Occurrence of pendelluft under pressure support ventilation in patients who failed a spontaneous breathing trial: an observational study

Abstract

Background

Pendelluft, the movement of gas within different lung regions, is present in animal models of assisted mechanical ventilation and associated with lung overstretching. Due to rebreathing of CO2 as compared to fresh gas, pendelluft might reduce ventilatory efficiency possibly exacerbating patient’s respiratory workload during weaning. Our aim was to measure pendelluft by electrical impedance tomography (EIT) in patients who failed a spontaneous breathing trial (SBT).

Methods

This is an observational study conducted in a general intensive care unit of a tertiary-level teaching hospital. EIT signal was recorded in 20 patients while pressure support (PS) ventilation was progressively reduced from clinical level (baseline) to 2 cmH2O, as in an SBT; four ventral-to-dorsal lung regions of interest were identified for pendelluft measurement. A regional gas movement (> 6 mL) occurring in a direction opposite to the global EIT signal was considered diagnostic for high pendelluft.

Results

Eight patients out of 20 (40%) were classified as high-pendelluft; baseline clinical characteristics did not differ between high- and low-pendelluft patients. At PS reduction, pendelluft and EtCO2 increased more in the high-pendelluft group (p < .001 and .011, respectively). The volume of gas subject to pendelluft moved almost completely from the ventral towards the dorsal lung regions, while the opposite movement was minimal (16.3 [10:32.8] vs. 0 [0:1.8] mL, p = .001). In a subgroup of patients, increased pendelluft volumes positively correlated with markers of respiratory distress such as increased respiratory rate, p0.1, and EtCO2.

Conclusions

Occult pendelluft can be measured by EIT, and is frequently present in patients failing an SBT. When present, pendelluft increases with the reduction of ventilator support and is associated with increased EtCO2, suggesting a reduction of the ability to eliminate CO2.

Background

Growing evidence suggests that the presence of spontaneous breathing during mechanical ventilation is associated with several advantages, but may be harmful at the same time, particularly during the acute phase [1,2,3]. One of the possible mechanisms of injury associated with spontaneous breathing and assisted ventilation is the development of pendelluft [1]. Pendelluft is a phenomenon known as the pendular movement of gas between different lung regions; classically it is described during controlled mechanical ventilation when regional heterogeneity in time constants (compliance * resistance) is present: after the tidal volume has been delivered, the gas moves from “faster” lung regions (shorter time constant) towards “slower” ones (longer time constant) [4, 5].

Recently, a different mechanism of pendelluft has been described in animal models and in a small patient series when switching from fully controlled to assisted mechanical ventilation with spontaneous breathing [6]. At the very beginning of inspiration, inflation of the dorsal regions due to uneven distribution of the negative pressure generated by the diaphragmatic pump resulted in a concomitant deflation of the ventral ones, consistent with a movement of gas within the lung (pendelluft). Consequently, this phenomenon may induce overdistension of dorsal regions even in the presence of fully protective “global” ventilator settings. Moreover, the gas subject to pendelluft moving within the lung will not contribute to gas exchange, possibly causing CO2 retention, resulting in wasted work of breathing and ventilatory inefficiency.

Switching from controlled mechanical ventilation to an assisted form of ventilation (e.g., pressure support, PS) is a common strategy to progressively increase patient’s work of breathing, for example during weaning [7]. During the weaning phase, patient effort progressively increases until all the inspiratory work is generated by the patient and none by the ventilator. Most patients do not present weaning problems, while others result difficult to wean; prolonged weaning is associated with worst clinical outcomes and increased costs [8,9,10,11]. Weaning failure is attributable to different causes, which include respiratory pump insufficiency, cardiovascular dysfunction or underlying infection [12].

We speculated that during the decrease of PS, a “vicious circle” might exist in which the increased effort causes pendelluft, thereby decreasing CO2 removal and further increasing breathing effort (Fig. 1). The purpose of this work is to assess the presence of pendelluft during the decrease of PS, measuring pendelluft volume and its relationship with clinically relevant variables by electrical impedance tomography (EIT), a non-invasive bedside monitoring technique allowing visualization of regional ventilation which gained increased interest in the recent years [13]. We focused on patients who failed at least one spontaneous breathing trial (SBT) since this population could be more at risk of weaning failure.

