Improvement on the linearity response of PandaX-4T with new photomultiplier tube bases

, , , , , , , , , , and

Published 10 May 2024 © 2024 The Author(s)
, , Citation Lingyin Luo et al 2024 JINST 19 P05021 DOI 10.1088/1748-0221/19/05/P05021

1748-0221/19/05/P05021

Abstract

With the expanding reach of physics, xenon-based detectors such as PandaX-4T in the China Jinping Underground Laboratory aim to cover an energy range from sub-keV to multi-MeV. A linear response of the photomultiplier tubes (PMTs) is required for both scintillation and electroluminescence signals. Through a dedicated bench test, we investigated the cause of the non-linear response in the Hamamatsu R11410-23 PMTs used in PandaX-4T. The saturation and suppression of the PMT waveform observed during the commissioning of PandaX-4T were caused by the high-voltage divider base. The bench test data validated the de-saturation algorithm used in the PandaX-4T data analysis. We also confirmed the improvement in linearity of a new PMT base design with three more low radioactivity capacitors at later dynodes, which will be used to upgrade the PMT readout system in PandaX-4T.

Export citation and abstract BibTeX RIS

Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.