Improving the SIENA performance using BEaST brain extraction

  • Kunio Nakamura ,

    Roles Conceptualization, Formal analysis, Methodology, Software, Writing – original draft

    nakamuk@ccf.org

    Affiliations McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal, Quebec, Canada, Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio, United States of America

  • Simon F. Eskildsen,

    Roles Conceptualization, Methodology, Resources, Software, Writing – review & editing

    Affiliation Centre of Functionally Integrative Neuroscience, Aarhus University, Aarhus, Denmark

  • Sridar Narayanan,

    Roles Supervision, Writing – review & editing

    Affiliations McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal, Quebec, Canada, NeuroRx Research, Montreal, Quebec, Canada

  • Douglas L. Arnold,

    Roles Supervision, Writing – review & editing

    Affiliations McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal, Quebec, Canada, NeuroRx Research, Montreal, Quebec, Canada

  • D. Louis Collins,

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    Affiliation McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal, Quebec, Canada

  • The Alzheimer's Disease Neuroimaging Initiative

Abstract

We present an improved image analysis pipeline to detect the percent brain volume change (PBVC) using SIENA (Structural Image Evaluation, using Normalization, of Atrophy) in populations with Alzheimer’s dementia. Our proposed approach uses the improved brain extraction mask from BEaST (Brain Extraction based on nonlocal Segmentation Technique) instead of the conventional BET (Brain Extraction Tool) for SIENA. We compared four varying options of BET as well as BEaST and applied these five methods to analyze scan-rescan MRIs in ADNI from 332 subjects, longitudinal ADNI MRIs from the same 332 subjects, their repeat scans over time, and OASIS longitudinal MRIs from 123 subjects. The results showed that BEaST brain masks were consistent in scan-rescan reproducibility. The cross-sectional scan-rescan error in the absolute percent brain volume difference measured by SIENA was smallest (p≤0.0187) with the proposed BEaST-SIENA. We evaluated the statistical power in terms of effect size, and the best performance was achieved with BEaST-SIENA (1.2789 for ADNI and 1.095 for OASIS). The absolute difference in PBVC between scan-dataset (volume change from baseline to year-1) and rescan-dataset (volume change from baseline repeat scan to year-1 repeat scan) was also the smallest with BEaST-SIENA compared to the BET-based SIENA and had the highest correlation when compared to the BET-based SIENA variants. In conclusion, our study shows that BEaST was robust in terms of reproducibility and consistency and that SIENA’s reproducibility and statistical power are improved in multiple datasets when used in combination with BEaST.

Introduction

Chronic brain atrophy is one of the pathologic hallmarks in neurological diseases such as multiple sclerosis and Alzheimer’s disease (AD). Advanced image processing of magnetic resonance imaging (MRI) allows in vivo quantification of brain atrophy. Of several sophisticated quantification algorithms available, SIENA (Structural Image Evaluation, using Normalization, of Atrophy) [1] from the FSL package [2], that has seen particularly wide use in multiple sclerosis research and clinical trials. The major advantages of the SIENA method are 1) skull-constrained registration, which corrects for incorrect pixel sizes [1] and reduces distortion artifacts [3], 2) halfway-space transformation whereby the images are equally interpolated, 3) measurement of edge shift using the first derivatives of the intensity profiles at the brain edge for sub-voxel accuracy, and 4) self-calibration to determine the ratio between surface area and volume to measure the percent volume change. SIENA is robust for different MRI slice thicknesses, has low transitivity error [1], and is statistically powerful [4]. SIENA has been used in many centers for research studies and clinical trials with varying protocols. On the other hand, SIENA is not commonly used in AD studies. One possible reason may be that some patients with AD exhibit extreme atrophy, and the FSL tools may not perform optimally in such cases. In fact, our previous work has shown that the performance of brain extraction in BET/FSL [5] worsened when applied to ADNI data compared to data from young normal subjects in the ICBM dataset in terms of Dice similarity coefficient (0.944 and 0.975), false positive rate (3.81 and 1.28), and false negative rate (2.71 and 0.45) [6]. In the same study, we demonstrated that the new method called BEaST (Brain Extraction based on nonlocal Segmentation Technique) improved the accuracy of brain extraction. The use of SIENA in AD studies may bridge these neurological conditions (such as multiple sclerosis and Alzheimer’s disease) and allow meta-analysis to compare or merge such datasets in the future. In the current study, we investigated the effect on SIENA measurements when applied in conjunction with BEaST-extracted brain masks. We 1) applied SIENA using various brain extraction approaches including BET and BEaST; 2) compared the extracted brain masks by measuring the overlaps between subjects and consistency within subjects; 3) compared SIENA scan-rescan reproducibility, and finally 4) compared the statistical power in terms of effect size and required sample size to detect significant changes using SIENA.

