Development of a Micro/Nano Probing System Using Double Elastic Mechanisms

1
School of Instrument Science and Opto-electronics Engineering, Hefei University of Technology, Hefei 230009, China
2
School of Mechanical Engineering, Dalian University of Technology, Dalian 116024, China
*
Author to whom correspondence should be addressed.
Sensors 2018, 18(12), 4229; https://doi.org/10.3390/s18124229
Submission received: 18 October 2018 / Revised: 28 November 2018 / Accepted: 28 November 2018 / Published: 2 December 2018
(This article belongs to the Special Issue Sensors for Precision Dimensional Metrology)

Abstract

:
To meet the requirement of high precision measurement of coordinate measurement machine system, a compact microprobe has been designed for 3D measurement in this paper. Aiming to reduce the influences of signal coupling during the probing process, the probe has been designed by adopting two elastic mechanisms, in which the horizontal and vertical motions of the probe tip can be separated by differential signals of quadrant photodetectors in each elastic mechanism. A connecting rod has been designed to transfer the displacement of the probe tip in vertical direction from lower to upper elastic mechanisms. The sensitivity models in horizontal and vertical directions have been established, and the sensor sensitivity has been verified through experiments. Furthermore, the signal coupling of three axes has been analyzed, and mathematical models have been proposed for decoupling. The probing performance has been verified experimentally.

1. Introduction

In the past 20 years, microprobes have been in rapid development to meet the enormous demand of the precision engineering industry [1,2,3]. In particular, to characterize a 3D component with micrometer to millimeter scale features and requiring nanometer to submicron precision is required, such as micro holes, micro lens, micro fluid channels and other microcomponents with high aspect ratio or transparent material. As a versatile metrology tool, coordinate measuring machines (CMMs) with a microprobing system are extensively used as prospective measuring instruments. With regard to the coordinate metrology, the probe system is a critical element that limits the overall accuracy of a measurement.
Many microprobes have been proposed in the field of micro metrology. The first recognizable microprobe for CMMs was developed by PTB (Physikalisch-Technische Bundesanstalt, Germany) in 1998 [4]. This probe used an optoelectronic system as the sensing device. The design concept of using optical sensors was then adopted by many research institutions like Eindhoven University of Technology [5], Korea Advanced Institute of Science and Technology [6], the Nation Institute of Standards and Technology [7], Harbin Institute of Technology [8] and Interstate University of Applied Sciences and Technology [9]. The probe with optical sensors usually chose fiber stylus to detect the images observed by the ball tip. Most of the optical probes couldn’t fulfil the requirement of high accuracy, due to the fact that their measuring accuracy could be easily influenced by the surface or the material of the measured piece, and the pixels of the image processing instrument were also limited. High resolution could be also obtained by confocal sensor [10], quadrant photo detector [11], Fizeau interferometer [12], white light interference [13,14,15] and other complex photoelectric sensing technology in the probe design. However, the assembly of these sensors is costly and the volume of the assembled probes are usually bulky, which is not suitable for the measurement of micro devices.
Other highly sensitive sensing devices with high response speeds have been adopted in the probe design. PTB developed a microprobe on the basis of a microfabricated silicon membrane, whose position was measured using diffused piezo-resistive sensors [16]. The system allowed measurement forces in the range of few micro-Newtons. The horizontal and vertical stiffness was approximately 2 mN/µm and 48 mN/µm, respectively. The approximate design of the contact probe that used piezo-resistive sensors typically required an elastic element with methods of microsystem technologies, such as the probe designs from the Braunschweig University of Technology in Germany [17,18], Eindhoven University of Technology in Netherlands [19,20], and Tianjin University in China [21]. On the one hand, the stiffness of the suspensions in these probes is limited by the manufacturing tolerances, and it is difficult to achieve equal stiffness in all three axes in practice. On the other hand, the resolution can be improved, but the cost will be high. NPL (National Physical Laboratory, London, UK) developed a microprobe that had a light structure and low probing force [22,23]. The motion of the probe tip was sensed by capacitive sensors. The high resolution of capacitive sensors can be exploited in the probe design. However, nonlinearity emerges during the measuring process, and the measurement range is limited. METAS (Federal Institute of Metrology, Bern, Switzerland) adopted an alternative approach that exhibited perfectly isotropic probing forces through a monolithic spring box mechanism with flexure hinges [24,25]. The movement in each direction could be measured using an inductor sensor. The stiffness was 20 mN/mm, and the repeatability was approximately 5 nm. The detection principle by using a vibrating element as the sensing device could also be adopted in the probe design, such as the UMAP (Ultrasonic Micro and Accurate Probe) developed by Mitutoyo Corporation [26,27]. The repeatability of the probe was reported to be 0.1 µm, which became commercially available. The NPL also proposed a probe using a three-legged elastic mechanism with two piezo drives to generate vibrations [28,29]. Thus, the amplitude of vibrations could be sensed by a piezoelectric sensor. The contact force of the vibrating probe was much lower than other normal probes.
For ultraprecision measurement of micro devices with large aspect ratio, our group has previously developed several microprobes. The contact scanning probe based on four DVD (Digital Versatile Disc) pickup heads was conducted in the initial probe design [10]. The elastic mechanism of the four DVD pickup heads probe was composed of a wire-suspended floating plate and connected to a fiber stylus. Later, to reduce the size of four DVD pickup heads probe, a confocal sensor modified from the Blu-Ray pickup head was developed [30]. The developed probe based on the confocal sensor was verified to have a measurement resolution of 1 nm. However, the volume of the disc probe was still considerably bulky, and the measurement range was limited by the utilized focus sensor. To reduce the size of the probe and expand the measurement range, a 2D angle sensor modified from a DVD pickup head and a miniature Michelson interferometer was developed for the probing system to measure the angle and displacement of the probe [31]. In several experiments, we found difficulties in maintaining a uniform tension of the suspending wires during the assembly. This problem can be solved by replacing the spring wire with a patterned leaf spring made of beryllium bronze sheet, which can yield favorable reproducibility [32]. However, the horizontal scanning range of this newly probe is limited due to the separation of two interfered beam spots and the decrease of fringe intensity when the probe is touched in the horizontal direction. In order to expand the measurement range, a parallel optical path was adopted using two separate laser diodes [33]. This new parallel optical path probe could reach a measurement range of ±20 μm both in horizontal and vertical directions. Then, we used a quadrant photodetector (QPD) as the sensing device to simplify the light path [34]. However, the adopted signal sensor cannot distinguish the different signals between the horizontal and vertical directions. And the sensitivity of the single sensor probe in vertical direction is lower than horizontal direction. The single sensor probe can only be used as a touch trigger probe. Aiming to increase the functionality and the practicality of the probe, a new structure design for the 3D contact scanning probe is conducted to increase the accuracy of the measurement results.
In this study, the probe principle based on double elastic mechanisms is proposed. The probe can reduce the signal coupling of three axes by structure design. The sensitivity model in the horizontal and vertical directions has been established to estimate the sensitivity of the sensor. The coupling of three axes has been analyzed, and the mathematic models have been proposed for decoupling. Furthermore, the probing performance has been verified through experiments.

