1 August 2013 Tiny endoscopic optical coherence tomography probe driven by a miniaturized hollow ultrasonic motor
Author Affiliations +
Abstract
We present an endoscopic probe for optical coherence tomography (OCT) equipped with a miniaturized hollow ultrasonic motor that rotates the objective lens and provides an internal channel for the fiber to pass through, enabling 360 deg unobstructed circumferential scanning. This probe has an outer diameter of 1.5 mm, which is ultra-small for motorized probes with an unobstructed view in distal scanning endoscopic OCT. Instead of a mirror or prism, a customized aspheric right-angle lens is utilized, leading to an enlargement of the numerical aperture and thus high transverse resolution. Spectral-domain OCT imaging of bio-tissue and a phantom are demonstrated with resolution of 7.5  μm(axial)×6.6  μm(lateral) and sensitivity of 96 dB.

1.

Introduction

Optical coherence tomography (OCT) is an emerging modality for obtaining cross-sectional images.1 As a noninvasive cross-sectional imaging technique with micrometer resolution, OCT has become very promising in the field of biomedical applications. Due to limitation of the light penetration, small and high-resolution probes for endoscopic OCT have been regarded as powerful tools in examining internal organs.2 There have been various designs of endoscopic OCT probes over the past decade,36 which can be divided into two groups based on the scanning modes: forward scanning and circumferential scanning. Among them, circumferential probes are utilized extensively for imaging within hollow tissues, such as arteries, esophagus, and colon.7 For the circumferential scanning probes, the scanning can be implemented in two ways based on their driving modes:8 proximal scanning or distal scanning.

One common design of proximal scanning probes is implemented by rotating the whole optical fiber at the distal end and the objective lens are fixed with the fiber.9,10 However, the fiber is irregularly twisted in the proximal scanning, causing unstable polarization distortion. The rotation may be nonuniform due to the irregular passageway, such as arterial bifurcation. Therefore proximal scanning suffers from the inevitable distortion and artifacts in the recorded OCT images.

In recent years, distal scanning is of high interest, which is mainly implemented by attaching the deflection prism to a micro-motor at the distal end. In this case, the whole fiber remains stationary, so the twisting and nonuniformity problems are avoided.11 Nevertheless, this design has two challenges. One is the wire-shadow effect resulting from the light blocked by power wires, making a portion of the image invisible. The other is the challenge of miniaturizing the probe due to the complex fabrication of a micro-motor. In fact, many novel technologies have emerged to solve those problems, such as microelectromechanical systems (MEMS)4,6,12 and squiggle motor.13 Due to the small size of the endoscopic probes, the numerical aperture of the objective lens is limited and thus the lateral resolution is lower than 10 μm4,6,813 in distal scanning probes. Meanwhile, to our knowledge, the current smallest distal scanning probe is 1.1 mm in diameter but suffers from wire-shadows,14 and the outer diameter of the reported smallest shadow-free distal scanning probe is 2.4 mm.13

In this article, we propose a new stationary-fiber ultra-small probe, which is equipped with a novel miniaturized hollow ultrasonic motor and a custom-designed objective lens to achieve high lateral resolution. The novel hollow ultrasonic motor enables the fiber to pass through its inside and permits having the fiber, the objective lens and the motor on the same side, enabling 360 deg imaging without any shadow resulted from light blocking by the power wires as in the traditional motor-driven endoscopic OCT. Instead of the traditional prism, an aspheric right-angle lens is utilized to be the last optics surface that optimized the size of the light spot. This probe achieves an outer diameter of 1.5 mm, which is ultra-small for shadow-free rotary probe in distal scanning endoscopic OCT.

2.

Device Design

The design scheme of our probe is demonstrated in Fig. 1, which can be divided into the ultrasonic motor and the rotary lens.

Fig. 1

Schematic of the stationary-fiber rotary scanning optical coherence tomography (OCT) probe driven by a hollow ultrasonic motor.