Fig. 1
figure 1

Vicious cycle due to pendelluft during the weaning phase. Illustration of the vicious cycle related with reduction of ventilator support in patients affected by the presence of pendelluft

Methods

This is a prospective observational study aimed at assessing the presence of a particular phenomenon such as pendelluft and generating new pathophysiological hypotheses. The local Institutional Review Board (Comitato Etico ASST Monza) approved the study protocol, and written informed consent was obtained as per local regulations.

Subjects

Patients admitted to the general intensive care unit (ICU) of the University hospital San Gerardo (Monza, Italy) were screened for inclusion criteria: age > 18 years, assisted mode of ventilation (PS is the first choice in our unit), “ready to wean” evaluation by the attending physicians (resolution of acute phase, PEEP ≤ 8 cmH2O, FiO2 ≤ 40%, light or no sedation, no need for vasopressors), failure of the standard spontaneous breathing trial performed in our ICU (defined as not tolerating PS = 2 cmH2O for at least 2 h without signs of respiratory distress). Exclusion criteria were: known severe chronic obstructive pulmonary disease (GOLD 4) [14], pneumothorax, impaired respiratory drive due to cerebral lesions, pneumonectomy, bronchial bleeding, presence of pleural catheters, rib fractures, thoracic wounds, limitations in patient’s mobilization (e.g., spinal fractures); enrollment in other competing study.

Study protocol

After enrollment, patients were monitored with an EIT system (PulmoVista 500, Dräger Medical GmbH, Lübeck, Germany) and recordings were started at clinically set PS level (baseline); demographic data were collected. PS level was then progressively reduced by 2-cmH2O steps, to reach PS = 2 cmH2O. All the other ventilator settings were left unchanged during the study as well as ventilator flow trigger. If the patients showed signs of marked respiratory distress, the PS level was increased back to baseline. We collected traces of the EIT signal, and recorded respiratory rate, tidal volume (Tv), Tv distribution by EIT [15], minute ventilation (Mv), mouth occlusion pressure at 100 ms (p0.1) [16], compliance of the respiratory system (Crs, measured by inspiratory hold) [17,18,19], end-tidal carbon dioxide (EtCO2), and long-term outcomes such as ICU survival, ICU stay, 28-day ventilator-free days. Rapid shallow breathing index (RSBi) was defined as respiratory rate/Tv [20].

Pendelluft volume measurement

EIT and ventilator waveforms were analyzed by Labchart, importing data at a 20-Hz rate (ADInstruments, Colorado Springs, CO). Data from a set of 20–30 representative breaths for each PS step were obtained by event-triggered averaging using the first point of positive ventilator airflow as triggering event [21]. Global EIT trace and four regions of interest (ventral-to-dorsal layers of the same thickness, ROI 1–4) were analyzed. The nadir of the global EIT trace was considered as the transition point from expiration to inspiration (T0); then, the nadirs of the four ROIs were identified together with the values in mL of each ROI at T0, using the ΔZ of the average tidal volume as a ΔZ/mL converting factor, as previously described [21].

We reasoned that pendelluft can be disclosed as a phase-shift of the regional EIT signal as compared to the global EIT trace in two distinct time-periods: before T0 and after T0. Before T0, the lung is still expiring and tracheal airflow is directed outward; ROIs inflating during expiration must gain gas from other ROIs that are deflating, indicating the pendelluft phenomenon (Fig. 2, panel a). Conversely, after T0 tracheal airflow is directed inward, and gas lost by late-deflating ROIs must be gained from the other ROIs that are inflating, indicating the pendelluft phenomenon as well (Fig. 2, panel b). We defined pendelluft volume as the sum of the volumes of gas inspired from the early-inflating ROIs during expiration (before T0) and the volume of gas expired from the late-deflating ROIs during inspiration (after T0). Pendelluft volume directed to the ventral lung regions was defined as the sum of gas volume gained by the ventral ROIs before T0 and the gas volume still to be deflated by the dorsal ROIs after T0; conversely, pendelluft volume directed to the dorsal lung regions was defined as the sum of gas volume gained by the dorsal ROIs before T0 and the gas volume still to be deflated by the ventral ROIs after T0.

Fig. 2
figure 2

Schematic representation of the pendelluft phenomenon occurring during assisted mechanical ventilation. Pendelluft (black arrows) was defined as the sum of the gas moving into early-inflating regions of interest during expiration (before the global electrical impedance tomography value reached its minimum value, T0 in the text, panel a) and the gas lost by late-deflating regions of interest during inspiration (after T0, panel b)

Pendelluft volume was compared with regional ventilation delay (RVD), as measured by the software included with the EIT monitor. RVD represents the percentage of lung zones which exhibit a delayed inflation as compared to the global EIT signal (inflation delay > 6% of the global inflating time, see Supplements for further RVD measurement details) [22].

Statistical analysis

Given the preliminary nature of the study and the lack of data available in the literature regarding incidence of pendelluft in this population, a formal sample size calculation was not performed and we chose a convenience sample size of 20 subjects. Patients were classified as high-pendelluft based on the presence of relevant pendelluft volumes at the lowest PS level; since no defined cut-off exists, we used the mean of the pendelluft volumes measured at baseline (6.3 mL).

Continuous variables were tested for normality by Shapiro–Wilk test. Comparisons between patients were performed by Mann–Whitney or independent sample t test; comparisons within the same patient were performed by Wilcoxon or paired-sample t test, as appropriate. Since some patients could not reach the 2-cmH2O PS level, changes in respiratory variables were considered at the minimum PS level. Correlations coefficients were expressed as Pearson’s r. Comparisons between two categorical variables were performed by Fisher’s exact test. A cut-off of p < 0.05 was considered for statistical significance. SPSS was used to perform statistical analyses (SPSS Inc. Chicago, IL).

Results

The median PS level of the 20 enrolled patients at baseline was 10 cmH2O (range 8–12 cmH2O); patients’ baseline conditions are reported in Table 1. Patients in the high-pendelluft group were eight, while the remaining 12 were classified as low-pendelluft. At baseline, patients in the high-pendelluft group already showed a pendelluft volume higher than the low-pendelluft group (p = 0.003), while the main respiratory variables did not differ. Seventeen patients tolerated the 2-cmH2O step for few minutes and recordings were present for all the study steps; one patient in the high-pendelluft and two in the low-pendelluft group developed respiratory distress before the 2-cmH2O step, and PS decrement was stopped at 4 cmH2O. Comparison between respiratory variables at baseline and at the lowest PS level is reported in Table 2.

Table 1 Baseline characteristics of study patients
Table 2 Changes of respiratory variables at the lower pressure support level

Average pendelluft volumes at the reduction of PS are depicted in Fig. 3. At the lowest PS level, pendelluft volume increased more in the high-pendelluft group, while remained almost stable in the low-pendelluft group (Table 2, Additional file 1: Figure S1 panel a). Similarly, the EtCO2 increase was more pronounced in the high- than in the low-pendelluft group at the reduction of ventilator support, and Tv reduction tended to be more marked in the high-pendelluft group (Table 2, Additional file 1: Figure S1 panel b and c). The increase in respiratory rate appeared more pronounced in the high-pendelluft group, but without any statistical significance (Additional file 1: Figure S1 panel d). Pendelluft volumes at the lowest PS level positively correlated with compliance of the respiratory system (p = .040 r = .463), but not with age, body mass index, SAPS II on admission, days of controlled mechanical ventilation before PS, PaO2/FiO2 ratio, negative inspiratory force or Tv distribution assessed by EIT; the correlation with compliance resulted significant in the high-pendelluft group (p = .014, r = .816), but not in the low-pendelluft (p = .207, Additional file 1: Figure S2). The average increase of pendelluft volume during reduction of PS levels correlated with average EtCO2 increase (p = .044 r = .495) and tended to correlate with average respiratory rate increase (p = .059 r = .429). In the high-pendelluft group, the average increase of pendelluft volume correlated with average increase of heart rate (p = .040, r = .730).

