1932

Abstract

Abstract

The fluid mechanics of artificial blood pumps has been studied since the early 1970s in an attempt to understand and mitigate hemolysis and thrombus formation by the device. Pulsatile pumps are characterized by inlet jets that set up a rotational “washing” pattern during filling. Strong regurgitant jets through the closed artificial heart valves have Reynolds stresses on the order of 10,000 dynes/cm2 and are the most likely cause of red blood cell damage and platelet activation. Although the flow in the pump chamber appears benign, low wall shear stresses throughout the pump cycle can lead to thrombus formation at the wall of the smaller pumps (10–50 cc). The local fluid mechanics is critical. There is a need to rapidly measure or calculate the wall shear stress throughout the device so that the results may be easily incorporated into the design process.

Loading

Article metrics loading...

/content/journals/10.1146/annurev.fluid.38.050304.092022
2006-01-21
2024-04-25
Loading full text...

Full text loading...

/content/journals/10.1146/annurev.fluid.38.050304.092022
Loading
/content/journals/10.1146/annurev.fluid.38.050304.092022
Loading

Data & Media loading...

  • Article Type: Review Article
This is a required field
Please enter a valid email address
Approval was a Success
Invalid data
An Error Occurred
Approval was partially successful, following selected items could not be processed due to error