RF Attenuation Characteristics for In Vivo Wireless Healthcare Chip

, , , , and

Published 8 July 2005 Copyright (c) 2005 The Japan Society of Applied Physics
, , Citation Tomohiro Yamada et al 2005 Jpn. J. Appl. Phys. 44 5275 DOI 10.1143/JJAP.44.5275

1347-4065/44/7R/5275

Abstract

We investigate a wireless communication system for an in vivo healthcare chip. In this paper, we present measured attenuation characteristics through the human body at several frequencies. In the measurement, we use physiological saline and fresh meat instead of a real human body. From the measured results, we found that 13.56 MHz has an attenuation of 47 dB and is suitable for the proposed system.