Article

Effect of Oxidization Temperatures and Aging on Performance of Carbonate Melt Oxidized Iridium Oxide pH Electrode

1
School of Civil Engineering, Qingdao University of Technology, Qingdao 266033, China
2
Guangdong Provincial Key Laboratory of Durability for Marine Civil Engineering, Shenzhen University, Shenzhen 518060, China
3
Shenzhen Durability Center for Civil Engineering, Shenzhen University, Shenzhen 518060, China
4
Cooperative Innovation Center of Engineering Construction and Safety in Shandong Blue Economic Zone, Qingdao 266033, China
*
Author to whom correspondence should be addressed.
Sensors 2019, 19(21), 4756; https://doi.org/10.3390/s19214756
Submission received: 27 September 2019 / Revised: 30 October 2019 / Accepted: 31 October 2019 / Published: 1 November 2019
(This article belongs to the Special Issue Metal Oxide Based Sensors)

Abstract

:
Iridium oxide pH electrodes employing the carbonate melt oxidation method were fabricated with oxidation temperatures of 750 °C, 800 °C and 850 °C, respectively. Scanning electron microscope (SEM) and atomic force microscope (AFM) images showed that the oxide film regularized with the increase in oxidation temperatures. The pH response, response time and long-term stability of the electrodes indicated that the electrodes made at 850 °C had the best performance. X-ray photoelectron spectra (XPS) surveys investigated the change in the electrodes’ chemical composition and element oxidation states at 850 °C, and the results showed that the relative content of Ir3+ had increased by 23.9%, and the Ir4+ and Ir6+ had decreased by 10.9% and 13%, respectively, in the surface oxide layer after one month of aging. However, the relative contents of Ir3+, Ir4+ and Ir6+ were almost constant for the inner oxide layer. Meanwhile, the XPS result also indicated that the outer oxide layer of the electrode had a higher hydration degree than the inner oxide layer.

1. Introduction

In general, pH is one of the most important parameters which has to be measured in many works of analytical chemistry research. For example, carbonation, corrosion, and acid attack of reinforced concrete structure are related to the pH value of the pore solution [1,2,3,4]. The pH value of normal concrete is usually between 12.5 and 13 because calcium hydroxide is one of the major cement hydration products. However, it will decrease due to carbonation, chloride ingress, or acid attack. Normally, reinforcing steel in concrete does not corrode because a compact and stable passive film always forms on the surface of the reinforcing steel in the highly alkaline concrete pore solutions. Carbonation of calcium hydroxide in concrete can reduce the pH of concrete to values less than 9 [5,6,7,8]. When the pH value in vicinity of the rebar is lower than 11, initiation of active corrosion will happen even without chloride ions near rebar [9]. Therefore, in situ pH measurements of hardened concrete is very important for durability assessment and maintenance of reinforced concrete. Glass membrane electrodes are the most common electrodes for measuring the pH value of solutions [10]. However, glass electrodes are unsuitable for use in strong alkaline solutions, as well as in biomedical, concrete, and food process monitoring, because of the alkaline error. Additionally, it is difficult to miniaturize them, and their glass substrate is fragile. Therefore, many researchers have developed alternative electrodes. Metal oxide electrodes are considered suitable for overcoming the aforementioned problems, due to their stability and ruggedness, their prompt response and their flexibility in terms of size. The theoretical background of using the metal oxide electrodes for measuring pH, relies on reversibility of their redox reaction in aqueous solutions. Several metal oxide electrodes have been developed by different researchers [11,12,13,14,15,16,17,18,19], such as IrOX, PbO2, OsO2, TiO2, WO3, PtO2, Sb2O3, and RuO2. Among them, IrOX is considered the most suitable material for pH electrodes used in cementitious material [11]. However, IrOX electrodes are the most developed electrodes amongst them for its high stability within a wide pH range, fast response time, even in harsh conditions, such as high pressure and high temperatures, or in aggressive environments [12,20,21]. There are many methods for fabricating IrOX electrodes for pH electrodes, including electrochemical growth [22,23], electrodeposition [24,25], sputtered coating [26,27], and thermal method [28]. However, the problem of electrode potential drift raised from these methods limits the widespread application of IrOX pH electrodes. Yao et al. [11] have reported a method for making IrOX electrodes by oxidizing the Ir substrate in molten carbonate. IrOX electrodes made in this way have excellent pH response characteristics and can maintain stability for at least 2.5 years. However, problems pertaining to electrode potential drift and lack of long-term stability remains, even when using the carbonate melt oxidation method to prepare IrOX electrodes [29,30,31]. Researchers have developed improved methods based on carbonate melt oxidation. For example, Pan et al. [30] used Li2CO3 and Na2O2 as the melt bath to oxidize iridium. The improved electrode has good reproducibility and a small drift potential. O’Malley et al. [31] have developed the vertical melt oxidation method which provides a more uniform reaction environment than previous methods. They also have investigated the effect of oxidation time on IrOX electrodes. Chen et al. [32] have investigated the impact of carbonate types on IrOX electrodes by using different carbonates as the oxidation media. Prior research has shown that the oxidation temperature, molten carbonate composition, oxidation time and hydration degree can influence the characteristics of IrOX electrodes. The mechanism of metal oxide film growth in molten carbonate has been investigated in molten carbonate fuel cells, and the oxidizing specie is O22−, which is produced from the reaction of dissolved O2 in melted CO32−. Moreover, under the same partial pressure of carbon dioxide, the diffusion coefficient of O22− in molten carbonate increases with the increasing oxidation temperatures [33].
Since the oxidation temperature is an important factor in melt-oxidized IrOX electrode fabrication, its influence on the electrode should not be ignored. However, the effect of oxidation temperatures on the performance of IrOX pH electrodes has received insufficient investigation. The melt point of lithium carbonate is about 723 °C, the fabrication temperature of this method must over 723 °C. And higher temperature can benefit to the stability of the electrode, but it can also accelerate the decomposition rate of the lithium carbonate, thus lead to the insufficient oxidation of iridium and adversely affect the stability of the electrode. If the temperature up to 900 °C, the oxidation time must be varied. Effect of oxidation time on performance of carbonate melt oxidized Iridium oxide pH electrode well be mentioned in another paper. In this paper, the carbonate melt oxidation method was adopted to fabricate IrOX electrodes, and three different oxidation temperatures (750 °C, 800 °C, 850 °C) were chosen to complete the fabrication. The influence of different oxidation temperatures on IrOX electrode performance was studied by combining the pH sensing characteristics and the morphology surveys. In order to explain the reason for the pH electrode potential changing over time, X-ray photoelectron spectrascopy (XPS) was performed to analyze the changes in chemical composition and the elemental oxidation state of the electrode fabricated at 850 °C, both before and after one month of aging in distilled water. Results will be helpful for preparation of solid pH sensor for in-situ pH measurements of hardened concrete.

