Title: Evaluation of BeO ceramic disks for thermally stimulated exoelectron dosimetry

Technical Report ·
DOI:https://doi.org/10.2172/7259378· OSTI ID:7259378

BeO ceramic disks were evaluated for application as dosimeters for both penetrating and non-penetrating radiation. Response to penetrating radiation was measured as thermoluminescence (TL) and to non-penetrating radiation as thermally stimulated exoelectron emission (TSEE). Field experiments demonstrated that both TSEE and TL responses from BeO can monitor diverse radiation fields. BeO disks in a passive dosimeter were found to be sensitive to a lower exposure level of 100 pCi-day/liter of radon. The depth of the more active exoelectron layer in BeO was found to be 4 ..mu..m. A second less active, exoelectron layer extends to a depth of at least 16 ..mu..m.