2. Probe Principle and Design

2.1. Principle of the Microprobe

The microprobe system is composed of two same optoelectronic sensors, two elastic mechanisms and a stylus, as shown in Figure 1. Each sensor consists of a laser diode (LD), a plane mirror and a quadrant photo detector (QPD). Each elastic mechanism consists of a patterned leaf spring and a floating plate with a plane mirror. A slender rod was designed to connect the two elastic mechanisms. The upper end of the connecting rod is designed as a hemispherical structure, so as to support the floating plate 2 in a non-constraint and point-contact way. The lower end of the connecting rod is installed on the floating plate 1 by a hollow frame in order to prevent blocking the light from LD 1. Figure 2 shows the photo of the elastic mechanisms and the rod. A laser beam from LD 1 is reflected by the plane mirror 1 in the lower elastic mechanism. The reflected beam will project on the photo surface of QPD 1. When a contact force is applied to the probe tip in horizontal, only the leaf springs 1 in the lower elastic mechanism will undergo an elastic deformation and then only the plane mirror 1 will have an angle along lateral axes. This motion will be simultaneously detected by QPD 1. Figure 3a depicts the sensing principle of the probe in this case. The light intensity detected by QPD 1 is converted into two voltage signals corresponding to the probe tip’s displacements in x and y directions (probe coordinate system). The two voltage signals are defined in Equations (1) and (2), respectively.
U x = k x × [ ( P a 1 + P b 1 ) ( P c 1 + P d 1 ) ] ,
U y = k y × [ ( P a 1 + P d 1 ) ( P b 1 + P c 1 ) ] ,
where P a 1 , P b 1 , P c 1 , P d 1 are the optical powers detected by the four photodiodes of QPD 1 respectively. k x , k y are constants.
Similarly, when a contact force is applied in vertical, the action on the probe tip will be transferred to the upper elastic mechanism through the connecting rod. The floating plate 1 will only have a translation in vertical since its angular displacement is limited by the centrally symmetrical leaf spring 1; the x signal from QPD 1 will have a new output that is a cross-talk error and can be reduced by the decoupling method in Section 4. The floating plate 2 will have both a translation in vertical and an angular displacement, the signal from QPD 2 corresponding to the u x signal of QPD1 will also have a new output, which is used as the sensing signal in vertical, the sensing principle is shown in Figure 3. Therefore, the voltage signal corresponding to probe tip’s displacement in z direction can be defined as follows:
U z = k z × [ ( P a 2 + P b 2 ) ( P c 2 + P d 2 ) ] .
where P a 2 , P b 2 , P c 2 , P d 2 are the optical powers detected by the four photodiodes of QPD 2 respectively. k z is a constant.