2. Experimental Program

2.1. Electrode Preparation

In this study, the carbonate melt oxidation method was used to fabricate the IrOX electrodes. In this method, a cleaned iridium wires (Φ0.5 mm, 99.8% purity, 10 mm length, purchased from Alfa Aesar (Tianjin) Chemical Co., Ltd., Tianjin, China) were placed in a gold foil lined alumina crucible and covered with Li2CO3 powder. The crucible was heated in oven with a heating rate of 10 °C/min until the designed temperature (750 °C, 800 °C, 850 °C) was reached. The temperature was then maintained for 5 h under air atmosphere. After cooling down to room temperature, 1 M HCl solution was used to dissolve the solid carbonate, followed by washing the oxidized wires with distilled water and then drying them at 120 °C for 12 h. Finally, the end of the oxidized wire (about 1 mm in length) was scraped off and the oxidized wire was spot-welded to a copper wire. The contact area was then covered with epoxy.

2.2. Open-Circuit Potential Measurement

The pH response characteristic of IrOX electrodes was examined by measuring the open circuit potentials in different test solutions. Three commercial pH buffer solutions with pH value of 4 (potassium hydrogen phthalate), 6.86 (mixed phosphate) and 9.18 (sodium tetraborate) were used to obtain a series of test solutions with different pH values (pH 1–13) by adding 1 M HCl or 1 M NaOH (all chemical reagents were purchased from Huadong Instrument, China). A glass pH electrode (E-201-C, INESA Scientific Instruments, China) was used to monitor the pH values of the test solutions. For each pH sensing points, we prepared a series of pH solutions and do pH testing one after another. Before each testing, the electrodes were rinsed with deionized water. The open circuit potential of the electrode was measured and recoded by an electrochemical workstation (VersaStat3, Princeton Applied Research, USA) with a platinum plate as the counter electrode and a saturated calomel electrode (SCE) as the reference electrode as shown in Figure 1. All potentiometric measurements were performed at room temperature (20 ± 3 °C), and the electrodes were kept in distilled water when they were not being tested.

2.3. Surface Analysis

Scanning electron microscopy (SEM) was performed with Quanta FEI 250. Atomic force microscopy (AFM) was conducted with Bruker Dimension Edge. The tests were conducted under ambient conditions with a scan size of 10 μm × 10 μm and scan rate is 10 μm/s. The silicon nitride tips (Bruker Company) and tapping mode were chosen in this study. X-ray photoelectron spectroscopy (XPS) was performed with Thermo ESCALAB 250XI by using the Al Kα (hν = 1486.6 eV) as its X-ray source. Xpspeak41 was used to process the data.

