Combined MR-assisted motion and partial volume effects corrections – impact on PET data quantification

Our goal in this study was to characterize the combined effect of MR-assisted motion correction (MC) and partial volume effects correction (PVEC) on the estimation of [11C]NNC112 binding potential (BP) in healthy volunteers.

29 subjects were scanned on the Siemens 3T MR-BrainPET scanner prototype. Emission data were acquired in list mode format for 90-minutes following the i.v. administration of ~8 mCi of [11C]NNC112. The head attenuation map was obtained from the MPRAGE data using an atlas-based method. Head motion estimates were derived from the MR data and used to correct the PET data in LOR space before image reconstruction [1]. PVEC was applied to the motion corrected data using the region-based voxel-wise (RBV) method [2] and regions of interest (ROIs) defined from the MPRAGE images using FreeSurfer and the measured point-spread function [3]. BPnd for each of the ROIs was estimated in PMOD using the simplified reference tissue kinetic (SRTM) model and the cerebellum as a reference tissue.

Maximum translations of up to 9 mm and rotations of up to 12 degrees have been observed in this group of subjects (Figure 1). Less variability in the tissue time activity curves (TACs) was noted after MC (the curves before and after MC for a representative subject are shown in Figure 2). The percentage changes in BPnd after MC and PVEC revealed both under- and overestimation in the ROIs analyzed (Figure 3). The cumulative effect exceeded 100% for some of the structures analyzed.

Figure 1
figure 1

Maximum translations (left) and rotations (right) measured in 29 healthy volunteers scanned for 90 minutes

Figure 2
figure 2

TACs in the left and right caudate and cerebellar cortices before (left) and after (right) MR-assisted motion correction

Figure 3
figure 3

Percent change in NNC112 BPnd after motion and partial volume effects corrections

Significant motion and PVE were observed in all the subjects, biasing the PET estimates. The combined effect is difficult to predict, depending on the size and location of the structure of interest and patient compliance. Without addressing these issues, the value of the BPnd’s derived from these data is questionable.

References

  1. Catana C, Benner T, van der Kouwe A, et al.: MRI-assisted PET motion correction for neurologic studies in an integrated MR-PET scanner. J Nucl Med 2011, 52: 154–161. 10.2967/jnumed.110.079343

    Article  PubMed Central  PubMed  Google Scholar 

  2. Thomas BA, Erlandsson K, Modat M, et al.: The importance of appropriate partial volume correction for PET quantification in Alzheimer's disease. Eur J Nucl Med Mol Imaging 2011, 38: 1104–1119. 10.1007/s00259-011-1745-9

    Article  PubMed  Google Scholar 

  3. Bowen SL, Byars LG, Michel CJ, Chonde DB, Catana C: Influence of the partial volume correction method on (18)F-fluorodeoxyglucose brain kinetic modelling from dynamic PET images reconstructed with resolution model based OSEM. Phys Med Biol 2013, 58: 7081–7106. 10.1088/0031-9155/58/20/7081

    Article  CAS  PubMed Central  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Catana, C., Chonde, D.B., Chen, K.T. et al. Combined MR-assisted motion and partial volume effects corrections – impact on PET data quantification. EJNMMI Phys 1 (Suppl 1), A38 (2014). https://doi.org/10.1186/2197-7364-1-S1-A38

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-1-S1-A38

Keywords