Materials and methods

Dataset

ADNI.

We downloaded “Original” images from ADNI datasets (S1). These images are not pre-processed and meant to better represent typical images in clinical and research studies where B1- or distortion- corrections are not performed. The inclusion criteria were: Field strength = 1.5-Tesla, Slice Thickness = 0.5–1.9 mm, Weighting = T1, Image Description = MP-RAGE, Subject Group = AD or normal control (NC), Visits = complete set of baseline, repeat baseline, and month 12. Cases with severe image artifacts were removed (n = 1). The remaining 139 AD patients (mean age (SD) = 75.2 (7.5) years; male proportion = 52%) and 193 normal controls (mean age (SD) = 77.2 (5.5) years; male proportion = 52%) were analyzed for percent brain volume change between baseline and year 1 and between baseline and repeat baseline. For each baseline scan, we also downloaded the repeat scan acquired on the same day. The MRI sequence is described elsewhere [7]

OASIS.

We also used the OASIS (Open Access Series of Imaging Studies) Longitudinal MRI Data in Nondemented and Demented Older Adults (OASIS2) [8]. We selected the patients who had the follow-up MRI between 365–1095 days (1–3 years). At each time-point in the OASIS study, 4 MP-RAGE scans were acquired and averaged together. Since typical studies would only acquire one MP-RAGE per time-point, we only used the first image for the study presented here. One subject was dropped due to an image inconsistency. Thus, the sample comprised 59 demented subjects (mean age (SD) = 74.9 (6.9) years; male proportion = 54%) and 64 non-demented subjects (mean age (SD) = 75.6 (8.6) years; male proportion = 30%).

Pre-processing

All images were converted to MINC format (Medical Imaging NetCDF, Montreal Neurological Institute, McGill University, Montreal, Quebec, Canada, http://www.mcgill.ca/bic/software-atlases/minc). All the pre-processing was performed using minc-toolkit (ITK4v2 branch, https://github.com/BIC-MNI/minc-toolkit). The pre-processing was similar to previous studies [911] and consisted of the following: 1) N3 correction [12], 2) de-noising using non-local means [13], 3) linear affine standard space registration [14] to nonlinear ICBM atlas (ICBM 2009c [15]), 4) intensity normalization based on histogram-matching, 5) brain extraction using BEaST (version 1.15, http://www.bic.mni.mcgill.ca/BEaST) [6], and 6) refinement of linear affine registration, N3 correction, and intensity normalization using the BEaST brain extraction mask. Importantly, the input to BEaST was N3-corrected, de-noised, and intensity-normalized image, resampled in standard space using 12-parameter affine registration. The BEaST library consisted of 10 samples from ICBM library [16], 10 samples from NIH study of normal brain development [17], and 60 samples from ADNI as described in the original study [6]. We left-right flipped the template images along the x-axis to artificially increase the sample for a total of 160 samples in the BEaST library. Of the 160 library images, BEaST preselects the 20 best matching samples for processing each image independently. The highest processing resolution of BEaST was 2mm with median filter set as a post-processing step. The mask was then brought back from standard space to the subject’s native space and resampled to the original resolution. Of 323 ADNI subjects, 39 cases were overlapping with images included in the library. For these 39 cases, we excluded the case from the library to avoid the potential advantage for BEaST.