3. Results and Discussion

3.1. Surface Morphology

The surface morphology of the IrOX electrodes fabricated at different oxidation temperatures are shown in Figure 2. The films appear different from the iridium oxide films made by the thermal decomposition method, which usually reveals the mud-cracked morphology [34]. Electrodes made at 750 °C showed tiny cracks and uneven particle size and distribution. The films become more uniform as oxidation temperature increased. The iridium oxide films made at 850 °C had a dense and homogenous structure. The cracked and uneven layer leading to potential pathways to the electrode matrix may have been detrimental to the pH sensing. Furthermore, the iridium oxide film made at 750 °C fell off the substrate, while the film made at 800 °C and 850 °C stuck to the iridium substrate; they were hard to scrape off, even with a knife. This indicates that the surface morphology can affect the adhesion of the oxide film; the cracked and uneven film is less adherent than the dense, uniform film.
Figure 3 shows the 3-D images of the surface of IrOX electrodes obtained by AFM. Obviously, the particle size and distribution of iridium oxide film made at different temperatures are different. The section roughness analysis from the AFM results are shown in Figure 1d, from which it is clear that the particle size of IrOX ranges from a few hundred nanometers to a few micros, and the morphology becomes more uniform with the increase in oxidation temperature.

3.2. Response Time

pH electrode response time has yet to be universally defined, either in the literature or in product specifications, as the environmental conditions are different [35]. In this paper, the response time was defined as the time needed for a potential change rate of less than 1 mV/min [36]. The electrodes’ response time characteristics were evaluated by continuously measuring the open-circuit potential (Figure 4). Each of the electrodes obtained the stable potentials at each pH level. The average response times for electrodes made at 750 °C, 800 °C and 850 °C at different pH levels were 85 s, 28 s and 19 s, respectively. The electrodes made at 800 °C had slow response rates in strong acidic solutions. The response time of the electrodes made at 750 °C were longer than those of the electrodes made at 850 °C and 800 °C. This may due to the cracked and uneven iridium oxide film needing more time to react and equilibrate with the solution [37].

3.3. pH Response Characteristics

The pH response characteristics of the IrOX electrodes fabricated at different oxidation temperatures were investigated for the pH range of 1–13 (Figure 5). Results showed that the electrodes had similar sensitivity, −52.7 mV/pH, −54.6 mV/pH, and −53.2 mV/pH for oxidation temperatures of 750 °C, 800 °C, and 850 °C, respectively. These results corroborate the value obtained by Chen [32], which was slightly lower than the theoretical value of −59 mV/pH. The E-pH curves had a linear relationship with a correlation coefficient over 0.998.

3.4. Long-Term Stability

The long-term stability of IrOX pH electrodes were measured by calibrating the electrodes three times within one month. During this month, the electrodes were stored in distilled water. Results are shown in Figure 6. All of the calibration curves maintained linearity, and the slope varied no more than 1 mV/pH throughout the test period. However, the intercept (E0) decreased 52 mV, 45.7 mV, and 13.4 mV for the electrodes made at 750 °C, 800 °C, and 850 °C, respectively. This aging process, characterized by the decrease in E0 over time, has been observed by several researchers for different IrOX electrode preparation methods [27,36,38,39,40]. Yao et al. [11] thought that the oxidation state and the hydration degree of the oxide film influence the long-term stability of the electrode.