JBO_18_8_086011_f001.png

The ultrasonic motor is an actuator utilizing piezoelectric material. In this case, our home-made hollow motor is 7 mm in length (with the rotor), 0.7 mm in side length and 36 mg in weight. Four pieces of piezoelectric plates are bonded to the sides of a brass square column (0.6×0.6×4mm3). At its center, an internal channel with a diameter of 250μm allows the fiber to pass through the whole motor and be fixed steadily with a protection needle (31-gauge). A tiny steel ring (0.8 mm outer diameter, 0.36 mm inner diameter, and 0.8 mm height) is bonded at the upper end of column to assemble the stator, which transfers torque to the magnet rotor (1 mm outer diameter, 0.4 mm inner diameter, and 1.2 mm height). When two sinusoidal signals with π/2 phase difference between adjacent signals are given to the motor with one end fixed, the piezoelectric plates undergo expansion or contracting sequentially, causing the stator to bend ultrasonic vibration,15 creating an orbital “hula hoop” motion either clockwise or anti-clockwise. The hollow rotor is then rotated by the friction when there is a pressure supplied by the magnetic force between the steel ring and magnet rotor. The phase shift between two channel waves reverses the direction of the orbit and hence changes the rotation direction of the rotor.

Compared with an electromagnetic motor, an ultrasonic motor possesses the following features: (1) 3 to 10 times larger output power density; (2) smaller responding time (microsecond); (3) no electromagnetic interference; (4) the position of the rotor is self-holding since there exists a large friction force between the rotor and stator. Ultrasonic motors have various, flexible and simple structures that make them the best candidates for micro-motors and MEMS application. Up to now, the ultrasonic motor has been applied in auto-focusing systems of cameras, precision instruments, automobiles, aerospace shuttles and medical instruments.

As shown in Fig. 1, the light emitted from the fiber is first collimated by an 8 deg angled gradient index lens (GRIN) (Towingroup, China), which is combined with the hollow rotor, and then deflected by the right-angled lens for focusing. The custom-designed right-angle lens is equivalent to a prism plus an aspheric lens, which is cut from a commercially available aspheric lens (LightPath) that is made of D-ZLAF52LA glass. The reflecting surface is coated with silver by evaporation to enhance its reflectivity. The tilt angle of the right-angle lens is 53 deg to reduce the back reflection. The rotor, GRIN and right-angle lens are glued by ultraviolet-curing epoxy (Norland Optical Adhesive 61, USA). So the circumferential scanning is realized by the rotation of the rotor.

In the traditional motorized distal scanning probe,10 there is spacing between the GRIN and deflection prism, leading to a length increase of the rigid part. Our design eliminates this spacing as the GRIN and right-angle lens are glued together, which further shortens the length of the rigid part. Currently, the length of rigid part of our probe is 14 mm, which can be easily shortened to 10 mm by utilizing a shorter catheter. Besides, the fiber, the objective lens and the motor are on the same side in our design, as shown in Fig. 1. The wires are not in the light path and thus do not disturb the imaging, providing 360 deg unobstructed view for endoscopic OCT.

The maximum diameter of the motor and lens is 1 mm. As shown in Fig. 2(a), this prototype probe is installed in a transparent tube (DURAN glass tube, USA) with an outer diameter of 1.5 mm and wall thickness of 0.1 mm. A drawing of the right-angle lens and GRIN lens is demonstrated in Fig. 2(b). A video of the rotation of the probe is available online (Video 1). The speed of rotation is nearly proportional to the driving voltage with a fluctuation of <2% in speed, as shown in Fig. 3(b). The speed, with a torque of 4μNm, is 2,000 rpm under the driving voltage of 7 Vrms and current of 3 mA. Low scanning speed of 2fps is also available at 1.6 Vrms.

Fig. 2

(a) Photo of the probe with an outer diameter of 1.5 mm. (Video 1, QuickTime, MOV, 10.4 MB) [URL: http://dx.doi.org/10.1117/1.JBO.18.8.086011.1] 650 nm laser light emitted from the probe is for demonstration. (b) A ZEMAX layout of the right-angle lens and GRIN lens.

JBO_18_8_086011_f002.png

Fig. 3

(a) Lateral cross-section of the Huygens point spread function (PSF) of the spot simulated in ZEMAX. (b) The rotation speed of probe vs. voltage of the driving signals. (c) and (d) Full width at half-maximum of the beam at different location in focal region, which is obtained by x and y cross-sections of the PSFs. Values are calculated in air.

JBO_18_8_086011_f003.png

The lateral cross-section of Huygens point spread function (PSF) of the probe is simulated by optical design software ZEMAX, as shown in Fig. 3(a). The theoretical lateral resolution turns out to be 5.6 μm according to the full width at half-maximum (FWHM) of PSF. The corresponding depth of field (DOF) is 80μm, which is designed for obtaining the image of adequate cellular layers. Considering the influence of the package catheter, the X, Y-waist of the beam with the catheter is simulated at different location in the focal region, as shown in Fig. 3(c) and 3(d). Currently the astigmatism caused by the catheter is not compensated. This also leads to an increase of the focal length by 0.1mm. The working distance is then chosen at 1.3mm with an X, Y-waist (FWHM) of 6.4 and 6.6μm, which is a compromise among working distance, resolution and DOF. The probe is currently used with a peritoneoscope that has an imaging channel of 1.8 mm in diameter. It also has the potential for vascular imaging if its outer diameter is further optimized.

3.

Results

To illustrate the performance of our probe, we integrate the probe with a spectral-domain optical coherence tomography (SD-OCT) system. A broadband semiconductor laser diode (Inphenix Inc., Livermore, CA) is used as light source that has a 45nm FWHM bandwidth centered at 853 nm, corresponding to an axial resolution of 7.1 μm16 in air as calculated by Δz=0.44λ2/Δλ. The charge-coupled diode (CCD) (AviiVAM2 CL, Camera Link line scan camera, e2v) contains 2048 pixels with the pixel size of 14×14μm2. The dispersion mismatch of the sample and reference arms is well compensated for numerically.

Experimental measurements are given to demonstrate the detailed performance of the probe. Figure 4(a) shows the axial PSFs at different depths when placing a mirror as the sample. The intensity of the axial PSF decreased 3 dB at the depth of 800μm. Calculated from the FWHM of mirror signal, the axial resolution of the system is 7.5 μm in air, which is in agreement with the axial resolution of 7.1 μm as calculated from the bandwidth of light source before. To obtain the size of the light spot, we measure the differentiation of relative intensity against lateral position of a sharp-edge of metal film placed at the working distance when the light spot moves over it, as shown in Fig. 4(c) and 4(d). A Gaussian fit of the light intensity gives a FWHM width of 6.5 μm. The lateral resolution of the probe, as measured with 20% to 80% width of the slope region of the intensity that is measured from the sharp-edge,17 is 6.6μm, close to the FWHM width of Gaussian fit. The total double-pass return intensity of the probe is 24%, measured by a mirror placed at the working distance. This intensity is strong enough for the CCD to run at an integration time of 50 μs for 512 A-scans. A circumferential scan image consists of 5120 A-scans, so the imaging rate is 2fps with an imaging region of 5120×1024 pixels. To measure the sensitivity of the system, the traditional method of inserting an neutral density filter into the beam path to attenuate the signal is not feasible, as the distance between the focal point and catheter is too small. A special sample is therefore made by optical matching gel (Fibkey, China) and BK7 glass. Fresnel reflection laws give that the reflectivity of their interface is 0.067%, which is also experimentally verified, corresponding to an attenuation of 36 dB. The measured signal-to-noise (S/N) ratio is 59.6 dB, as shown in Fig. 4(b), which means the system sensitivity is 96dB. The band (centered around 320 μm) to the right of the signal from the test sample may be caused by the internal coherence from the small spacing of 0.4mm between the fiber and the GRIN lens.

Fig. 4

(a) Axial PSFs obtained by the probe from the working distance. (b) The return signal of 59.6 dB from a sample made of optical-matching gel and BK7 glass. (c) The differentiation of the reflection intensity from a sharp-edge and Gaussian fit. (d) Measurement of the lateral resolution defined as the 20%–80% bandwidth of the curve as 6.6μm.

JBO_18_8_086011_f004.png

A phantom was imaged with the probe, as shown in Fig. 5(a), which was made by coiling several layers of translucent tape on a glass tube with an outer diameter of 4 mm and wall thickness of 0.1 mm. The thickness of the tape is 50μm. Currently, post processing is implemented to synchronize the rotation. A trigger from the motor is possible in the future to synchronize the rotation and data-acquisition. The layers of the phantom can be distinguished clearly in the image. The arrow indicates the beginning of the inner first layer. The back reflections from catheter and glass tube are weakened by the mentioned tile angle. As shown in Fig. 5(b) and 5(c), ex vivo imaging of chicken trachea tissue gives the structure of the mucous membrane, the under-layer and the cartilage. The scan depth is about 1 mm. It is clear that all the images are unobstructed 360 deg views with no shadow effect. Due to limited penetration and strong scattering of 850nm and relative low depth of focus of the objectives, the quality of image is not good. This may be improved with higher S/N, DOF extending algorithm, proper index matching and dispersion compensation.

Fig. 5

OCT images obtained with the probe. (a) The phantom of translucent tape. (b) Ex vivo chicken trachea tissue. (c) Part of the enlarged image of the dashed region in Fig. 5(b).

JBO_18_8_086011_f005.png

4.

Conclusions

In conclusion, a novel stationary-fiber rotary endoscopic probe for SD-OCT is proposed. This probe provides an unobstructed 360 deg view for circumferential imaging with a spatial resolution of 7.5μm(axial)×6.6μm(lateral) and sensitivity of 96dB. Ex vivo images of chicken trachea and a phantom are shown to demonstrate the performance. The probe has its outer diameter of 1.5 mm, which is ultra-small for shadow-free motorized probe in distal scanning endoscopic OCT.

In this case, the configuration of a hollow motor and right-angle lens has been proved to be a simple, robust and feasible idea for the design of an ultra-small distal scanning probe. The hollow ultrasonic motor not only minimizes the size of entire probe but also reduces the length of rigid part, providing great advantage in the potential applications like vascular endoscopic imaging. There is still room for improvement in design, for this work is in experimental stage. In fact, with a new type of hollow rotor that does not limit the size of aperture at the bottom of GRIN lens, the actual working distance can be reduced, which is more convenient for practical purposes. More biocompatible catheters like the polycarbonate or Pebex tube will be used in our next version. For the convenience in practice, the tip of endoscopy will be sealed by a short transparent catheter with radius designed to match the working distance. This tip will be put in contact with the surface of the lesion, insuring that the tissue is in focus. These results and discussions imply that this design is promising in the future development of endoscopic OCT for clinical applications.

Acknowledgments

This work was supported in part by the National Natural Science Foundation of China under Grant 61227807, by the Ministry of Science and Technology of China under contracts 2006AA02Z472, 001CB510307, and 2009CB929400, by the Ministry of Education of China Grant 20130002110079 for Doctoral Program, and by the Tsinghua Initiative Scientific Research Program Grant 2013THZ02-3. The first two authors contribute equally to this paper.

References

1. 

D. Huanget al., “Optical coherence tomography,” Science, 254 (5035), 1178 –1181 (1991). http://dx.doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

2. 

G. J. Tearneyet al., “In vivo endoscopic optical biopsy with optical coherence tomography,” Science, 276 (5321), 2037 –2039 (1997). http://dx.doi.org/10.1126/science.276.5321.2037 SCIEAS 0036-8075 Google Scholar

3. 

P. R. Herzet al., “Micromotor endoscope catheter for in vivo, ultrahigh-resolution optical coherence tomography,” Opt. Lett., 29 (19), 2261 –2263 (2004). http://dx.doi.org/10.1364/OL.29.002261 OPLEDP 0146-9592 Google Scholar

4. 

K. H. Kimet al., “Two-axis magnetically-driven MEMS scanning catheter for endoscopic high-speed optical coherence tomography,” Opt. Express, 15 (26), 18130 –18140 (2007). http://dx.doi.org/10.1364/OE.15.018130 OPEXFF 1094-4087 Google Scholar

5. 

K. Aljasemet al., “Integrated three-dimensional scanner for endoscopic optical coherence tomography,” Proc. SPIE, 7716 77162P (2010). http://dx.doi.org/10.1117/12.851644 PSISDG 0277-786X Google Scholar

6. 