SIENA

We used SIENA (FSL version 5.0, https://fsl.fmrib.ox.ac.uk/fsl/) on CentOS 6.3 on the supercomputer Guillimin from McGill University (http://www.hpc.mcgill.ca). We measured the percent brain volume change (PBVC) in three ways: (1) between the baseline scan and follow-up (year-1) (original PBVC), (2) between baseline scan and baseline repeat scan for cross-sectional scan-rescan comparison, and (3) between repeat-scan at baseline and repeat-scan at year 1 (repeat PBVC) for longitudinal scan-rescan comparison. We compared five variations of SIENA as described next and summarized in Table 1 and Fig 1. We ran the analysis completely automatically.

thumbnail
Fig 1. Flowchart diagram of the processing pipelines: The BET brain masks and masked brain images are discarded for the BEaST pipeline and replaced by BEaST brain masks and masked images.

https://doi.org/10.1371/journal.pone.0196945.g001

1. Vanilla-SIENA.

As a comparator, we applied the SIENA with “-m” option, which performs standard-space masking as well as BET, as a previous study has shown the benefit of standard space masking [18]. While typical research centers do not perform Vanilla-SIENA, this version serves as a traditional reference, and less advanced centers still use this option.

2. Popescu-SIENA.

To compare an improved brain extraction method within the FSL framework, we applied the options suggested by Popescu et al. [19]. In their study, different BET options were compared with respect to the manual segmentation using multi-center MRIs from multiple sclerosis patients. As described, we performed the neck-cropping using the field-of-view region-of-interest (ROI) created in MNI space. We also used the BET [5] options “-B” and “-f 0.1,” where “-B” corrects the bias field and cleans up the neck and “-f” is a fractional intensity threshold that would lead to a larger brain outline at the bottom with smaller f-value.

3. Cover-SIENA.

We also applied the options suggested by Cover et al. [20]. In that study, “-B -f 0.2” appears to be the optimal parameter for analysis of 1.5-Tesla MRIs in ADNI. We used the same neck-cropping as in Popescu-SIENA and changed the BET option. The only difference between Popescu-SIENA was the f-option.

4. CBET-SIENA.

We used an option “-c” to define the center of gravity, which was calculated from a predefined coordinate in the standard ICBM space (0,-12,4) between left and right thalami and the affine standard-space registration matrix used to map the MRI into standard space. There was no neck-cropping. There were no additional options for BET. This version serves as a reference to BEaST-SIENA.

5. BEaST- SIENA.

In the proposed pipeline, we replaced the BET brain mask with the BEaST brain mask. BEaST was previously validated in terms of accuracy to be superior to BET and the VBM8 toolbox of SPM8 [6]. In this pipeline, we ran BET only to produce the skull mask since SIENA requires the skull mask for registration. We again used the “-c” option to set the center of gravity, using the same coordinate as in CBET-SIENA. There was no neck-cropping.

Assessment of methods

Mask comparison.

To detect gross brain masking errors, we generated average masks in standard space using the following procedure: we 1) nonlinearly registered all baseline MR images to the ICBM 2009c template using ANTS (http://stnava.github.io/ANTs) [21], 2) transformed and resampled the brain masks from the native space to standard space using the transformation from ANTS, 3) computed the voxel-by-voxel mean and standard deviation of the masks from each subject for each time-point, and 4) repeated the process for each of the five SIENA variations (Vanilla, Popescu, Cover, CBET, and BEaST). ANTS was applied to the full image without brain extraction. We did not use the brain mask to avoid introducing the mask’s influence on the performance of nonlinear registration. Note that BEaST was applied independently for baseline and follow-up image. In other words, there was no affine registration between baseline and follow-up scans in generating BEaST masks so that the comparison of masks is fair across the pipelines. We visually inspected the general accuracy of nonlinear registration. This test was only intended to show gross errors in brain extraction since there is no gold standard for each subject, and since the definition of true brain extraction may be different for BET and BEaST. We also counted the number of failures, defined by “incomplete analysis” whereby SIENA did not produce a valid value, which happened due to poor brain extraction and/or poor registration.