Fig. 3
figure 3

Pendelluft volumes at the reduction of ventilator support. Mean pendelluft volume remained low at the reduction of pressure support ventilation (PSV) level in the low-pendelluft group (filled circles), while increased in the high-pendelluft group (empty circles)

Considering the high-pendelluft group, the volume of gas subject to pendelluft moved almost completely from the ventral towards the dorsal lung regions (Figs. 4 and 5, Table 2). Pendelluft volume was unevenly distributed in the four ventral-to-dorsal ROIs, and the pendelluft phenomenon was observed mainly during inspiration rather than at end-expiration (Fig. 2b and Additional file 1: S3). Analyzing within-patient values, in some high-pendelluft patients pendelluft volumes positively correlated with markers of respiratory distress such as the increase of respiratory rate, the increase of p0.1, the increase in EtCO2 (Fig. 6).

Fig. 4
figure 4

Pendelluft gas movement direction in the high-pendelluft group. In the high-pendelluft group, the movement of gas subject to pendelluft was almost entirely directed towards the dorsal regions of interest, while the amount directed towards the ventral ones was minimal (PSV, pressure support ventilation)

Fig. 5
figure 5

Pendelluft measurement and distribution in patient of the high-pendelluft group (single breath, unfiltered signal) due to early inflation of the dorsal regions and late deflation of the ventral ones. Minimum impedance values of the dorsal regions of interest (ROI 3 and 4) occurred during expiration (EXP), while in the ventral ones (ROI 1 and 2) occurred during inspiration (INSP). Total pendelluft volume was calculated as the sum of the volumes (red bars) inflated during expiration (ROI 3 and 4) and deflated during inspiration (ROI 1 and 2). Processed EIT imaging (relative to minimum) indicated that inspiration in the dorsal ROIs started when ventral ones were still expiring (expiration in purple color); the phenomenon was not immediately evident with standard EIT imaging. Pendelluft gas moved from the ventral towards the dorsal lung regions (boxed image)

Fig. 6
figure 6

Association between pendelluft volume change during the study and markers of increased inspiratory effort in exemplary patients of the high-pendelluft group. In some patients of the high-pendelluft group, pendelluft volume increase from baseline was associated with markers of increased inspiratory effort such as increased respiratory rate (panel a) or increased mouth occlusion pressure at 100 ms (p0.1, panel b), or ventilatory inefficiency such as increased end-tidal carbon dioxide (EtCO2, panel c). Significant correlations are marked with asterisks (*)

Pendelluft volumes measured at the lowest PS level positively, albeit loosely, correlated with RVD in the high-pendelluft group (p = .04, r = .69, RVD range 18–47%), but not in the low-pendelluft group (p = .65, RVD range 0–41%, Fig. 7). While in the high-pendelluft group a high RVD value was associated with relevant pendelluft volumes, in the low-pendelluft group high RVD values were present despite the low pendelluft values (representative patient analyses in Additional file 1: Figure S4).

Fig. 7
figure 7

Correlation between pendelluft volume and regional ventilation delay. At the lowest pressure support ventilation (PSV) level, pendelluft volume positively correlated with the percentage of lung zones affected by ventilation delay (RVD) in the high-pendelluft group (p = .04, r = .69, empty circles), but not in the low-pendelluft group (p = .65, filled circles). While in the high-pendelluft group a high regional ventilation delay value was associated with the presence of pendelluft, some patients in the low-pendelluft group showed high regional ventilation delay values but low pendelluft volumes

ICU mortality, length of ICU stay and 28-day ventilator-free days did not differ between high- and low-pendelluft groups.

Discussion

In the present study, we show that pendelluft was present in eight out of 20 patients (40%) and it was possible to detect and quantify it by EIT monitoring. The reduction of PS led to a doubling of baseline pendelluft volume; the phenomenon was almost entirely due to gas movement from the ventral to the dorsal regions. At the lowest PS level, the group of patients with high pendelluft volumes showed a greater increase in EtCO2; in a subset of these patients, the increase in pendelluft volume significantly correlated with markers of respiratory distress.

