Abstract
With the expanding reach of physics, xenon-based detectors such as PandaX-4T in the China Jinping Underground Laboratory aim to cover an energy range from sub-keV to multi-MeV. A linear response of the photomultiplier tubes (PMTs) is required for both scintillation and electroluminescence signals. Through a dedicated bench test, we investigated the cause of the non-linear response in the Hamamatsu R11410-23 PMTs used in PandaX-4T. The saturation and suppression of the PMT waveform observed during the commissioning of PandaX-4T were caused by the high-voltage divider base. The bench test data validated the de-saturation algorithm used in the PandaX-4T data analysis. We also confirmed the improvement in linearity of a new PMT base design with three more low radioactivity capacitors at later dynodes, which will be used to upgrade the PMT readout system in PandaX-4T.
Export citation and abstract BibTeX RIS
Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.