Cross-sectional scan-rescan error.

First, the reproducibility of brain extracted masks was compared among BEaST and the four BET variations using the cross-sectional ADNI scan-rescan data acquired on the same day. We calculated absolute volume difference, sensitivity, specificity, Jaccard similarity index, and mean surface distance between the scan and rescan brain extraction masks. All measurements were made in native space with linear rigid-body registration without the use of brain extraction masks. Thus, we used the same linear registration matrix for all 5 versions. We also created absolute difference maps between the scan and rescan brain masks for each subject and averaged them in the nonlinearly resampled standard space for each SIENA method.

To evaluate the impact on reproducibility of SIENA, we measured the percent brain volume difference and its absolute value for each of the 5 variations of SIENA.

Effect size and sample size.

To evaluate the quantitative improvement on statistical power, we calculated the effect size to detect a 25% change in PBVC over a 12-month period for each method, for each dataset. For ADNI, PBVC was measured between baseline scan and year-1 scan (original PBVC). The OASIS PBVC values were annualized since the MRI intervals were not consistent across subjects. The effect size was calculated using the standard equation as follows: Eq 1 where the denominator is a pooled standard deviation, σ is the group PBVC standard deviation, and μ is the group PBVC mean obtained from longitudinal data. This equation includes the effect of normal aging thus the hypothesized treatment effect acts on the difference between AD and healthy controls; that is, a 100% treatment effect leads to patients having the same atrophy rate as healthy controls. We then calculated the required sample size per arm from the effect size according to the Eq 2: Eq 2 where SDp is the pooled standard deviation, u = 0.842 for 80% power, v = 1.96 to test at 0.05 significance level, Ceil is the round-up operator, and TxE is the predefined treatment effect (25%), which is assumed to take immediate and constant effect. All available PBVC values were included in the analysis.

Longitudinal scan-rescan error.

Using the ADNI rescan data acquired over time, we measured PBVC between the baseline repeat and year-1 repeat (repeat PBVC). We compared the repeat PBVC with the original PBVC between baseline scan and year-1 scan and calculated the mean difference and mean absolute difference in PBVC as well as their correlation and regression (slope and intercept).

Results

Fig 2 shows the qualitative evaluations of methods. BEaST appears to have the smallest variability in average mask obtained from 910 masks in 332 ADNI and 123 OASIS cases (top). BEaST also had consistent segmentation within subjects from the scan-rescan pairs from 332 ADNI cases acquired on the same day (bottom).

thumbnail
Fig 2.

Top four rows: twelve transverse slices through the voxel-by-voxel average of 910 masks generated for each method (from left to right: Vanilla-SIENA, Popescu-SIENA, Cover-SIENA, CBET-SIENA, and BEaST-SIENA), on spectral colormap. The white and black are the areas with complete agreement while colored voxels show variable degree of disagreement shown on the colorbar. BEaST masks overall show the smallest variability compared to other methods. Bottom four rows: The average difference map from scan-rescan dataset for each method. The difference map was created in the native space using SIENA’s flirt registration matrix. Then the difference maps were nonlinearly transformed into the common ICBM space for averaging. The range for all difference maps was set to a constant; note the darker intensity for BEaST, indicating smaller error.

https://doi.org/10.1371/journal.pone.0196945.g002

Table 2(A) shows the comparison of scan-rescan brain extraction masks. BEaST performed better in all the comparisons. Note for all measures (sensitivity, specificity, Jaccard index, and mean surface distance), the worst quartile from BEaST outperformed the best quartile of other methods. Note also that the volume difference in Fig 3A and the absolute volume difference in Fig 3B contain no outliers for BEaST, which is not the case for the other methods. There are few very low specificity cases in Vanilla-SIENA (n = 3) and CBET-SIENA (n = 1), and thus the means are similar to the 25-percentile value. Table 2B, 2C and 2D shows scan-rescan percent brain volume difference and absolute values from SIENA. BEaST-SIENA had the smallest standard deviation in scan-rescan percent brain volume difference among the 5 methods. The smallest mean and median absolute values were obtained by BEaST-SIENA, followed by CBET-SIENA, and Cover-SIENA. The absolute scan-rescan percent difference for BEaST-SIENA was significantly different from all other versions of SIENA (p < 0.0187, paired tests).

thumbnail
Fig 3.

