Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Reply to comments on “Absolute linearity measurements on a PbS detector in the infrared”

Not Accessible

Your library or personal account may give you access

Abstract

Some issues on the method used by Zhang et al. [Appl. Spectrosc. 51, 576 (1997)] to correct for the nonlinearity of a HgCdTe detector in Fourier transform spectrometry measurements are discussed in response to comments by Zhang. Other problems that must be taken into consideration before the effects of detector nonlinearity of HgCdTe detectors are fully accounted for in FT spectrometry are highlighted.

© 2007 Optical Society of America

Full Article  |  PDF Article
More Like This
Comments on "Absolute linearity measurements on a PbS detector in the infrared"

Zhuomin M. Zhang
Appl. Opt. 46(25) 6483-6484 (2007)

Absolute linearity measurements on a PbS detector in the infrared

Evangelos Theocharous
Appl. Opt. 45(11) 2381-2386 (2006)

Absolute linearity measurements on HgCdTe detectors in the infrared region

Evangelos Theocharous, Juntaro Ishii, and Nigel P. Fox
Appl. Opt. 43(21) 4182-4188 (2004)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All Rights Reserved