2.2. Uniform Stiffness

The stiffness of the probe should be isotropic and controlled to less than 1 mN/μm [31]. To obtain an appropriate stiffness, two patterned leaf springs and floating plates as Figure 1 and Figure 2 shows were designed. Simulation analysis using ANSYS software was performed to design and optimized the elastic mechanisms since it is difficult to calculate the mechanical model of the complex leaf spring. Several parameters of the elastic mechanism affect the stiffness of the probe directly, such as material, diameters, widths, thickness and the length of the stylus. Beryllium bronze was used to fabricate the leaf springs because of its excellent elasticity. According to the specification of the selling stylus, the length of the stylus was determined as 10 mm. The remained parameters were tried, analyzed and optimized repeatedly using ANSYS software until achieving a uniform stiffness of less than 1 mN/μm in 3D. Most of the appropriate parameters of the elastic mechanisms are shown in Figure 4 and the others are listed in Table 1. The simulation results using the optimized parameters are shown in Figure 5. From Figure 5, we can see that the probe tip will have a displacement of 1.16 μm and 1.12 μm in horizontal and vertical respectively when a 1 mN force is applied. The simulation results (Figure 5 shows) indicate that the probe tip has a uniform stiffness of 0.86 mN/μm in lateral directions and a quite consistent stiffness of 0.89 mN/μm in vertical direction. Mode analysis was also conducted by ANSYS software to investigate the dynamic characteristics of the probe, the results are shown in Figure 6. The first order natural frequency of the probe should be high than 60 Hz [31]. From Figure 6 we can find that the natural frequencies of the probe are 164.5 Hz (first order), 203.9 Hz (second order), 309.7 Hz (third order) respectively, which is high enough for the probe.

2.3. Sensitivity Analysis

In the sensing unit, the reflected beam will project on the QPD at the center of the active area. The displacement of the light spot induced by the mirror deflection can be detected when a contact force is applied. Figure 7 exhibits the detecting principle in horizontal. The upper elastic mechanism is not affected by the probe tip’s motion in this case, because the slender rod supports the floating plate 2 in a non-constraint and point-contact way and θ is smaller than 7.2″ ( δ h < 20 μm, l = 10 mm). We can obtain that the caused vertical displacement of floating plate 2 is less than 1 nm when the probe tip is pushed 20 μm in horizontal. The light beam with an incident angle is reflected by the plane mirror 1 and focused onto the surface of the QPD 1. The focused spot is shifted from point A to point B on the surface of the QPD 1. From the geometrical relationship of the light path, we have:
A B ¯ = p × tan 2 θ ,
sin θ = δ h / l ,
where p is the distance between the QPD surface and the reflecting point on the plane mirror; l is the length of the stylus. The output voltage U h is proportional to the displacement of the focused spot and U h = k h × A B ¯ ( k h is the constant and can be calibrated). A relationship between displacement δ h and sensitivity S h can be described by the following equation:
S h = U h δ h ,
Because θ is a considerably small angle, the approximations sin θ θ , tan 2 θ 2 θ can be made, then S h can be described as follows:
S h = U h δ h = k h × p × tan 2 θ l × sin θ 2 k h × p l ,
Similarly, Figure 8 illustrates the deflection and sensing principle of the probe when a contact force is applied along vertical direction. Mirror 2 has a translation in vertical and an angular displacement, which cause an optical spot shift of C B ¯ and A C ¯ . With respect to the geometry depicted in Figure 8, we can obtain the total shift of the optical spot A B ¯ = A C ¯ + C B ¯ . In Δ O A C and Δ B C D , we have:
A C = p × tan 2 θ ,
C B ¯ = C D ¯ / cos 2 θ = F H ¯ / cos 2 θ ,
From the geometrical relationship of the optical path, we can derive F G ¯ δ v n × tan ϕ (because ϕ is considerably small); thus, in Δ O F H and Δ O F G , it can be obtained that:
F H ¯ = O F ¯ × sin [ π ( π 2 α 2 ϕ ) ] = O F ¯ × sin 2 ( α + ϕ ) ,
O F ¯ = F G ¯ sin α = δ v sin α ,
Here, α is the angle between incident light and plane mirror. Because all the reflect angles are small in practice and U v = k v × A B ¯ , the displacement δ v and sensitivity S v can be written as
S v = U v δ v = k v × [ p × tan 2 ϕ + δ v × sin ( 2 α + 2 ϕ ) sin α × cos 2 ϕ ] δ v ,
Since ϕ is small, α + ϕ α , cos 2 ϕ 1 , tan ϕ ϕ δ v / n (n is the distance from the reflected point of the light to the fixed point of the leaf spring). Equation (12) can be expressed as:
S v = U v δ v 2 k v × ( m n + cos α ) ,
Let Equation (7) is considered equal to Equation (13) to determine the uniform sensitivity, thus yielding
cos α = p l p n .
On the basis of the constrained conditions of uniform sensitivity and stiffness in three dimensions, the remained parameters have also been optimized as α = 45 ° , l = 10 mm, t = 1 mm, n = 14 mm, and p = 24 mm.

3. Experiments

Figure 9 shows the photos of the probe and an experimental setup to investigate the probe’s performance. The probe was assembled in a frame of a stand. A 2 mm × 2 mm square hole formed by four 0 grade gauge blocks was used to contact the probe tip in lateral directions. The thicknesses of the gauge blocks are 2 mm, 2 mm, 1.6 mm and 1.7 mm respectively. The gauge blocks with 2 mm thickness are clamped by the other two. Two parallel surfaces of square hole are the reference planes of the gauge block, but the other two are not. Therefore, the probe tip can be contacted in a lateral direction by the two reference planes of the square hole. The contacting direction can be changed by rotating the shaft of the probe, assisting with a round dial with an accuracy of 0.5 degree. A 2D linear stage was used to change the initial position of the square hole manually. A high-precision 3D nano-positioning stage (Physik Instrument (PI), model P561.3 CD with 2 nm repeatability and 100 μm travel, Karlsruhe, Germany) was used as a displacement reference.

3.1. Drift

Stability is a key parameter of the probe, which needs to be tested. Considering that most of the components can be measured within half an hour, the drift of the probe was tested about 30 min in this study. Figure 10a depicted the drift of the probe in 30 min when the temperature was not controlled. The drift reaches about 60 nm, which caused by the temperature fluctuation. A similar test was also conducted when the probe was covered by a constant temperature chamber developed by ourselves [35]. Figure 10b demonstrated the results when the temperature’s fluctuation was controlled within less than 0.13 °C (Figure 11 shows). The drift is approximately 23 nm in 30 min, which can be reduced furthermore if a more stable temperature was achieved.

3.2. Repeatability

The behavior of single-point probing was also tested through the experimental setup as Figure 9 shows. The probe tip was contacted by the reference planes of the high-accuracy gauge blocks driven by the PI stage along the X+, X−, Y+, Y−, and Z directions respectively. For each touching action, the probe tip was pushed 1 µm, the output signals were recorded and then the gauge blocks were moved back. The procedure was repeated 10 times in each point. Figure 12 exhibited the repeatability measurement results of the probe. The maximum standard deviation is 18.2 nm within the measurement range of ± 10 × ± 10 × 10 µm (X × Y × Z).

