Preliminary Results of Vacuum Pressure Measurement by Laser Ionization Method

, and

Copyright (c) 1987 The Japan Society of Applied Physics
, , Citation Shingo Ichimura et al 1987 Jpn. J. Appl. Phys. 26 L573 DOI 10.1143/JJAP.26.L573

1347-4065/26/5A/L573

Abstract

The possibility of vacuum pressure measurement by ionizing gaseous atoms or molecules with laser radiation was studied. Xe ions were successfully detected after irradiation by a focused Xe-Cl excimer laser, indicating that non-resonant 4 photon ionization was realized with the present system. It was confirmed that Xe signal intensity, as measured by a quadrapole mass analyzer, changed in proportion to Xe gas pressure from 10-5 Torr to 10-9 Torr. Xe isotopes were also measured and the results correlated well with their natural abundance.