Liquid-Filled Proportional Counter

Richard A. Muller, Stephen E. Derenzo, Gerard Smadja, Dennis B. Smith, Robert G. Smits, Haim Zaklad, and Luis W. Alvarez
Phys. Rev. Lett. 27, 532 – Published 23 August 1971
PDFExport Citation

Abstract

We report properties of single-wire proportional chambers and multiple-wire ionization chambers filled with liquid xenon. Proportional multiplication is seen for anode fields > 106 V/cm (anode-cathode voltage >1 kV for a 3.5-μm-diam anode wire). The efficiency for detecting ionizing radiation is ≅100%. The time resolution of the chambers is ±107 sec. We attained a spatial resolution of ±15 μm with a multiwire ionization chamber, considerably better than that of any other real-time particle detector.

  • Received 30 June 1971

DOI:https://doi.org/10.1103/PhysRevLett.27.532

©1971 American Physical Society

Authors & Affiliations

Richard A. Muller

  • University of California Space Sciences Laboratory, Berkeley, California 94720

Stephen E. Derenzo, Gerard Smadja, Dennis B. Smith, Robert G. Smits, Haim Zaklad, and Luis W. Alvarez

  • Lawrence Radiation Laboratory, University of California, Berkeley, California 94720

References (Subscription Required)

Click to Expand
Issue

Vol. 27, Iss. 8 — 23 August 1971

Reuse & Permissions