Yonsei Med J. 2004 Aug;45(4):748-750. English.
Published online Feb 17, 2009.
Copyright © 2004 The Yonsei University College of Medicine
Brief Communication

Management of Detachment of Pilot Balloon During Intraoral Repositioning of the Submental Endotracheal Tube

Kyung Bong Yoon,1 Byung Ho Choi,2 Hye Sook Chang,1 and Hyun Kyo Lim1
    • 1Department of Anesthesiology and Pain Medicine, Yonsei University Wonju College of Medicine, Wonju, Korea.
    • 2Department of Oral and Maxillofacial Surgery, Yonsei University Wonju College of Medicine, Wonju, Korea.
Received October 10, 2003; Accepted April 23, 2004.

Abstract

Submental endotracheal intubation for surgery was used as an alternative to nasotracheal intubation in patients with craniomaxillofacial injury. Generally extubation was performed in the operation room by pulling the tube through the submental incision site. When extubation is not indicated, intraoral indwelling is preferred to submental intubation. We report a case of a 35-year-old male patient with multiple facial bone fractures. At the end of the surgery, we noticed the oropharyngeal edema, and so the submental intubation was converted into a standard orotracheal intubation. During that procedure, the pilot balloon was accidentally detached from the endotracheal tube. The situation was managed by cutting a pilot tube from a new, unused endotracheal tube and connecting it to the intubated tube using a needle connector.

Keywords
Submental intubation; maxillofacial injuries; endotracheal tube


Metrics
Share
PERMALINK