Scan-rescan comparison of brain extraction masks (Vanilla-BET, Popescu-BET, Cover-BET, CBET, and BEaST) in terms of (a) percent volume difference, (b) absolute percent volume difference, (c) Jaccard similarity index, and (d) mean surface distance. The colored box plot shows extreme values and interquartile range; the red crosses show the outliers with 2.5 SD limit; and the black lines indicate mean and its 95% confidence interval.

https://doi.org/10.1371/journal.pone.0196945.g003

thumbnail
Table 2. Scan-rescan percent brain volume difference, number of failures, and scan-rescan percent brain volume difference.

https://doi.org/10.1371/journal.pone.0196945.t002

Table 3 shows the effect size for all methods estimated from the ADNI and OASIS datasets. The actual mean (SD) interval between the baseline and follow-up of 1.08 (0.07) for ADNI and 1.82 (0.40) for OASIS. BEaST-SIENA had the largest effect size: 41% and 27% larger, respectively for ADNI and OASIS, compared to Vanilla-SIENA.

thumbnail
Table 3. Mean (SD) annualized percent brain volume change (PBVC), effect size, number of failures, and required sample size per arm for each method and dataset.

https://doi.org/10.1371/journal.pone.0196945.t003

Longitudinal scan-rescan experiment showed that BEaST-SIENA had the smallest mean absolute difference (0.11%) compared to other methods, best correlated with original PBVC values with the regression line closest to slope of 1 and intercept of 0 (Table 4). The effect size determined from the repeat dataset was overall lower than those calculated from the original PBVC but the trend that BEaST-SIENA performed optimally with CBET-SIENA closely next was similar to the scan-dataset.

thumbnail
Table 4. Repeat PBVC results and its comparison to original PBVC.

https://doi.org/10.1371/journal.pone.0196945.t004

Discussion and conclusion

Our study not only compared our results with those obtained using the default options of SIENA, but also with previously published “optimized” parameters [19, 20]. We have shown the improved reproducibility of BEaST brain extraction mask over the BET-derived brain masks in the scan-rescan dataset (Table 2, Fig 2 bottom and Fig 3), and better longitudinal consistency in the ADNI and OASIS datasets (Fig 2 top). The study also demonstrates the robust impact of BEaST brain masks on the measurement of whole brain atrophy using SIENA in ADNI and OASIS datasets (Tables 3 and 4).

The cross-sectional scan-rescan brain extraction comparisons in Fig 3 and Table 2 show that BEaST’s reproducibility is higher than the variants of BET studied here. Both BEaST and Popescu-BET masks were the least biased with a percent volume difference of <0.1%, However at 0.3%, the standard deviation of BEaST percent volume difference was almost 5 times smaller than Popescu-BET (1.4%), indicating more reproducible and robust brain masks for BEaST. Cover-BET performed worse, with a bias of -3.1% and a standard deviation of 5.8% for percent volume difference in the scan-rescan tests. For sensitivity and Jaccard index, the lowest 25 percentile of BEaST values was greater than the each of the BET variant 75 percentile values.

Better brain masks are likely to translate to improved scan-rescan SIENA measurements. Table 2C shows that all methods yield similar numbers with Popescu-SIENA giving the smallest bias for scan-rescan SIENA measurements. The absolute PBVC is also very similar between techniques, with CBET-SIENA and BEaST-SIENA yielding the smallest standard deviations.

