Transcatheter mitral valve repair in a patient with prior atrial septal defect occlusion

Hiroki Niikura1, MD; Marcus Burns1, DNP; Richard Bae1, MD; Paul Sorajja1, MD

Figure 1. Transcatheter mitral valve repair after prior MitraClip therapy and atrial septal defect closure. A) Transoesophageal echocardiography (TEE) showing prior atrial septal occluder (ASO) and no space posterior to the ASO for safe transseptal puncture. B) Fluoroscopy showing a transseptal puncture through an ASO (arrow). C) Fluoroscopy showing predilatation with 12 × 40 mm balloon (arrow). D) TEE with three-dimensional (3D) imaging shows a steerable guide catheter through the ASO (arrow). E) Fluoroscopy showing deployment of an additional clip (arrow) through the ASO. F) TEE showing trivial MR (arrows) after additional clips. G) 3D-TEE showing new iatrogenic defect (arrow). H) TEE showing residual shunting (arrow). I) Fluoroscopy showing placement of the AMPLATZER Vascular Plug II (arrow). J) Transthoracic echocardiography showing no residual shunting (arrow). LA: left atrium; LV: left ventricle; RA: right atrium

An 83-year-old man presented with symptomatic, recurrent severe mitral regurgitation (MR) two years after a previous transcatheter mitral valve repair with the MitraClip® device (Abbott Vascular, Santa Clara, CA, USA), in which closure of a large iatrogenic atrial septal defect (iASD) with a 28 mm AMPLATZER™ atrial septal occluder (ASO; Abbott Vascular) was performed. Following a multidisciplinary Heart Team evaluation, a second MitraClip procedure was undertaken.

Initial procedural imaging demonstrated no space posterior to the ASO for a safe transseptal puncture, necessitating crossing of the ASO for the placement of the 24 Fr steerable guide catheter (SGC) to enable the second MitraClip procedure (Figure 1A). Using an NRG® C0 Transseptal Needle (Baylis Medical, Toronto, ON, Canada) with 10 watts of radiofrequency (RF) energy and an 8 Fr Mullins sheath, a crossing location on the ASO with a height of 4.5 cm to the mitral valve was utilised (Figure 1B). A 230 cm Inoue wire (Toray, Tokyo, Japan) was inserted into the Mullins sheath, followed by balloon dilatation of the ASO occluder segment in escalating sizes. However, the SGC was unable to pass across the ASO until dilatation with a 12×40 mm Atlas® GOLD balloon (Bard, Tempe, AZ, USA) and passage of an Inoue dilator (Toray) (Figure 1C, Figure 1D). Transcatheter repair with two additional MitraClip NTR clips was performed, leading to significant reduction in MR (Figure 1E, Figure 1F). Although a new iatrogenic defect was created through the previous ASO, this was not closed initially due to the possible need for additional MitraClip therapy, as well as the potential for unloading of left atrial hypertension (Figure 1G, Figure 1H).

The patient returned one month later with heart failure (NYHA Class III) symptoms. A 14 mm AMPLATZER™ Vascular Plug II (AVP II; Abbott Vascular) was used to close the new iASD completely using intracardiac echocardiography and fluoroscopic guidance, leading to clinical improvement (Figure 1I, Figure 1J). The MR reduction remained stable (grade 1+), and the patient had only mild dyspnoea (NYHA Class II) at one-year follow-up. Also, the AVP II was well seated with a minimal residual shunt and the patient was without haemolytic anaemia despite continued anticoagulation therapy.

To our knowledge, the present case represents the first experience of successful transcatheter therapy for residual MR after MitraClip therapy and ASD closure. These techniques may also be applicable for other left atrial interventions (e.g., appendage closure, pulmonary vein isolation) that may be required after atrial septal closure. While closure of an ASD with occluder devices can pose challenges for repeat access to the left atrium, this case demonstrates the feasibility of crossing such devices to perform beneficial therapy.

Conflict of interest statement

R. Bae reports being on the speakers’ bureau of Abbott Vascular. P. Sorajja has received research grants and consulting fees from and is on the speakers bureau of Edwards Lifesciences, Medtronic, Boston Scientific, and Abbott Structural, has equity in and is a consultant for Admedus, and has received consulting fees from Gore and Cardionomic. The other authors have no conflicts of interest to declare.

Supplementary data

To read the full content of this article, please download the PDF.

Volume 16 Number 15
Feb 5, 2021
Volume 16 Number 15
View full issue


Key metrics

On the same subject

INTERVENTIONAL FLASHLIGHT

10.4244/EIJ-D-17-00789 Feb 2, 2018
Residual “inter-clip” regurgitation due to a partial detachment, treated with AMPLATZER Vascular Plug II implantation
Danduch L et al
free

Image – Interventional flashlight

10.4244/EIJ-D-18-00794 Dec 20, 2018
Percutaneous treatment of anterior mitral leaflet perforation using an AMPLATZER Septal Occluder
Tiburtius C et al
free

10.4244/EIJV15I6A87 Aug 9, 2019
Patching residual leaks following a MitraClip procedure
Kar S and Sharma R
free

EXPERT REVIEW

10.4244/EIJ-D-17-00505 Sep 24, 2017
Mitral valve: repair/clips/cinching/chordae
Vahanian A et al
free

MITRAL VALVE INTERVENTIONS

10.4244/EIJV12SYA14 Sep 18, 2016
The MitraClip system: strategies for optimal patient selection and optimised results
Grasso C and Ince H
free

EXPERT REVIEW

10.4244/EIJ-D-18-00511 Sep 7, 2018
Transcatheter mitral valve repair: review of the clinical evidence
Taramasso M et al
free
Trending articles
337.53

State-of-the-Art Review

10.4244/EIJ-D-21-00904 Apr 1, 2022
Antiplatelet therapy after percutaneous coronary intervention
Angiolillo D et al
free
284.93

State-of-the-Art Review

10.4244/EIJ-D-21-00695 Nov 19, 2021
Transcatheter treatment for tricuspid valve disease
Praz F et al
free
240.7

State of the art

10.4244/EIJ-D-21-01117 Sep 20, 2022
Recanalisation of coronary chronic total occlusions
Di Mario C et al
free
226.28

State-of-the-Art Review

10.4244/EIJ-D-21-00426 Dec 3, 2021
Myocardial infarction with non-obstructive coronary artery disease
Lindahl B et al
free
209.85

State-of-the-Art Review

10.4244/EIJ-D-21-01034 Jun 3, 2022
Management of in-stent restenosis
Alfonso F et al
free
167.9

Expert review

10.4244/EIJ-D-21-00690 May 15, 2022
Crush techniques for percutaneous coronary intervention of bifurcation lesions
Moroni F et al
free
150.28

State-of-the-Art

10.4244/EIJ-D-22-00776 Apr 3, 2023
Computed tomographic angiography in coronary artery disease
Serruys PW et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved