Real-Time Monitoring of Stem Cells by Diamond-Based Impedance Sensors †

1
Institute of Physic, Czech Academy of Sciences v.v.i., Cukrovarnická 10, 162 00 Prague 6, Czech Republic
2
Faculty of Electrical Engineering, Czech Technical University in Prague, Technická 2, 166 27 Prague, Czech Republic
3
Institute of Physiology, Czech Academy of Sciences v.v.i., Videnska 1083, 14220 Prague 4, Czech Republic
4
Faculty of Biomedical Engineering, Czech Technical University in Prague, nám. Sítná 3105, 272 01 Kladno, Czech Republic
5
Faculty of Civil Engineering, Czech Technical University in Prague, 166 29 Prague, Czech Republic
*
Author to whom correspondence should be addressed.
Presented at the Eurosensors 2017 Conference, Paris, France, 3–6 September 2017.
Proceedings 2017, 1(4), 515; https://doi.org/10.3390/proceedings1040515
Published: 23 August 2017
(This article belongs to the Proceedings of Proceedings of Eurosensors 2017, Paris, France, 3–6 September 2017)

Abstract

:
Cell-based impedance spectroscopy is a promising label-free method for electrical monitoring of cell activity. Here we present a diamond-based impedance sensor with built-in gold interdigitated electrodes (IDEs) as a promising platform for simultaneous electrical and optical monitoring of adipose-derived stem cells (ASC). The impedance spectra were collected in a wide frequency range (from 100 Hz to 50 kHz) for 27 h of cultivation in chambers designed for static cultivation. Absolute impedance spectra were analyzed in terms of measurement frequencies and cell properties monitored by high-resolution digital camera.

1. Introduction

Bioelectrical sensors are of high interest due to non-invasive label-free techniques for in vitro monitoring of biological events and due to simplicity and fast response in real time. Presently, there are two analytic approaches which employ either optical or electronic signal processing [1]. Both these approaches reached state of the art in a transducer type which converts a stimulus-induced cellular response into the quantifiable signal (i.e., biosensor signal). From the broad family of electronic systems, impedance measurements seem to be one of the simplest and still a powerful method for monitoring cellular signals during the cell cultivation. As presented in many reports, the monitored impedance signal is sensitive not only to ionic currents but also to cell growth stages (i.e., cell attachment, spreading, shape, proliferation, differentiation and communication) [2].
Diamond is proposed as a promising material for life science and regenerative medicine due to its biocompatibility, chemical stability and favorable combination of optical, mechanical and electrical properties. In addition, its surface can be covalently terminated by specific atoms or molecules which control the cell occupation, adhesion and cell differentiation [3]. In our previous studies, we have already introduced an optically transparent surface-conductive diamond thin film as a functional layer in impedance sensors for biological studies [4] and recognition of gas and chemical molecules [5,6]. In this study, an impedance sensor with the functional diamond layer deposited on gold interdigitated electrodes is employed to electrically and optically monitor the activity of adipose-derived stem cells.

2. Materials and Methods

The transducer of the sensor is represented by metal composite interdigitated electrodes (IDEs) (10 nm Ti and 80 nm Au) evaporated on a quartz substrate (Corning XG) (Figure 1a). The width/gap of electrode periodicities was 200 µm. Samples were coated with approx. 100 nm thick diamond layer using a linear antenna pulsed microwave plasma chemical vapor deposition system [7]. The deposition conditions were as follows: microwave power 2 × 1700 W, pressure 0.1 mbar, gas mixture 200/5/20 sccm of H2/CH4/CO2, temperature 400 °C and process time 50 h.
The impedance measurements were performed in chambers designed for the static cultivation (Figure 1b,c) with integrated unit for optical monitoring of cells during their cultivation. The cultivation area was 0.7 cm2 with 500 µL reservoir for medium. The adipose-derived stem cells (ASC) (i.e., human cells isolated from lipoaspirate obtained by liposuction) were seeded into the chamber with an initial density approx. 28,000 cells/cm2 in a DMEM medium with 10% of fetal bovine serum (FBS). Then the chambers were installed into time-lapse microscope system and connected to vector impedance analyzer (Bode 100, OMICRON electronics GmbH). The impedance spectra were collected in a frequency range from 100 Hz to 50 kHz for 27 h cultivation.
Simultaneously to these experiments, reference measurements with the same cultivation conditions were performed with an xCELLigence sensor (Roche Applied Science) which uses circular-shape Au IDE structures [2]. This system provides the so-called cell index defined as (Zi-Z0)/const, where Zi is the impedance measured at time i and Z0 is the starting impedance. Here the measurement frequencies are limited to three discrete frequencies: 10 kHz, 25 kHz, and 50 kHz.

3. Results and Discussion

Figure 2a shows SEM image of a diamond film grown on the active sensor area. The whole sensor area was homogenously covered with the diamond thin film with thickness approx. 100 nm. The brighter regions represent the diamond film grown on Au IDEs and darker regions correspond to the film grown on the quartz substrate. The diamond character of the deposited film was confirmed by Raman spectra measured at two different positions: on a quartz substrate and Au layer (i.e., on interdigitated electrodes). Both Raman spectra were similar and had typical features of CVD diamond films [8]. The spectra consist of a sharp diamond-peak centered at ~1332 cm−1 which is attributed to sp3 carbon bonds. Broad bands centered at 1350 cm−1 (D-band) and at ~1590 cm−1 (G-band) are related to sp2 carbon bonds which are commonly localized at grain boundaries (Figure 2b).
Figure 3 shows normalized absolute impedance spectra (|Z|) of the diamond coated Au sensor during ASC cells cultivation in the medium. Spectra are plotted for frequencies varied from 500 Hz to 50 kHz. Three different regions can be distinguished. In the first region (up to ~2.5 h), a sharp increase of |Z| followed by a short decrease is observed at low frequencies (see detail view in Figure 3b). For the cultivation time from 2.5 to 11 h (region II.), the normalized impedance spectra saturated and have a constant value. In the last region (time > 11 h), the normalized impedance increased monotonously reaching nearly the same value after 25 h (Figure 3a).
The rapid increase of |Z| at the beginning of measurements (region I) should be attributed to the temporary period, during which the cells are added to the cultivation medium, they are falling down and are deposited onto the sensor active area. During this temporary period, they can migrate on and start to adhere to the substrate. If the initial number of added cells is enough low and there is still reaming a space between them on the sensor area, the cells will proliferate, causing a gradual yet steady increase in |Z|. When the cells reach confluence, the |Z| value is at a constant value (region II), reflecting the fact that the monitored active sensor area, which is in touch with bulk media, is fully covered. The increase of |Z| in the region III should reflect change in the cultivation medium (e.g., due to the medium evaporation or the change in pH caused by cell metabolic activity, etc.).
The only difference between |Z| measured at different frequencies was observed in the first region related to the peak intensity (at ~1.4 h). As abovementioned, this |Z| peak maximum is related to the fully covered sensing surface by cells [9]. While at low frequencies (i.e., 500 Hz or 1 kHz, Figure 3a) this peak was significant, its intensity decreases with increasing frequencies and disappeared at high frequencies (50 kHz) (Figure 3b). At low frequencies, most of the electrical current flow through the solution channels under and between adjacent cells, while at higher frequencies, current capacitively couples and directly flow through the insulating cell membranes. It is known that the high-frequency impedance measurements are preferably affected by the cell coverage, whereas the low frequency measurement reflects changes in the space either under or between the cells [9,10]. Thus, the decrease of the |Z| observed at lower frequencies in the region II should be related to the changed cell morphology, cell adhesion or even it should be also caused by the cell delamination from the diamond surface. To note, reference measurements performed by the xCelligence system (not shown here) showed similar dependences as measurements provided on the diamond-based impedance sensors.

4. Conclusions

We employed an impedance sensor type using the functional diamond layer deposited on metal interdigitated electrodes for monitoring the adhesion, growth and metabolic activity of adipose-derived stem cells during their cultivation. The gold IDEs built-in thin diamond layer (100 nm) reveal a promising platform for simultaneous electrical and optical monitoring of cell events on functionally terminated surfaces. Wide frequency range measurements (100 Hz to 50 kHz) with time dependent impedance monitoring correlated with time-lapse optical microscopy allowed us to provide studies of the cell-cell and cell substrates interactions and investigate electrochemical processes at the electrode-electrolyte interface.

Acknowledgments

This work was supported by the Ministry of Health of the Czech Republic (grant No. AZV 15-33018A).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Fang, Y. Label-Free Biosensors for Cell Biology. Int. J. Electrochem. 2011, 2011, 1–16. [Google Scholar] [CrossRef]
  2. Guan, N.; Deng, J.; Li, T.; Xu, X.; Irelan, J.T.; Wang, M.-W. Label-free monitoring of T cell activation by the impedance-based xCELLigence system. Mol. Biosyst. 2013, 9, 1035. [Google Scholar] [CrossRef] [PubMed]
  3. Verdanova, M.; Rezek, B.; Broz, A.; Ukraintsev, E.; Babchenko, O.; Artemenko, A.; Izak, T.; Kromka, A.; Kalbac, M.; Hubalek Kalbacova, M. Nanocarbon Allotropes-Graphene and Nanocrystalline Diamond-Promote Cell Proliferation. Small 2016, 12, 2499–2509. [Google Scholar] [CrossRef] [PubMed]
  4. Izak, T.; Novotna, K.; Kopova, I.; Bacakova, L.; Rezek, B.; Kromka, A. H-terminated diamond as optically transparent impedance sensor for real-time monitoring of cell growth: H-terminated diamond as optically transparent impedance sensor. Phys. Status Solidi B 2013, 250, 2741–2746. [Google Scholar] [CrossRef]
  5. Davydova, M.; Stuchlik, M.; Rezek, B.; Larsson, K.; Kromka, A. Sensing of phosgene by a porous-like nanocrystalline diamond layer with buried metallic electrodes. Sens. Actuators B Chem. 2013, 188, 675–680. [Google Scholar] [CrossRef]
  6. Stehlik, S.; Izak, T.; Kromka, A.; Dolenský, B.; Havlík, M.; Rezek, B. Sensitivity of Diamond-Capped Impedance Transducer to Tröger’s Base Derivative. ACS Appl. Mater. Interfaces 2012, 4, 3860–3865. [Google Scholar] [CrossRef] [PubMed]
  7. Potocký, Š.; Čada, M.; Babchenko, O.; Ižák, T.; Davydova, M.; Kromka, A. Perspectives of linear antenna microwave system for growth of various carbon nano-forms and its plasma study: Perspectives of linear antenna microwave system. Phys. Status Solidi B 2013, 250, 2723–2726. [Google Scholar] [CrossRef]
  8. Prawer, S.; Nemanich, R.J. Raman spectroscopy of diamond and doped diamond. Philos. Trans. R. Soc. Math. Phys. Eng. Sci. 2004, 362, 2537–2565. [Google Scholar] [CrossRef] [PubMed]
  9. Applied Biophysics. Available online: http://www.biophysics.com/ecis-theory.php (accessed on 20 June 2017).
  10. Ižák, T.; Szabó, O.; Bačáková, L.; Kromka, A. Diamond Functional Layers for Cell-based Impedance Spectroscopy. Procedia Eng. 2016, 168, 614–617. [Google Scholar] [CrossRef]
Figure 1. (a) Schematic top- and cross-sectional draws of impedance sensor with built-in gold interdigitated electrodes covered with the diamond layer; (b) Cultivation chamber and (c) exploded 3D view of the chamber.
Figure 1. (a) Schematic top- and cross-sectional draws of impedance sensor with built-in gold interdigitated electrodes covered with the diamond layer; (b) Cultivation chamber and (c) exploded 3D view of the chamber.
Proceedings 01 00515 g001
Figure 2. (a) SEM image of diamond film and (b) corresponding normalized Raman spectra of the diamond layer deposited on Au IDE and the quartz substrate, respectively.
Figure 2. (a) SEM image of diamond film and (b) corresponding normalized Raman spectra of the diamond layer deposited on Au IDE and the quartz substrate, respectively.
Proceedings 01 00515 g002
Figure 3. (a) Normalized absolute impedance curves measured by diamond-coated built-in Au IDE sensor at various frequencies from 500 Hz to 50 kHz for ASC cells cultivated up to 27 h and (b) detailed view of selected curves (frequencies from 2 kHz to 50 kHz) at the beginning of measurements (up to 2.5 h). (The width/gap of electrodes was 200 μm).
Figure 3. (a) Normalized absolute impedance curves measured by diamond-coated built-in Au IDE sensor at various frequencies from 500 Hz to 50 kHz for ASC cells cultivated up to 27 h and (b) detailed view of selected curves (frequencies from 2 kHz to 50 kHz) at the beginning of measurements (up to 2.5 h). (The width/gap of electrodes was 200 μm).
Proceedings 01 00515 g003
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Procházka, V.; Matějka, R.; Ižák, T.; Szabó, O.; Štěpanovská, J.; Bačáková, L.; Kromka, A. Real-Time Monitoring of Stem Cells by Diamond-Based Impedance Sensors. Proceedings 2017, 1, 515. https://doi.org/10.3390/proceedings1040515

AMA Style

Procházka V, Matějka R, Ižák T, Szabó O, Štěpanovská J, Bačáková L, Kromka A. Real-Time Monitoring of Stem Cells by Diamond-Based Impedance Sensors. Proceedings. 2017; 1(4):515. https://doi.org/10.3390/proceedings1040515

Chicago/Turabian Style

Procházka, Václav, Roman Matějka, Tibor Ižák, Ondrej Szabó, Jana Štěpanovská, Lucie Bačáková, and Alexander Kromka. 2017. "Real-Time Monitoring of Stem Cells by Diamond-Based Impedance Sensors" Proceedings 1, no. 4: 515. https://doi.org/10.3390/proceedings1040515

Article Metrics

Back to TopTop