Abstract
The penumbra has a major impact on obtaining uniformity of isodose distributions in radiation therapy. The penumbra phenomena of intensity-modulated radiation therapy (IMRT) techniques using multi-leaf collimators (MLCs) has an impact on the dose distributions in the border of the target volumes and the MLC. The aim of this study was to determine the impact of high photon energy (6 MV, 10 MV) on the penumbra for various depths and field sizes by using the Pencil Beam Convolution algorithms (eclipse 8.6) and self-developing Gafchromic™ EBT2 film. For dose calculations and EBT2 measurements, we used an acryl phantom with dimensions 20 × 20 × 20 cm3. The 200 cGy dose was delivered to the central depth (10 cm) of the acryl phantom. The result of this study was that increased energy, field size and depth are rise to an increased penumbra (20% ∼ 80%) width. For a 6 MV photon energy, the penumbra widths (20%–80%) at 1.5 cm, 5 cm, and 10 cm depths were 4.2 mm, 4.4 mm, and 5.7 mm for the eclipse calculations and 2.9 mm, 4.1 mm, and 4.2 mm for the EBT2 film measurements for 10 × 10 cm2 field sizes, respectively. For a 10 MV photon energy, the penumbra widths were 4.5 mm, 4.7 mm, and 6.2 mm for eclipse calculations and 4.1 mm, 4.6 mm, and 4.9 mm for EBT2 film measurements, respectively. As the field size was changed to 3 cm, 5 cm, 7 cm, 10 cm, and 15 cm, the penumbra widths changed to 5.1 mm, 5.3 mm, 5.6 mm, 5.9 mm, and 6.1 mm for eclipse calculations and 2.9 mm, 3.3 mm, 3.6 mm, 4.2 mm, and 5.1 mm for EBT2 measurements, respectively, for 10 cm depths for 6 MV photon energies. In this study, compared to the 10 MV photon energy, the 6 MV photon energy was preferred in treatments such as the 3D conformal radiation therapy and the IMRT for critical organs near the target volume.
Similar content being viewed by others
References
F. M. Khan, The Physics of Radiation Therapy, 4th ed. (Lippincott Williams and Wilkins, Baltimore, 2010), p. 185.
R. C. Miller, J. A. Bonner and R. W. Kline, Int. J. Radiat. Oncol. Biol. Phys. 41, 707 (1998).
L. Wang, E. Yorke, G. Desobry and C. S. Chui, J. Appl. Clin. Med. Phys. 3, 51 (2002).
T. Cheung, M. J. Butson and P. K. N. Yu, Med. Phys. 33, 2912 (2006).
S. W. Park, H. J. Kim, K. B. Kim, D. H. Lee, Y. H. Ji, R. N. Lee and S. I. Kwon, J. Korean Phys. Soc. 55, 694 (2009).
O. A. Garcia-Garduno, M. A. Celis, J. M. Larraga-Gutierrez, S. Moreno-Jimenez, A. Martinez-Davalos and M. Rodriguez-Villafuerte, J. Appl. Clin. Med. Phys. 9, 90 (2008).
F. M. Khan et al., Med. Phys. 18, 1 (1991).
P. Metcalfe, T. Kron, A. Elliott, T. Wong and P. Hoban, Med. Phys. 20, 1439 (1993).
M. J. Butson, P. K. N. Yu and T. Cheung, Phys. Med. Biol. 48, N247 (2003).
M. R. Arnfield, K. Otto, V. R. Aroumougame and R. D. Alkins, Med. Phys. 32, 12 (2005).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Oh, S.A., Kang, M.K., Yea, J.W. et al. Study of the penumbra for high-energy photon beams with Gafchromic™ EBT2 films. Journal of the Korean Physical Society 60, 1973–1976 (2012). https://doi.org/10.3938/jkps.60.1973
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.3938/jkps.60.1973