3.5. XPS

In order to further assess the potential drift mechanisms of the electrodes fabricated by the carbonate melt oxidation method, XPS were performed to investigate the chemical composition and element oxidation state change of the electrodes fabricated at 850 °C. This was done both before and after one month of aging. Deconvolution of Ir4f spectra was conducted considering the spin-obit doublet binding energy splitting of 2.95 eV. The peak ratio between Ir4f7/2 and Ir4f5/2 was constrained to 4:3. Several groups deconvoluted the Ir4f spectra, and an array of line shapes and interpretations of present species have been proposed [40,41,42,43,44,45]. In this investigation, the Ir4f spectra was deconvoluted into three species (Figure 7). The binding energies and the relative contents of different iridium species are presented in Table 1. The species present at 62 eV, 61.25 eV, 61.89 eV and 61.12 eV of Ir4f7/2 corresponded the existence of Ir3+ [40,45]. The species present at 63 eV, 62.49 eV, 62.87 eV, and 62.09 eV of Ir4f7/2 reflected the existence of Ir4+ [40,45]. The species present at 63.84 eV, 63.65 eV, 63.53 eV, and 63.5 eV of Ir4f7/2 reflected the existence of Ir6+ [46]. The table shows that the Ir4f binding energies attributed to Ir3+ and Ir4+ species on the electrode before argon-ion etching were significantly higher (0.75–0.77 eV) than the corresponding values after argon-ion etching. This phenomenon was probably due to the outer oxide layer having a higher hydration degree than the inner oxide layer [47]. The relative contents of the three iridium species show that for the freshly prepared electrodes, the Ir4+ and Ir6+ contents decreased by 14.6% and 16%, respectively. Meanwhile, the Ir3+ content increased by 30.6% after argon-ion etching. This indicated that the outer oxide layer had a higher average oxidation state than the inner oxide layer. This may have been due to the outer oxide layer covering the inner oxide layer, thus preventing the inner oxide layer from being further oxidized by the strong oxidizing species in melt carbonate. In contrast, for the electrodes that were aged for one month, the Ir3+, Ir4+ and Ir6+ contents were relatively stable, both before and after etching. The Ir3+ content also increased by 23.9%, while the Ir4+ and Ir6+ contents decreased by 10.9% and 13.3%, respectively, compared to the freshly prepared electrodes and the electrodes that had been aged for a month.
According to an earlier report on O1s deconvolution of IrOX [40,41,47,48], three oxygen species may contribute to the intensity of O1s spectra. The range of binding energies for oxide is 530 ± 0.5 eV, for hydroxides it is 531.4 ± 0.5 eV, and for bound water it is 533 ± 1 eV. In our experiment, the deconvolution of O 1s spectra (Figure 8 and Table 2) shows three peaks. This indicates that the O1s spectra may contain all three kinds of oxygen. Analysis of the relative content of different oxygen species has found that the hydroxide species contributes most to the overall O1s signal of electrodes, both before and after one month of aging. However, the oxide (lattice oxidized O2−) became the major species after the ion etching process. This confirmed the different hydration degrees between the outer oxide layer and the inner oxide layer. It also indicated that the freshly prepared iridium oxide film was hydrated, due to the treatment of HCl solution and distilled water in the fabrication process.
The atomic ratios of Ir3+/Ir4+ and O1s/Ir4f in our experiment are summarized in Table 3. The Ir3+/Ir4+ atomic ratio increased 70% and O1s/Ir4f decreased 60% after one month of aging. This indicated that the Ir3+ complex species have an O/Ir ratio lower than that of the Ir4+ species. Huang et al. [40] have suggested that the pH response reaction is affected by the Ir3+/Ir4+ ratio and the hydration of the oxide film. Thus, the Ir6+ species have been regarded as the “defect structures”, and they have proposed the pH response reaction mechanism shown in Equations (1) and (2). In these equations, Hf + and Hs+ are the H+ ions inside the hydrated oxide film and in the solution, respectively. Additionally, the Ir3+/Ir4+ ratio and the hydration of the oxide film can affect the value of the intercept, but cannot influence the sensitivity of the electrodes. This is consistent with the results of our experiment:
2 { [ IrO 2 ( OH ) 2 · 2 H 2 O ] 2 · 2 H f + } + 2 e + 2 H s + [ Ir 2 O 3 ( OH ) 3 · 3 H 2 O ] 3 · 3 H f + + 3 H 2 O
E = E 0 + 2.303 RT 2 F log { [ IrO 2 ( OH ) 2 · 2 H 2 O ] 2 · 2 H f + } 2 · [ H s + ] 2 [ Ir 2 O 3 ( OH ) 3 · 3 H 2 O ] 3 · 3 H f +    = E 0 + 2.303 RT 2 F log { [ IrO 2 ( OH ) 2 · 2 H 2 O ] 2 · 2 H f + } 2 [ Ir 2 O 3 ( OH ) 3 · 3 H 2 O ] 3 · 3 H f +    + 2.303 RT F log [ H s + ] = E 0 , 2.303 RT F pH = E 0 , 59.16 pH

4. Conclusions

SEM and AFM images showed that the oxide film regularized in terms of particle size and distribution as the oxidation temperature increased. Sub-Nernstian responses of −52.7 mV/pH, −54.6 mV/pH, and −53.2 mV/pH have been obtained corresponding to 750 °C, 800 °C, and 850 °C, respectively. The better performance in response time and long-term stability was manifest in better quality oxide film made at 850 °C than at 800 °C or 750 °C. The pH response characteristics of the electrodes fabricated at the three different oxidation temperatures maintained linearity and the slope changed little, but the intercept decreased during the one-month aging process. The XPS results showed that the element Ir of the oxide film existed in the form of Ir 3+, Ir4+, and Ir6+. The relative contents of Ir3+ increased by 23.9%, while Ir4+ and Ir6+ decreased by 10.9% and 13.3% in the surface oxide film after one month of aging. An argon-ion etching experiment on the electrodes, both before and after one month of aging, resulted in the different hydration degrees along the depth of the oxide films. It also showed that the hydration effect occurred primarily on the surface of the oxide film.

Author Contributions

Investigation: T.G., Z.D., and Z.L.; methodology: T.Z. and Z.J.; software: B.D.; writing—original draft: P.W.; writing—review and editing: P.W.