A larger effect size means that smaller sample sizes are required to show a statistically significant difference between groups. The methods that consistently increased the effect size compared to Vanilla-SIENA were Cover-SIENA, CBET-SIENA and BEaST-SIENA, with the largest effect sizes achieved by BEaST-SIENA in both ADNI and OASIS datasets. BEaST-SIENA had 41% larger effect size than Vanilla-SIENA in ADNI data. BEaST-SIENA and Cover-Siena were robust without any processing failures, while the three other BET-derived SIENA pipelines failed to compute PBVC for some datasets in the ADNI cohort. There were no failures in the OASIS cohort. Cover-SIENA, CBET-SIENA and BEaST-SIENA result in the smallest required sample sizes with CBET-SIENA and BEaST-SIENA yielding almost the same sample size in the ADNI cohort, and Cover-SIENA and BEaST-SIENA yielding almost the same sample size in the OASIS cohort.

Our study did not evaluate the accuracy of PBVC measurements. The accuracy of BEaST has been studied in the original development and validation manuscript [6]. Creation of a gold standard for PBVC measurements is extremely difficult [22] as it requires manual segmentation of the brain parenchyma by multiple experts in high-resolution baseline and follow-up MRIs. Given the potential inter- and intra-rater variability in manual segmentation, it is not clear that such a gold standard would be good enough to quantify the differences between PBVC measures since the manual variability could be of the same magnitude as the errors and inaccuracies of the SIENA procedures using either BEaST- or BET-derived masks.

We did not investigate the effect of parameters directly used in SIENA. Since brain extraction is one the first processing steps in the SIENA pipeline, we expected the choice of brain mask would have the largest influence on the outcome of PBVC. We also did not investigate the sensitivity of templates in the BEaST library, as this was previously studied, and the use of n = 20 preselected templates appear to be sufficiently accurate [6]. We also did not assess the brain masks individually, reject results, or re-analyze data, which may take place in real analyses and in clinical trials. However, the robustness demonstrated by the scan-rescan measures (sensitivity, specificity, Jaccard, and means surface distance), average mask, the average difference mask, and greater statistical power suggests that such individual assessment may not be necessary with the BEaST-SIENA method.

Our results also highlight the variable performance of SIENA when BET is used for brain extraction. The published optimized method (Cover-SIENA) performed well and had similar effect sizes compared to the BEaST-SIENA pipeline in both ADNI and OASIS datasets. This high quality performance may be attributed to the fact that SIENA combines two masks (with an OR-operation) before brain tissue segmentation, thus reducing the risk of failure in case one brain mask is partially missing brain tissue. We have found BEaST to be very robust on over 3,000 datasets in other combined studies [9, 2326].

The improved sample size estimation using BEaST-SIENA compared to the second best method (Cover-SIENA) was not large for ADNI dataset (7%, Table 3). However, our results show that SIENA results are very sensitive to BET options as changing the option “-f” from 0.1 to 0.2 changed the mean PBVC in the OASIS dataset by 25%, i.e., from -0.330 to -0.413 in non-demented group. Previous studies have attempted to optimize the parameters for BET in multiple sclerosis and AD studies [1820] and resulted in varying parameters. Therefore, it may be difficult in prospective multi-center studies to predict optimal parameters for BET for both accuracy and effect size. The ranking of algorithms differed in BET-based SIENA pipelines between the ADNI and OASIS datasets, which may reflect the sensitivity of SIENA to brain extraction and emphasize its importance.

The ADNI standard analysis list contains 195 normal subjects and 133 AD patients for Complete 1 Year Visits [27]. Our cohort does not exactly match this list; one normal subject was removed due to poor quality, one normal subject is no longer available as of March 16, 2015, and 6 AD patients were added because Month-6 scans were missing while Screening and Year-1 were available.

In conclusion, BEaST was robust in two large studies (ADNI and OASIS), consistent across and within subjects, and led to more reproducible and sensitive brain volume change measurements than conventional BET when applied in SIENA.

Supporting information

S2 File. SIENA results for ADNI and OASIS.

https://doi.org/10.1371/journal.pone.0196945.s002

(XLSX)

Acknowledgments

The authors thank Calcul Québec. The computations were made on the supercomputer Guillimin from McGill University, managed by Calcul Québec and Compute Canada. The operation of this supercomputer is funded by the Canada Foundation for Innovation (CFI), NanoQuébec, RMGA, the Fonds de recherche du Québec—Nature et technologies (FRQ-NT) and a FRSQ-Pfizer Innovation grant.

