Abstract
Graphenic carbon (GC) x-ray transmission windows for EDX and XRF applications with a high transparency for x-rays below 2.5 keV have been fabricated on 6 inch wafers with a CMOS-compatible CVD process. GC windows with an open diameter of 7.4 mm and a thickness of 770 nm withstand up to 6.5 bars of differential pressure. A high transmissivity of 40 % for fluorine Kα (0.677 keV) radiation is demonstrated for a GC thickness of 650 nm. The GC membranes outperform beryllium (Be) windows, in terms of higher x-ray transmission and better mechanical stability while avoiding the toxicity of Be. Optical profilometry has been employed to visualize a large deformation of the GC layer during the window fabrication. This seems to limit the thickness of the GC windows that can currently be fabricated. A two-step growth process can overcome these limitations and windows with a thickness of up to 6 µm have been realized.
Similar content being viewed by others
References
S. Huebner, N. Miyakawa, S. Kapser, A. Pahlke, and F. Kreupl, IEEE Trans. Nucl. Sci. 62 (2), 588 (2015).
S. Huebner, N. Miyakawa, A. Pahlke, and F. Kreupl, Phys. Status Solidi B 252 (11), 2564 (2015).
F. Scholze and M. Procop, X-Ray Spectrom. 34 (6), 473 (2005).
P. Torma and H. Sipila, IEEE Trans. Nucl. Sci. 60 (2), 1311 (2013).
F. Kreupl, MRS Proceedings 1303, mrsf10-1303-y02-01-b2-0 (2011)
B. L. Henke, E. M. Gullikson, and J. C. Davis, At. Data Nucl. Data Tables 54, (2), 181 (1993).
O. L. Blakslee, D. G. Proctor, E. J. Seldin, G. B. Spence, and T. Weng, J. Appl. Phys. 41 (8), 3373 (1970).
C. A. Klein, J. Appl. Phys. 88 (9), 5487 (2000).