Abstract

The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of the analytical units and then an evaluation in routine work conditions. The operational characteristics of the system were also studied.The first stage included a photometric study: dependent on the absorbance, the inaccuracy varies between +0.5% to -0.6% at 405 nm and from -5.6% to 10.6% at 340 nm; the imprecision ranges between -0.22% and 0.56% at 405 nm and between 0.09% and 2.74% at 340 nm. Linearity was acceptable, apart from a very low absorbance for NADH at 340 nm; and the imprecision of the serum sample pipetter was satisfactory.Twelve serum analytes were studied under routine conditions: glucose, urea urate, cholesterol, triglycerides, total bilirubin, creatinine, phosphate, iron, aspartate aminotransferase, alanine aminotransferase and gamma-glutamyl transferase.The within-run imprecision (CV%) ranged from 0.67% for phosphate to 2.89% for iron and the between-run imprecision from 0.97% for total bilirubin to 7.06% for iron. There was no carryover in a study of the serum sample pipetter. Carry-over studies with the reagent and sample pipetters shows some cross contamination in the iron assay.