJ-D-17-00437 Feb 20, 2018
Catheter-based therapies in acute pulmonary embolism
Schultz J et al
free
12.1

Clinical research

10.4244/EIJ-D-23-00264 Sep 18, 2023
Histological examination of renal nerve distribution, density, and function in humans
Struthoff H et al
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved