1 April 1974 Coded Aperture Gamma-Ray Imaging with Stochastic Apertures
A. Z. Akcasu, R. S. May, G. F. Knoll, W. L. Rogers, K. F. Koral, L. W. Jones
Author Affiliations +
Abstract
A number of schemes employing coded apertures are currently under investigation for in vivo imaging of gamma-emitting tracers. These schemes are based on stationary aperture codes, which depend upon having a cleanly peaked spatial auto-correlation function to facilitate faithful recovery of the source distribution. The image is recovered by performing a spatial integration optically, digitally, or by some other suitable means. This paper describes an alternate method based on time modulated apertures with digital image reconstruction. In contrast to the stationary aperture, the time modulated aperture requires no spatial integration to achieve faithful reproduction of the source distribution. This solves the problem of false structure arising from out-of-focus sources. Good 3-dimensional point response functions free of side lobes have been obtained and it is expected that this approach will yield high quality clinical images.
A. Z. Akcasu, R. S. May, G. F. Knoll, W. L. Rogers, K. F. Koral, and L. W. Jones "Coded Aperture Gamma-Ray Imaging with Stochastic Apertures," Optical Engineering 13(2), 132117 (1 April 1974). https://doi.org/10.1117/12.7971683
Published: 1 April 1974
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Coded apertures

Gamma ray imaging

Stochastic processes

Image restoration

Modulation

Image quality

In vivo imaging