RADIOISOTOPES
Online ISSN : 1884-4111
Print ISSN : 0033-8303
ISSN-L : 0033-8303
蛍光ガラス線量計 (SC-1) の高エネルギー重イオンに対する応答
保田 浩志
著者情報
ジャーナル フリー

1999 年 48 巻 10 号 p. 645-647

詳細
抄録

Relative efficiency of a radiophotoluminescent glass (RPLG, product name: SC-1) was examined forHe, C, Ne, Si, Ar, andFebeams and γ-rays from60Coand137Cssources. Although dose dependency was not seen under 10-100mGy-water exposure, a strong LET dependency was observed. The RPL efficiency decreased almost proportionally with the logarithm of the infinite LET in water from 0.5 to 200keVμm-1.These results suggest that the value of RPLG obtained as γ-ray equivalent should underestimate the absorbed dose where heavy charged particles are mixed like in space. It is expected, however, to utilize RPLG for the evaluation of radiation quality by combining other detectors having different LET dependencies.

著者関連情報
© 社団法人 日本アイソトープ協会
前の記事 次の記事
feedback