References

  1. 1. Smith SM, Zhang Y, Jenkinson M, Chen J, Matthews PM, Federico A, et al. Accurate, robust, and automated longitudinal and cross-sectional brain change analysis. Neuroimage. 2002;17(1):479–89. Epub 2002/12/17. pmid:12482100.
  2. 2. Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TE, Johansen-Berg H, et al. Advances in functional and structural MR image analysis and implementation as FSL. Neuroimage. 2004;23:S208–S19.
  3. 3. Caramanos Z, Fonov VS, Francis SJ, Narayanan S, Pike GB, Collins DL, et al. Gradient distortions in MRI: characterizing and correcting for their effects on SIENA-generated measures of brain volume change. Neuroimage. 2010;49(2):1601–11. Epub 2009/08/18. pmid:19682586.
  4. 4. Altmann DR, Jasperse B, Barkhof F, Beckmann K, Filippi M, Kappos LD, et al. Sample sizes for brain atrophy outcomes in trials for secondary progressive multiple sclerosis. Neurology. 2009;72(7):595–601. pmid:19005170
  5. 5. Smith SM. Fast robust automated brain extraction. Human brain mapping. 2002;17(3):143–55. Epub 2002/10/23. pmid:12391568.
  6. 6. Eskildsen SF, Coupe P, Fonov V, Manjon JV, Leung KK, Guizard N, et al. BEaST: brain extraction based on nonlocal segmentation technique. Neuroimage. 2012;59(3):2362–73. Epub 2011/09/29. pmid:21945694.
  7. 7. Jack CR Jr., Bernstein MA, Fox NC, Thompson P, Alexander G, Harvey D, et al. The Alzheimer's Disease Neuroimaging Initiative (ADNI): MRI methods. Journal of magnetic resonance imaging: JMRI. 2008;27(4):685–91. pmid:18302232; PubMed Central PMCID: PMC2544629.
  8. 8. Marcus DS, Fotenos AF, Csernansky JG, Morris JC, Buckner RL. Open Access Series of Imaging Studies: Longitudinal MRI Data in Nondemented and Demented Older Adults. Journal of cognitive neuroscience. 2009;22(12):2677–84. pmid:19929323
  9. 9. Nakamura K, Guizard N, Fonov VS, Narayanan S, Collins DL, Arnold DL. Jacobian integration method increases the statistical power to measure gray matter atrophy in multiple sclerosis. NeuroImage: Clinical. 2014;4:10–7. http://dx.doi.org/10.1016/j.nicl.2013.10.015.
  10. 10. Aubert-Broche B, Fonov VS, Garcia-Lorenzo D, Mouiha A, Guizard N, Coupe P, et al. A new method for structural volume analysis of longitudinal brain MRI data and its application in studying the growth trajectories of anatomical brain structures in childhood. Neuroimage. 2013;82:393–402. pmid:23719155.
  11. 11. Eskildsen SF, Coupé P, Fonov VS, Pruessner JC, Collins DL. Structural imaging biomarkers of Alzheimer's disease: predicting disease progression. Neurobiology of aging. 2015;36, Supplement 1:S23–S31. http://dx.doi.org/10.1016/j.neurobiolaging.2014.04.034.
  12. 12. Sled JG, Zijdenbos AP, Evans AC. A nonparametric method for automatic correction of intensity nonuniformity in MRI data. Medical Imaging, IEEE Transactions on. 1998;17(1):87–97.
  13. 13. Coupe P, Yger P, Prima S, Hellier P, Kervrann C, Barillot C. An Optimized Blockwise Nonlocal Means Denoising Filter for 3-D Magnetic Resonance Images. IEEE transactions on medical imaging. 2008;27(4):425–41. pmid:18390341
  14. 14. Collins DL, Neelin P, Peters TM, Evans AC. Automatic 3D intersubject registration of MR volumetric data in standardized Talairach space. Journal of computer assisted tomography. 1994;18(2):192–205. Epub 1994/03/01. pmid:8126267.
