Skip to main content

A case report on epicardial ultrasonography of coronary anastomoses using a stabilizing device without the use of ultrasound gel

Abstract

Background

Intraoperative epicardial ultrasonography of coronary artery bypass graft anastomoses is a procedure used for anatomical quality assessment of peripheral anastomoses during coronary artery bypass grafting. However, it may be difficult to keep the ultrasound transducer in steady contact with the anastomoses on the beating heart without causing any deformation. Furthermore, we are not aware of any sterile ultrasound gel approved for application into the pericardial space.

Case presentation

We report a method using a stabilizing connecting device for an ultrasound transducer to be used for visualization of coronary anastomoses without application of ultrasound gel during on-pump coronary bypass surgery.

Conclusion

Use of a stabilizing device and coagulated blood from the patient as an alternative for ultrasound gel facilitates peroperative ultrasonography of coronary anastomoses. The procedure provides surgeons with non-deformed echocardiographic longitudinal and transverse images of all parts of the anastomoses.

Trial registration

The patient participated in a still ongoing clinical feasibility study: Trial registration: ClinicalTrials.gov ID: NCT02919124; Registered September 29, 2016.

Peer Review reports

Background

Intraoperative quality assessment of peripheral coronary anastomoses during coronary artery bypass grafting (CABG) has come into increased focus as it offers surgeons the opportunity to detect technical failures mandating revision of the bypass graft before the chest closure. Failures that may be detected are e.g. twisting of the graft, anastomosis stenosis, a narrow toe of the anastomosis, kinked graft etc. [1, 2].

Transit-time flow measurement (TTFM) during CABG is well described and has been used for approximately two decades for functional graft assessment [3, 4]. Recently, use of high-frequency intraoperative epicardial ultrasonography of coronary anastomoses has been shown to be complementary to TTFM improving the diagnostic accuracy of intraoperative graft assessment [5]. Optimal acoustic and stable contact between the ultrasound probe and the anastomoses to be examined are needed without the risk of distortion of the anastomoses.

We are not aware of any ultrasound gel approved for application into the pericardial space during cardiac surgery in humans. The objective of this case report is to describe a method for ultrasound imaging of coronary artery anastomoses during on-pump CABG using the patient’s coagulated blood instead of ultrasound gel together with a stabilizing device.

Case presentation

A male, 58 years of age, was referred for an elective CABG procedure due to stable angina pectoris. Five weeks earlier, he suffered a ST-elevation acute myocardial infarction due to three-vessel coronary artery disease. Coronary angiography showed an occluded right coronary artery (RCA) and significant stenoses located to the obtuse marginal branch (OM) and the left anterior descending coronary artery (LAD). The occluded RCA was treated with emergent percutaneous balloon angioplasty and stenting within 6 h after the myocardial infarction developed. The patient experienced a temporary 3o-atrio-ventricular block, but was discharged in sinus rhythm 4 days later after an otherwise uneventful course. Following a heart team discussion, it was decided to perform surgical revascularisation of the left sided coronary arteries 4 weeks later. The left ventricular ejection fraction was 50%, and no valvular diseases were diagnosed by transthoracic echocardiography. Logistic EuroSCORE II was 0.79%. After surgery, the patient was discharged following an uneventful course.

Surgical procedure

Surgery was performed on-pump. Routine thoracic surgical and anaesthetic procedures were employed, and transoesophageal echocardiography was available. A saphenous vein graft was harvested endoscopically and anastomosed end-to-side to the OM with the proximal anastomosis on the ascending aorta. A left internal mammary artery (LIMA) pedicel graft was anastomosed to the LAD. Cold blood cardioplegia was administered twice in the aortic root during 29 min of aortic cross-clamping. Perfusion time was 70 min.

High-resolution epicardial ultrasonography

The Medistim VeriQ™ System with a 15 MHz ultrasonography probe was used (Medistim A/S, Oslo, Norway). The ultrasonography probe was connected to a disposable stabilizing device (Echoclip, Aalborg University Hospital, Denmark) (Fig. 1). This device allows the anastomosis to be imaged without applying pressure on the graft when stabilising the heart with the stabiliser. The surgeon should make sure that the graft is located in the cavity of the stabilizer in order to avoid obstruction of the graft flow during the scanning procedure. The echoclip device is created such that slippery excess coagulated blood and ultrasound gel (if this will be approved in the future) are allowed to escape from the cavity during scanning. The echoclip device, which is not commercially available yet, is a disposable article, which comes in different sizes.

Fig. 1
figure 1

Drawing of the echoclip stabilizing device

Instead of gel we used coagulated blood obtained from the patient before heparinization in order to obtain acoustic contact with the anastomosis. Coagulated blood was placed in the cavity of the stabilizer designed to receive a part of the graft and the anastomosis to be examined (Fig. 2). The peripheral vein graft anastomosis was validated during cross-clamp, while infusing cold blood cardioplegia directly into the vein graft. A rotating plate to which the imaging probe was secured allowed imaging of the coronary anastomosis both the longitudinal (Fig. 3) and transverse plane (Fig. 4). The LIMA anastomosis was validated after the cross-clamp was released while still on pump. Less than 5 min was used to obtain images of the heel, the central portion and the toe of the anastomoses in both the longitudinal and transverse planes. The coagulated blood was sucked away from the pericardial space after the imaging procedure. TTFM was performed in addition to epicardial ultrasonography.

