A method for in vivo detection of abnormal subepidermal tissues based on dielectric properties

Abstract

In this paper, a convenient and noninvasive scanning imaging method using microwave frequency band for abnormal subepidermal tissues detection is proposed in the aim to improve diagnosis and prognosis in clinical environments. This method is based on the reflective detection technology with coaxial probe that is used to measure the dielectric properties of tissues. An improved equivalent circuit and simulated annealing algorithm (SA) were used in this work to analyze the dielectric properties of tissues. Computational simulations incorporating a simplified model of subcutaneous hemorrhage described in this work were used to evaluate this method. The dielectric properties data of tissues in the model of simulation is derived from the literature. The simulation results demonstrated the potential of this method to detect abnormal subepidermal tissues conveniently and expose them in the image accurately.