3.3. Resolution

The resolution of the probe in its measurement range is an important parameter that determines the precision of the whole CMM system. An experiment was conducted to further verify the measurement resolution of the probe by using the experimental setup as Figure 9 shows. At the beginning of the experiment, the stage was controlled to drive the gauge block to push the probe tip and then reverted to the initial position. The probe was pushed forward three steps and then retracted. The moving step was set from 10 nm to 5 nm with an interval of 1 nm in sequence and the resolution could be determined by the minimum clear step. Figure 13 illustrates that the resolution of the probing system is better than 5 nm.

4. Decoupling of the Cross-Talk Errors

Cross-talk errors usually exist because of the misalignment between probe coordinate systems and CMM stage coordinate systems, which should be calibrated and decoupled. Multiple regression methods can be applied and a transfer matrix can be derived to transform the outputs of the probe ( U x , U y , U z ) to the displacements of the probe tip (X, Y, Z) in a CMM stage coordinate system [36].
( X 1 Y 1 Z 1 X 2 Y 2 Z 2 X n Y n Z n ) = ( U x 1 U y 1 U z 1 1 U x 2 U y 2 U z 2 1 U x n U y n U z n 1 ) ( a 1 a 2 a 3 b 1 b 2 b 3 c 1 c 2 c 3 d 1 d 2 d 3 ) ,
In order to calculate the transfer matrix, a test was conducted in X+, X, Y+, Y and Z direction respectively with an interval of 1 μm. The experimental setup as Figure 9 shows and a similar measuring process of repeatability test was also used. In order to minimize the influence of the random errors, 10 tests were repeated in each position and the average values shown in Figure 14 were used. Figure 14 shows that the probe has a basically uniform sensitivity of 0.3 V/μm. The difference could be caused by the manufacturing and assembling errors.
Substituting the test data shown in Figure 14 into Equation (15), we can obtain the following transfer matrix by applying the least-squares methods.
X ^ 0 = ( a 10 a 20 a 30 b 10 b 20 b 30 c 10 c 20 c 30 d 10 d 20 d 30 ) = ( 2.44965 0.032749 1.10654 2.41337 3.002049 4.181964 1.199704 0.09759 2.06242 0.0019 0.003539 0.1431 ) ,
The accuracy of the estimated values can also be obtained from the error matrix obtained by the least-squares method [37]:
σ = ( 5.1 4.0 6.2 5.3 4.1 6.3 4.1 3.2 4.9 5.9 4.6 7.1 )
The above calibration experiments were conducted again to verify the effectiveness of the obtained transfer matrix. The probe was tested one time in each position. The errors of the decoupling model in each axis shown in Figure 15a can be obtained by comparing the displacements of the PI stage with the values calculated by the transfer matrix. Figure 15a illustrates that the probe error caused by the decoupling model was less than ±50 nm in each direction. Figure 15b–d presents the decoupled results of the probe when the probe was contacted along X, Y and Z directions respectively. The cross-talk errors have been reduced from several micrometers to 390 nm. The nonlinearity of the line X in Figure 15b, line Y in Figure 15c and line Z in Figure 15d are 0.28%, 0.19% and 0.42% respectively.

5. Conclusions

This study presents an innovative probe with low interference between horizontal and vertical measurement by adopting two elastic mechanisms. A connecting rod with a hemispherical end was used to transfer vertical displacement of the probe tip to the upper elastic mechanism. The horizontal motions of the probe tip can only be detected by the lower elastic mechanism and its sensor, while the upper elastic mechanism and its sensor can only detect the vertical motions of the probe tip. The probe has been designed, analyzed and optimized. The simulation results verified that the probe has a uniform stiffness in 3D. The experimental results show that the probe has a repeatability of 20 nm, a resolution of 5 nm, an analog measurement range of larger than ±10 μm, and a stability of 23 nm within 30 min when the fluctuation of the environment temperature was controlled within 0.13 °C. The crosstalk errors of the probe have been compensated. The probe can be used in the coordinate measurement machine and any other precision measurement systems. A reference sphere will be tested by the developed probe to evaluate its 3D accuracy in future.

Author Contributions

R.-J.L. and K.-C.F. conceived the probe. R.-J.L. and P.X. wrote the manuscript. P.X. designed and analyzed the probe. P.X. and P.-Y.W. performed the experiments and analyzed the data. R.-J.C. edited the manuscript. Q.-X.H. provided the experimental program.