To our knowledge, this is the first clinical study measuring the gas volume subject to pendelluft during spontaneous assisted ventilation, a hidden phenomenon often difficult to recognize from standard ventilator waveforms. The pendelluft phenomenon is classically disclosed in paralyzed patients by an end-inspiratory occlusion maneuver [23]. Recently, EIT availability at the bedside allowed researchers to show the presence of pendelluft during assisted mechanical ventilation: in animal models, a vigorous inspiratory effort caused movement of gas at the beginning of inspiration from ventral to dorsal zones, which are more subject to diaphragm activity, but no quantitative measurement was performed [6]. Of note, none of our patients’ characteristics at baseline was associated with the presence of relevant pendelluft volumes, suggesting that direct monitoring is necessary for its detection. As described in the present study, the pendelluft phenomenon occurred at the moment of cycling from expiration to inspiration. The phenomenon requires a regional monitoring tool with high temporal resolution, making EIT an ideal choice. To remove cardiac-related impedance changes, we used an approach based on event-triggered averaging, a filtering technique that preserves the actual EIT waveform better than standard low-pass filtering [21].

Since gas subject to pendelluft comes from within the lung, it contains higher CO2 and lower O2 concentrations as compared to fresh gas, and contributes less to pulmonary gas exchange; moving such a gas from one lung region to another results in a waste of energy and reduced ventilatory efficiency, two crucial factors during ventilator weaning. The different response to PS reduction in the high- as compared to the low-pendelluft group and the correlation with markers of respiratory distress in some of the high-pendelluft patients suggest that the chosen cut-off properly identifies the relevance of the phenomenon. However, due to the exploratory nature of our study, we cannot provide any indication about a specific cut-off for future studies.

The detection of pendelluft at baseline and the correlation with respiratory system compliance suggest an association with some intrinsic lung characteristics or with altered respiratory muscles/lung interaction; when workload is increased and respiratory muscles activity is higher, the pendelluft phenomenon directed towards the dorsal zones is then exacerbated. Reduction of PS is associated with redistribution of tidal volume to the dorsal regions, due to higher diaphragmatic activity [15]; we speculate that increased diaphragmatic activity causes pendelluft in some patients at the very beginning of inspiration, leading to an intra-pulmonary movement of gas from the ventral to the dorsal zones, which are more exposed to the diaphragm stretching forces.

While the impact of pendelluft on some respiratory parameters was detectable, the measured pendelluft volumes were relatively small (on average, 5.4% of the Tv at the lowest PS level in the high-pendelluft group, maximum value 8.4%). Previous literature showed that a theoretical pendelluft volume < 2% of the tidal volume can be concentrated in a poorly ventilated region, reaching up to 13% of the volume entering the region and therefore leading to notable effects [4]. Pendelluft volumes were possibly underestimated because the analyzed four ROIs were considered as homogeneous lung zones; the application of smaller ROIs might lead to a higher sensitivity disclosing a higher within-lung movement of gas.

Pendelluft appeared a phenomenon distinct from RVD as known from the literature (i.e., a slower regional inflation as compared to the global signal) [22, 24]. We showed that when pendelluft is present RVD is also present: when a region starts inflation later, it is likely delayed as compared to the entire lung. However, the opposite is not necessarily true, as suggested by the lack of correlation between pendelluft volume and RVD in the low-pendelluft group: some regions may show slow inflation without the presence of pendelluft. Therefore, an RVD measurement tool is not adequate to measure pendelluft, due to excessive false positives.

