IFCC Guideline for sampling, measuring and reporting ionized magnesium in plasma

  • International Federation of Clinical Chemistry and Laboratory Medicine (IFCC): IFCC Scientific Division Committee on Point of Care Testing: Mohammed C. Ben Rayana , Robert W. Burnett , Arthur K. Covington , Paul D'Orazio , Niels Fogh-Andersen , Ellis Jacobs , Wolf R. Külpmann , Katsuhiko Kuwa , Lasse Larsson , Andrzej Lewenstam , Anton H.J. Maas , Gerhard Mager , Jerzy W. Naskalski , Anthony O. Okorodudu , Christoph Ritter and Andrew St John

Abstract

Analyzers with ion-selective electrodes (ISEs) for ionized magnesium (iMg) should yield comparable and unbiased results for iMg. This IFCC guideline on sampling, measuring and reporting iMg in plasma provides a prerequisite to achieve this goal [in this document, “plasma” refers to circulating plasma and the forms in which it is sampled, namely the plasma phase of anticoagulated whole blood (or “blood”), plasma separated from blood cells, or serum].

The guideline recommends measuring and reporting ionized magnesium as a substance concentration relative to the substance concentration of magnesium in primary aqueous calibrants with magnesium, sodium, and calcium chloride of physiological ionic strength. The recommended name is “the concentration of ionized magnesium in plasma”. Based on this guideline, results will be approximately 3% higher than the true substance concentration and 4% lower than the true molality in plasma.

Calcium ions interfere with all current magnesium ion-selective electrodes (Mg-ISEs), and thus it is necessary to determine both ions simultaneously in each sample and correct the result for Ca2+ interference. Binding of Mg in plasma is pH-dependent. Therefore, pH should be measured simultaneously with iMg to allow adjustment of the result to pH 7.4.

The concentration of iMg in plasma may be physiologically and clinically more relevant than the concentration of total magnesium. Furthermore, blood-gas analyzers or instruments for point-of-care testing are able to measure plasma iMg using whole blood (with intact blood cells) as the sample, minimizing turn-around time compared to serum and plasma, which require removal of blood cells.

Clin Chem Lab Med 2008;46:21–6.


Corresponding author: Dr. Niels Fogh-Andersen, Department of Clinical Biochemistry, Herlev Hospital, Herlev Ringvej 75, 2730 Herlev, Denmark Phone: +45-4-4883313, Fax: +45-4-4883311,

Received: 2006-11-9
Published Online: 2007-07-30
Published in Print: 2008-01-01