Funding

This work was funded by the National Natural Science Foundation of China (Grant No. 51,675,157 and 51805136), the Fundamental Research Funds for the Central Universities (JZ2018HGTA0211 and JZ2017HGBZ0943).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Alshehri, A.; Kraft, M.; Gardonio, P. Two-mass MEMS velocity sensor: Internal feedback loop design. IEEE Sens. J. 2013, 13, 1003–1011. [Google Scholar] [CrossRef]
  2. Woźniak, A.; Dobosz, M. Influence of measured objects parameters on CMM touch trigger probe accuracy of probing. Precis. Eng. 2005, 29, 290–297. [Google Scholar] [CrossRef]
  3. Weckenmann, A.; Estler, T.; Peggs, G.; McMurtry, D. Probing systems in dimensional metrology. CIRP Ann. 2004, 53, 657–684. [Google Scholar] [CrossRef]
  4. Guilun, J.; Schwenke, H.; Trapet, E. Opto-tactile sensor for measuring small structures on coordinate measuring machines. Proc. ASPE 1998, 18, 25–28. [Google Scholar] [CrossRef]
  5. Van Vliet, W.P.; Schellekens, P.H. Development of a fast mechanical probe for coordinate measuring machines. Precis. Eng. 1998, 22, 141–152. [Google Scholar] [CrossRef]
  6. Kim, S.W.; Yang, D.Y. New design of precision CMM based upon volumetric phase-measuring interferometry. CIRP Ann. 2001, 50, 357–360. [Google Scholar] [CrossRef]
  7. Muralikrishnan, B.; Stone, J.A.; Stoup, J.R. Fiber deflection probe for small hole metrology. Precis. Eng. 2006, 30, 154–164. [Google Scholar] [CrossRef]
  8. Cui, J.W.; Li, L.; Tan, J.B. Optical fiber probe based on spherical coupling of light energy for inner-dimension measurement of microstructures with high aspect ratios. Opt. Lett. 2011, 36, 4689–4691. [Google Scholar] [CrossRef]
  9. Ettemeyer, A. New three-dimensional fiber probe for multisensory coordinate measurement. Opt. Eng. 2012, 51, 1502. [Google Scholar] [CrossRef]
  10. Fan, K.C.; Cheng, F.; Wang, W.; Chen, Y.J.; Lin, J.Y. A scanning contact probe for a micro-coordinate measuring machine (CMM). Meas. Sci. Technol. 2010, 21, 054002. [Google Scholar] [CrossRef]
  11. Enami, K.; Kuo, C.C.; Nogami, T. Development of nano-probe system using optical sensing. In Proceedings of the IMEKO-XV World Congress, Wien, Austria, 25–28 September 2000. [Google Scholar]
  12. Liebrich, T.; Knapp, W. New concept of a 3D-probing system for micro-components. CIRP Ann. 2010, 59, 513–516. [Google Scholar] [CrossRef] [Green Version]
  13. Yang, W.; Liu, X.; Lu, W.; Yu, N.; Chen, L.; Zhou, L.; Chang, S. A novel white light interference based AFM head. J. Lightwave Technol. 2017, 35, 3604–3610. [Google Scholar] [CrossRef]
  14. Lei, Z.; Liu, X.; Zhao, L.; Chen, L.; Li, Q.; Yuan, T.; Lu, W. A novel 3D stitching method for WLI based large range surface topography measurement. Opt. Commun. 2016, 359, 435–447. [Google Scholar] [CrossRef]
  15. Lei, Z.; Liu, X.; Chen, L.; Lu, W.; Chang, S. A novel surface recovery algorithm in white light interferometry. Measurement 2016, 80, 1–11. [Google Scholar] [CrossRef]
  16. Buetefisch, S.; Dai, G.; Danzebrink, H.U. Novel design for an ultra high precision 3D micro probe for CMM applications. Procedia Eng. 2010, 5, 705–712. [Google Scholar] [CrossRef] [Green Version]
  17. Peiner, E.; Balke, M.; Doering, L. Tactile probes for dimensional metrology with microcomponents at nanometre resolution. Meas. Sci. Technol. 2008, 19, 064001. [Google Scholar] [CrossRef]
  18. Tibrewala, A.; Phataralaoha, A.; Büttgenbach, S. Development, fabrication and characterization of a 3D tactile sensor. J. Micromech. Microeng. 2009, 19, 125005. [Google Scholar] [CrossRef]