Funding

This research was funded by the support of the National Natural Science Foundation of China (51608286, 51420105015, U1706222), the Natural Science Foundation of Shandong province (2019GSF110006), the Guangdong Provincial Key Laboratory of Durability for Marine Civil Engineering, Shenzhen Durability Center for Civil Engineering, Shenzhen University (GDDCE 18-9), and the Source Innovation Program of Qingdao City (17-1-1-13-jch).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Liu, X.; Niu, D.; Li, X.; Lv, Y.; Fu, Q. Pore Solution pH for the Corrosion Initiation of Rebars Embedded in Concrete under a Long-Term Natural Carbonation Reaction. Appl. Sci. 2018, 8, 128. [Google Scholar] [CrossRef]
  2. Femenias, Y.; Angst, U.; Moro, F.; Elsener, B. Development of a novel methodology to assess the corrosion threshold in concrete based on simultaneous monitoring of pH and free chloride concentration. Sensors 2018, 18, 3101. [Google Scholar] [CrossRef] [PubMed]
  3. Gunasekara, C.; Law, D.; Bhuiyan, S.; Setunge, S.; Ward, L. Chloride induced corrosion in different fly ash based geopolymer concretes. Constr. Build. Mater. 2019, 200, 502–513. [Google Scholar] [CrossRef]
  4. Gu, L.; Bennett, T.; Visintin, P. Sulphuric acid exposure of conventional concrete and alkali-activated concrete: Assessment of test methodologies. Constr. Build. Mater. 2019, 197, 681–692. [Google Scholar] [CrossRef]
  5. Chang, C.F.; Chen, J.W. The experimental investigation of concrete carbonation depth. Cem. Concr. Res. 2006, 36, 1760–1767. [Google Scholar] [CrossRef]
  6. Yoon, I.S.; Çopuroğlu, O.; Park, K.B. Effect of global climatic change on carbonation progress of concrete. Atmos. Environ. 2007, 41, 7274–7285. [Google Scholar] [CrossRef]
  7. Ji, Y.S.; Wu, M.; Ding, B.; Liu, F.; Gao, F. The experimental investigation of width of semi-carbonation zone in carbonated concrete. Constr. Build. Mater. 2014, 65, 67–75. [Google Scholar] [CrossRef]
  8. Liu, X.; Niu, D.; Li, X.; Lv, Y. Effects of Ca (OH)2–CaCO3 concentration distribution on the pH and pore structure in natural carbonated cover concrete: A case study. Constr. Build. Mater. 2018, 186, 1276–1285. [Google Scholar] [CrossRef]
  9. Behnood, A.; Van Tittelboom, K.; De Belie, N. Methods for measuring pH in concrete—A review. Constr. Build. Mater. 2016, 105, 176–188. [Google Scholar] [CrossRef]
  10. Galster, H. pH Measurement: Fundamentals, Methods, Applications, Instrumentation; VCH: Weinheim, Germany, 1991. [Google Scholar]
  11. Yao, S.; Wang, M.; Madou, M. A pH electrode based on melt-oxidized iridium Oxide. J. Electrochem. Soc. 2001, 148, 29–36. [Google Scholar] [CrossRef]
  12. Wang, M.; Yao, S.; Madou, M. A long-term stable iridium oxide pH electrode. Sens. Actuators B Chem. 2002, 81, 313–315. [Google Scholar] [CrossRef]
  13. Lu, Y.; Wang, T.; Cai, Z.; Cao, Y.; Yang, H.; Duan, Y.Y. Anodically electrodeposited iridium oxide films microelectrodes for neural microstimulation and recording. Sens. Actuators B Chem. 2009, 137, 334–339. [Google Scholar] [CrossRef]
  14. Elsen, H.A.; Monson, C.F.; Majda, M. Effects of electrodeposition conditions and protocol on the properties of iridium oxide pH sensor electrodes. J. Electrochem. Soc. 2009, 156, 1–6. [Google Scholar] [CrossRef]
  15. Franz, M.; Beyer, J.B. Development of Ti/PbO2/Sb2O3 electrode as indicator electrode for pH measurements and conductometric titrations in aqueous solutions. J. Pharma. Biol. Sci. 2014, 13, 1653–1654. [Google Scholar]
  16. Fog, A.; Buck, R.P. Electronic semiconducting oxides as pH sensors. Sens. Actuators 1984, 5, 137–146. [Google Scholar] [CrossRef]
  17. Zhang, W.D.; Xu, B. A solid-state pH sensor based on WO3-modified vertically aligned multiwalled carbon nanotubes. Electrochem. Commun. 2009, 11, 1038–1041. [Google Scholar] [CrossRef]