In the high-pendelluft group, the pendelluft phenomenon was associated with lower ventilatory efficiency and higher distress. While minute ventilation was unchanged, EtCO2 increased more in the high-pendelluft group, possibly because of reduced CO2 removal due to inefficient intra-pulmonary gas mixture or increased respiratory muscles metabolism [25,26,27]. Then, a trend towards greater reduction of tidal volumes was recorded, possibly due to a partial wasting of patient’s inspiratory effort for pendelluft gas movement, as described in an obese patient ventilated with insufficient PEEP [28]. Lastly, in some patients the increase in pendelluft volumes positively correlated with markers of respiratory distress such as increased respiratory rate and p0.1. Of note, a higher respiratory rate associated with reduced Tv leads to increased workload, due to higher dead space/tidal volume ratios, further affecting ventilatory efficiency [29]. The reduction of Tv and the increase of respiratory rate are known risk factors for weaning failure; our data raise the hypothesis that pendelluft leading to higher respiratory workload might represent an adjunctive cause of difficulty in weaning [30].

The correlation between the volume of pendelluft and respiratory system compliance might suggest that, in patients with higher compliance, pendelluft could be due to an interaction between diaphragm and lungs with longer (or heterogeneous) time constants, defined as the product of compliance and resistance. In our study protocol, we did not measure airways resistance values. Future studies could elucidate the connection between airway resistance and pendelluft occurrence.

We enrolled patients failing a spontaneous breathing trial reasoning that these patients could be more at risk of difficult weaning and that pendelluft occurrence might be higher in such patients than in easily extubated ones. The restriction of the enrollment to such patients limits the generalizability of the results, but allowed us to focus on subjects for whom even a small increase in ventilatory workload could be relevant. We restricted our investigation to reduction of PS levels; however, during weaning PEEP is also lowered and low PEEP levels are associated with pendelluft exacerbation in animal studies [31]. Future studies might investigate the occurrence of the pendelluft phenomenon in other patient populations, at different PEEP levels and whether pendelluft is to any extent specific to difficult weaning. Another limitation is that our method is based on off-line measurement of pendelluft volumes after EIT data download and analysis; however, the same process could be easily integrated in the current EIT monitors becoming available at the bedside to set the ventilator properly, possibly reducing pendelluft negative effects. Finally, the presence of pendelluft was not associated with relevant long-term outcomes; our study was certainly not powered to disclose such associations, and we believe that long-term outcomes may be explained only in part by single specific factors.

Conclusions

In our study, EIT technology disclosed the presence of relevant pendelluft volumes during reduction of PS in a fair amount of patients failing a spontaneous breathing trial. Pendelluft was mainly due to the movement of gas from the ventral to the dorsal zone, and appeared associated with increased respiratory workload.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author after local Institutional Review Board (Comitato Etico ASST Monza) approval.

Abbreviations

PS:

Pressure support

EIT:

Electrical impedance tomography

ICU:

Intensive care unit

Tv:

Tidal volume

Mv:

Minute ventilation

p0.1:

Mouth occlusion pressure at 100 ms

Crs :

Compliance of the respiratory system

EtCO2 :

End-tidal carbon dioxide

ROI:

Region of interest

RVD:

Regional ventilation delay

References

  1. Yoshida T, Nakahashi S, Nakamura MAM, Koyama Y, Roldan R, Torsani V, et al. Volume-controlled ventilation does not prevent injurious inflation during spontaneous effort. Am J Respir Crit Care Med. 2017;196:590–601.

    Article  Google Scholar 

  2. Brochard L, Slutsky A, Pesenti A. Mechanical ventilation to minimize progression of lung injury in acute respiratory failure. Am J Respir Crit Care Med. 2017;195:438–42.

    Article  Google Scholar 

  3. Bellani G, Pesenti A. Assessing effort and work of breathing. Curr Opin Crit Care. 2014;20:352–8.

    Article  Google Scholar 

  4. Greenblatt EE, Butler JP, Venegas JG, Winkler T. Pendelluft in the bronchial tree. J Appl Physiol. 1985;2014(117):979–88.

    Google Scholar 

  5. Lumb A. Causes of time dependence of pulmonary elastic behaviour. Nunn’s Applied Respiratory Physiology. Amsterdam: Elsevier; 2016. p. 24.