  19. Pril, W. Development of High Precision Mechanical Probes for Coordinate Measuring Machines. Ph.D. Thesis, Eindhoven University of Technology, Eindhoven, The Netherlands, 2002. [Google Scholar]
  20. Haitjema, H.; Pril, W.O.; Schellekens, P.H.J. Development of a silicon-based nanoprobe system for 3-D measurements. CIRP Ann. 2001, 50, 365–368. [Google Scholar] [CrossRef]
  21. Yuan, L. Study of MEMS Piezo-Resistance 3D Micro Tactile Probe and Application in NMM. Ph.D. Thesis, Tianjing University of Technology, Tianjing, China, 2007. [Google Scholar]
  22. Murphy, J.; Leach, R. The design of coordinate measuring probe for characterising truly three-dimensional micro-structures. In Proceedings of the 4th Euspen International Conference: European Society for Precision Engineering and Nanotechnology, Glasgow, UK, 31 May–2 June 2004. [Google Scholar]
  23. Peggs, G.N.; Lewis, A.J.; Oldfield, S. Design for a compact high-accuracy CMM. CIRP Ann. 1999, 48, 417–420. [Google Scholar] [CrossRef]
  24. Meli, F.; Küng, A.; Thalmann, R. Ultraprecision micro-CMM using a low force 3D touch probe. Meas. Sci. Technol. 2005, 18. [Google Scholar] [CrossRef]
  25. Thalmann, R.; Meli, F.; Küng, A. State of the art of tactile micro coordinate metrology. Appl. Sci. 2016, 6, 150. [Google Scholar] [CrossRef]
  26. Masuzawa, T.; Hamasaki, Y.; Fujino, M. Vibroscanning method for nondestructive measurement of small holes. CIRP Ann. 1993, 42, 589–592. [Google Scholar] [CrossRef]
  27. Kim, B.; Masuzawa, T.; Bourouina, T. The vibroscanning method for the measurement of micro-hole profiles. Meas. Sci. Technol. 1999, 10, 697. [Google Scholar] [CrossRef]
  28. Claverley, J.D.; Leach, R.K. A vibrating micro-scale CMM probe for measuring high aspect ratio structures. Microsyst. Technol. 2010, 16, 1507–1512. [Google Scholar] [CrossRef]
  29. Claverley, J.D.; Georgi, A.; Leach, R.K. Modelling the interaction forces between an ideal measurement surface and the stylus tip of a novel vibrating micro-scale CMM probe. In Proceedings of the 5th International Precision Assembly Seminar, Chamonix, France, 14–17 February 2010; Volume 315, pp. 131–138. [Google Scholar]
  30. Li, R.J.; Fan, K.C.; Huang, Q.X.; Qian, J.Z.; Gong, W.; Wang, Z.W. Design of a large-scanning-range contact probe for nano-coordinate measurement machines. Opt. Eng. 2012, 51, 081503. [Google Scholar] [CrossRef]
  31. Li, R.J.; Fan, K.C.; Miao, J.W.; Huang, Q.X.; Tao, S.; Gong, E.M. An analogue contact probe using a compact 3D optical sensor for micro/nano coordinate measuring machines. Meas. Sci. Technol. 2014, 25, 094008. [Google Scholar] [CrossRef]
  32. Li, R.J.; Wang, P.Y.; Li, D.D.; Fan, K.C.; Liu, F.F.; Chen, L.J.; Huang, Q.X. Precision Manufacturing of Patterned Beryllium Bronze Leaf Springs via Chemical Etching. Appl. Sci. 2018, 8, 1476. [Google Scholar] [CrossRef]
  33. Li, R.J.; Fan, K.C.; Huang, Q.X.; Zhou, H.; Gong, E.M.; Xiang, M. A long-stroke 3D contact scanning probe for micro/nano coordinate measuring machine. Precis. Eng. 2016, 43, 220–229. [Google Scholar] [CrossRef]
  34. Li, R.J.; Xiang, M.; He, Y.X.; Fan, K.C.; Cheng, Z.Y.; Huang, Q.X.; Zhou, B. Development of a high-precision touch-trigger probe using a single sensor. Appl. Sci. 2016, 6, 86. [Google Scholar] [CrossRef]
  35. Feng, J.; Li, R.J.; Fan, K.C.; Zhou, H.; Zhang, H. Development of a low-cost and vibration-free constant-temperature chamber for precision measurement. Sens. Mater. 2015, 27, 329–340. [Google Scholar] [CrossRef]
  36. Park, J.; Kwon, K.; Cho, N. Development of a coordinate measuring machine (CMM) touch probe using a multi-axis force sensor. Meas. Sci. Technol. 2006, 17, 2380. [Google Scholar] [CrossRef]