  15. 15. Fonov V, Evans A, McKinstry R, Almli C, Collins D. Unbiased nonlinear average age-appropriate brain templates from birth to adulthood. NeuroImage. 2009;47:S102.
  16. 16. Mazziotta JC, Toga AW, Evans A, Fox P, Lancaster J. A Probabilistic Atlas of the Human Brain: Theory and Rationale for Its Development: The International Consortium for Brain Mapping (ICBM). NeuroImage. 1995;2(2, Part A):89–101. http://dx.doi.org/10.1006/nimg.1995.1012. pmid:9343592
  17. 17. Evans AC. The NIH MRI study of normal brain development. Neuroimage. 2006;30(1):184–202. pmid:16376577
  18. 18. Battaglini M, Smith SM, Brogi S, De Stefano N. Enhanced brain extraction improves the accuracy of brain atrophy estimation. NeuroImage. 2008;40(2):583–9. pmid:18255315
  19. 19. Popescu V, Battaglini M, Hoogstrate WS, Verfaillie SCJ, Sluimer IC, van Schijndel RA, et al. Optimizing parameter choice for FSL-Brain Extraction Tool (BET) on 3D T1 images in multiple sclerosis. NeuroImage. 2012;61(4):1484–94. pmid:22484407
  20. 20. Cover KS, van Schijndel RA, Popescu V, van Dijk BW, Redolfi A, Knol DL, et al. The SIENA/FSL whole brain atrophy algorithm is no more reproducible at 3T than 1.5T for Alzheimer׳s disease. Psychiatry Research: Neuroimaging. 2014;224(1):14–21. pmid:25089020
  21. 21. Avants BB, Epstein CL, Grossman M, Gee JC. Symmetric diffeomorphic image registration with cross-correlation: evaluating automated labeling of elderly and neurodegenerative brain. Medical image analysis. 2008;12(1):26–41. Epub 2007/07/31. pmid:17659998; PubMed Central PMCID: PMCPMC2276735.
  22. 22. de Bresser J, Portegies MP, Leemans A, Biessels GJ, Kappelle LJ, Viergever MA. A comparison of MR based segmentation methods for measuring brain atrophy progression. NeuroImage. 2011;54(2):760–8. pmid:20888923
  23. 23. Nakamura K, Brown RA, Narayanan S, Collins DL, Arnold DL, Alzheimer's Disease Neuroimaging I. Diurnal fluctuations in brain volume: Statistical analyses of MRI from large populations. Neuroimage. 2015;118:126–32. pmid:26049148.
  24. 24. Voskuhl RR, Wang H, Wu TC, Sicotte NL, Nakamura K, Kurth F, et al. Estriol combined with glatiramer acetate for women with relapsing-remitting multiple sclerosis: a randomised, placebo-controlled, phase 2 trial. Lancet neurology. 2016;15(1):35–46. pmid:26621682.
  25. 25. Nakamura K, Vollmer TL, Gorfine T, Knappertz V, Arnold DL, editors. Effect of laquinimod on gray matter and white matter atrophy in relapsing-remitting multiple sclerosis: analysis of the BRAVO phase III trial. 2014 Joint ACTRIMS-ECTRIMS Meeting (MSBoston 2014); 2014.
  26. 26. Nakamura K, Brown RA, Araujo D, Narayanan S, Arnold DL. Correlation between brain volume change and T2 relaxation time induced by dehydration and rehydration: Implications for monitoring atrophy in clinical studies. NeuroImage: Clinical. 2014;6(0):166–70. pmid:25379428; PubMed Central PMCID: PMC4215533.
  27. 27. Wyman BT, Harvey DJ, Crawford K, Bernstein MA, Carmichael O, Cole PE, et al. Standardization of analysis sets for reporting results from ADNI MRI data. Alzheimer's & Dementia. 2013;9(3):332–7.