    Google Scholar 

  6. Yoshida T, Torsani V, Gomes S, De Santis RR, Beraldo MA, Costa EL, et al. Spontaneous effort causes occult pendelluft during mechanical ventilation. Am J Respir Crit Care Med. 2013;188:1420–7.

    Article  Google Scholar 

  7. Brochard L, Rauss A, Benito S, Conti G, Mancebo J, Rekik N, et al. Comparison of three methods of gradual withdrawal from ventilatory support during weaning from mechanical ventilation. Am J Respir Crit Care Med. 1994;150:896–903.

    Article  CAS  Google Scholar 

  8. Beduneau G, Pham T, Schortgen F, Piquilloud L, Zogheib E, Jonas M, et al. Epidemiology of weaning outcome according to a New definition. The WIND Study. Am J Respir Crit Care Med. 2017;195:772–83.

    Article  Google Scholar 

  9. Sellares J, Ferrer M, Cano E, Loureiro H, Valencia M, Torres A. Predictors of prolonged weaning and survival during ventilator weaning in a respiratory ICU. Intensive Care Med. 2011;37:775–84.

    Article  Google Scholar 

  10. Funk GC, Anders S, Breyer MK, Burghuber OC, Edelmann G, Heindl W, et al. Incidence and outcome of weaning from mechanical ventilation according to new categories. Eur Respir J. 2010;35:88–94.

    Article  Google Scholar 

  11. Jubran A, Lawm G, Kelly J, Duffner LA, Gungor G, Collins EG, et al. Depressive disorders during weaning from prolonged mechanical ventilation. Intensive Care Med. 2010;36:828–35.

    Article  Google Scholar 

  12. Perren A, Brochard L. Managing the apparent and hidden difficulties of weaning from mechanical ventilation. Intensive Care Med. 2013;39:1885–95.

    Article  Google Scholar 

  13. Pesenti A, Musch G, Lichtenstein D, Mojoli F, Amato MBP, Cinnella G, et al. Imaging in acute respiratory distress syndrome. Intensive Care Med. 2016;42:686–98.

    Article  Google Scholar 

  14. Global initiative for COPD. Pocket guide to COPD diagnosis, management and prevention, 2019 https://goldcopd.org/wp-content/uploads/2018/11/GOLD-2019-v1.7-FINAL-14Nov2018-WMS.pdf. Accessed 3 Apr 2020.

  15. Mauri T, Bellani G, Confalonieri A, Tagliabue P, Turella M, Coppadoro A, et al. Topographic distribution of tidal ventilation in acute respiratory distress syndrome: effects of positive end-expiratory pressure and pressure support. Crit Care Med. 2013;41:1664–73.

    Article  Google Scholar 

  16. Whitelaw WA, Derenne JP, Milic-Emili J. Occlusion pressure as a measure of respiratory center output in conscious man. Respir Physiol. 1975;23:181–99.

    Article  CAS  Google Scholar 

  17. Bellani G, Grassi A, Sosio S, Foti G. Plateau and driving pressure in the presence of spontaneous breathing. Intensive Care Med. 2019;45:97–8.

    Article  Google Scholar 

  18. Bellani G, Grassi A, Sosio S, Gatti S, Kavanagh B, Pesenti A, et al. Driving pressure is associated with outcome during assisted ventilation in acute respiratory distress syndrome. Anesthesiology. 2019;131(3):594–604.

    Article  Google Scholar 

  19. Foti G, Cereda M, Banfi G, Pelosi P, Fumagalli R, Pesenti A. End-inspiratory airway occlusion: a method to assess the pressure developed by inspiratory muscles in patients with acute lung injury undergoing pressure support. Am J Respir Crit Care Med. 1997;156:1210–6.

    Article  CAS  Google Scholar 

  20. Yang KL, Tobin MJ. A prospective study of indexes predicting the outcome of trials of weaning from mechanical ventilation. N Engl J Med. 1991;324:1445–50.

    Article  CAS  Google Scholar 

  21. Coppadoro A, Eronia N, Foti G, Bellani G. Event-triggered averaging of electrical impedance tomography (EIT) respiratory waveforms as compared to low-pass filtering for removal of cardiac related impedance changes. J Clin Monit Comput. 2019. https://doi.org/10.1007/s10877-019-00348-2.