  37. Markovsky, I.; Van Huffel, S. Overview of total least-squares methods. Signal Process. 2007, 87, 2283–2302. [Google Scholar] [CrossRef] [Green Version]
Figure 1. Structure of the probe.
Figure 1. Structure of the probe.
Sensors 18 04229 g001
Figure 2. Photo of the elastic mechanisms and the rod.
Figure 2. Photo of the elastic mechanisms and the rod.
Sensors 18 04229 g002
Figure 3. Sensing principle of the probe when the probe tip has a displacement (a) in horizontal (b) in vertical.
Figure 3. Sensing principle of the probe when the probe tip has a displacement (a) in horizontal (b) in vertical.
Sensors 18 04229 g003
Figure 4. Dimensions of the leaf spring 1 (left), leaf spring 2 (middle) and the floating plate 1 and 2 (right), (unit: mm).
Figure 4. Dimensions of the leaf spring 1 (left), leaf spring 2 (middle) and the floating plate 1 and 2 (right), (unit: mm).
Sensors 18 04229 g004
Figure 5. Simulation results when a 1 mN force is applied to the probe tip: (a) in horizontal; (b) in vertical.
Figure 5. Simulation results when a 1 mN force is applied to the probe tip: (a) in horizontal; (b) in vertical.
Sensors 18 04229 g005
Figure 6. Mode analysis results of the probe: first order (left), second order (middle), third order (right).
Figure 6. Mode analysis results of the probe: first order (left), second order (middle), third order (right).
Sensors 18 04229 g006
Figure 7. Sensing principle in horizontal direction.
Figure 7. Sensing principle in horizontal direction.
Sensors 18 04229 g007
Figure 8. Sensing principle in vertical direction.
Figure 8. Sensing principle in vertical direction.
Sensors 18 04229 g008
Figure 9. The probe and experimental setup.
Figure 9. The probe and experimental setup.
Sensors 18 04229 g009
Figure 10. Drifts of the probe: (a) the temperature was not controlled; (b) the temperature was controlled.
Figure 10. Drifts of the probe: (a) the temperature was not controlled; (b) the temperature was controlled.
Sensors 18 04229 g010
Figure 11. Temperature in the chamber.
Figure 11. Temperature in the chamber.
Sensors 18 04229 g011
Figure 12. Results of the repeatability measurement.
Figure 12. Results of the repeatability measurement.
Sensors 18 04229 g012
Figure 13. Results of the resolution test for the probe.
Figure 13. Results of the resolution test for the probe.
Sensors 18 04229 g013
Figure 14. Test results of the probe in X (a), Y (b) and Z (c) direction.
Figure 14. Test results of the probe in X (a), Y (b) and Z (c) direction.
Sensors 18 04229 g014
Figure 15. Results of the decoupling process: (a) residual errors; (b) in X direction; (c) in Y direction; (d) in Z direction.
Figure 15. Results of the decoupling process: (a) residual errors; (b) in X direction; (c) in Y direction; (d) in Z direction.
Sensors 18 04229 g015
Table 1. Parameters of the probe.
Table 1. Parameters of the probe.
ParametersValue
Material of the leaf springBeryllium–copper alloy
Young’s modulus of the leaf spring (Gpa)130
Material of the floating plateAluminum alloy
Weight of the floating plate (g)1.5
Thickness (mm)2
Young’s modulus of the floating plate (Gpa)71
Material of the stylusTungsten stylus with a ruby ball tip
Length of the stylus (mm)10
Diameter of the probe ball (mm)1
Young’s modulus of the stylus (Gpa)193
Distance between the two leaf springs (mm)34

Share and Cite

MDPI and ACS Style

Li, R.-J.; Xu, P.; Wang, P.-Y.; Fan, K.-C.; Cheng, R.-J.; Huang, Q.-X. Development of a Micro/Nano Probing System Using Double Elastic Mechanisms. Sensors 2018, 18, 4229. https://doi.org/10.3390/s18124229

AMA Style