Fig. 2
figure 2

Coagulated blood placed in the cavity of the stabilizing device designed to receive the graft at the site of the coronary anastomosis

Fig. 3
figure 3

Epicardial ultrasonography of the left internal mammary artery anastomosis to the left anterior descending coronary artery in the longitudinal plane

Fig. 4
figure 4

Epicardial ultrasonography of the left internal mammary artery anastomosis in the transverse plane

Discussion and conclusions

A combination of TTFM and high-frequency epicardial ultrasonography of peripheral bypass anastomoses facilitate quality assessment of coronary anastomoses offering a possibility for graft revision before closure of the chest [5]. However, surgeons have to overcome a learning curve especially in relation to the use of epicardial ultrasonography of peripheral coronary anastomoses. Experienced surgeons have indicated that the basics can be learned in about 10–20 grafts [1, 2].

We have described a reproducible and simple procedure to perform epicardial ultrasonography of peripheral coronary bypass anastomoses with use of a stabilizing device and with the use of the patients coagulated blood instead of ultrasound gel. Thus, the patients coagulated blood may be used until sterile ultrasound gel is approved for use in the pericardial space during cardiac surgery. It is a limitation, that this procedure can only be performed if coagulated blood is collected prior to heparinization. It remains to be proven, if coagulated blood can be used instead of gel without the use of a stabilizing device both during on-and off-pump cases and whether the procedure can be used during off-pump surgery. Development of a handle attached to the echoclip device will probably be needed if ultrasound imaging is to be performed during off-pump surgery on the backside of the heart, and if imaging is need after administration of protamine.

In conclusion, use of a stabilizing device and coagulated blood from the patient as an alternative for ultrasound gel facilitates peroperative ultrasonography of coronary anastomoses. The procedure provides surgeons with non-deformed echocardiographic longitudinal and transverse images of all parts of the anastomoses.

Abbreviations

CABG:

Coronary artery bypass grafting

LAD:

Left anterior descending coronary artery

LIMA:

Left internal mammary artery

OM:

Obtuse marginal branch

RCA:

Right coronary artery

TTFM:

Transit-time flow measurement

References

  1. Kieser TM. Graft quality verification in coronary artery bypass graft surgery: how, when and why? Curr Opin Cardiol. 2017;32:722–36.

    Article  Google Scholar 

  2. Kieser TM, Taggart DP. The use of intraoperative graft assessment in guiding graft revision. Ann Cardiothorac Surg. 2018;7:652–62.

    Article  Google Scholar 

  3. D’Ancona G, Karamanoukian HL, Ricci M, Schmid S, Bergsland J, Salerno TA. Graft revision after transit Jtime flow measurement in off-pump coronary artery bypass grafting. Eur J Cardiothorac Surg. 2000;17:287–93.

    Article  Google Scholar 

  4. Di Giammarco G, Marinelli D, Foschi M, Di Mauro M. Intraoperative graft verification in coronary surgery. J Cardiovasc Med. 2017;18:295–304.

    Article  Google Scholar 

  5. Di Giammarco G, Canosa C, Foschi M, Rabozzi R, Marinelli D, Masuyama S, et al. Intraoperative graft verification in coronary surgery: increased diagnostic accuracy adding high-resolution epicardial ultrasonography to transit-time flow measurement. Eur J Cardiothorac Surg. 2014;45:e41–5.

    Article  Google Scholar 

Download references

Acknowledgements

We thank all clinical personnel involved in the clinical management of the patient.

Funding

JJA has a free Medistim VeriQ™ System with 15 MHz ultrasonography probes available for research purposes from Medistim, Denmark. Echoclip stabilizing devices to be used for a feasibility study are produced for research purposes by Medistim A/S, Norway.

Availability of data and materials

All data presented are available from the corresponding author on reasonable request.

Author information

Authors and Affiliations

Authors

Contributions

JJA: Author of the text. DN and ASJ: Collected data and critically reviewed the text. ASJ: Analysed the ultrasonographic images. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Jan Jesper Andreasen.

Ethics declarations

Ethics approval and consent to participate

Use of the echoclip stabilizing device and use of coagulated blood from the patient was approved by the Ethics committee of the Nord Denmark Region (N-20160012) and by the Danish Medicines Agency (reference number 2016020479), and the patient gave his informed written consent to participate in a clinical study (ClinicalTrials.gov ID: NCT02919124; Registered September 29, 2016).

Consent for publication

The patient gave written consent to anonymous publication of the data.

Competing interests

The corresponding author is a co-inventor of the echoclip device, and the North Denmark Region holds a patent of the echoclip device.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Andreasen, J.J., Nøhr, D. & Jørgensen, A.S. A case report on epicardial ultrasonography of coronary anastomoses using a stabilizing device without the use of ultrasound gel. J Cardiothorac Surg 14, 59 (2019). https://doi.org/10.1186/s13019-019-0882-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-019-0882-2

Keywords