  18. Wang, M.; Ha, Y. An electrochemical approach to monitor pH change in agar media during plant tissue culture. Biosens. Bioelectron. 2007, 22, 2718. [Google Scholar] [CrossRef]
  19. Tanumihardja, E.; Olthuis, W.; Berg, A. Ruthenium Oxide pH Sensing for Organs-On-Chip Studies. Multidiscip. Digit. Publ. Inst. Proc. 2018, 2, 709. [Google Scholar] [CrossRef]
  20. Katsube, T.; Lauks, I.R.; van der Spiegel, J.; Zemel, J.N. High temperature and high pressure pH sensors with sputtered iridium oxide films. Jpn. J. Appl. Phys. 1983, 22, 131–132. [Google Scholar] [CrossRef]
  21. Hitchman, M.L.; Ramanathan, S. Potentiometric determination of proton activities in solutions containing hydrofluoric acid using thermally oxidized iridium electrodes. Analyst 1991, 116, 1131–1133. [Google Scholar] [CrossRef]
  22. Cherevko, S.; Geiger, S.; Kasian, O.; Mingers, A.; Mayrhofer, K.J. Oxygen evolution activity and stability of iridium in acidic media. Part 2.—Electrochemically grown hydrous iridium oxide. J. Electroanal. Chem. 2016, 774, 102–110. [Google Scholar] [CrossRef]
  23. Olthuis, W.; Robben, M.A.M.; Bergveld, P.; Bos, M.; Van der Linden, W.E. pH sensor properties of electrochemically grown iridium oxide. Sens. Actuators B Chem. 1990, 2, 247–256. [Google Scholar] [CrossRef] [Green Version]
  24. Prats-Alfonso, E.; Abad, L.; Casañ-Pastor, N.; Gonzalo-Ruiz, J.; Baldrich, E. Iridium oxide pH sensor for biomedical applications. Case urea-urease in real urine samples. Biosens. Bioelectron. 2012, 39, 163–169. [Google Scholar] [CrossRef] [PubMed]
  25. Kim, T.Y.; Yang, S. Fabrication method and characterization of electrodeposited and heat-treated iridium oxide films for pH sensing. Sens. Actuators B Chem. 2014, 196, 31–38. [Google Scholar] [CrossRef]
  26. Katsube, T.; Lauks, I.; Zemel, J.N. pH-sensitive sputtered iridium oxide films. Sens. Actuators 1981, 2, 399–410. [Google Scholar] [CrossRef]
  27. Kuo, L.M.; Chou, Y.C.; Chen, K.N.; Lu, C.C.; Chao, S. A precise pH microsensor using RF-sputtering IrO2 and Ta2O5 films on Pt-electrode. Sens. Actuators B Chem. 2014, 193, 687–691. [Google Scholar] [CrossRef]
  28. Fierro, S.; Kapałka, A.; Frey, O.; Koudelka, M.; Comninellis, C. Microelectrode-array of IrO2, prepared by thermal treatment of pure Ir. Electrochem. Commun. 2010, 12, 587–591. [Google Scholar] [CrossRef]
  29. Smiechowski, M.F.; Lvovich, V.F. Iridium oxide sensors for acidity and basicity detection in industrial lubricants. Sens. Actuators B Chem. 2003, 96, 261–267. [Google Scholar] [CrossRef]
  30. Pan, Y.W.; Ying, Y.E.; Han, C.H. An improved approach of pH electrode preparation for application in the deep-sea environment. Acta Oceanol. Sin. 2010, 32, 73–79. [Google Scholar]
  31. O’Malley, M.J. Production and isolation of pH sensing materials by carbonate melt oxidation of iridium and platinum. J. Mater. Chem. 2012, 22, 7782–7790. [Google Scholar] [CrossRef]
  32. Chen, D.C.; Zheng, J.C.; Fu, C.Y. Preparation of Ir/IrOx pH electrode based on melt-oxidation and its response mechanism investigation. Trans. Nonferrous Met. Soc. China 2003, 13, 1459–1464. [Google Scholar]
  33. Dave, B.B.; White, R.E.; Srinivasan, S.; Appleby, A.J. Impedance analysis for oxygen reduction in a lithium carbonate melt: Effects of partial pressure of carbon dioxide and temperature. J. Electrochem. Soc. 1993, 140, 2139–2144. [Google Scholar] [CrossRef]
  34. Hou, Y.Y.; Hu, J.M.; Liu, L.; Zhang, J.Q.; Cao, C.N. Effect of calcination temperature on electrocatalytic activities of Ti/IrO2, electrodes in methanol aqueous solutions. Electrochim. Acta 2006, 51, 6258–6267. [Google Scholar] [CrossRef]
  35. Huang, W.D.; Cao, H.; Deb, S.; Chiao, M.; Chiao, J.C. A flexible pH sensor based on the iridium oxide sensing film. Sens. Actuators A Phys. 2011, 169, 1–11. [Google Scholar] [CrossRef]