    Article  PubMed  Google Scholar 

  22. Wrigge H, Zinserling J, Muders T, Varelmann D, Gunther U, von der Groeben C, et al. Electrical impedance tomography compared with thoracic computed tomography during a slow inflation maneuver in experimental models of lung injury. Crit Care Med. 2008;36:903–9.

    Article  Google Scholar 

  23. Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care. 2014;59:1773–94.

    Article  Google Scholar 

  24. Muders T, Luepschen H, Zinserling J, Greschus S, Fimmers R, Guenther U, et al. Tidal recruitment assessed by electrical impedance tomography and computed tomography in a porcine model of lung injury*. Crit Care Med. 2012;40:903–11.

    Article  Google Scholar 

  25. Shikora SA, Bistrian BR, Borlase BC, Blackburn GL, Stone MD, Benotti PN. Work of breathing: reliable predictor of weaning and extubation. Crit Care Med. 1990;18:157–62.

    Article  CAS  Google Scholar 

  26. Manthous CA, Hall JB, Kushner R, Schmidt GA, Russo G, Wood LD. The effect of mechanical ventilation on oxygen consumption in critically ill patients. Am J Respir Crit Care Med. 1995;151:210–4.

    Article  CAS  Google Scholar 

  27. Bellani G, Foti G, Spagnolli E, Milan M, Zanella A, Greco M, et al. Increase of oxygen consumption during a progressive decrease of ventilatory support is lower in patients failing the trial in comparison with those who succeed. Anesthesiology. 2010;113:378–85.

    Article  Google Scholar 

  28. Teggia Droghi M, De Santis Santiago RR, Pinciroli R, Marrazzo F, Bittner EA, Amato MBP, et al. High positive end-expiratory pressure allows extubation of an obese patient. Am J Respir Crit Care Med. 2018;198:524–5.

    Article  Google Scholar 

  29. Hedenstierna G. The effect of respiratory frequency on pulmonary function during artificial ventilation. a review. Acta Anaesthesiol Scand. 1976;20:20–31.

    Article  CAS  Google Scholar 

  30. Tobin MJ, Perez W, Guenther SM, Semmes BJ, Mador MJ, Allen SJ, et al. The pattern of breathing during successful and unsuccessful trials of weaning from mechanical ventilation. Am Rev Respir Dis. 1986;134:1111–8.

    CAS  PubMed  Google Scholar 

  31. Yoshida T, Roldan R, Beraldo MA, Torsani V, Gomes S, De Santis RR, et al. Spontaneous effort during mechanical ventilation: maximal injury with less positive end-expiratory pressure. Crit Care Med. 2016;44:e678–88.

    Article  Google Scholar 

Download references

Acknowledgements

We are grateful to Maddalena Teggia-Droghi, MD, for careful manuscript revision and the very useful suggestions provided.

Funding

Department funds, University of Milan-Bicocca.

Author information

Authors and Affiliations

Authors

Contributions

GB and GF designed the study; AG, GC, FR, NE, AB acquired the data; AC analyzed the data; AC and GB drafted the manuscript; GF and RF substantively revised the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Giacomo Bellani.

Ethics declarations

Ethics approval and consent to participate

All procedures performed in the present study were in accordance with the ethical standards of the local Institutional Ethical Committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The local Institutional Review Board (Comitato Etico ASST Monza) approved the study protocol, and written informed consent was obtained as per local regulations.

Consent for publication

Not applicable.

Competing interests

GB, AB and GF received lecturing fees from Draeger Medical.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary information

Additional file 1.

A pdf file containing supplemental Methods and Figures (S1–S4).

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Coppadoro, A., Grassi, A., Giovannoni, C. et al. Occurrence of pendelluft under pressure support ventilation in patients who failed a spontaneous breathing trial: an observational study. Ann. Intensive Care 10, 39 (2020). https://doi.org/10.1186/s13613-020-00654-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13613-020-00654-y

Keywords