X. Muet al., “Compact MEMS-driven pyramidal polygon reflector for circumferential scanned endoscopic imaging probe,” Opt. Express, 20 (6), 6325 –6339 (2012). http://dx.doi.org/10.1364/OE.20.006325 OPEXFF 1094-4087 Google Scholar

7. 

G. J. Tearneyet al., “Three-dimensional coronary artery microscopy by intracoronary optical frequency domain imaging,” J. Am. Coll. Cardiol, 1 (6), 752 –761 (2008). http://dx.doi.org/10.1016/j.jcmg.2008.06.007 JACCDI 0735-1097 Google Scholar

8. 

Z. Yaqoobet al., “Methods and application areas of endoscopic optical coherence tomography,” J. Biomed. Opt., 11 (6), 063001 (2006). http://dx.doi.org/10.1117/1.2400214 JBOPFO 1083-3668 Google Scholar

9. 

G. J. Tearneyet al., “Scanning single-mode fiber optic catheter-endoscope for optical coherence tomography,” Opt. Lett., 21 (7), 543 –545 (1996). http://dx.doi.org/10.1364/OL.21.000543 OPLEDP 0146-9592 Google Scholar

10. 

H. L. Fuet al., “Flexible miniature compound lens design for high-resolution optical coherence tomography balloon imaging catheter,” J. Biomed. Opt., 13 (6), 060502 (2008). http://dx.doi.org/10.1117/1.3037340 JBOPFO 1083-3668 Google Scholar

11. 

P. H. Tranet al., “In vivo endoscopic optical coherence tomography by use of a rotational microelectromechanical system probe,” Opt. Lett., 29 (11), 1236 –1238 (2004). http://dx.doi.org/10.1364/OL.29.001236 OPLEDP 0146-9592 Google Scholar

12. 

L. Liuet al., “Miniature endoscopic optical coherence tomography probe employing a two-axis microelectromechanical scanning mirror with through-silicon vias,” J. Biomed. Opt., 16 (2), 026006 (2011). http://dx.doi.org/10.1117/1.3533323 JBOPFO 1083-3668 Google Scholar

13. 

S. Changet al., “Stationary-fiber rotary probe with unobstructed 360° view for optical coherence tomography,” Opt. Lett., 36 (22), 4392 –4394 (2011). http://dx.doi.org/10.1364/OL.36.004392 OPLEDP 0146-9592 Google Scholar

14. 

T. Wang, “Intravascular optical coherence tomography imaging at 3200 frames per second,” Opt. Lett., 38 (10), 1715 –1717 (2013). http://dx.doi.org/10.1364/OL.38.001715 OPLEDP 0146-9592 Google Scholar

15. 

Z. Tieying, “A cylindrical rod ultrasonic motor with 1 mm diameter and its application in endoscopic OCT,” Chin. Sci. Bull., 50 (8), 826 –830 (2005). http://dx.doi.org/10.1007/BF03183686 CSBUEF 1001-6538 Google Scholar

16. 

N. Zhanget al., “Spectral-domain optical coherence tomography with a Fresnel spectrometer,” Opt. Lett., 37 (8), 1307 –1309 (2012). http://dx.doi.org/10.1364/OL.37.001307 OPLEDP 0146-9592 Google Scholar

17. 

A. Duboiset al., “High-resolution full-field optical coherence tomography with a Linnik microscope,” Appl. Opt., 41 (4), 805 –812 (2002). http://dx.doi.org/10.1364/AO.41.000805 APOPAI 0003-6935 Google Scholar
CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Tianyuan Chen, Ning Zhang, Tiancheng Huo, Chengming Wang, Jing-gao Zheng, Tieying Zhou, and Ping Xue "Tiny endoscopic optical coherence tomography probe driven by a miniaturized hollow ultrasonic motor," Journal of Biomedical Optics 18(8), 086011 (1 August 2013). https://doi.org/10.1117/1.JBO.18.8.086011
Published: 1 August 2013
Lens.org Logo
CITATIONS
Cited by 33 scholarly publications.
Advertisement
Advertisement
Back to Top