  36. Huang, F.; Jin, Y.; Wen, L.; Mu, D.; Cui, M. Effects of Thermal Oxidation Cycle Numbers and Hydration on IrOx pH Sensor. J. Electrochem. Soc. 2013, 160, 184–191. [Google Scholar] [CrossRef]
  37. Park, J.; Kim, M.; Kim, S. Surface renewable nano-iridium oxide polymeric composite pH electrodes. Sens. Actuators B Chem. 2014, 204, 197–202. [Google Scholar] [CrossRef]
  38. Hitchman, M.L.; Ramanathan, S. Evaluation of iridium oxide electrodes formed by potential cycling as pH probes. Analyst 1988, 113, 9–35. [Google Scholar] [CrossRef]
  39. Kinoshita, K.; Madou, M.J. Electrochemical measurements on Pt, Ir, and Ti oxides as pH probes. Macromolecules 1984, 131, 1089–1094. [Google Scholar] [CrossRef]
  40. Huang, F.; Jin, Y.; Wen, L. Investigations of the hydration effects on cyclic thermo-oxidized Ir/IrOx Electrode. J. Electrochem. Soc. 2015, 162, 337–343. [Google Scholar] [CrossRef]
  41. Atanasoska, L.; Gupta, P.; Deng, C.; Warner, R.; Larson, S.; Thompson, J. XPS, AES, and electrochemical study of iridium oxide coating materials for cardiovascular stent application. Ecs Trans. 2009, 16, 37–48. [Google Scholar]
  42. Hall, H.Y.; Sherwood, P.M.A. X-ray photoelectron spectroscopic study of the iridium electrode system. J. Electroanal. Chem. Interfacial Electrochem. 1984, 15, 135–152. [Google Scholar] [CrossRef]
  43. Pfeifer, V.; Jones, T.E.; Vélez, J.V.; Massué, C.; Greiner, M.T.; Arrigo, R.; Teschner, D.; Girgsdies, F.; Scherzer, M.; Allan, J.; et al. The electronic structure of iridium oxide electrodes active in water splitting. Phys. Chem. Chem. Phys. 2016, 18, 2292–2296. [Google Scholar] [CrossRef] [PubMed]
  44. Casella, I.G.; Contursi, M.; Toniolo, R. Anodic electrodeposition of iridium oxide particles on glassy carbon surfaces and their electrochemical/SEM/XPS characterization. J. Electroanal. Chem. 2014, 736, 147–152. [Google Scholar] [CrossRef]
  45. Da Silva, L.A.; Alves, V.A.; De Castro, S.C.; Boodts, J.F.C. XPS study of the state of iridium, platinum, titanium and oxygen in thermally formed IrO2+ TiO2+ PtOX, films. Colloids Surf. A Physicochem. Eng. Asp. 2000, 170, 119–126. [Google Scholar] [CrossRef]
  46. Kim, K.S.; Sell, C.D.; Winograd, N. Electrochemical Society Softbound Proceedings Series; Breiter, M.W., Ed.; The Electrochemical Society: Princeton, NJ, USA, 1974. [Google Scholar]
  47. Kötz, R.; Neff, H.; Stucki, S. Anodic Iridium Oxide Films: XPS-studies of oxidation state changes and O2-evolution. J. Electrochem. Soc. 1984, 131, 72–77. [Google Scholar] [CrossRef]
  48. Augustynski, J.; Koudelka, M.; Sanchez, J.; Conway, B.E. ESCA study of the state of iridium and oxygen in electrochemically and thermally formed iridium oxide films. J. Electroanal. Chem. Interfacial Electrochem. 1984, 160, 233–248. [Google Scholar] [CrossRef]
Figure 1. Schematic picture of the pH testing setup.
Figure 1. Schematic picture of the pH testing setup.
Sensors 19 04756 g001
Figure 2. SEM images of IrOX electrodes fabricated at 750 °C (a), 800 °C (b), and 850 °C (c).
Figure 2. SEM images of IrOX electrodes fabricated at 750 °C (a), 800 °C (b), and 850 °C (c).
Sensors 19 04756 g002
Figure 3. AFM images of IrOX electrodes fabricated at 750 °C (a), 800 °C (b), 850 °C (c), and section roughness analysis (d).
Figure 3. AFM images of IrOX electrodes fabricated at 750 °C (a), 800 °C (b), 850 °C (c), and section roughness analysis (d).
Sensors 19 04756 g003
Figure 4. Response time, measured in different pH test solutions for IrOX electrodes fabricated at 750 °C (a), 800 °C (b), and 850 °C (c).
Figure 4. Response time, measured in different pH test solutions for IrOX electrodes fabricated at 750 °C (a), 800 °C (b), and 850 °C (c).
Sensors 19 04756 g004
Figure 5. pH response characteristics of IrOX electrodes fabricated at different oxidation temperatures.
Figure 5. pH response characteristics of IrOX electrodes fabricated at different oxidation temperatures.
Sensors 19 04756 g005
Figure 6. Long-term stability of IrOX pH electrodes fabricated at 750 °C (a), 800 °C (b), and 850 °C (c).
Figure 6. Long-term stability of IrOX pH electrodes fabricated at 750 °C (a), 800 °C (b), and 850 °C (c).
Sensors 19 04756 g006
Figure 7. X-ray photoelectron spectra of Ir4f for IrOX electrodes fabricated at 850 °C, (a) freshly prepared, (b) sample (a) after argon-ion etching for 90 s, (c) after aging for one month in distilled water, (d) sample (c) after argon-ion etching.
Figure 7. X-ray photoelectron spectra of Ir4f for IrOX electrodes fabricated at 850 °C, (a) freshly prepared, (b) sample (a) after argon-ion etching for 90 s, (c) after aging for one month in distilled water, (d) sample (c) after argon-ion etching.
Sensors 19 04756 g007
Figure 8. X-ray photoelectron spectra of O1s for IrOX electrodes fabricated at 850 °C: (a) freshly prepared; (b) sample; (a) after argon-ion etching for 90 s: (c) after aging for one month in distilled water, (d) sample (c) after argon-ion etching for 90 s.
Figure 8. X-ray photoelectron spectra of O1s for IrOX electrodes fabricated at 850 °C: (a) freshly prepared; (b) sample; (a) after argon-ion etching for 90 s: (c) after aging for one month in distilled water, (d) sample (c) after argon-ion etching for 90 s.
Sensors 19 04756 g008
Table 1. Binding energies and relative contents of different iridium species, as observed from the Ir4f X-ray photoelectron spectra of electrodes fabricated at 850 °C.
Table 1. Binding energies and relative contents of different iridium species, as observed from the Ir4f X-ray photoelectron spectra of electrodes fabricated at 850 °C.
ElectrodeSpeciesBinding Ir4f7/2 (eV) EnergiesRelative Content (%)
Freshly preparedIr3+6213.1
Ir4+6346.3
Ir6+63.8440.6
Freshly prepared and etchedIr3+61.2543.7
Ir4+62.4931.7
Ir6+63.6524.6
Aged for one monthIr3+61.8937
Ir4+62.8735.4
Ir6+63.5327.6
Aged for one month and etchedIr3+61.1239.5
Ir4+62.0933.2
Ir6+63.527.4
Table 2. Binding energies and relative contents of different oxygen species, as observed from the O 1s X-ray photoelectron spectra of electrodes fabricated at 850 °C.
Table 2. Binding energies and relative contents of different oxygen species, as observed from the O 1s X-ray photoelectron spectra of electrodes fabricated at 850 °C.
ElectrodeSpeciesBinding Energies of Ir4f7/2 (eV)Relative Content (%)
Freshly preparedOxide529.7820.9
Hydroxide531.2767.4
H2O532.7411.7
Freshly prepared and etchedOxide529.662.1
Hydroxide531.1725.8
H2O532.0612.1
Aged for one monthOxide529.5516.8
Hydroxide530.949.2
H2O532.7434
Aged for one month and etchedOxide529.953.8
Hydroxide531.3232.4
H2O532.1513.8
Table 3. Atomic-ratio of Ir3+/Ir4+ and O1s/Ir4f.
Table 3. Atomic-ratio of Ir3+/Ir4+ and O1s/Ir4f.
ElectrodeIr3+/Ir4+O1s/Ir4f
Freshly prepared0.37.4
Freshly prepared and etched1.43.9
Aged for one month16.8
Aged for one month and etched1.22.1

Share and Cite

MDPI and ACS Style

Wang, P.; Guo, T.; Zhao, T.; Du, Z.; Jin, Z.; Dong, B.; Li, Z. Effect of Oxidization Temperatures and Aging on Performance of Carbonate Melt Oxidized Iridium Oxide pH Electrode. Sensors 2019, 19, 4756. https://doi.org/10.3390/s19214756

AMA Style

Wang P, Guo T, Zhao T, Du Z, Jin Z, Dong B, Li Z. Effect of Oxidization Temperatures and Aging on Performance of Carbonate Melt Oxidized Iridium Oxide pH Electrode. Sensors. 2019; 19(21):4756. https://doi.org/10.3390/s19214756

Chicago/Turabian Style

Wang, Penggang, Tengfei Guo, Tiejun Zhao, Zhenxing Du, Zuquan Jin, Biqin Dong, and Zhe Li. 2019. "Effect of Oxidization Temperatures and Aging on Performance of Carbonate Melt Oxidized Iridium Oxide pH Electrode" Sensors 19, no. 21: 4756. https://doi.org/10.3390/s19214756

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop